Hanna Instruments HI 981404N User manual

Instruction Manual
(HI 981404N and HI 981405N)
pH & TDS/EC Monitors
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
HI 981404N and HI 981405N are warranted for two years
against defects in workmanship and materials when used for their in-
tended purpose and maintained according to instructions. The probes
are warranted for a period of six months. This warranty is lim-
ited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments.
Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using
the meter.
If you need additional technical information, do not hesi-
tate to e-mail us at [email protected].
These instruments are in compliance with the directives.
Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each meter is supplied complete with:
•HI 1286 pH electrode
•HI 1283 grounding probe
• Conductivity probe
HI 7634 for HI 981404N & HI 7632 for HI 981405N
• Calibration screwdriver
• 12 Vdc power adapter and instructions
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
HI 981404N (pH and TDS) and HI 981405N (pH and EC)
are specially designed to meet the needs of simple continuous
monitoring.
You can simply install the meter above the sample to be
tested for continuous measurement.
The HI 1286 gel-filled pH electrode features a waterproof
sheath to protect BNC connector against humidity and
vapours. Moreover, the unique design of the electrode
provides longer life in aggressive solutions.
The meter is equipped with two large LCDs for easy-to-read
measurements even from a distance.
Measurements are accurate and the meters can be cali-
brated at one or two points for pH, and at a single point
for TDS and EC.
You no longer need to worry about battery changes because
these units run without interruption on 12 Vdc power
supply.
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1. Liquid Crystal Displays
2. “pH 4” calibration trimmer
3. “pH 7” calibration trimmer
4. BNC connector
5. Protective sheath
6. HI 1286 pH electrode
7. HI 1283 grounding probe
8. HI 7632 or HI 7634 probe (depending on model)
9. 12 Vdc power adapter
10. Power supply connector
11. EC or TDS calibration trimmer
Common Specifications
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@ 25°C/77°F) ±0.2 pH
Calibration Manual, 1 or 2 point
Probes HI 1286 pH electrode (included) and
HI 1283 grounding probe (fixed)
Environment 0 to 50°C (32 to 122°F); RH max 95%
Power Supply 12 Vdc adapter (included)
Dimensions 160 x 110 x 35 mm (6.3 x 4.3 x 1.4”)
Weight 300 g (10.6 oz.)
HI 981404N only
Range 0 to 1990 mg/L (ppm)
Resolution 10 mg/L (ppm)
Accuracy (@ 25°C/77°F) ±2% f.s.
TDS Factor 0.7 ppm = 1 µS/cm
Temperature Automatic, 5 to 50°C (41 to 122°F)
Compensation TDS only
Calibration Manual, 1 point
TDS Probe HI 7634 (fixed)
HI 981405N only
Range 0.00 to 9.99 mS/cm
Resolution 0.01 mS/cm
Accuracy (@ 25°C/77°F) ±2% f.s.
Temperature Automatic, 5 to 50°C (41 to 122°F)
Compensation EC only
Calibration Manual, 1 point
EC Probe HI 7632 (fixed)
SPECIFICATIONS
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instrument in residential areas could cause unaccept-
able interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching
the bulb at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance. To avoid electrical shock, do not use this instrument when voltage at the
measurement surface exceeds 24 Vac or 60 Vdc. To avoid damage or burns, do not perform
any measurement in microwave ovens.
CE DECLARATION OF CONFORMITY
ACCESSORIES
HI 1286 Double junction, plastic body pH electrode
with 2 m (6.6’) cable and BNC connector
HI 1283* Stainless steel grounding probe with 2 m
(6.6’) cable
HI 7632* Conductivity probe with 2 m (6.6’) cable for
HI 981405N
HI 7634* Conductivity probe with 2 m (6.6’) cable for
HI 981404N
HI 70004P pH 4.01 buffer, 20 mL sachet, 25 pcs
HI 70007P pH 7.01 buffer, 20 mL sachet, 25 pcs
HI 70010P pH 10.01 buffer, 20 mL sachet, 25 pcs
HI 7004M pH 4.01 buffer solution, 230 mL bottle
HI 7007M pH 7.01 buffer solution, 230 mL bottle
HI 7010M pH 10.01 buffer solution, 230 mL bottle
HI 7004L pH 4.01 buffer solution, 500 mL bottle
HI 7007L pH 7.01 buffer solution, 500 mL bottle
HI 7010L pH 10.01 buffer solution, 500 mL bottle
HI 70031P 1.41 mS/cm solution, 20 mL sachet, 25 pcs
HI 70442P 1500 ppm solution, 20 mL sachet, 25 pcs
HI 7031M 1.41 mS/cm standard solution, 230 mL bottle
HI 70442M 1500 ppm standard solution, 230 mL bottle
