Hontzsch UVATP User manual

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 1/18
Operating Instructions
Transducer UVATP
in AS102 housing
Contents
1Safety Symbols .........................................................................................................................2
2Safety Precautions .....................................................................................................................2
3Intended Use.............................................................................................................................3
4Operating Safety .......................................................................................................................3
5Scope of Delivery.......................................................................................................................3
5.1 Description, Type Plates.....................................................................................................4
6Technical Specifications ..............................................................................................................4
6.1 Conformity with Standards .................................................................................................4
6.2 Storage Conditions ............................................................................................................4
6.3 Operating Conditions .........................................................................................................5
6.4 Housing and Connectors ....................................................................................................5
6.5 Electrical Data ..................................................................................................................5
7Installation................................................................................................................................6
7.1 Block Diagram ..................................................................................................................6
7.2 Wiring Diagrams ...............................................................................................................7
7.2.1 Shielding contact at cable bush ................................................................................7
7.2.2 Power supply .........................................................................................................7
7.2.3 VA Sensor input (separate transducers only) .............................................................8
7.2.4 Temperature sensor Pt100 ......................................................................................8
7.2.5 Absolute pressure sensor 4 … 20 mA ........................................................................8
7.2.6 Analog outputs A1 and A2 .......................................................................................9
7.2.7 Relay output..........................................................................................................9
7.2.8 RS232 Interface...................................................................................................10
7.2.9 LCD Display.........................................................................................................11
7.3 Tips on Electromagnetic Compatibility (EMC) ...................................................................... 11
7.4 Connection Cable ............................................................................................................12
8Functional Description .............................................................................................................. 12
9Settings..................................................................................................................................14
10 Performance Test.....................................................................................................................14
11 Initial Operation.......................................................................................................................14
12 Operation ............................................................................................................................... 14
13 Shut-down, Dismantling ........................................................................................................... 14
14 Inspection...............................................................................................................................15
15 Maintenance............................................................................................................................ 15
16 Meaning of LEDs ......................................................................................................................15
17 Calibration .............................................................................................................................. 15
18 Troubleshooting.......................................................................................................................16
19 Returns ..................................................................................................................................17
20 Disposal.................................................................................................................................. 17
21 Declaration of Conformity, Declaration of Incorporation ................................................................18

_________________________________________________________________________________________________
________________________________________________________________________________________
2/18 U398_UVATP_B_e_221219
1Safety Symbols
Warning!
Failure to observe the instructions can result in serious injury and damage to property!
Important notice!
Non-observance can result in serious damage to the equipment or performance restriction!
2Safety Precautions
Danger to life, risk of injury and damage to material or property.
Read the Operating Instructions carefully before initial operation.
Observe general safety precautions as well as those included in various sections of these Operating
Instructions.
Hazard risks:
-non-observance of the Operating and Safety Instructions
-modifications to the device by the customer
-handling the device outside the specified operating conditions
-handling the transducers outside the specified operating conditions
-use of unsuitable power supplies and peripheral devices
-improper use of the device
Prevention of voltage hazards:
-use only the dedicated adapter plug for the mains supply
-make sure that the PC is correctly connected to the mains (earthed safety socket, earthing) when
using a USB connection
-when connecting analog outputs or inputs to peripheral devices make sure that these are correctly
connected to the mains (earthed safety socket, earthing)
Danger when installing the sensors in pressurized pipelines:
-sensors for use in pressurized pipelines are to be inserted or retracted only in depressurized conditions;
non-observance may result in serious injuries to personnel
-when installing or removing under pressure, the appropriate protective equipment must be used,
e.g. ball valve and probe guide pieces with chain guard or spindle probe guide pieces

