Hitachi L51K User manual

i
MN1-6268 Rev.2
L51K Probe
Instruction Manual
Specication
MN1-6268 Rev. 2
Notes for operators and responsible maintenance personnel
★
Please read through this Instruction Manual as well as the separate Instruction
Manual “Safety (MN1-5984)” and “Cleaning, Disinfection and Sterilization
(MN1-6000)“ carefully prior to use.
★
Keep this Instruction Manual together with the ultrasound diagnostic instru-
ment for any future reference.
© Hitachi, Ltd. 2016, 2017. All rights reserved.

ii
MN1-6268 Rev.2

iii
MN1-6268 Rev.2
Introduction
This is the instruction manual for L51K probe. The probe is available by connecting to Hitachi’s ultrasound diagnostic
instrument. It is operated by grasping with a forceps and can be mainly used for observation of human internal organs
during surgery.
Prior to use, read this manual as well as the separate instruction manual “Safety” in which information for safe use is
provided.
The probe bears the CE mark but the mark is valid only when the probe is connected to the ultrasound diagnostic
instrument bearing the CE mark.
Symbols used in this document
Safety information is classied into Danger ,Warning Caution, and Note according to the level of hazard. Those terms
are used in the safety information provided to prevent hazards and injuries to the operator or patients.
Danger
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury to the
operator or patient.
Warning
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury to the
operator or patient.
Caution
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to
the operator or patient, or property damage only.
Note
Indicates a strong request concerning an item that must be observed in order to prevent damage or
deterioration of the equipment and also to ensure that it is used efciently.
The type of safety information is indicated by the symbols below.
This symbol means that attention is required.
This symbol means that the described action is prohibited.
This symbol means that the described action is mandatory.

iv
MN1-6268 Rev.2
This instruction manual contains 4 pages of front matter and 12 pages of the main content.
CONTENTS
1. General Information...................................................................................................................................1
1-1. Intended use...................................................................................................................................................................1
1-2. Classication of ME equipment....................................................................................................................................1
1-3. Standard components ....................................................................................................................................................1
1-4. Options ..........................................................................................................................................................................2
2. Specications and Parts name....................................................................................................................3
2-1. Specications.................................................................................................................................................................3
2-2. Name of each parts........................................................................................................................................................4
2-3. Forceps...........................................................................................................................................................................5
3. Preparations before use .............................................................................................................................7
3-1. Visual check ..................................................................................................................................................................7
3-2. Conrmation of cleaning and sterilization...................................................................................................................7
3-3. Operation check.............................................................................................................................................................7
3-4. Probe insertion check. ..................................................................................................................................................7
4. Operation....................................................................................................................................................9
4-1. Insertion of the probe ....................................................................................................................................................9
4-2. Operation.......................................................................................................................................................................9
4-3. Pulling out the probe ...................................................................................................................................................10

-1-
MN1-6268 Rev.2
1. General Information
General information for the probe is provided below.
1-1. Intended use
This probe is intended for use by trained personnel (doctor, sonographer,etc.) for the diagnostic ultrasound evaluation
during robotic/non-robotic intra-operative and laparoscopic procedures.
Please refer to the ultrasound diagnostic instrument instruction manual used with this probe for the probe intended use
information.
Regarding with the connectable instrument, please refer to section 2-1. Specications of this manual.
Warning
Do not use this equipment for other than its intended use.
Otherwise it could cause burns or other injuries to the patient or operator.
1-2. Classication of ME equipment
This probe is classied as follows according to IEC60601-1.
Please refer to the section 2-1 for the range of applied part, the part treated as applied part, and the range of IPX7.
• Classication based on the degree of protection against electric shock ........ Type BF applied part
• Classication for protection against ingress of liquids .................................. IPX7 (Watertight equipment)
• Operation mode............................................................................................... Continuous operation
• Method of sterilization.................................................................................... Refer to the separate instruction manual
“Cleaning, Disinfection and Sterilization”
1-3. Standard components
The standard components of L51K probe are as follows.
L51K Probe ··························································· 1 set
Storage case ··························································· 1 set
Instruction Manual
• Specication (MN1-6268) ······································· 1 copy
• Safety (MN1-5984) ··············································· 1 copy
• Cleaning, Disinfection and Sterilization (MN1-6000) ······ 1 copy