HI 7031L 1.41 mS/cm standard solution, 500 mL bottle
HI 70442L 1500 ppm standard solution, 500 mL bottle
HI 70300M Storage solution, 230 mL bottle
HI 70300L Storage solution, 500 mL bottle
HI 7061M Cleaning solution, 230 mL bottle
HI 7061L Cleaning solution, 500 mL bottle
HI 710005 12 Vdc power adapter, US plug
HI 710006 12 Vdc power adapter, European plug
HI 710012 12 Vdc power adapter, Australian plug
HI 710013 12 Vdc power adapter, Southern Africa plug
HI 710014 12 Vdc power adapter, UK plug
* To be replaced by authorized technical personnel only
ISTCOMBON R1 05/05
OPERATIONAL GUIDE
pH ELECTRODE CONNECTION
In order to protect the instrument against vapors and
humidity, the BNC connector is shielded behind a waterproof
sheath.
• Slide the protective sheath down, connect the pH elec-
trode to the BNC connector and then slide the protective
sheath back up. For maximum waterproof protection,
make sure the connector is completely covered.
• Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH elec-
trodes and they dissolve when rinsed with water.
TAKING MEASUREMENTS
• Turn the meter on by connecting the 12 Vdc power
adapter to the meter and the mains.
• Remove the protective cap from the pH electrode.
• Immerse the tips (4cm/1½”)
of pH electrode, conductiv-
ity and grounding probes
into the sample.
For better accuracy, the
probes should not touch or
stand close to the vessel’s
walls or bottom.
• The two LCDs will show the
pH and EC (or TDS) values.
Any initial variation may
be due to the pH electrode
conditioning and EC/TDS
temperature compensation.
Allow the readings to sta-
bilize.
pH ELECTRODE MAINTENANCE
• When not in use, rinse the electrode with water and store
it with a few drops of storage (HI 70300) or pH7
(HI 7007) solution in the protective cap. Always replace
the protective cap after use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
• If the electrode has been left dry, soak the tip in a
storage (HI 70300) or pH7 (HI 7007) solution over-
night to reactivate it.
The instrument must be recalibrated whenever:
a) After cleaning or replacing the probe
b) Where high accuracy is required
c) At least once a month
pH CALIBRATION
Pour small quantities of pH 7.0 (HI 7007) and pH 4.0
(HI 7004) or pH 10.0 (HI 7010) solution into two clean
beakers.
For accurate calibration use two beakers for each buffer
solution, the first one for rinsing the tip of the electrode and
the second one for calibration.
• Turn the meter on.
• Remove the protective cap, rinse and immerse pH
electrode and gorunding probe in pH 7.0 buffer solu-
tion. Stir gently and wait for the reading to stabilize.
Note: The pH electrode should be submerged approximately
4 cm (1½”) in the solution.
• With the calibration screwdriver, adjust the “pH 7”
trimmer until the LCD shows “pH 7.0”.
•Rinse and immerse pH electrode and grounding probe
in pH 4.0 (or pH 10.0) buffer. Stir gently.
CALIBRATION
• Wait a couple of minutes and then adjust the “pH 4”
trimmer until the LCD shows “pH 4.0” (or “pH 10.0”).
The pH calibration is now complete.
CONDUCTIVITY CALIBRATION
• Turn the meter on.
• Pour a small quantity of the proper calibration solution
in a beaker (HI 7031 for HI 981405N and HI 70442 for
HI 981404N). If possible, use plastic beakers to mini-
mize any EMC interference.
• Immerse the conductivity probe in the solution, making
sure that the metal pins are completely submerged.
Note: For best accuracy, the probe body does not touch nor
stand close to the side walls of the beaker.
• Wait for a couple of minutes for thermal equilibrium to
be reached and the temperature to be compensated.
• Tap the probe gently on the bottom, then shake it to
make sure no air bubbles have remained trapped.
• With the supplied screwdriver, adjust the calibration trim-
mer until the display shows 1.41 mS/cm (HI 981405N) or
1500 ppm (HI 981404N).
Calibration is now complete.
• To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it at least once a
month. Immerse the tip of the electrode in HI 7061
cleaning solution for half an hour and then rinse it with
tap water.
CONDUCTIVITY PROBE MAINTENANCE
The following cleaning procedure is recommended at least
once a month:
• Immerse the tip of the probe in HI 7061 cleaning
solution for half an hour.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning, rinse the probe with tap water.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments PRIMO 5 User manual

