Capacitive touchscreen
Channel
Settings
Back key Home key Menu key
Users System
Settings Measurement
Settings Log Recall Help
• return to a previous hierarchical menu level
• access to measurement screen & configured profile
• access to:
`Users – login & rights configuration
`System Settings – system configuration, connectivity & printing
`Measurement Settings – measurement configuration
`Log Recall – logged measurement data
`Help – support
• from measurement screen, tap to access measurement menu
Electrode usage guidelines & Calibration tips
• Remove protective cap and rinse with water to eliminate salt build up.
• Check correct electrolyte level. Refill when level falls below ½”
from the fill hole opening. Keep the fill hole open (1) during
calibration and measurement.
• To refresh the the electrode, soak in HI70300 Storage solution
for at least 30 minutes.
• Keep the glass bulb (3) and junction (2) hydrated to maintain a
quick response time.
Store with a few drops of storage solution in the protective cap.
• Keep the junction (2) immersed during calibration/measurement.
Calibration tips
• Ensure electrode connection to correct meter input.
• Rinse with the calibration solution intended for calibration next.
Discard rinsing solution.
• Place electrode and temperature sensor in calibration solution.
Use a rinse beaker and a separate calibration beaker for each buffer.
• On first use, tap Clear Calibration to clear all buffers and start new.
pH measurement tips
• Only work with recently calibrated electrodes.
• Use the arm holder for electrode transfer in and out of containers.
• Rinse with purified water between samples.
• Pour two beakers of sample: use one to rinse the sensor and the
second for measurement (limit contamination).
• Gently stir the sample to ensure a representative measurement.
• Keep the fill solution topped off to permit the fill solution flow
through the ceramic junction and maintain a stable reference signal.
• Record measured data once the reading is stable.
Logging in & Creating a new account
1. Power the meter and tap .
2. Tap followed by .
3. Tap Edit Users to enter Account Management.
4. Tap to enable Account Creation and Logins. Tap to return.
5. Tap Logout to enter Users screen. Tap the plus symbol avatar.
6. Input user name and tap .
7. Enter password and tap . Re-enter password to confirm.