Happy Electronics Alert Stresslocator Oximeter User manual

Alert Stresslocator Oximeter – User's guide version 2
StressLocator & Sleep as Android
Alert Stresslocator Oximeter
User's Guide
1/6

Alert Stresslocator Oximeter – User's guide version 2
Introduction .......................................................................................................................................... 2
Disclaimer ............................................................................................................................................. 2
Compatibility ........................................................................................................................................ 2
Package content .................................................................................................................................... 2
pecifications ....................................................................................................................................... 3
Product description ............................................................................................................................... 3
Measurement Principle ..................................................................................................................... 3
Advice and warnings ............................................................................................................................ 4
First start ............................................................................................................................................... 4
Directions for use ............................................................................................................................. 4
Display Introduction ......................................................................................................................... 5
Operating button explanation ........................................................................................................... 5
Function Description ........................................................................................................................ 5
Usage with applications ........................................................................................................................ 6
Usage with the tressLoctor app ...................................................................................................... 6
Usage with the leep as Android app ............................................................................................... 6
Cleaning and Disinfection .................................................................................................................... 6
Cleaning ............................................................................................................................................ 6
Disinfection ...................................................................................................................................... 6
Introduction
Congratulations on purchasing Alert tresslocator Oxieter. Your new product measures blood
oxygenation and heart rate, it is a comfortable, light-weight, portable device for non-invasive, on
the spot measuring, able to transmit the data to your smartphone via bluetooth. There are 2
inteligent alarms - one for PO2 and the other for heart rate decrease, which will wake you up,
when your testing results are beyond the normal range! This guide will help you set up and operate
the device.
Disclaimer
The Alert tresslocator Oxieter is no professional medical device and should be used by healthy
individuals and only after a proper introduction.
Compatibility
The Alert tresslocator Oxieter is designed to work with Android devices running the operating
system Android 4.3 or higher that are quipped with bluetooth, compatible with tressLocator and
leep as Android apps.
Package content
Alert tresslocator Oxieter
2x soft foam sensor
charging cable
user's guide
2/6

Alert Stresslocator Oximeter – User's guide version 2
Specifications
Bluetooth
version 4.0
BLE - Bluetooth Low Energy
Measurement Range
po2: 35~100 %
Accuracy: ± 2 % (80~100 %); ± 3 % (70~79 %)
Pulse Rate: 25~250 BPM
Accuracy: ± 2 bpm
Environment
Temperature
Working: 5~40 °C
torage:- 10~50 °C
Humidity
Working: 15~80%
torage: 10~90%
Product description
The Alert tresslocator Oxieter measures blood oxygenation and heart rate. The device is optical-
based - two beams of different vawelenght of light screen the blood vessels and capillaries of your
finger; more precisely of the nail tip, and that through the finger sensor.
Results showing 96% - 99% are considered to be normal. Greater height above sea level and other
factors may affect the measurement results.
Measurement Principle
Operation principle of the instrument is Photoelectric Oxyhemoglobin Inspection Technology is
adopted in accordance with Capacity Pulse canning and Recording Technology, so that two beams
of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto
human nail tip through perspective clamp finger-type sensor. Then measured signal can be obtained
by a photosensitive element, information acquired through which will be shown on two groups of
LEDs through process in electronic circuits and microprocessor.
Arterial oxygen saturation is measured by a method called pulse oximetry. It is a continuous, non-
invasive method based on the different spectra absorption of hemoglobin and oxyhemoglobin
(called spectrophotometer principle). It measures how much light, sent from light sources on the
other side.
iagram of Operation Principle
3/6

Alert Stresslocator Oximeter – User's guide version 2
Advice and warnings
Do not use the oximeter nearby sources of strong magnetic resonance.
Do not use the device and a blood pressure monitor with a cuff on the same arm at the same
time.
To ensure proper functioning, keep the device clean.
For precise measuring, your finger must be clean.
The following factor may reduce the accuracy of the measurement:
oblinking or very strong light
ostrong transmitters (e.g. television, radio) may affect precesion of the oximeter
olow perfusion
olow hemoglobin level
ohypotension, severe vasoconstriction, severe anemia or hypothermia
Use the pO2 sensor delivered with the device or a sensor which the manufacturer has
approved. As a user, you are responsible for checking the compatibility.
Do not use in flammable environment.
Do not use in an environment that does not fulfill the operating and storage requirements.
The Alert tresslocator Oxieter is fragile, try to avoid any strong pressure, knock or
vibration that could damage it.
The device is not made of any health damaging materials.
The package of the device is recyclable and must be handeled in accordance with local laws.
All unrecyclable parts of the tressLocator Oximeter must be handeled in accordance with
local laws.
First start
Directions for use
1. Plug the sensor into the body of the oximeter
2. Turn the device on by pressing the power button
3. Fix the sensor on patient's finger as shown on the image below
4/6

Alert Stresslocator Oximeter – User's guide version 2
Display Introduction
Operating button explanation
Power on: Long touch “POWER/FUNCTION” button, machine starts up and enter main menu.
Power off: If the sensor is dropped off for 10 seconds, the machine will automatically power off.
Function Description
When the product is powered on (as shown in Figure 5), short touch the
“POWER/FUNCTION” button, the main interface will be reversed and the function of
buzzer will be activated. (as shown in Figure 6)
hort touch the “POWER/FUNCTION” button again, the main interface will be reversed to
the obverse side and the function of buzzer is still available. (as shown in Figure 7)
If you want to shut down the buzzer, short touch the “POWER/FUNCTION” button and the
main interface will be reversed, the buzzer will be shut down at the same time. hort touch
the “POWER/FUNCTION” button again, the main interface will be reversed to the obverse
side and the buzzer still remains unavailable.
The product will automatically power off when no finger is in the device for longer than 10
seconds. (as shown in Figure 8)
Long touch the “POWER/FUNCTION” button, to enter the main menu or to select or to
switch the option
hort touch the “POWER/FUNCTION” button, to go trough the menu
Figure 5 Figure 6 Figure 7 Figure 8
5/6

Alert Stresslocator Oximeter – User's guide version 2
Usage with applications
Usage with the StressLoctor app
1. Download and install the tressLocator app from Google Play
2. Turn the device on
3. Go to the settings menu of your Android device
4. elect the “bluetooth” option
5. witch “bluetooth” to “on”
6. It is NOT possible to pair BLE tresslocator, appropriate connection is made directly in the
app
7. tart the app
8. elect “ tart” in the main menu
9. In the Device menu, select “ BLE tressLocator ”, put it on your finger and start
measurement
Usage with the Sleep as ndroid app
1. Download the leep as Android app from Google Play
2. Go to the settings menu of your Android device
3. elect the “bluetooth” option
4. witch “bluetooth” to “on”
5. It is NOT possible to pair BLE tresslocator, appropriate connection is made directly in the
app
6. Enable Settings – Wearables – Pulse Oximeter, put the device on your finger and start
sleep tracking
If you have any problem with this app please send us a report and we will help you solve it
Cleaning and Disinfection
Cleaning
1. Clean the product with cotton or soft cloth moistened with water.
2. After cleaning, wipe off the water with a soft cloth.
3. Allow the product to air dry.
Disinfection
The recommended disinfectants include: 70% ethanol, 70% isopropanol, glutaraldehyde (2%)
solution disinfectants.
1. Clean the product as instructed above.
2. Disinfect the product with cotton or soft cloth moistened with one of the recommended
disinfectants.
3. After disinfection, be sure to wipe off the disinfectant left on the product with a soft cloth
moistened with water.
4. Allow the product to air dry.
6/6
Table of contents