Hitachi HF2000 TEM User manual

1
HITACHI HF2000 TEM
Operation Manual
- Alignment and Operation -
Dear users,
To help you have a smooth and successful TEM observation, I wrote up this basic operation manual.
Systematic theoretical training plus hands-on practical training may be the best way to understand the
details of this operation instruction for HF2000 TEM.
There are some notes in this instruction. Please read and follow them carefully, and please make sure you
follow what I told you ask before you touch the TEM if you do not know the consequence of your action.
Takamichi MIYAZAKI / June 2009
Contents
1. Startup
2. Set the sample (loading)
3. Applying High Voltage
4. Prepare for beam alignment
5. Correct condenser astigmatism
6. Z-height adjustment
7. Voltage center
8. Image focus and Objective lens astigmatism correction
9. Selected area electron diffraction
10. Bright field imaging
11. Photographing
12. CCD camera
13. Sample tilt
14. Astigmatism correction of beam tilt
15. High resolution imaging
16. Microbeam diffraction
17. Sample exchange (unloading)
18. Shutdown
19. Film exchange

2
1. TEM startup
•Check the comments in the TEM log book. Enter your name,
laboratory and contact information on the log book before any
action on the TEM.
•Make sure that TEM vacuum is good. At the vacuum state
(right panel), SPEC, COL, and CAM indicators should be
VAC position (all green), and GUN indicator is at EVAC
position (green flashing).
•Make sure that TEM is set at standby positions:
a) OBJ. TEMP. MONITOR: 210C and below
b) Objective, selected area and condenser apertures are out (position 0).
c) GV (Gun Valve) is Closed
d) ACC. VOLTAGE: off; HV: on at 200kV
e) Sample position: X = 0, Y = 0; Sample tilts: 0
f) Bright/Dark field selector set at Bright field.
g) Microscope is at [ANA1] mode.
h) Magnification: 1000K – 1500K
i) Normally, the sample holder stays TEM. It is situated at position A
j) Lens Current (C1, C2, OBJ) : not 0
k) Unexposed film number: 50 FULL
Film number: 1 0
•After confirmation of the standby positions, enter chiller temperature and pressure, COMPR pressure,
HV tank pressure, and vacuum state on the log book.
Position A
EVAC
Selected Area
diffraction aperture
Objective aperture
Correct condenser
a
p
erture
X-ray detector
a
p
erture
Right main panel
Vacuum state

3
2. Set the sample (loading)
Normally, the sample holder stays in TEM. It is situated at position A (see in the following diagram).
Note: Objective aperture must be out during sample holder loading / unloading.
•Turn the switch on goniometer to Air. Wait till SPEC is at Air position (red light on vacuum state) and
pull the holder out.
•The sample is securely seated in the holder
•Loading sample holder into TEM
Follow the route indicated in the diagram below to load
your sample holder into the TEM.
1) Insert the sample holder horizontally along and
press it, then turn the switch to EVAC.
2) When electron sound rings, turn the holder clockwise till it stops.
3) Let the sample holder go into the TEM horizontally along and smoothly. The sample
holder must be under control.
4) Turn the sample holder counter-clockwise till it stops
5) Let the holder go into the TEM fully.
B位置
C位置
A位置
(EVAC⇔
A
IR)
B
位置
A位置
B位
置
Bposition
Aposition
Cposition
Aposition
Bposition
EVAC / AIR

4
3. Applying HV
•Apply HV when vacuum is good
•Make sure that the applying HV conditions are set at following state:
V0 VOLTAGE 200kV
RRATIO 5.5
I1 CURRENT 30µA
•Press the [FE] bottom once (left panel) and press [FLASH] when it’s flashing. IF current rise up, if the
IF current is lower than 0.30 mA, repeatedly press [FLASH] bottom till IF current is > 0.30mA.
•After IF current exceeds 0.30mA, wait till V0 VOLTAGE rises up to 200kV. Then V1 and V2 Voltage
rises gradually up to the setting value (watch the ACC VOLT on the CRT monitor).
•Enter the V1 Voltage and IF current on the log book (Yellow book).
********************************************************************************
When the emission current drops below 10µA, press [I1C] and recalibrate extraction voltage to raise the
emission current to 30µA. The beam wander off during the process but return by the end of the I1C state.
Repeat alignment procedure because the changed extraction voltage affects the caustic.
********************************************************************************
4. Prepare for beam alignment
•Press [ANA1] to change selected mode: Analysis →TEM mode.
•Reset switch (Red square button)
[I1C]
[FLASH]
Left main panel

