manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Measuring Instrument
  8. •
  9. Hanna Instruments HI 991003 User manual

Hanna Instruments HI 991003 User manual

Dear Customer,
Thank you for choosing a Hanna Instruments
product.
Please read this instruction manual care-
fully before using the instrument. If you need
additional technical information, do not hesi-
tate to e-mail us at [email protected]
This instrument is in compliance with the
directives.
All Hanna Instruments meters are warranted
for two years against defects in workmanship
and materials when used for their intended pur-
pose and maintained according to instructions.
The probes are warranted for a period of
six months.
This warranty is limited to repair or replacement
free of charge. Damage due to accidents, mis-
use, tampering or lack of prescribed maintenance
are not covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of pur-
chase, serial number and the nature of the prob-
lem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instru-
ments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service depart-
ment and then send it with shipment costs pre-
paid. When shipping any instrument, make sure it
is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent of
the copyright owner.
WARRANTY
PRELIMINARY EXAMINATION
Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your dealer.
Note: Save all packing material until you
are sure that the instrument func-
tions correctly. All defective items
must be returned in the original
packing together with the supplied
accessories.
HI 991003 is a portable pH/pH-mV/ORP/
temperature meter, with unique Sensor
Check feature, that allows the user to de-
termine the electrode status at any time.
Main features also include: extended mea-
surement ranges; waterproof and compact
casing; large dual-line LCD; low battery de-
tection; automatic pH calibration at one or
two points with two memorized buffer sets;
selectable temperature unit (°C or °F).
The pH/ORP electrode has a built-in tem-
perature sensor for simultaneous pH (or
mV) and temperature readings, and also
contains a mini amplifier to render mea-
surements impervious to noise and elec-
trical interferences.
Each meter is supplied complete with
•HI 1297D pH/ORP/temperature probe with
DIN connector and 1 m (3.3’) cable
•calibration & cleaning solution sachets
•3x1.5V AA alkaline batteries
• instruction manual
• rugged carrying case.
GENERAL DESCRIPTION
SPECIFICATIONS
Range (*) -2.00 to 16.00 pH
±825 mV (pH-mV)
±1999 mV (ORP)
-5.0 to 105.0°C / 23.0 to 221.0°F
Resolution 0.01 pH
1 mV
0.1°C / 0.1°F
Accuracy ±0.02 pH
(@20°C/68°F) ±2 mV
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140°F; ±2°F outside
TypicalEMC ±0.02 pH
Deviation ±2 mV
±0.2°C or ±0.4°F
Temperature Automatic for pH readings
Compensation
pHCalibration Automatic, 1 or 2 point
with 2 sets of memorized buffers
(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)
mV Calibration Factory calibrated
Probe (included) HI 1297D
amplified pH/ORP/temperature
probewithDIN connector and 1 m (3.3’)cable
Battery Type 3x 1.5VAA/IECLR6
Battery Life Approx. 1500 hours
Auto-off After 8 minutes of non-use
Environment 0 to 50°C (32 to 122°F);
RH max 100%
Dimensions
150 x 80 x 36 mm (5.9 x 3.1 x 1.4”)
Weight 210 g (7.4 oz.)
(*) the temperature range is limited to 80°C (176°F) if
using the HI1296D or HI1297D probes.
To clean the meter, use water only.
HI 1297D Combination amplified pH/
ORP/temperature probe with
DIN and 1 m (3.3’) cable
HI 1296D Combination amplified pH/
temperature probe with DIN
connector and 1 m (3.3’) cable
HI 7004L pH 4.01 buffer solution, 500 mL
HI 7006L pH 6.86 buffer solution, 500 mL
HI 7007L pH 7.01 buffer solution, 500 mL
HI 7009L pH 9.18 buffer solution, 500 mL
HI 7010L pH 10.01 solution, 500 mL
HI 7020L 200/275mVtestsolution, 500 mL
HI 70300L Storage solution, 500 mL
HI 7061L Cleaning solution, 500 mL
HI 710007 Blue shockproof rubber boot
HI 710008 Orangeshockproofrubberboot
HI 721312 Rugged carrying case
HI 76405 Electrode holder
ACCESSORIES
LCD DESCRIPTION
IST991003R1 07/05
1. Automatic Temperature Compensation
indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display
www.hannainst.com
Instruction Manual
HI 991003
PortablepH/mVMeter
withSensorCheck
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
To connect the probe
With the meter turned off, connect the HI 1297D
probetotheDINsocketonthetopofthemeterby
aligningthe pins and pushingin the plug. Tighten
the nut to ensure a good connection. Remove
