Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
6
Solution 2: Replace the connecting wire between main board and driver board
and check if machine returns to normal.
Possible Cause 2: driver board is damaged.
Solution: Replace the driver board.
1.5Error code 0XE3” after turn on the machine.
Possible Cause 1: Memory board encounters interference that cause incorrect
writing.
Solution: Recover factory default setting as per instructions in user manual.
Note: If the infusion set being used is not supplied by IV pump manufacturer,
please recalibrate its accuracy and pressure value after recover to factory default
setting.
Possible Cause 2: Main board is damaged.
Solution: Replace the main board.
2. Abnormal During Operation
2.1 Green indicator light sometimes is on and sometimes is off; or error alarm for
air bubble even correct installation of IV set.
Possible Cause 1: The IV set is too thin or too soft (the normal accepted diameter
of IV set is 3.5---4.5mm)
Possible Cause 2: ir bubble sensor is damaged.
Solution: Replace air bubble sensor.
1). unplug the red black wire and white yellow wire. Of air bubble sensor.
2). Remove the screw on air bubble sensor
3). Open pump door
4). Use tweezers from the position shown in Diagram11 to push off the sensor.
5). Install new air bubble sensor. Insert two cable as shown in Diagram 11.