Hontzsch UVATP User manual

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 1/17
Transducer UVATP
in AS102 Housing
Operating Instructions

__________________________________________________________________________________________________
_____________________________________________________________________________________
2/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
Contents
1
Safety Symbols
2
Proper Use in Compliance with Regulations
3
Operating Safety
4
Scope of Delivery
4.1
Description, Type Plates
5
Technical Specifications
5.1
Conformity with Standards
5.2
Storage Conditions
5.3
Operating Conditions
5.4
Housing and Connectors
5.5
Electrical Data
6
Installation
6.1
Block Diagram
6.2
Wiring Diagrams
6.2.1
Cable shielding connection in the cable glands
6.2.2
Power supply
6.2.3
Temperature sensor Pt100
6.2.4
Absolute pressure sensor 4...20 mA
6.2.5
Analog outputs A1 and A2
6.2.6
Relay output
6.2.7
VA sensor input (separate transducers only)
6.2.8
RS232 interface
6.2.9
LCD display
6.3
Tips on Electromagnetic Compatibility (EMC)
6.4
Connection Cable
7
Functional Description
8
Settings
9
Performance Check
10
Initial Operation
11
Operation
12
Shut-down, Dismantling
13
Inspection
14
Maintenance
15
Calibration
16
Troubleshooting
17
Returns
18
Disposal
19
Declaration of Conformity, Manufacturer’s Declaration

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 3/17
1 Safety symbols
Warning! Failure to observe the instructions can result in serious injury and damage
to property!
Important notice!
Non-observance can result in serious damage to the equipment or performance
restriction!
2 Proper Use according to Regulations
The UVATP transducer is for mass flow output by measuring the actual flow rate, the operating
temperature and the absolute pressure. It is designed for industrial application.
The AS102 is an aluminium housing in protection class IP65.
The manufacturer is not liable for damage caused by improper use and/or non-compliance with the
regulations.
Do not carry out any structural modifications to the transducers.
Always follow the instructions on the type plate, especially the information regarding supply voltage.
The UVATP can be integrated in the AS102 connection housing of measuring tubes VA TP DN ExactFlow II
or is available as a separate transducer in AS102 housing.
3 Operating Safety
All steps described below must be carried out by qualified personnel only!
Please read the Operating Instructions carefully before unpacking the equipment!
Safety can only be guaranteed if the equipment is operated in accordance with the regulations.
Improper handling can result in serious injury and damage to property.
4 Scope of Delivery
- UVATP integrated in AS102 housing of VA TP DN ExactFlow II measuring tube
or as separate transducer in AS102 housing
- LCD display 2x16 digit in the cover
- Operating Instructions, Data Sheet
- CD-ROM with PC configuration software UCOM (optional)
- Interface cable RS232 for PC COM port connection (optional)
- USB adapter in addition to the interface cable RS232 (optional)
Please check that everything listed in the Delivery Note / Technical Data Sheet is included in the delivery.

__________________________________________________________________________________________________
_____________________________________________________________________________________
4/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
4.1 Description, Type Plates
Remove the cover to display one of the following type plates:
S/N. : ... : serial number
Type : UVATP : for vortex VA measuring tubes, temperature, pressure
Power : 24 V DC : supply voltage 24 V direct voltage current
12 V DC : supply voltage 12 V direct voltage current
Output A1 : precision 4 ... 20 mA : analog output with high precision current 4 ... 20 mA
precision 0 ... 10 V : analog output with high precision voltage 0 ... 10V
Output A2 : fast 4 ... 20 mA : analog output with short time constant current 4 ... 20 mA
fast 0 ... 10 V : analog output with short time constant voltage 0 ... 10V
Input : VADi : for vortex VA measuring tubes (separate transducers only)
RS232 : RJ22 plug : for serial PC interface
5 Technical Specifications
5.1 Conformity with Standards
The UVATP is manufactured according to the best available technology, is both safe and reliable
and complies with the relevant regulations, EU directives and standards.
5.2 Storage Conditions
Storage temperature: -30 bis +70 °C
5.3 Operating Conditions
Ambient temperature during operation with LCD display: -20 to +50 °C
protection class: IP65
mounting attitude: no restrictions, preferably with cable bush facing down.
separate UVATP integrated UVATP