-2-
MN1-6268 Rev.2
1-4. Options
The following options are available for this probe.
• Reprocessing by liquid detergent, disinfectant or sterilant
Whole the probe is able to immerge into the liquids by putting the connector of the ultrasound probe to the
waterproof case WP-001 as below table 1.
Precautions about the waterproof case, please refer to the instruction manual.
Table 1 Accessory for reprocessing by liquid detergent, disinfectant or sterilant
Product Name Product No.
Waterproof case WP-001

-3-
MN1-6268 Rev.2
2. Specications and Parts name
The specications and the name of each part are provided below.
2-1. Specications
Application: Intraoperative diagnosis
Type of patient contact: Intraoperative
Connectable instruments: ARIETTA Precision
NOTE:
At the time of publication of this manual, the connectable diagnostic ultrasound
instrument or instrument software version available with this probe is different for each
country, please refer to the instrument instruction manual or contact your local Hitachi
representative.
Field of view: 13mm
Frequency: 8.5 MHz
Cable length: 3.0 m
Service life: 3 years
Applied part: Probe tip including ultrasonic radiation part, see the section 2-2
Part treated as applied part: Cable up to 0.2 m length from the probe tip
IPX7 range: See Figure 1 (In case that not putting the waterproof case to the ultrasound probe
connector)
In case that putting the waterproof case to the ultrasound probe connector, whole the
probe from the tip of the ultrasound probe to the connector with Waterproof Case WP-001
is IPX7. range
Measurement accuracy: Refer to the instruction manual of the ultrasound diagnostic instrument
External dimensions: See Figure 1
Cable Length
8.5
Applied part
25
9.4
IPX7 range
1000
Remark: Unless otherwise specied in the gure, the tolerance for the dimension is ±10%.
Figure 1 External View
Unit: mm
9.4 ± 5%

-4-
MN1-6268 Rev.2
2-2. Name of each parts
The name of each part is shown in Figure 2 and the explanation for each part is listed in Table 2.
Ultrasound
radiation part
Ultrasound
radiation part
FinFin
Probe tipProbe tip
ConnectorConnector
CableCable
Front markFront mark
Front markFront mark
Figure 2 Name of each parts
Table 2 Name of each part and its explanation
Name Explanation
Ultrasonic radiation part Ultrasound is radiated from this part. The electronic
linear transducer is integrated underneath this part.
Front mark The cable side corresponds to the side of the orientation
mark on the image.
Fin This part is gripped with a forceps during operation.
Cable Cable transfers electric input/output signals.
Connector The connector is connected to the ultrasound diagnostic
instrument.
Caution
Do not pull, bend, twist, or apply excessive force to the cable.
The probe may malfunction due to cable disconnection.
Do not subject the ultrasonic radiation part to hard impact.
The impact may cause damage to the transducer, and that results in noise or no echo in the image.
In most cases, the ultrasonic radiation part itself is not damaged because the part is made of elastic
material.

-5-
MN1-6268 Rev.2
2.3. Forceps
Forcepses listed in the table 3 are recommended for using with the probe.
If you are unable to purchase the forceps, please contact our sales ofce.
Table 3 Forceps
Product No. Manufacturer
AdTec® mini PO305R B.Braun Aesculap®
ProGrasp™ Forceps 420093 Intuitive Surgical, Inc.
Caution
Regarding the precautions of the forceps, please refer to its instruction manual.

-6-
MN1-6268 Rev.2

-7-
MN1-6268 Rev.2
3. Preparations before use
This chapter describes preparations needed to use the probe safely. Please prepare the probe prior to each use by
following the instructions below.
3-1. Visual check
Visually check the probe tip, ultrasonic radiation part, cable, and connector.
If any holes, indentations, abrasion, cracks, deformation, looseness, discoloration, or other abnormalities are found, do
not use the probe.
3-2. Conrmation of cleaning and sterilization
Conrm that the probe is certainly cleaned and sterilized. The degree of reprocessing depends on the intended use.
Please refer to the separate instruction manual “Cleaning, Disinfection and Sterilization“ for cleaning and sterilization
procedure.
3-3. Operation check
Connect the probe to the ultrasound diagnostic instrument and check that the displayed scan type and frequency
correspond to those of the probe. Check also that there is no abnormality in the image.
3-4. Probe insertion check
Insert the probe into the trocar and make sure the probe can be smoothly inserted/removed.
Remark: Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to connect the
probe and information displayed on the monitor.
If the probe is operated in still air, brightness on the top of the image may be non uniform, but this does not
affect the performance of the probe.
Warning
Make preparations prior to each use.
The operator and the patient may be injured if the equipment has any abnormality.
If any abnormality is found in the equipment, stop using it and contact our ofce written on the back
cover.