5
5. Correct condenser astigmatism(Magnification : 100K ~ 200K)
Be sure to use the correct condenser astigmatism for TEM mode:
•Open the GV and make sure you hear it click open.
•Turn the Brightness control (left panel) to crossover, and center the beam with the Brightness
Centering knobs (right panel).
•Using the condenser stigmator (right sub panel) make the caustic as round or axial symmetry. And
using the 3rd order stigmator (left sub panel) makes the caustic as symmetry. The 3rd order correction
will be slight and the controls should be close to the center positions.
•The correctly aligned caustic will appear as a circle with Star in TEM mode. The image is shown
below.
•Insert condenser aperture (First aperture) and
minimize the swinging beam using the condenser
X/Y adjustment screws when passing through
crossover.
3rd Cond Stig
2nd Cond Stig
3rd Cond stigmator
2nd Cond stigmator

6
6. Z-height adjustment
•Move edge of sample and set the objective lens current to 5.72 using focus control. Focus the sample
using the Z-height adjustment screw on the specimen holder entry port.
Note: An objective lens current of 5.72 is the optimal setting for minimum spherical aberration.
You may also focus sample using wobblers method or minimum contrast method for the image
focus.
7. Voltage center
•Move edge of sample.
•Turn on [HVM] (HV MODULATION, right panel). The image feature may oscillate.
•Adjust BEAM TILT (right sub panel) to minimize the vibration of the feature.
•Turn off [HVM] (right panel).
•Repeat the alignment procedure a few times because the beam tilt and objective stigmatism correction
have disturbed the condenser stigmatism corrections. In general, completing the alignment procedure
twice will be sufficient.
Z-height adjustment

7
左サブパネル
左サブパネル
8. Astigmatism correction of objective lens
•Reset the OBJ STIGM-X and Y control knobs (stigmator)
on the left sub panel.
*This returns setting to the stigmator to preset level.
•Set a magnification 2 or 3 times higher than used for
photographing.
•Turn the focus control knob clockwise so as to obtain
slightly over-focused image, and a fringe appears outside
the collodion hole.
•Adjust the OBJ STIGM-X and Y control knobs so that the width of the fringe outside the collodion
hole becomes uniform.
•Manipulate the focus control knob so as to approach just focus from over-focus in order to narrow the
fringe width.
•Find a as granularity a amorphous as possible and adjust the OBJ STIGM-X and Y control knobs so
that granularity a amorphous becomes finer granularity as shown in the figure below.
9. Image focus and Objective lens astigmatism correction
(Including Selected area electron diffraction (SAED))
This is to tell you how to correct objective lens astigmatism using amorphous film which is useful in
recording a high resolution image.
•Center a small hole / particle / the sharp edge of sample.
•Insert and center a suitable Selected Area (SA) aperture.
•Press [DIFF] (left panel) and set the camera length to 0.80m using Magnification knob. If the central
spot (transmitted wave) is still disk, focus the pattern by adjusting Diffraction Spot (left panel).
Note: Open the beam to form a parallel illumination on the sample by turning Brightness
clockwise. Diffraction spots will become weaker and smaller. If the central spot is disk instead of
a spot, adjusting Diffraction Spot is needed.
Left sub panel