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 5/17
5.4 Housing and Connectors
Protection class : IP65
Material : aluminium
External dimensions : L/W/H = 150/100/80 mm
Cable entry point : nickel-plated brass screwed cable glands
for cable diameter 5 ... 10 mm
with cable shielding connections
Connections : 'push in' PCB terminals
for wire cross-section 0.14 ... 1.5 mm²
No tools necessary for core connection –simply insert strands
(twisted or with end sleeve) into the terminal.
To separate strands apply pressure to the terminal release spring using
a pen or screwdriver.
5.5 Electrical Data
Supply voltage:
24 V DC (20 ... 27 V DC), power < 5 W
alternatively:
12 V DC (10 ... 17 V DC), power < 5 W
The mains supply is electrically isolated from the UVATP inputs and outputs.
Input flow v/VA: (separate transducers only)
for Höntzsch vortex flow sensors VA as measuring tube (VA Di ...),
resolution 0.125 Hz.
Input temperature T:
for Pt100 temperature probe in 4-wire system,
acquisition time constant 2 s,
resolution 0.1 K.
Input pressure P:
for absolute pressure sensors 4 .. 20 mA, 12 .. 36 V in 2-wire system,
acquisition time constant 0.125 s,
resolution 1 hPa,
initial and terminal values configurable.
Analog outputs A1 and A2:
optionally, the analog outputs can be electrically isolated with each other and from the inputs using an
additional isolating amplifier.
Analog output A1 precision:
high precision analog output,
update every 0.125 s,
time constant 4 s, with frequency hopping > 25 % 2 s,
time constants up to 99 s are configurable.
4 ... 20 mA = 0 ... x kg/h, (NV/t)
terminal value x configurable / resistance max. 500 Ohm,
alternatively:
0 ... 10 V = 0 ... x kg/h, (NV/t)
terminal value x configurable / impedance 1 kOhm

__________________________________________________________________________________________________
_____________________________________________________________________________________
6/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
Analog output A2 fast:
analog output with short time constant,
update every 0.125 s, optionally 0.065 s
time constant 0.125 s
4 ... 20 mA = 0 ... x kg/h, (NV/t)
terminal value x configurable / resistance max. 500 Ohm,
alternatively:
0 ... 10 V = 0 ... x kg/h, (NV/t)
terminal value x configurable / impedance 1 kOhm
Relay: (potential-free normally open contact), max. 300 mA / 27 V DC,
configurable as limit value m/t (NV/) or mass pulse (standard flow rate pulse)
(see Section 7, Functional Description)
Optional raw signal output: BNC connector for VA frequency signal without linearization of the
Höntzsch vortex flow sensors with 5V level (TTL signal).
RS232 interface:
communication with UCOM (see Functional Description)
9600 Baud, 8Bit, no parity, 2 stop bits, Xon/Xoff
Connection for LCD display:
flat ribbon cable with 10-pin connector
(do not connect or disconnect when power is on)
6 Installation
Authoritative here are the European installation regulations EN 60079-10: 1996 ff, as well as the general
engineering regulations and these Operating Instructions.
6.1 Block Diagram
Transducer UVATP
terminals
16 .. 20
for separate
transducer
only

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 7/17
6.2 Wiring Diagrams
Electrical connection must be carried out according to the appropriate wiring diagram. Faulty connection
can cause damage to persons and destruction of the electronics.
Do not install or wire up the transducer under mains voltage. Non-compliance can cause damage to
persons and destruction of the electronics.
In this connection and depending on the configuration of the equipment, one of the following wiring
diagrams must be observed. Wiring diagrams for measuring systems in customer-specific design
will be supplied separately.
6.2.1 Cable shielding connection in the cable glands
6.2.2 Power supply
Before connecting please check that the power supply is within the specification. Remove the housing
cover of the UVATP to reveal the type plate with all relevant information.
P= protective earth

__________________________________________________________________________________________________
_____________________________________________________________________________________
8/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
6.2.3 Temperature sensor Pt100
6.2.4 Absolute pressure sensor 4 ... 20 mA
The initial value (4 mA) and the terminal value (20 mA) can be configured with the PC software UCOM via
the RS232 interface. Customer-specific programmed values can be found in the accompanying documents.
6.2.5 Analog outputs A1 and A2
All relevant information can be found on the type plate.
The terminal value of the analog outputs can be configured with the PC software UCOM via the RS232
interface. Customer-specific programmed values can be found in the accompanying documents.
Output signals are electrically isolated from the power supply. Optionally, both 4 ... 20 mA outputs can be
isolated with each other and from the inputs using an additional isolating amplifier.
Pt100 in 4-wire system
Absolute pressure sensor 4...20 mA 2-wire
system
4 ... 20 mA
resistance max. 500 Ohm
optional 0 ... 10 V
impedance 1 kOhm