the protective cap from the probe before taking
any measurements.
To turn the meter ON
Press the ON/OFF/MODE button until the dis-
play lights up.At start-up, all the LCD segments
are displayed for 1 second, then the percent in-
dication of the remaining battery life is shown for
another second (E.g. % 100 BATT). The meter
then enters the normal measuring mode.
Note: If the display needs to be checked, keep
the ON button pressed while turning the meter
on. The meter will display all segments as
long as the button is pressed.
To select the measurement range
While in normal measurement mode, press the
SET/HOLD button to select pH, pH-mV or ORP
(mV) measurement range.
Note: “pH-mV” indicates the mV value equiva-
lent to the pH.
To freeze the display
While in measurement mode, press the SET/
HOLD button until HOLD appears on the sec-
ondary display; the reading will be frozen on the
LCD (E.g. pH 5.73 HOLD). Press any button to
return to normal mode.
To turn the meter OFF
While in normal measurement mode, press the
ON/OFF/MODE button. OFF will appear on the
secondary display. Release the button.
Note: The meter is provided with an acoustic
signal feature, which can be disabled using
the switch located in the battery compartment.
Note: When the meter detects the absence of a
temperature probe at its input, the Automatic
Temperature Compensation is turned off, and
the meter uses a default value of 25°C (77°F)
for the temperature measurement and com-
pensation. In this condition, the secondary
LCD shows 25.0°C (77.0°F) blinking. When a
probe is connected, the meter automatically
returnsto theATC mode, theATCtagisturned
on, and the temperature is shown on the sec-
ondary display.
OPERATIONAL GUIDE
• Make sure the meter has been calibrated be-
fore use.
• If the probe is dry, soak it in HI 70300 storage
solution for one hour to reactivate it.
• Set the meter to pH range.
• Place the probe into the sample to be tested.
• Stir briefly and wait until the symbol on the
top left of the LCD is turned off.
• The LCD shows the pH value (automatically
compensated for temperature) and the tem-
perature of the solution.
• If measurements are taken in different
samples successively, rinse the probe tip thor-
oughly to eliminate cross-contamination. After
cleaning, rinse the probe tip with some of the
sample to be measured.
pH calibration
• PressandholdtheON/OFF/MODEbuttonuntil
OFF on the secondary display is replaced by
CAL.
• Release the button. The LCD enters the cali-
bration mode, displaying “pH 7.01 USE” (or
“pH 6.86 USE” if the NIST buffer set was se-
lected). After 1 second the meter activates the
automatic buffer recognition feature. If a valid
buffer is detected, then its value is shown on
the primary display, and REC appears on the
secondarydisplay.Ifno validbufferisdetected,
the meter keeps the USE indication active for
12seconds,andthenitreplacesitwithWRNG,
indicating that the sample being measured is
not a valid buffer.
• For a
single-point calibration
with buffers pH
4.01, 9.18 or 10.01, the meter automatically
accepts the calibration when the measure-
ment is stable; the meter will show on the
primary display the accepted buffer, with the
message “OK 1” on the secondary display,
and an audible signal is produced.After 1 sec-
ond, the meter automatically returns to the
normal mode.
Ifasingle-point calibration with buffers pH 7.01
or 6.86 is desired, then after the calibration
point has been accepted the ON/OFF/MODE
button must be pressed in order to return to
the normal measuring mode. After the button
is pressed, the meter shows "7.01" (or "6.86")
- "OK 1", and an audible signal is produced.
pH MEASUREMENT & CALIBRATION
Setup mode allows the selection of the tem-
perature unit and of the pH buffer set.
To enter the Setup mode, press and hold the
ON/OFF/MODE button until CAL on the sec-
ondary display is replaced by TEMP and the
currenttemperatureunit(E.g.TEMP °C). Then:
•
for°C/°Fselection
,useSET/HOLD.Afterthe
temperature unit has been selected, press
ON/OFF/MODE to enter the buffer set se-
lection mode; press ON/OFF/MODE twice
to return to the normal measuring mode.
•
to change the calibration buffer set
, after
setting the temperature unit, the meter will
show the current buffer set: “pH 7.01 BUFF”
(for 4.01/7.01/10.01) or “pH 6.86 BUFF” (for
4.01/6.86/9.18). Change the set with SET/
HOLD, then press ON/OFF/MODE to return
to normal mode.
METER SETUP
The meter displays the remaining battery per-
centage when turned on. When the level is
below 5%, the