Hanna Instruments

Hanna Instruments PRIMO 5 User manual

Hanna Instruments HI84529 User manual

Hanna Instruments

Hanna Instruments HI84529 User manual

Hanna Instruments HI97713 User manual

Hanna Instruments

Hanna Instruments HI97713 User manual

Hanna Instruments HI 96759C User manual

Hanna Instruments

Hanna Instruments HI 96759C User manual

Hanna Instruments HI 717 User manual

Hanna Instruments

Hanna Instruments HI 717 User manual

Hanna Instruments HI1006-18 User manual

Hanna Instruments

Hanna Instruments HI1006-18 User manual

Hanna Instruments HI99111 User manual

Hanna Instruments

Hanna Instruments HI99111 User manual

Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS User manual

Hanna Instruments

Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS User manual

Hanna Instruments HI98194 User manual

Hanna Instruments

Hanna Instruments HI98194 User manual

Hanna Instruments HI 93124 User manual

Hanna Instruments

Hanna Instruments HI 93124 User manual

Hanna Instruments HI 981402 User manual

Hanna Instruments

Hanna Instruments HI 981402 User manual

Hanna Instruments HI727 User manual

Hanna Instruments

Hanna Instruments HI727 User manual

Hanna Instruments HI 762 User manual

Hanna Instruments

Hanna Instruments HI 762 User manual

Hanna Instruments HI 9812-0 User manual

Hanna Instruments

Hanna Instruments HI 9812-0 User manual

Hanna Instruments Gro Line HI981030 User manual

Hanna Instruments

Hanna Instruments Gro Line HI981030 User manual

Hanna Instruments HI 991001 User manual

Hanna Instruments

Hanna Instruments HI 991001 User manual

Hanna Instruments HI97753 User manual

Hanna Instruments

Hanna Instruments HI97753 User manual

Hanna Instruments HI96101C User manual

Hanna Instruments

Hanna Instruments HI96101C User manual

Hanna Instruments HI 84533 User manual

Hanna Instruments

Hanna Instruments HI 84533 User manual

Hanna Instruments HI 95729C User manual

Hanna Instruments

Hanna Instruments HI 95729C User manual

Hanna Instruments HI 9811-5 User manual

Hanna Instruments

Hanna Instruments HI 9811-5 User manual

Hanna Instruments HI 96753C User manual

Hanna Instruments

Hanna Instruments HI 96753C User manual

Hanna Instruments HI6321 User manual

Hanna Instruments

Hanna Instruments HI6321 User manual

Hanna Instruments HI 5421 User manual

Hanna Instruments

Hanna Instruments HI 5421 User manual

Popular Measuring Instrument manuals by other brands

Oval EL4101 Key operations guide

Oval

Oval EL4101 Key operations guide

TOOLCRAFT LDM 50 U operating instructions

TOOLCRAFT

TOOLCRAFT LDM 50 U operating instructions

PCB Piezotronics IMI SENSORS EX637 Series Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS EX637 Series Installation and operating manual

Wet Labs ac-s user guide

Wet Labs

Wet Labs ac-s user guide

TSI Instruments 9535A Operation and service manual

TSI Instruments

TSI Instruments 9535A Operation and service manual

BMS ETA-0832 Operational manual

BMS

BMS ETA-0832 Operational manual

Tektronix Keithley 7711 instructions

Tektronix

Tektronix Keithley 7711 instructions

MG MG600S instructions

MG

MG MG600S instructions

Keysight Technologies 34136A user guide

Keysight Technologies

Keysight Technologies 34136A user guide

Heyuan Intelligence DZ81-MA3UI5C user manual

Heyuan Intelligence

Heyuan Intelligence DZ81-MA3UI5C user manual

Nalco TRASAR 2 Procedure guide

Nalco

Nalco TRASAR 2 Procedure guide

IFM Electronic efector300 SD5000 operating instructions

IFM Electronic

IFM Electronic efector300 SD5000 operating instructions

Adeunis RF WATERLEAK SPOT quick start guide

Adeunis RF

Adeunis RF WATERLEAK SPOT quick start guide

TSI Instruments DP-CALC 5825 Operation and service manual

TSI Instruments

TSI Instruments DP-CALC 5825 Operation and service manual

Extech Instruments MW12 user guide

Extech Instruments

Extech Instruments MW12 user guide

Emtec BioProTT FlowTrack SL user manual

Emtec

Emtec BioProTT FlowTrack SL user manual

Pro's Kit MT-3102 user manual

Pro's Kit

Pro's Kit MT-3102 user manual

PPI ODPM-48 operating manual

PPI

PPI ODPM-48 operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.