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 9/17
6.2.6 Relay output
The normally open contact is shown in rest
position (relay coil off).
The function of the relay output and the corre-
sponding setting parameter can be configured
with the PC software UCOM via the RS232 inter-
face. Customer-specific settings can be found in
the documentation.
6.2.7 VA sensor input (separate transducers only)
In the case of sensor housings with screwed cable glands, the shielding of the sensor connection cable is
connected in the cable glands.
6.2.8 RS232 interface
The RS232 interface connection is below left next to the connecting terminals.
The RJ22 plug of the PC connecting cable is plugged in to the socket (see Fig. 1). PC connection follows at
a COM port or using an optional USB adapter.
Abb. 1: PC connection with RJ22 plug with cover open
VA Sensor
S= shielding
F1 = VA signal
G= VA mass
V+ = VA power supply

__________________________________________________________________________________________________
_____________________________________________________________________________________
10/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
6.2.9 LCD display
The flat ribbon cable with 10-pin connector should not be plugged in or out when live! Risk to
persons and equipment!
Visible are the readout potentiometer for the LCD display contrast and the ST1 and ST2 jumpers (see un-
der 7 –Functional Description).
6.3 Tips on Electromagnetic Compatibility (EMC)
Keep all connecting cables as short as possible.
With cable routes longer than 30 m or with strong electromagnetic disturbance along the cable route
between sensor and evaluation unit, the use of a double-shielded cable is recommended: to do this
lay inner shielding to one side of the evaluation unit and lay outer shielding on both sides over a large
area with low-impedance connection on the sensor and evaluation unit or on the control cabinet
Do not loop the cable!
Lay free strands at both ends on protection potential!
Lay cables as close to ground as possible, as for example side panels, mounting plates or steel
girders.
When using frequency converters there is a risk of interference by HF transient emissions.
Therefore, decouple the power input of the frequency converter via an interference filter
against active transient emissions. This also increases the passive interference resistance of the
equipment.
Use only shielded motor cables with shield support on both sides between motors and converters.
spatially separate cables which emit interference from measuring lines and evaluation units. If
necessary, lay measuring lines in a metallic tubular cable protection.
Ensure large area and low resistance connection of metallic parts in control cabinets, such as
subracks with control electronics or subplates.
Relays, contactors and electro valves in the same electric circuit, are to be wired with spark
quenching combinations or surge-limiting components.
Abb. 2: LCD display connection with cover open

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 11/17
Lay the shielding of analog signal lines on one side only –preferably on the evaluation unit - and low
resistive. Twist unshielded lines to counteract balanced interference to source terminals.
Lay the shielding of digital signal lines on both sides across a wide area. In the case of potential
differences between these points it is advisable to lay a separate equipotential bonding conductor.
Preferably provide shielded connector for connections to connector cable separation points.
If using terminals, these should be in a HF-shielded housing, in which EMC-compatible cable bushes
should be used. Connect outer shielding of the connector cables to the cable bush.
6.4 Connection cable
For connecting between sensor and separate transducer UVATP LiYCY cable with simple copper braiding
screen can be used if the cable path is short and there is marginal electromagnetic interference. Over
longer distances or with a high rate of electromagnetic interference, a double-screened cable of type
LiYCY-CY should be used.
VA sensor 3 x 0.25 mm2
Max. permissible conductor resistance per strand is 15 Ohm.
Conductor resistances for finely stranded conductors:
approx. 79 Ohm/km with wire cross section 0.25 mm2
approx. 39 Ohm/km with wire cross section 0.50 mm2
approx. 26 Ohm/km with wire cross section 0.75 mm2
7 Functional Description
UVATP transducers are for mass flow output or standard flow rate output by measuring the actual flow
rate, the operating temperature and the absolute pressure.
The signal frequency coming from the flow sensor and proportional to the actual flow rate signal is
converted into a linear analog output signal 4-20 mA or 0-10 V taking temperature and absolute
pressure into account (see Electrical Data, 5.5).
The analog terminal value is configurable.
4 ... 20 mA = 0 ... x kg/h (0… x Nm³/h)
alternatively:
0 ... 10 V = 0 ... x kg/h (0… x Nm³/h)
The relay output (normally open contact) can be configured for 1 of 2 different functions:
1. as limit value for the mass flow (standard flow rate):
mass flow (NV/t) < or = limit value: relay contact in rest position
mass flow (NV/t) > limit value: relay contact in operating position
2. as quantity pulse for the quantity measurement:
max. pulse repetition frequency 1 Hz per mass unit (standard flow rate),
configurable, e. g. 1 pulse per 1, 10 or 100 kg
pulse duration 0.5 s
In the case of measuring systems with a maximum mass flow rate up to 6500 kg/h, the output in the dis-
play and at the analog output follows with a decimal place (e.g. 6489.7 kg/h) and with higher maximum
mass flow rate without a decimal place (e.g. 24585 kg/h).