symbol on the bottom left
of the LCD blinks to indicate a low battery con-
dition.Ifthebatterylevel is lowenoughtocause
erroneous readings, the Battery Error Preven-
tion System (BEPS) turns the meter off.
Unscrew the 4 screws located on the back of
the meter and carefully replace the 3 AA bat-
teries located in the battery compartment,
while paying attention to their polarity. Reat-
tach the back making sure that the gasket is
in place and tighten the 4 screws.
BATTERY REPLACEMENT
• Make sure the probe is connected.
• Set the meter to ORP (mV) range.
• Place the probe into the sample to be tested,
stir briefly and wait until the symbol on the
top left of the LCD is turned off.
• The meter will show the ORP (mV) value and
the temperature of the sample.
Note: The ORP range is factory calibrated.
Note:ATCtagturnsoffbecausethemVreadings
are not temperature compensated.
ORP MEASUREMENT
SENSOR CHECK
The Sensor Check feature allows the user to
check the electrode status at any time. This is
possible by setting the meter to pH-mV range.
The offset value is the reading of the electrode
immersedinpH7buffer(@25°C/77°F).Ifthisread-
ing is outside the range ±40 mV, the electrode is
considered "poor" and it is recommended to re-
place it with a new one.
Theslopevalueoftheelectrodeisthedifference
between readings in pH7 and in pH4 buffers.
Foranewelectrodethisisabout174mV(@25°C/
77°F). With time, the slope value decreases.
When it reaches the value of about 150 mV, the
electrode is considered "poor", and it is recom-
mended to replace it with a new one.
Note: Always recalibrate the meter after replac-
ing the electrode.
After1second,themeter automatically returns
to the normal measuring mode.
• For a
two-point calibration
, place the probe in
pH7.01(orpH 6.86)buffer.Afterthecalibration
point has been accepted, the “pH 4.01 USE”
message appears. The message is held for
12 seconds, unless a valid buffer is recog-
nized. If no valid buffer is recognized, then the
WRNG message is shown. If a valid buffer
(pH 4.01, pH 10.01 or pH 9.18) is detected,
then the meter completes the calibration pro-
cedure. When the buffer is accepted, the LCD
shows the accepted value with the “OK 2”
message on the secondary display. The meter
then returns to the normal measuring mode.
Note: When the calibration procedure is com-
pleted, the CAL tag is turned on.
Toquitcalibrationandtoresetto the default
• After entering the calibration mode and before
the first point is accepted, it is possible to quit
the procedure and return to the last calibration
data by pressing the ON/OFF/MODE button.
The secondary LCD displays ESC for 1 sec-
ond and the meter returns to normal mode.
• To reset to the default values and clear a pre-
vious calibration, press the SET/HOLD button
after entering the calibration mode and before
the first point is accepted.
The secondary LCD displays CLR for 1 sec-
ond, the meter resets to the default calibration
and the “CAL” tag on the LCD disappears.
Note: To ensure reliable readings, the electrode
must be cleaned with cleaning solution be-
fore measuring the offset and the slope.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI739 User manual

Hanna Instruments

Hanna Instruments HI739 User manual

Hanna Instruments HALO 2 HI9810392 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810392 User manual

