Westlab M-100F User manual

SILICONE
UNI-STOPPERS
CODE 663-393 TO 663-395
INSTRUCTION
MANUAL
Before Use 1
1.Nomenclature 2
2.Operation 3
2-1 Set the Specimen Slide 3
2-2 Set Illumination 3
2-3 Adjust Focus 3
2-4 Adjust Diaphragm 3
2-5 Choose the Objective 3
2-6 Change the Lamp 3
3.General Specification 4
4.Parameter/Technical Terms 5
5.Troubleshooting 7
*Please read these instructions
carefully before use.
.
WESTLAB
EXTREME MONOCULAR
MICROSCOPE,
RECHARGEABLE, LED
USER MANUAL
CODE 664-089C

1. OPERATION
(1.) As the microscope is a precision instrument, always handle
it with care, avoiding impact or abrupt movement during
transportation.
2. MAINTENANCE
(1.) All glass surfaces must always be kept clean. Fine dust on
the optical surface should be blown off by means of a hand
blower or gently wiped off with a soft lens tissue. Carefully
wipe off oil or fingerprints on the lens surfaces with tissue
moistened with a small amount of 3:7 mixture of alcohol
and ether.
(2.) Do not use organic solution to wipe the surface of the other
components. These parts, especially the plastic parts,
should be cleaned with a neutral detergent.
(3.) Do not take down or assemble it yourself.
(4.) After use, cover the microscope with the dustcover provided,
and keep it in a dry and clean place for preventing rust.
(2.) Do not let the microscope BE in the sunlight directly. Keep
it in a dry and clean place. Avoid high temperature and acute
shaking. Following environment is required: Indoor
temperature: 0°C~40°C, Max relative humidity: 85%.
(3.) Avoid impacting the definition of the image, do not leave
dust and fingerprints on the lens surfaces.
(4.) Before using, examine to ensure that the power supply voltage
is consistent with the rating voltage.
BEFORE USE
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
1

WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
2
1. NOMENCLATURE
Iris

The microscope is housed in a moulded styrofoam container.
First take the container out of the carton, and lay the container
on its side. Open the container carefully and don’t let the optical
items drop down, avoiding them being damaged. Check carefully
to ensure the arm and accessories are intact.
Install the objective into the microscope nosepiece from the lowest
magnification to the highest, in a clockwise direction from the rear.
Insert the eyepiece into the eyepiece tube.
2-1 Set the specimen slide
Place the specimen to be studied on a glass slide, and fix it with
the slide-holder of the mechanical stage. If we use the movable
specimen holder, fix the specimen by slide-holder of the
movable specimen holder. Adjust its position by the switch of
the movable specimen holder.
2-2 Set illumination
2-3 Adjust focus
Adjust the coarse-focusing-knob to bring the slide into focus.
Then lock the limit-stop-screw to avoid impact between the
objective and slide. Adjust the fine-focusing-knob to get the
image sharper and clearer.
2-4 Adjust diaphragm
Turn the disc diaphragm to select a aperture to get the
background brightness suitable.
2-5 Choose the objective
Turn the nosepiece to choose the objective. Objective selected
should be set vertically right to the slide. Generally, first use the
objective 4X to focus to reveal general structural image. Then
use the high power objective to reveal smaller details.
2-6 Change the lamp
Before changing the lamp, first pull the plug out off the electrical
socket and wait for a while until the lamp cools down to avoid
being burnt. Then screw off the base lens in an anti-clockwise
direction and insert a new lamp and screw on the base lens.
For the microscope with built-in electrical illuminator, insert
the plug of the main cable into the power socket and turn on
the light switch to get the specimen illuminated.
2. OPERATION
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
3

Lamp
Illumination
Parts Specifications M-100F M-100FL
Methylene Chloride
WF10X/18mm
WF10X/18mm with pointer
WF10X/18mm with reticle
WF15X/13mm
WF20X/10mm
• •
Install
195 Objective
4X/0.10
10X/0.25
40XS/0.65
60XS/0.85
100XS/1.25(oil)
• •
• •
• •
Model
H5X/ H6X
H10X
H12.5X
H15X/ H16X
Huygens Eyepiece
Dual viewing head
Monocular head 360°rotating
Binocular head 360°rotating
Metal base and stand
Binocular head 360°rotating
Metal base and stand
110mmX120mm
Built-in movable stage
125mmX130mm
Triple nosepiece
Quadruple nosepiece
Single N.A.0.65
AbbeN.A.1.25
Blue/Yellow/Green
ф42mm mirror
Built-in illuminator
Five-aperture disc diaphragm
Iris diaphragm
Tungsten bulb 115V/20W
Halogen bulb 230V/20W
45°View Head
Stand
Polarized Set
Nosepiece
Stage
Movable White
Specimen Holder
Clips
Filter
Condenser
Diaphragm
• •
• •
• •
• •
• •
• •
• •
• •
•
•
•
4
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
BIOLOGICAL MICROSCOPE GENERAL
SPECIFICATIONS (FOR OPTIONS)