8
•Insert objective aperture into the central spot that should be centered by using the objective X/Y
adjustment screws.
•Press [ZOOM] to seen a bright field image.
•Pull out SA aperture
•Correct the astigmatism by adjusting fine focus and OBJ STIG X and Y controls if necessary.
(Using minimum contrast method and amorphous contrast)
10. Bright fielding imaging
•Find an interesting sample area.
•Focus the beam and sample to the crossover.
•Press [DIFF]. (see Sec.9, SAED)
Note: If the central spot is not at the viewing center, center it using the INTER ALIGN X and Y knobs
(left sub panel). If the spot is not round, make it round using INTER STIG X and Y knobs (left sub
panel).
•Insert the objective aperture into the beam by turning the aperture handle clockwise.
•Press [ZOOM] to see a bright field image.
•Adjust fine focus and OBJ STIG if necessary.
•Adjust Brightness so that Auto exposure time is between 2-5 seconds.
•Cover the viewing window
•Press [Photo] to expose the film and record a bright field image.
電子回折図形
明視野像
格子(多波干渉)像
Bright fielding imaging
Electron diffraction pattern
Electron diffraction pattern
Lattice imaging

9
11. Photographing for film
Usage of auto exposure time set mode
Time mode provides a function for automatically setting the exposure time so as to obtain optimum
photographic density on a film in response to a determined brightness.
•Set the exposure time setting knob at AUTO.
•Press the PHOTO key to start film exposure.
Manual mode
•Set a desired exposure time by turning the EXPOSURE TIME control knob. The set exposure time is
displayed at the bottom right on the CRT.
•Optimum exposure time lines for
Bright field image: 1 to 4 seconds
Dark field image: 12 to 30 seconds
Selected area diffraction pattern: 0.5 to 1 seconds
Nano beam diffraction pattern: 8 to 16 seconds
•Press the PHOTO key to start recording
12. CCD camera
•Turn on the power to TV monitor and control.
•Open the beam by turning Brightness clockwise. Brightness will become weaker.
Note: Do not use CCD camera for diffraction pattern, low-magnification image and direct beam.
•Cover the viewing window.
•Press [STOP] and [PHOTO].

10
13. Sample tilt
•Press [DIFF]. (see Sec.9, SAED)
•Adjust the sample tilt with electron beam direction along the zone axis.
14. Astigmatism correction of beam tilt
•Turn the WOBBLER ANGELE control on the right main panel fully clockwise.
•Press the WOBBLER ON/OFF switch.
•When astigmatism is not corrected, rotating motion in a specific direction is observed. Adjust the
[BTX] and [BTVX] controls on the left sub panel so that non-directional image rotation is observed as
shown below.
•Turn the WOBBLER ANGELE control counterclockwise by 3 steps. When astigmatism is not
corrected, adjust the [BTY] and [BTVY] controls on the left sub panel so that non-directional image
rotation is observed as shown below.
FOOT
RIGHT
FOOT
LEFT
HAND
LEFT
HAND
RIGHT
FOOT
RIGHT
HAND
LEFT

11
15. High resolution imaging
•Find an interesting sample area.
•Focus the beam and sample to the crossover.
•Press [DIFF] (see Sec.9, SAED)
•Set the sample tilt along zone-axis by seeing SAED.
•Insert the largest objective aperture into the beam by turning the aperture handle clockwise.
•Press [ZOOM] to see a high resolution image.
•Turn Magnification up to 300K – 700K.
•Adjust fine focus and OBJ STIG if necessary. (see Sec.8)
•Adjust Brightness so that Auto exposure time is between 2-5 seconds.
•Cover the viewing window.
•Press [Photo] to expose the film and record a bright field image.
16. Microbeam diffraction
•Insert condenser aperture (Second aperture) and minimize the swinging beam using the condenser X/Y
adjustment screws when passing through crossover.
•Sample tilt (see Sec.9, SAED)
•Change the condenser aperture to forth aperture. And minimize the swinging beam using the
condenser X/Y adjustment screws when passing through crossover.
•Adjust the objective focus, and confirm to select an interesting area.
•Press [DIFF]
17. Exchange the sample (unloading)
•Make sure that GV should be closed. Please remember that the unloading without closing GV cause a
lot of problems, FE-tip’s fatal damage in a worst-case scenario.
•Unloading sample holder from TEM
Follow step 5 to take it out.
2) Close GV and pull the sample holder till it stops.
3) Turn the sample holder clockwise till it stops.
4) Pull the sample holder till it stops.