__________________________________________________________________________________________________
_____________________________________________________________________________________
12/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
Self diagnosis according to NAMUR NE43 specifications:
No error: yellow LED off
green LED on (actual flow rate = 0)
or
green LED flashes (actual flow rate > 0)
Error: yellow LED on and at the analog output
- output 4-20 mA : < 3.6 mA
- output 0-10 V : < -0.2 V
The following are monitored: power supply, data logging, sensor interface, parameter settings
(see under 16: Troubleshooting)
PC serial port RS232:
for modification of calibration data and setting parameters.
For this purpose remove cover.
Plug PC connector cable (optional) with RJ22 into the socket in the transducer (see Fig. 1, Page 9).
Connect other end of cable to PC COM port.
If a USB connection is required, then an optional USB / RS232 interface converter must be inserted.
Changes to the settings can now take place after starting the PC programme UCOM (optional)
Jumper settings (see photograph section 6.2.9) and effects on display, analog output
and counter
LCD display Exactflow II - jumper settings
Jumper
Jumper
LCD display
1. row
LCD display
2. row
analog output
(qantity counter)
ST1
ST2
m/s
A
BV/t
temperature
pressure
m/t
(m)
m/s
B
m/t
temperature
pressure
m/t
(m)
m³/h
A
BV/t
temperature
pressure
NV/t
(NV)
m³/h
B
NV/t
temperature
pressure
NV/t
(NV)
While changing the jumper settings the display and the analog output change accordingly.
The number of the analog end value remains but refers to the newly selected unit.
8 Settings
The setting parameters are readable and alterable using the UCOM software.
Customer-specific settings can be found in the accompanying documents.
For Operating Instructions Software UCOM see Document U385.

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 13/17
9 Performance Check
Sensor not connected
(Please pay attention to 6.2.2 Power supply and 6.2.5 Analog outputs)
After connecting the supply voltage the green and yellow LED light up.
The analog output provides a value between 3.4 mA and 3.6 mA with current output or between –0.3 V
and –0.2 V with voltage output.
10 Initial Operation
Sensor connected
(Please pay attention to 6.2.2 Power supply, 6.2.3 and 6.2.4 sensor inputs and 6.2.5 Analog outputs)
After connecting the supply voltage the green LED lights up and the yellow LED is off.
No flow at sensor: the green LED is permanently on; the analog output provides a value of 4 mA with
current output and 0 V with voltage output.
Flow at sensor: the green LED flashes; the analog output supplies an analog value deviating from the
zero flow conditions (see above).
11 Operation
(Please pay attention to 5.3 Operating Conditions and 5.5 Electrical Data)
12 Shut-down, Dismantling
Before disconnecting the cable, please ensure that the supply voltage is switched off before
disconnecting the cable.
13 Inspection
Check the LEDs, (see under 7 Functional Description, Self diagnosis)

__________________________________________________________________________________________________
_____________________________________________________________________________________
14/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
14 Maintenance
Use only residue-free drying cleaning agents which are compatible with the housing materials.
15 Calibration
The UVATP has an excellent long-term stability. However, it makes sense to have it calibrated at regular
intervals. The time lapse between calibrations depends on the individual operating conditions and the
tolerable measurement error. Therefore, as a precautionary measure, shorter calibration cycles are
recommended in the beginning. As a general rule and under 'normal' operating conditions we suggest a
cycle of approx. 3 - 5 years.
The transducer must be returned to the manufacturer for calibration (see under 17).