-8-
MN1-6268 Rev.2
Caution
Do not use the probe if the displayed scan type and frequency do not correspond to those of the probe.
Incorrect acoustic output can result in burns or other injuries to the patient. Contact our ofce written
on the back cover.

-9-
MN1-6268 Rev.2
4. Operation
This chapter describes the operation of the probe.
4-1. Insertion of the probe
Gently insert the tip of the probe into the insertion opening of the trocar.
4-2. Operation
Grasp the n with a forceps and place the ultrasonic radiation part of the probe onto the inner organ surface during
surgery. An image of the region of interest is displayed on the monitor of the ultrasound diagnostic instrument. For
details on displaying and adjusting the image, refer to the documentation supplied with the ultrasound diagnostic
instrument.
Warning
Use an appropriate size trocar.
If an appropriate size trocar is not used, the probe may be damaged when inserted into the trocar. Use
a trocar with a diameter of 10 mm or more. Please refer to 3-4 for preparations before use.
Do not insert and operate the probe forcibly, as this may cause injury to a patient.
Insert the probe without excessive force on the probe and the cable. If the probe cannot be inserted
smoothly, stop using the probe.
Do not grasp the cable with forceps.
The probe may malfunction due to cable disconnection. Please operate by grasping the n with forceps
always.

-10-
MN1-6268 Rev.2
Caution
Do not operate the probe with excessive force.
Use with excessive force could result in injury to the patient.
Scan for minimum time necessary at the lowest possible acoustic output.
Acoustic output may affect the patient’s internal tissues.
For details about the acoustic output, please refer to the documentation supplied with the ultrasound
diagnostic instrument.
Do not touch the connector terminal pin of the probe.
Electrostatic discharge may result in malfunction of the probe.
Do not touch the probe connector of the ultrasound diagnostic instrument and the patient at the same
time. It can cause electric shock to the patient.
When it is used in conjunction with a high-frequency surgical instrument, do not place the probe and
operator’s hands in the path of high-frequency current.
The probe could be damaged, and the patient, examiner or operator could receive burns.Operate
this instrument with caution, paying attention to the positions of the counter electrode plates and the
connecting cord relative to the probe.
4-3. Pulling out the probe
Gently pull out the probe from the trocar.
Warning
When removing the probe, pull out the cable slowly and carefully without bending or twisting.
Do not remove the probe forcibly, as this may cause injury to a patient.
If the probe is not removed smoothly, operate it carefully and conrm endoscopically that there is no
abnormality.
Clean and sterilize the probe if blood is adhered to the probe, otherwise there is a risk of infection

-11-
MN1-6268 Rev.2
This page intentionally left blank

-12-
MN1-6268 Rev.2
US Importer: Hitachi Healthcare Americas Corporation
Address: 1959 Summit Commerce Park, Twinsburg, Ohio 44087
Distributor
MN1-6268 Rev. 2 ’17.08.31
■Manufacturer
Hitachi, Ltd.
2-16-1, Higashi-Ueno, Taito-ku, Tokyo, 110-0015, Japan
■Contact
+81-3-6284-3668
http://www.hitachi.com/businesses/healthcare/index.html
Overseas Offices:
Hitachi Medical Systems GmbH
Otto-von-Guericke-Ring 3 D-65205 Wiesbaden, Germany
EU Importer: Hitachi Medical Systems Europe Holding AG
Address: Sumpfstrasse 13 CH-6300 Zug, Switzerland
Table of contents
Other Hitachi Measuring Instrument manuals

Hitachi
Hitachi EUP-C514 User manual

Hitachi
Hitachi UST-52105 User manual

Hitachi
Hitachi UST-676P User manual

Hitachi
Hitachi U-5100 User manual

Hitachi
Hitachi C22K User manual

Hitachi
Hitachi C22T User manual

Hitachi
Hitachi RES670 Quick start guide

Hitachi
Hitachi UST-9133 User manual

Hitachi
Hitachi U-2810 User manual

Hitachi
Hitachi 902 User manual

Hitachi
Hitachi ZA3000 Series User manual

Hitachi
Hitachi LAB-X5000 User manual

Hitachi
Hitachi VC34 User manual

Hitachi
Hitachi L53K User manual

Hitachi
Hitachi Relion 670 Series User manual

Hitachi
Hitachi U-2910 User manual

Hitachi
Hitachi HT7700 User manual

Hitachi
Hitachi SML44 User manual

Hitachi
Hitachi EUP-L75 User manual

Hitachi
Hitachi Hi8 VM-H70E User manual