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 3/18
3Intended Use
-The UVATP transducer is for mass flow or standard flow rate output by measuring the actual flow rate,
the operating temperature and the absolute pressure.
-It is designed for industrial application.
-The AS102 is an aluminium housing in protection class IP65.
-The manufacturer is not liable for damage caused by improper use and/or non-compliance with the
regulations.
-Do not carry our any structural modifications to the transducers.
-Always follow the instructions on the type plate, especially the information regarding supply voltage.
-The UVATP can be integrated in the AS102 connection housing of measuring tubes VA TP DN ExactFlow II
or is available as a separate transducer in AS102 housing.
4Operating Safety
All steps described below must be carried out by qualified personnel only!
Please read the Operating Instructions carefully before unpacking the equipment!
Safety can only be guaranteed if the equipment is operated in accordance with the regulations. Improper
handling can result in serious injury and damage to property.
5Scope of Delivery
-UVATP integrated in AS102 housing of VA TP DN ExactFlow II measuring tube or as separate transducer
in AS102 housing
-LCD display 2x16 digit in the cover
-Operating Instructions, Data Sheet
-CD-ROM with PC configuration software UCOM VTP (optional)
-Interface cable RS232 for PC COM port connection (optional)
-USB adapter in addition to the interface cable RS232 (optional)
Please check that everything listed in the Delivery Note / Technical Data Sheet is included in the delivery.

_________________________________________________________________________________________________
________________________________________________________________________________________
4/18 U398_UVATP_B_e_221219
5.1 Description, Type Plates
Remove the cover to display one of the following type plates:
S/N. : ... : serial number
Type : UVATP : for vortex VA measuring tubes, temperature, pressure
Power : 24 V DC : supply voltage 24 V direct voltage current
12 V DC : supply voltage 12 V direct voltage current
Output A1 : precision 4 ... 20 mA : analog output with high precision current 4 ... 20 mA
precision 0 ... 10 V : analog output with high precision voltage 0 ... 10 V
Output A2 : fast 4 ... 20 mA : analog output with short time constant current 4 ... 20 mA
fast 0 ... 10 V : analog output with short time constant current 0 ... 10 V
Input : VADi : for vortex VA measuring tubes (separate transducers only)
RS232 : RJ22 plug : for serial PC interface
6Technical Specifications
6.1 Conformity with Standards
The transducer UVATP is manufactured according to the best available technology, is safe and reliable and
comply with the relevant regulations, EU directives and standards.
6.2 Storage Conditions
Storage temperature: -30 to +70 °C
separate UVATP integrated UVATP

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 5/18
6.3 Operating Conditions
ambient temperature
of connection housing
in use : -25 ... +60 °C
protection class : IP65
mounting attitude : no restrictions, preferably with cable bush facing down
6.4 Housing and Connectors
Protection class (housing) : IP65
Material : aluminium
External dimensions : L/W/H = 150/100/80 mm
Cable entry point : nickel-plated brass screwed cable glands
for cable diameter 5 ... 10 mm
with cable shielding connections
Connections : 'push in' PCB terminals
for wire cross-section 0.14 ... 1.5 mm²
No tools necessary for core connection –simply insert strands
(twisted or with end sleeve) into the terminal.
To separate strands apply pressure to the terminal release spring
using a pen or screwdriver.
6.5 Electrical Data
Supply voltage,
mains supply : 24 V DC (20 ... 27 V DC), power < 5 W
alternatively
12 V DC (10 ... 17 V DC), power < 5 W
The mains supply is electrically isolated from the UVATP inputs and outputs.
Input flow v/VA : (separate transducers only)
for Höntzsch vortex flow sensors as measuring tube (VA Di …),
resolution 0.125 Hz
Input temperature T : for Pt100 temperature probe in 4-wire-system, acquisition time constant
2 s, resolution 0.1 K
Input pressure p : for absolute pressure sensors 4 … 20 mA, 12 … 36 V in 2-wire-system,
acquisition time constant 0.125 s, resolution 1 hPa, initial and terminal
value configurable
Analog outputs A1 and A2 : optionally, the analog outputs can be electrically isolated with each
other and from the inputs using an additional isolating amplifier
Analog output A1 (precision): high precision analog output, update every 0.125 s, time constant 4 s,
with frequency hopping > 25 % 2 s, time constant up to 99 s are con-
figurable
4 ... 20 mA = 0 ... x kg/h, (NV/t)
terminal value x configurable / resistance max. 500 Ohm,
alternatively:
0 ... 10 V = 0 ... x kg/h, (NV/t)
terminal value x configurable / impedance 1 kOhm