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 15/17
16 Troubleshooting
Fault
Cause
Troubleshooting
green LED does not
light up
no power supply
check connecting cable; measure voltage
at connecting terminals
transducer electronics faulty
return to factory
green LED does not
flash despite flow
VA sensor cable break or short
circuit
check terminals;
check cable for continuity and replace if
necessary
transducer electronics or sensor
faulty
return to factory
yellow LED lights up,
green LED flashes with
flow, analog output =
error
(<3.6 mA or < -0.2V)
parameter error
check parameter with UCOM software;
save new checksum
(or return to factory)
temperature probe or connection
cable faulty
check terminals;
check cable for continuity and replace if
necessary or return to factory
absolute pressure sensor or cable
faulty
check terminals;
check cable for continuity and replace if
necessary or return to factory
transducer electronics faulty
return to factory
no measured value
sensor contaminated
clean sensor according to instructions
profile factor set at 0.000
set profile factor at '1.000' with volumet-
rically calibrated measuring tubes
measured value too
low
sensor contaminated
clean sensor according to instructions
profile factor set too low
set profile factor at '1.000' with volumet-
rically calibrated measuring tubes
input/output section too short
change sensor position, improve flow
conditions with a flow rectifier
rotational flow
reposition sensor in flow direction, install
flow rectifier
reduced acoustic coupling in the
sensor elements as a result of
intense vibration or powerful
impact
return sensor to factory for performance
test
burden at current output greater
than specified in the Technical
Data Sheet resulting in correct
output values in the lower range
and no longer increasing values
at the top end of the measuring
range
reduce resistance
incorrect scaling of analog output
check setting and amend if necessary
measured value too
high
profile factor set too high
set profile factor at '1.000' with volumet-
rically calibrated measuring tubes
incorrect scaling of analog output
check setting and amend if necessary
EMC problem
see 6.3

__________________________________________________________________________________________________
_____________________________________________________________________________________
16/17 U398_UVATP_B_e_190425
Operating Instructions Transducer UVATP
in AS102 Housing
17 Returns
No special measures need to be taken for non-integrated UVATP transducers.
Clean associated sensors or those with integrated transducer thoroughly before returning to the
manufacturer (see cleaning instructions for the sensor).
A hazard warning or declaration of harmlessness for substances which have come into contact with
or infiltrated the sensor, must be submitted with all returns. If adhesion of hazardous substances on the
surface of the equipment cannot be ruled out, a detailed description of safety measures to be taken when
handling the equipment must be submitted.
18 Disposal
The customer should assume the duty to dispose of the equipment at his own expense and according to
statutory provisions (e.g. ElektroG in Germany).

__________________________________________________________________________________________________
___________________________________________________________________________________________________
U398_UVATP_B_e_190425 17/17
19 Declaration of Conformity
____________________________________________________________________________
We, Höntzsch GmbH
Gottlieb-Daimler-Straße 37
D-71334 Waiblingen
bearing sole responsibility, hereby declare that the product
Transducer UVATP in AS102 Housing
referred to in this declaration, is in conformity with the following standards or normative documents:
Provisions of the Directive
Reference and date of issue
2014/30/EU: Electromagnetic Compatibility
EN 61000-6-4
EN 61000-6-2
2014/68/EU: Pressure Equipment Directive
2006/42/EC: Safety of Machinery
EN 61010
Waiblingen, 25.04.2019 Jürgen Lempp / Managing Director
Höntzsch GmbH
Gottlieb-Daimler-Straße 37
D-71334 Waiblingen
Tel: +49 7151 / 17 16-0
Internet: www.hoentzsch.com
Subject to alterations
Other manuals for UVATP
1
Table of contents
Other Hontzsch Transducer manuals
Popular Transducer manuals by other brands

Iskra
Iskra MI 436 user manual

Electro-Voice
Electro-Voice DH3 Technical specifications

Pulsar Measurement
Pulsar Measurement HART dBi Series installation manual

M-system
M-system LTCE instruction manual

Endress+Hauser
Endress+Hauser Cerabar PMP23 operating instructions

MESSKO
MESSKO Pt-MU operating instructions

Airmar
Airmar PM265LH Owner's guide & installation instructions

Phoenix Contact
Phoenix Contact MACX MCR-CONF user manual

Thyracont
Thyracont VSR53D operating instructions

Airmar
Airmar R299 Owner's guide and installation instructions

Setra Systems
Setra Systems 280G operating instructions

Absolute Measurement
Absolute Measurement CLT instruction manual