12
5) Turn the sample holder counterclockwise till it stops.
6) Turn the switch on goniometer on Air. Wait till SPEC is at Air position (red light on
vacuum state) and pull the sample holder out.
18. TEM shutdown and film exchange (Shutdown procedure)
•Close GV
•Press [FE] twice to shutdown high voltage
•Set Objective, selected area and condenser apertures out (position 0).
•Press [ANA1], and turn Magnification up to 1000K-1500K to keep lenses warm and stable, and
prevent any water vapor condensation.
•Set sample shift: X = Y =0, sample tilts = 0
•Take the sample out of the sample holder (see II, III)
Note: Carefully unload the sample holder and set the holder A position after removing sample.
19. Film exchange
•Wear gloves.
•Turn off all lights (room light, CRT monitor, panel light, CCD monitor) except the safelight.
•Turn the switch of the desiccators down to Air position.
•After a few minutes, open the desiccators door, take out full film magazine and prepare empty
receiver.
•Open the cover of the camera chamber and turn the CAMERA EVAC switch to AIR.
Note: Air is leaked to the camera chamber. Wait for a few minutes until air leak complete. Upon
completion of air leak, the AIR (red) lamp of VACUUM STATE is lit (right panel) and the buzzer
sounds.
•Open the front door of the camera chamber.
•Take out the receiver. And take out the film magazine from inside the receiver. Pull out the guide rail
Position A
(EVAC →AIR)

13
while raising the stopper, and the film magazine will come out.
•Load a new full film magazine into the camera chamber.
•Load an empty receiver.
•Close the front door of the camera chamber.
•Turn the CAMERA EVAC switch to EVAC. Usually after about 10 minutes. EVAC (green) lamp of
VACCUM STATE lights up.
•Loading of new film
1) Take out exposed film and the light shielding plate
2) Place film in the magazine.
3) After loading the films, don’t forget to insert the light shielding plate
•Open the desiccator’s door and introduce the full film magazine.
•Turn the knob to EVAC.
•Reset film number to 50 on the CRT using the keypad. Listen for the shielding plate to properly feed
through into the film receiver.
•Change the FILM NUMBER to 0 using the keypad.
20. Final check procedure
•Turn off all lights (room light, CRT monitor, panel light).
•Complete the log book.
•Fill final check list.

14
EMSA形式で保存
ファイル名を入力
HITACHI HF2000 TEM
Operation Manual
- EDX analysis in zoom mode -
1. Start up
•Adjust the beam alignment in [ZOOM] mode.
•Insert the X-ray detector.
•Insert condenser aperture (2nd aperture) and minimize the swinging beam using the condenser X/Y
adjustment screws when passing through crossover.
•Remove the any objective aperture and turn off the column pressure gauge filament by pressing the
“FILAMENT” switch.
Notice: not use any objective apertures when the detector is in! The objective aperture will generate
hard x-rays and burn out the detector.
2. The EDS procedure
•Make sure to turn off the column pressure gauge and insert condenser aperture (2nd aperture).
•Turn on the power switch of the EDS workstation.
-Screen-
•Click “Spectral Display” icon.
•Select “Acquire” →“Parameter”
→“Set up”. Set the max. energy to 20.48 keV.
•Find region of interest and focus it properly with
brightness knob.
Notice: When the dead time is too high (over 30%),
select smaller condenser aperture, or
acquire spectrum from thinner region.
•Depress on the quick button.
•Depress , when the intensity reaches certain intensity.
•Enter the file name, save the spectrum as “EMSA”.
Enter the file name
Save as EMSA

15
3. Peak identification (Label peak)
•Open the “Analyze” menu and select “Manual Peak identification“
•Click the expand button (+) to display the set of markers for the selected element.
4. Quantification
•Open the “Analyze” menu and select “Quant Periodic Table”.
•Choose the element and lines for quantification, and set “Correction method” to “MBTS”.
K-ratio: Fit is performed and intensity ratios are reported.
ZAF: Adjusts the theoretical spectrum for average atomic number, absorption, and
fluorescence factors, which influence spectrum differences between elements in pure and
composition. Used in metallurgical SEM applications.
Proza: Same as above. Used on SEM applications, especially for light elements in heavy matrix.
Bence-Albee: Use in SEM applications, especially for mineral samples
MBTS: Correction assumes that there is no absorption of x-rays within the sample. Used in TEM
applications.
•Press “Quantify” button.
右クリックをすると
周期律表が出る
ピークにラベル付け
Clicking right lets you view a
list of periodic table and
element line. Select whether the
element is absent or present.
Label peak