_________________________________________________________________________________________________
________________________________________________________________________________________
6/18 U398_UVATP_B_e_221219
Analog output A2 (fast) : analog output with short time constant, update every 0.125 s, optionally
0.065 s, time constant 0.125 s
4 ... 20 mA = 0 ... x kg/h, (NV/t)
terminal value x configurable / resistance max. 500 Ohm,
alternativ:
0 ... 10 V = 0 ... x kg/h, (NV/t)
terminal value x configurable / impedance 1 kOhm
Relay : (potential-free normally open contact), max. 300 mA / 27 V DC, config-
urable as limit value m/t (NV/t) or mass pulse (standard flow rate pulse)
(see under 8 Functional Description)
Optional raw signal output : BNC connector for VA frequency signal without linearization of the
Höntzsch vortex flow sensors with level 5 V (TTL signal)
RS232 interface : communication with PC programme UCOM VTP
(see under 8 Functional Description)
9600 Baud, 8 Bit, no parity, 2 stop bits, Xon/Xoff
Connection for LCD display : flat ribbon cable with 10-pin cable socket
Do not plug in or out when live!
7Installation
The current European Specifications for Assembly, the recognised standards of good practice and this
Operating Instructions apply.
7.1 Block Diagram
Transducer UVATP
terminals
16 .. 20
for separate
transducer
only

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 7/18
7.2 Wiring Diagrams
Electrical connection must be carried out according to the appropriate wiring diagram.
Faulty connection can cause damage to persons and destruction of the electronics.
Do not install or wire up the transducer under mains voltage. Non-compliance can cause damage to
persons and destruction of the electronics.
In this connection and depending on the configuration of the equipment, one of the following wiring dia-
grams must be taken into account. Wiring diagrams for measuring systems in customer-specific design will
be supplied separately.
7.2.1 Shielding contact at cable bush
7.2.2 Power supply
Before connecting please check that the power supply is within the specification.
Remove the housing cover of the UVATP to reveal the type plate with all relevant information.
P= protective earth

_________________________________________________________________________________________________
________________________________________________________________________________________
8/18 U398_UVATP_B_e_221219
7.2.3 VA Sensor input (separate transducers only)
In the case of sensor housings with screwed cable glands, the shielding of the sensor connection cable is
connected in the cable glands.
7.2.4 Temperature sensor Pt100
7.2.5 Absolute pressure sensor 4 … 20 mA
The initial value (4 mA) and the terminal value (20 mA) can be configured with the PC software UCOM VTP
via the RS232 interface. Customer-specific programmed values can be found in the accompanying
documents.
Absolute pressure sensor 4...20 mA 2-wire
system
Pt100 in 4-wire system
VA Sensor
S= shielding
F1 = VA signal
G= VA mass
V+ = VA power supply

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 9/18
7.2.6 Analog outputs A1 and A2
All relevant information can be found on the type plate.
The terminal value of the analog outputs can be configured with the PC software UCOM VTP via the RS232
interface. Customer-specific programmed values can be found in the accompanying documents.
Output signals are electrically isolated from the power supply. Optionally, both 4 ... 20 mA outputs can be
isolated with each other and from the inputs using an additional isolating amplifier.
7.2.7 Relay output
The normally open contact is shown in rest
position (relay coil off).
The function of the relay output and the
corresponding setting parameter can be config-
ured with the PC software UCOM VTP via the
RS232 interface. Customer-specific settings can
be found in the documentation.
4 ... 20 mA
resistance max. 500 Ohm
optional 0 ... 10 V
impedance 1 kOhm

_________________________________________________________________________________________________
________________________________________________________________________________________
10/18 U398_UVATP_B_e_221219
7.2.8 RS232 Interface
The RS232 interface connection is below left next to the connecting terminals.
The RJ22 plug of the PC connecting cable is plugged in to the socket (see Fig. 1). PC connection follows at
a COM port or using an optional USB adapter.
Fig. 1: PC connection with RJ22 plug with cover open

