manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hawk
  6. •
  7. Water Pump
  8. •
  9. Hawk HK-1001 User manual

Hawk HK-1001 User manual

Other Hawk Water Pump manuals

Hawk Doppler Microwave Series User manual

Hawk

Hawk Doppler Microwave Series User manual

Hawk HFR Series User manual

Hawk

Hawk HFR Series User manual

Hawk s1 User manual

Hawk

Hawk s1 User manual

Hawk XLT IEX Series User manual

Hawk

Hawk XLT IEX Series User manual

Hawk Hawk-i1 User manual

Hawk

Hawk Hawk-i1 User manual

Popular Water Pump manuals by other brands

Neptun NPHW 5500 operating instructions

Neptun

Neptun NPHW 5500 operating instructions

Sensidyne Gilian BDX-II Operation & service manual

Sensidyne

Sensidyne Gilian BDX-II Operation & service manual

Draper 47808 instructions

Draper

Draper 47808 instructions

Trolla 12820 manual

Trolla

Trolla 12820 manual

Sunfab SVH Series installation guide

Sunfab

Sunfab SVH Series installation guide

WilTec 50740 Operation manual

WilTec

WilTec 50740 Operation manual

Fieldmann FVC 5015 EK user manual

Fieldmann

Fieldmann FVC 5015 EK user manual

Everbilt EFSUB5-122HD Use and care guide

Everbilt

Everbilt EFSUB5-122HD Use and care guide

Graco Husky 3300e manual

Graco

Graco Husky 3300e manual

esotec 101018 operating manual

esotec

esotec 101018 operating manual

Becker BASIC VASF 2.80/1-0.AC230 operating instructions

Becker

Becker BASIC VASF 2.80/1-0.AC230 operating instructions

Sykes AmeriPumps GP100M Operation and maintenance instructions

Sykes AmeriPumps

Sykes AmeriPumps GP100M Operation and maintenance instructions

DUROMAX XP WX Series user manual

DUROMAX

DUROMAX XP WX Series user manual

BRINKMANN PUMPS SBF550 operating instructions

BRINKMANN PUMPS

BRINKMANN PUMPS SBF550 operating instructions

Franklin Electric IPS Installation & operation manual

Franklin Electric

Franklin Electric IPS Installation & operation manual

Xylem e-1532 Series instruction manual

Xylem

Xylem e-1532 Series instruction manual

Milton Roy PRIMEROYAL instruction manual

Milton Roy

Milton Roy PRIMEROYAL instruction manual

STA-RITE ST33APP owner's manual

STA-RITE

STA-RITE ST33APP owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
Service Manual
(i
nfusion Pump & enteral feeding pump
)
V0.3
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
1
1. Abnormal After Turn on The Infusion Pump
1.1 The alarm light on and indicating use battery” after turn on the machine:
Possible Cause 1: C power cable is damaged or not plugged in.
Solution 1: plug in the power cable with proper force.
Solution 2: replace power cable. (as per diagram 1)
Diagram 1
Possible Cause 2: The connection between unit board and driver board is poor
contact / loosing/ damaged.
Solution: Open the front and rear shell of machine; check the 5-pin connecting
wire between power board and driver board. Reconnect the wire or replace it. ( s
shown in Diagram 2 and 3).
.
Diagram 2
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
2
Diagram 3
Possible Cause 3: The driver board fuse was burnt.
Solution: Use the multimeter to check if the fuse is connected; if no, use the same
specification of fuse ( 2 63V) for replacement. Location of fuse as shown in
Diagram 4.
Diagram 4
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
3
1.2 AC/DC indicator light, LCD screen and machine top indicator light are not on,
and there is no boot tone after turn on the machine.
Possible Cause 1: The connecting wire between main board and driver board is
loosed or damaged.
Solution 1: Open back shell of machine, reinsert the connecting wire between main
board and driver board and check if machine returns to normal. (as shown in
Diagram 5)
Diagram 5
Solution 2: Replace the connecting wire between main board and driver board to
check if machine returns to normal.
Diagram 6
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
4
Possible Cause 2: The 5-pin connecting wire between power board and driver
board is loosing or damaged.
Solution: Please check the 5-pin connecting wire, reinserting or replacing it.
Possible Cause 3: certain part of driver board is damaged.
Solution: Replace the driver board to see if machine returns to normal ( as shown
in Diagram 8)
Diagram 7
Diagram 8
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
5
1.3 AC/DC indicator light is on, but there is one or two abnormal between LCD
screen, top indicator light, boot tone after turn on the machine.
Possible Cause: Corresponding part of main board is damaged.
Solution: Replace the main board and check if machine returns to normal.
Diagram 9 ( main board)
Diagram 10
1.4Error code 0XE0” after turn on the machine.
Possible Cause 1: The connecting wire between main board and driver board is
poor contact or damaged.
Solution 1: Open back shell of machine, reinsert connecting wire between main
board and driver board and check if machine returns to normal.
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
6
Solution 2: Replace the connecting wire between main board and driver board
and check if machine returns to normal.
Possible Cause 2: driver board is damaged.
Solution: Replace the driver board.
1.5Error code 0XE3” after turn on the machine.
Possible Cause 1: Memory board encounters interference that cause incorrect
writing.
Solution: Recover factory default setting as per instructions in user manual.
Note: If the infusion set being used is not supplied by IV pump manufacturer,
please recalibrate its accuracy and pressure value after recover to factory default
setting.
Possible Cause 2: Main board is damaged.
Solution: Replace the main board.
2. Abnormal During Operation
2.1 Green indicator light sometimes is on and sometimes is off; or error alarm for
air bubble even correct installation of IV set.
Possible Cause 1: The IV set is too thin or too soft (the normal accepted diameter
of IV set is 3.5---4.5mm)
Possible Cause 2: ir bubble sensor is damaged.
Solution: Replace air bubble sensor.
1). unplug the red black wire and white yellow wire. Of air bubble sensor.
2). Remove the screw on air bubble sensor
3). Open pump door
4). Use tweezers from the position shown in Diagram11 to push off the sensor.
5). Install new air bubble sensor. Insert two cable as shown in Diagram 11.
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
7
Diagram 11
Possible Cause 3) air bubble detection circuit on the motherboard damage
Maintenance Step 1> contact the supplier .
2.2 Error code 0XE1” or OXE2”during operation.
OEX1 / OEX2 alarm means the wire connecting Hall sensor and pump body is loosing.
We need to replace and re-install Hall connecting wire.
Solution:
Step 1: Open the pump shell. Take out the old Hall connecting wire.
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
8
Step 2: Insert the new Hall connecting wire
Shenzhen Hawk Optical Electronic Instrument Co.,Ltd.
9
Step 3: Use solid sols to fix the connecting wire. The procedure to replace the Hall
connecting wire is finished.
2.3 Always indicating low battery.
Possible Cause 1: The battery is aging and can not recharge.
Solution: replace the battery
1). Remove the screw as shown in diagram 13, take off batter cover
2). Install new battery
3) Pull out the old battery, install a new battery with tweezers (as shown in Diagram
14)
4). Install battery cover, tighten the screws
Diagram 13