4. PARAMETER/TECHNICAL TERMS
4-1 OBJECTIVE
DIN
achromatic
objective
195mm
4x
10x
40x
60x
100x
0.10
0.25
0.65
0.85
1.25
45
45
45
45
45
Red
Yellow
Light blue
Deep blue
White
Air
Air
Air
Air
Cedar
oil
Type Magnification Medium
Numerical
aperture
(N.A)
Parfocal
distance
(mm)
Magnification
market
(color ring)
4-2 EYEPIECE
4-3 ELECTRIC CERTIFICATION
4-4 PARAMETER
(1) Total magnification:
20X~1600X
(2) Field of view:
ф0.08~ф4.5mm
(3) Mechanical tube length:
160mm
(4) Object to primary image distance:
195mm
100V~120V power supply : 100V~120V±10%, 50/60Hz
Lamp: 110V/20W tungsten lamp
These units have UL certification.
Magnification 10x 15x 20x 5x/6x 10x 12.5x 15x/16x
Field of view
(mm) ф18 ф13 ф11 ф15 ф12 ф10 ф8
Type Wide field Huygens
5
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C

4. PARAMETER/TECHNICAL TERMS
4-5 TECHNICAL TERMS
(1) Total magnification=(magnification of objective) X (magnification
of eyepiece)
(2) Field of view=(line field of view of the eyepiece selected)
÷(magnification of the objective selected)
(3) N.A.=n·sinα (max), N.A.is very important parameter which
marks the features of the objective and condenser. The “n”
is the refractive index of the medium (air or immersion oil)
between the cover glass of the objective and the specimen.
The “α” is the half of the aperture angle. The N.A.is bigger,
the resolution of the objective is higher.
(4) Object to primary image distance: the distance between the
object plane to the primary image plane. The conjugate
distance is fixed.
(5) Mechanical tube length: The distance between the objective
shoulder and the ocular shoulder.
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
6

5. TROUBLESHOOTING
Optics
(1)The side of
the field of
view is dark or
not even
The nosepiece is not in the right
position
Turn the nosepiece
into the right position
Clean the lens
Stain or dust has accumulated on the
condenser, objective, eyepieces,
base, lens
Symptom Cause
(2)Stain or
dust is
observed in
the field
of view
Stains have accumulated on the
specimen
Clean the specimen
Clean the lens
Stains have accumulated on the lens
(3)Unclear
image
No cover glass on the specimen slide Add the cover glass
The cover glass is not standard Use a standard cover
glass with thickness
0.17mm
Remedy
Make it face up
Clean thoroughly
Use immersion oil
Get rid of the air bubble
Use a correct one
Adjust the iris
diaphragm
Clean lens
Adjust the condenser
The specimen faces down
The immersion oil has accumulated
on the dry objective
The immersion oil is not used for oil
objective 100XR
Air bubble in the immersion
Use wrong immersion oil
The aperture is not opened to
correct size
Stains or dust has accumulated on
the lens in the inlet of the head
The condenser is not in the right
position
4)One side of
the field of view
is dark or the
image moves
while focusing
The specimen slide is not fixed Fix with clips
Turn the nosepiece
into the right position
The nosepiece is not in the right
position
(5)The field of
view is not
bright enough
The iris diaphragm is not bigger
enough
Adjust the iris
diaphragm
(6)The image
color is not true
No filter is used Use correct filter
Adjust the condenser
Clean lens
The condenser is not in the right
position
Stain or dust has accumulated on the
condenser, objective, eyepieces,
base lens
7
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C

8
5. TROUBLESHOOTING (CONT)
Optics
Electrics
(1) The image
is not focused
while using
high power
objective
The cover glass faces down Place cover glass
to face up
Use a standard cover
glass with thickness
of 0.17mm
The cover glass is not standard
Symptom Cause
(2) The
objective
touches the
cover glass
when turning
the nosepiece
The cover glass faces down Put the cover glass
to face up
Use a standard cover
glass with thickness
of 0.17mm
The cover glass is not standard
(3) Cannot
move the
slide smoothly
The slide is not fixed correctly Adjust it correctly
Remedy
Tighten it
The movable specimen holder is not
fixed properly
(2)The bulb
burnt out
usually
The voltage is too high Use correct power
supply
(3)The fuse
burnt out
usually
The bulb will burn out soon
The wire doesn’t connect all right
Replace with a new
one
Connect correctly
Replace with a
correct one
Use a wrong bulb
(1)The bulb
does not work
No power supply Check the connection
of the power cable
Tighten it
The bulb is not inserted correctly
Replace it
The bulb burnt out
Replace it
The fuse burnt out
order and delivery details at the end
of the day.
ONLINE
www.westlab.com
Order online with our easy Express
Order form for fast, secure ordering.
PHONE
1877 822 1455
Speak to one of our
friendly, efficient staff
from 8.00am to 5.00pm Monday to
Friday.
EMAIL
sales@westlab.com
24 hours, 7 days a week convenience.
FAX
604 538 5203
Order via fax and we will confirm your
ORDERING WITH
WESTLAB IS EASY...
WESTLAB
EXTREME
MONOCULAR
MICROSCOPE,
RECHARGEABLE,
LED
USER MANUAL
CODE 664-089C
This manual suits for next models