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 11/18
7.2.9 LCD Display
The flat ribbon cable with 10-pin connector should not be plugged in or out when live! Risk to
persons and equipment!
Visible are the readout potentiometer for the contrast of the LCD display, the reset button for the quantity
counter, the ST1 and ST2 jumpers (see under 8 Functional Description).
7.3 Tips on Electromagnetic Compatibility (EMC)
Keep all connecting cables as short as possible.
With cable routes longer than 30 m or with strong electromagnetic disturbance along the cable route
between sensor and evaluation unit, the use of a double-shielded cable is recommended: to do this
lay inner shielding to one side of the evaluation unit and lay outer shielding on both sides over a large
area with low-impedance connection on the sensor and evaluation unit or on the control cabinet
Do not loop the cable!
Lay free strands at both ends on protection potential!
Lay cables as close to ground as possible, as for example side panels, mounting plates or steel
girders.
When using frequency converters there is a risk of interference by HF transient emissions.
Therefore, decouple the power input of the frequency converter via an interference filter
against active transient emissions. This also increases the passive interference resistance of the equip-
ment.
Use only shielded motor cables with shield support on both sides between motors and converters.
spatially separate cables which emit interference from measuring lines and evaluation units. If
necessary, lay measuring lines in a metallic tubular cable protection.
Ensure large area and low resistance connection of metallic parts in control cabinets, such as
subracks with control electronics or subplates.
Relays, contactors and electro valves in the same electric circuit, are to be wired with spark
Fig. 2: LCD display connection with cover open

_________________________________________________________________________________________________
________________________________________________________________________________________
12/18 U398_UVATP_B_e_221219
quenching combinations or surge-limiting components.
Lay the shielding of analog signal lines on one side only –preferably on the evaluation unit - and low
resistive. Twist unshielded lines to counteract balanced interference to source terminals.
Lay the shielding of digital signal lines on both sides across a wide area. In the case of potential
differences between these points it is advisable to lay a separate equipotential bonding conductor.
Preferably provide shielded connector for connections to connector cable separation points.
If using terminals, these should be in a HF-shielded housing, in which EMC-compatible cable bushes
should be used. Connect outer shielding of the connector cables to the cable bush.
7.4 Connection Cable
For connecting between sensor and separate transducer UVATP LiYCY cable with simple copper braiding
screen can be used if the cable path is short and there is marginal electromagnetic interference. Over
longer distances or with a high rate of electromagnetic interference, a double-screened cable of type
LiYCY-CY should be used.
VA sensor 3 x 0.25 mm²
Max. permissible conductor resistance per strand is 15 Ohm.
Conductor resistances for finely stranded conductors:
approx. 79 Ohm/km with wire cross section 0.25 mm2
approx. 39 Ohm/km with wire cross section 0.50 mm2
approx. 26 Ohm/km with wire cross section 0.75 mm2
8Functional Description
UVATP transducers are for mass flow output or standard flow rate output by measuring the actual flow
rate, the operating temperature and the absolute pressure.
The signal frequency coming from the flow sensor and proportional to the actual flow rate signal is con-
verted into a linear analog output signal 4-20 mA or 0-10 V taking temperature and absolute pressure
into account (see under 6.5 Electrical Data).
The analog terminal value is configurable.
Current output : 4 ... 20 mA = 0 ... x kg/h (0 … x Nm³/h)
alternatively:
Voltage output : 0 ... 10 V = 0 ... x kg/h (0 … x Nm³/h)
The relay output (normally open contact) can be configured for 1 of 2 different functions:
1. as limit value for the mass flow (standard flow rate):
mass flow (NV/t) < or = limit value : relay contact in rest position
mass flow (NV/t) > limit value : relay contact in operating position
2. as quantity pulse for the quantity measurement:
max. pulse repetition frequency 1 Hz per mass unit (standard flow rate unit), configurable,
e.g. 1 pulse per 1, 10 or 100 kg, pulse duration 0.5
In the case of measuring systems with a maximum mass flow rate up to 6500 kg/h, the output in the dis-
play and at the analog output follows with a decimal place (e.g. 6489.7 kg/h) and with higher maxi-mum
mass flow rate without a decimal place (e.g. 24585 kg/h).