Hanna Instruments HI 93706 User manual

Hanna Instruments

Hanna Instruments HI 93706 User manual

Hanna Instruments HI98194 User manual

Hanna Instruments

Hanna Instruments HI98194 User manual

Hanna Instruments HI97706 User manual

Hanna Instruments

Hanna Instruments HI97706 User manual

Hanna Instruments HI83303 User manual

Hanna Instruments

Hanna Instruments HI83303 User manual

Hanna Instruments 96708 User manual

Hanna Instruments

Hanna Instruments 96708 User manual

Hanna Instruments Checker HI717 User manual

Hanna Instruments

Hanna Instruments Checker HI717 User manual

Hanna Instruments Foodcare HI981035 User manual

Hanna Instruments

Hanna Instruments Foodcare HI981035 User manual

Hanna Instruments HI 2214 User manual

Hanna Instruments

Hanna Instruments HI 2214 User manual

Hanna Instruments HI96705C User manual

Hanna Instruments

Hanna Instruments HI96705C User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI97715 User manual

Hanna Instruments

Hanna Instruments HI97715 User manual

Hanna Instruments HI 96701C User manual

Hanna Instruments

Hanna Instruments HI 96701C User manual

Hanna Instruments HI96733C User manual

Hanna Instruments

Hanna Instruments HI96733C User manual

Hanna Instruments HI 96734C User manual

Hanna Instruments

Hanna Instruments HI 96734C User manual

Hanna Instruments HI98164 User manual

Hanna Instruments

Hanna Instruments HI98164 User manual

Hanna Instruments HI 95759 User manual

Hanna Instruments

Hanna Instruments HI 95759 User manual

Hanna Instruments HI 96721C User manual

Hanna Instruments

Hanna Instruments HI 96721C User manual

Hanna Instruments HI96704C User manual

Hanna Instruments

Hanna Instruments HI96704C User manual

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments HI84531-02 User manual

Hanna Instruments

Hanna Instruments HI84531-02 User manual

Hanna Instruments 96711C User manual

Hanna Instruments

Hanna Instruments 96711C User manual

Hanna Instruments HI747 User manual

Hanna Instruments

Hanna Instruments HI747 User manual

Popular Measuring Instrument manuals by other brands

Promax 6 manual

Promax

Promax 6 manual

GreenValley LiAir 50N user guide

GreenValley

GreenValley LiAir 50N user guide

Aqualytic AL100 instruction manual

Aqualytic

Aqualytic AL100 instruction manual

Procom REX Series installation guide

Procom

Procom REX Series installation guide

red lion PAXLTC quick start guide

red lion

red lion PAXLTC quick start guide

Promax TV EXPLORER II user manual

Promax

Promax TV EXPLORER II user manual

AutoTest BAMbino user manual

AutoTest

AutoTest BAMbino user manual

EGAmaster 65507 operating instructions

EGAmaster

EGAmaster 65507 operating instructions

Check-line TI-25DL-MMX-EXT operating instructions

Check-line

Check-line TI-25DL-MMX-EXT operating instructions

Aqua Medic Marin Control Digital Operation manual

Aqua Medic

Aqua Medic Marin Control Digital Operation manual

Kyoritsu Electrical Instruments Works, Ltd. 5202 instruction manual

Kyoritsu Electrical Instruments Works, Ltd.

Kyoritsu Electrical Instruments Works, Ltd. 5202 instruction manual

BZB Gear BG-AVTPG-4K user manual

BZB Gear

BZB Gear BG-AVTPG-4K user manual

Berger + Schröter LW800 GOLF PRO user manual

Berger + Schröter

Berger + Schröter LW800 GOLF PRO user manual

HYDACELECTRONIC HIT 1500 Protocol manual

HYDACELECTRONIC

HYDACELECTRONIC HIT 1500 Protocol manual

Spot-on 39240 product manual

Spot-on

Spot-on 39240 product manual

In-situ Partech 228531 Operator's manual

In-situ

In-situ Partech 228531 Operator's manual

Unit UTS3000B/A Series user manual

Unit

Unit UTS3000B/A Series user manual

Digi-Sense 20250-65 user manual

Digi-Sense

Digi-Sense 20250-65 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.