16
5. Shut down
•Depress the “Retract” button to retract the X-ray detectors.
•Close the spectra display window.
•Shutdown the Windows software.
MBTSを選択
定量分析に適切な
ラインを選択
Select “MBTS”
Choose the elements and
lines for quantification

17
HITACHI HF2000 TEM
Operation Manual
- Alignment in Analysis mode
1. Start up
•Execute the beam alignments in zoom mode. (See the manual above)
•Place the beam through the hole in the sample and increase magnification to 150kX. Remove all
apertures, center the beam, and go to crossover.
•Depress [ANA-1] and reset.
2. Correct condenser astigmatism(Magnification : 100K ~ 200K)
•Using the condenser stigmators (right sub panel) make the caustic as round as possible and using the
3rd order stigmators (left sub panel) make the caustic as symmetrical as possible.
•The 3rd order correction will be slight and the controls should be close to the center positions. Be sure
to use the correct 3rd order stigmators for the selected mode: TEM and ANA mode.
•The correctly aligned caustic will appear as a circle with the Star of David in the center in TEM mode
and it has the appearance of the Mercedes-Benz star in Analysis mode.

18
3. Voltage center of condenser lens
Note the direction of the caustic movement when going to
crossover.
•Use the beam tilt controls (right sub panel) to shift the
caustic in the same direction of the movement and
brightness centering controls (right panel) to return the
caustic to the center of the phosphor screen.
•Correct both 1st and 3rd order condenser astigmatism.
Repeat step 3 until caustic does not move with
directionality when passing through crossover. This
correction requires significant beam tilting if the previous
operator used the TEM alignment procedure.
•Insert condenser aperture (3rd or 4th) and centering by minimizing the tails of intensity surrounding the
central bright spot using by the condenser X/Y adjustment screws.

19
HITACHI HF2000 TEM
Operation Manual
- STEM imaging and element mapping -
1. Start up
•Move area of interest to the crosshairs on the phosphor screen and focus.
•Adjust the z-height until the sample and correct objective astigmatism using the stigmators on the left
sub panel.
•Insert the X-ray detector.
•Insert condenser aperture (2nd aperture) and minimize the swinging beam using the condenser X/Y
adjustment screws when passing through crossover.
•Remove the any objective aperture and turn off the column pressure gauge filament by pressing the
“FILAMENT” switch.
2. Beam alignments in STEM mode (Magnification: 100K ~ 200K)
•Depress the [STEM] (left main panel).
•Insert the “D-STEM DET” on left side of column.
•Depress the [IMAGE SHIFT], centering the black circle on CRT monitor using image shift.
•Adjust the FOCUS, CONTRAST, and STIGMA.
3. The element mapping
•Make sure to turn off the column pressure gauge and insert condenser aperture (2nd aperture).
•Turn on the power switch of the EDS workstation.
-Screen-
•Click “Spectral Display” icon.

20
•Select “Acquire” →“Parameter” →“Set up”. Set the max. energy to 20.48 keV.
•Find region of interest.
Notice: When the dead time is too high (over 30%), select smaller condenser aperture, or acquire
spectrum from thinner region.
•Open the “Analyze” menu and select “Manual Region of Interest …“
•Choose the element and lines for element mapping
•Open the “Image Display” and select [Xray1: Map1] on left cell.
Table of contents
Other Hitachi Microscope manuals

Hitachi
Hitachi TM3030Plus Specification sheet

Hitachi
Hitachi TM3030 Plus User manual

Hitachi
Hitachi McDATA Sphereon 4500 User manual

Hitachi
Hitachi S-4800 User manual

Hitachi
Hitachi TM3030Plus Specification sheet

Hitachi
Hitachi HT7800 User manual

Hitachi
Hitachi S-4100 User manual

Hitachi
Hitachi TM4000 Specification sheet

Hitachi
Hitachi 3400-N Guide

Hitachi
Hitachi S-4800 User manual