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 13/18
Self diagnosis according to NAMUR NE43 specifications:
For analog output 4 … 20 mA:
No error: analog output = 4 mA (flow velocity = 0)
or analog output > 4 mA (flow velocity > 0)
Error : analog output < 3.6 mA
Bei Analogausgang 0 … 10 V:
No error : analog output = 0 V (flow velocity = 0)
or analog output > 0 V (flow velocity > 0)
Error : analog output < -0.2 V
Monitoring of power supply, data logging, sensor interface, parameter settings (see under 18 Trouble-
shooting)
PC serial port RS232
for modification of calibration data and setting parameters.
For this purpose remove cover.
Plug PC connector cable (optional) with RJ22 into the socket in the transducer UVATP (see Fig. 1, see un-
der 7.2.8 RS232 Interface)
Connect other end of cable to PC COM port.
If a USB connection is required, then an optional USB / RS232 interface converter must be inserted.
Changes to the settings can now take place after starting the PC programme UCOM VTP (optional) (see
under Fehler! Verweisquelle konnte nicht gefunden werden. Settings).
Jumper settings (see Fig. 2 under 7.2.9 LCD Display) and effects on display, analog output and counter
LCD display ExactFlow II –jumper settings
Jumper
Jumper
LCD display
1. row
LCD display
2. row
Analog output
(qty. counter)
ST1
ST2
m/s
A
BV/t
temperature
pressure
m/t
(m)
m/s
B
m/t
temperature
pressure
m/t
(m)
m³/h
A
BV/t
temperature
pressure
NV/t
(NV)
m³/h
B
NV/t
temperature
pressure
NV/t
(NV)
While changing the jumper settings the display and the analog output change accordingly.
The number of the analog end value remains but refers to the newly selected unit m/t or NV/t.
If an error occurs, „EEE“ is displayed in row 1. (see under 18 Troubleshooting)

_________________________________________________________________________________________________
________________________________________________________________________________________
14/18 U398_UVATP_B_e_221219
9Settings
The setting parameters are readable and alterable using the UCOM VTP software.
Die kundenspezifischen Einstellungen dieser Parameter sind im Parameter-Ausdruck dargestellt,
der den Lieferdokumenten beiliegt.
For Operating Instructions Software UCOM VTP see document U441.
10 Performance Test
Sensor not connected:
(see under Fehler! Verweisquelle konnte nicht gefunden werden. Power supply)
(see under 7.2.6 Analog outputs A1 and A2)
After connecting the supply voltage the green and yellow LED light up.
The analog output provides a value between 3.4 mA and 3.6 mA with current output or between –0.3 V
and –0.2 V with voltage output (see under 8 Functional description)
11 Initial Operation
Sensor connected:
(see under Fehler! Verweisquelle konnte nicht gefunden werden. Power supply)
(see under 7.2.3, 7.2.4 and 7.2.5 Sensor inputs)
(see under 7.2.6 Analog outputs A1 and A2)
After connecting the supply voltage the green LED lights up and the yellow LED is off.
No flow at sensor: the green LED is permanently on; the analog output provides a value of 4 mA with cur-
rent output and 0 V with voltage output (see under 8 Functional Description)
Flow at sensor: the green LED flashes; the analog output supplies an analog value deviating from zero flow
conditions (see above).
12 Operation
(see under Fehler! Verweisquelle konnte nicht gefunden werden. Operating Conditions)
(see under 6.5 Electrical Data)
13 Shut-down, Dismantling
Before disconnecting the cable, please ensure that the supply voltage is switched off.

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 15/18
14 Inspection
Check the LEDs, (see under 8 Functional Description, self diagnosis)
15 Maintenance
Use only residue-free drying cleaning agents which are compatible with the housing materials.
Any repair work is to be carried out solely by Höntzsch GmbH & Co. KG.
16 Meaning of LEDs
LED green
LED yellow
Description
off
off
no power supply
An
off
power supply ok,
no error,
no flow
flashing
off
power supply ok,
no error,
flow detected
on
on
power supply ok,
parameter error and/or sensor error,
no flow
flashing
on
power supply ok,
parameter error,
flow detected
17 Calibration
The UVATP has an excellent long-term stability. However, it makes sense to have it calibrated at regular
intervals. The time lapse between calibrations depends on the individual operating conditions and the tol-
erable measurement error. Therefore, as a precautionary measure, shorter calibration cycles are recom-
mended in the beginning. As a general rule and under 'normal' operating conditions we suggest a cycle of
approx. 3 - 5 years.
The transducer must be returned to the manufacturer for calibration (see under 19 Returns)

