This hot plate includes an integrated magnetic stirrer, ideal for light duty use in a laboratory.
The dual controls allow you to set heat output and stirring speed independently. The hot
plate reaches a temperature of over 250°C (480°F) with an output of 180 watts. The
magnetic stirrer is adjustable from 0 to up to about 1400 rpm and suitable for mixing up to
500ml of solution in a container of 5” diameter or less.
Components
The following items are included:
•Base with power switch, two control knobs, and LED indicators
•Teflon-coated magnetic stir bar (1 inch length)
•Power cord
•Support stand (2-piece support rod, right angle clamp, and thermometer holder)
Operation
To set up the support stand:
Screw the two rods together, creating one support rod. Make sure they fit tightly. Insert the
slotted end of the rod into the plastic holder on the back of the base. Slip the right angle clamp onto the support rod and
tighten the knob. Slide in the thermometer holder and tighten the other knob. The height of the thermometer holder can be
adjusted by moving the clamp up or down on the support.
To use the hot plate:
1. Make sure the knobs are turned off. The white mark on each knob should be pointing down.
2. Connect the power cord to the back of the base. The base should be set on a hard surface, like a countertop or table.
3. Plug the other end of the power cord in to a standard outlet. Make sure the outlet is safe, and the power cord is not
twisted.
4. Place the beaker, flask or other vessel containing material to be heated on the hot plate.
WARNING: Do not operate the hot plate without a liquid filled vessel on it and do not let the vessel boil dry.
Use without a heat sink will cause the heating element to fail and void your warranty.
5. Flip the power switch on the right side of the base. The line (–) indicates on, and the circle (o) indicates off.
WARNING: Be careful with the hot plate turned on. Do NOT touch the heating surface, or other parts of the
unit except the control knobs, as it will result in a burn.
WARNING: Unit is not sealed. Avoid spilling liquid on the base as it will damage the unit. Avoid using the unit
near flammable vapors or to heat flammable liquid as it may result in fire or explosion. Disconnect from
power before replacing fuse or trouble-shooting.
6. Rotate the right knob in a clockwise direction. At maximum power the hot plate will take around 20-30 minutes to boil
a large beaker of water. The plate surface temperature can reach over 250°C when the knob is turned to maximum
power.
WARNING: Do not use the unit for more than 4 hours on maximum heat. Do not use the unit for more than 8
hours heating continuously. If the unit malfunctions, let it cool completely, then begin operation again.
7. Adjust the knob to keep the desired heat input, using a thermometer to check the temperature of the solution.
8. To turn off the hot plate, rotate the knob counter-clockwise until you hear a click and the white mark is pointing down.