_________________________________________________________________________________________________
________________________________________________________________________________________
16/18 U398_UVATP_B_e_221219
18 Troubleshooting
Fault
Cause
Troubleshooting
green LED does not
light up
no power supply
check connecting cable; measure voltage at
connecting terminals
transducer electronics faulty
return to factory
green LED does not
flash despite flow
VA sensor cable break or short
circuit
check terminals;
check cable for continuity and replace if
necessary
transducer electronics or
sensor faulty
return to factory
yellow LED lights
up, green LED
flashes with flow,
analog output =
error
(<3.6 mA or < -
0.2V)
parameter error
check parameter with UCOM software;
save new checksum
(or return to factory)
temperature probe or
connection cable faulty
check terminals;
check cable for continuity and replace if
necessary or return to factory
absolute pressure sensor or
cable faulty
check terminals;
check cable for continuity and replace if
necessary or return to factory
transducer electronics faulty
return to factory
no measured value
sensor contaminated
clean sensor according to instructions
profile factor set at 0.000
set profile factor at '1.000' with volumetrically
calibrated measuring tubes
measured value
too low
sensor contaminated
clean sensor according to instructions
profile factor set too low
set profile factor at '1.000' with volumetrically
calibrated measuring tubes
input/output section too short
change sensor position, improve flow conditions
with a flow rectifier
rotational flow
reposition sensor in flow direction, install flow
rectifier
reduced acoustic coupling in
the sensor elements as a re-
sult of intense vibration or
powerful impact
return sensor to factory for performance test
burden at current output
greater than specified in the
Technical Data Sheet resulting
in correct output values in the
lower range and no longer
increasing values at the top
end of the measuring range
reduce resistance
incorrect scaling of analog
output
check setting and amend if necessary
measured value
too high
profile factor set too high
set profile factor at '1.000' with volumetrically
calibrated measuring tubes
incorrect scaling of analog
output
check setting and amend if necessary
EMC problem
see remark regarding electromagnetic compati-
bility (EMC) (see under 7.3)

_________________________________________________________________________________________________
______________________________________________________________________________________
U398_UVATP_B_e_221219 17/18
19 Returns
No special measures need to be taken for non-integrated UVATP transducers.
Clean associated sensors or those with integrated transducer thoroughly before returning to the (see
cleaning instructions for the sensor). A hazard warning or declaration of harmlessness for substances
which have come into contact or infiltrated the sensor, must be submitted with all returns. If adhesion of
hazardous substances on the surface of the equipment cannot be ruled out, a detailed description of safety
measures to be taken when handling the equipment must be submitted.
20 Disposal
The customer should assume the duty to dispose of the equipment at his own expense and according to
statutory provisions (e.g. ElektroG in Germany).

_________________________________________________________________________________________________
________________________________________________________________________________________
18/18 U398_UVATP_B_e_221219
21 Declaration of Conformity, Declaration of Incorporation
Waiblingen, 19.12.2022 Jürgen Lempp / Managing Director
Höntzsch GmbH & Co. KG
Gottlieb-Daimler-Straße 37
D-71334 Waiblingen
Tel: +49 7151 / 17 16-0
E-Mail info@hoentzsch.com
Internet www.hoentzsch.com
Subject to alterations
We, Höntzsch GmbH & Co. KG
Gottlieb-Daimler-Str. 37
D-71334 Waiblingen
bearing sole responsibility, hereby declare that the product
Transducer UVATP in AS102 Housing
referred to in this declaration, is in conformity with the following standards or norma-
tive documents:
Provisions of the Directive
Reference and date of issue
2014/30/EU: Electromagnetic Compatibility
EN 61000-6-4
EN 61000-6-2
2014/68/EU: Pressure Equipment Directive
2006/42/EC: Safety of Machinery
Safety requirements for electrical equipment for
measurement, control, and laboratory use e
EN 61010
Other manuals for UVATP
1
Table of contents
Other Hontzsch Transducer manuals