Preface
Limited Warranty and Limited Responsibility
Safety Instruction
Preliminary Acquaintance with Infrared Thermal Imager
Overview
Cleaning of the Products
Lens Maintenance
Charging of the Battery and Description
Performance Index
Product Description
1. Instruction to Structure
2. Display Description
3. Key Description
Initial Operation
Introduction to the Menu
1.“Image Overlapping” Submenu
1.1. Description of Image Overlapping
1.2. Application of Image Overlapping
2. Introduction to “Image” Sub-menu
2.1. View Image
2.2. Delete Images
3. Introduction to “Video” Submenu
3.1. View Video
23
24
25
26
27
28
28
30
28
28
29
33
29
24
21
22
22
28
31
3.2. Delete Video
4. Introduction to “Color Palette” Sub-menu
4.1. Color Palette Description
4.2. Application of Color Palette
5. Introduction to “Emissivity” Sub-menu
5.1. Emissivity Description
5.2. Emissivity Setting
5.3. The Emissivity Value of Common Materials
6. Introduction to “Setting” Sub-menu
6.1. Auto Shutdown Setting
6.2. Intensity Settings
6.3. Language Settings
6.4. Unit Setting
6.5. Tempertature Range Setting
6.6. Time Format Setting
6.7. Time Setting
6.8. Enable/Disable of the Highest and Lowest Temperature Cursor
6.9. Wireless WIFI on and off
Thermal Imaging Image Analysis Software Operation Guide

1 2
Preface
Limited Warranty and Limited Responsibility
Respected users:
How are you! Thank you for purchasing our instrument. In order to use it in
a better manner, we remind you to read the instruction for use carefully
before use and keep it properly so as to look up in the future use.
Dongguan Xintai Instrument Co.,Ltd. guarantees that the product enjoys
warranty service of one year from the purchase date.
But the warranty is not applicable to fuse and disposable battery or the
damage caused by improper operation, accidents, negligence, misuse,
remodeling, pollution and abnormal condition of the operation environment.
The distributor has no right to provide any warranty in the name of
Dongguan Xintai Instrument Co.,Ltd.
Dongguan Xintai Instrument Co.,Ltd. authorizes the retail dealers to
provide new products or unused products to final customers but does not
authorize them to provide guarantee with wider scope or different contents.
Only the products that are purchased through distributors authorized by
Dongguan Xintai Instrument Co.,Ltd. or products that the purchaser paid
according to proper international price can enjoy the warranty support
of our company. When the product bought in a country is sent to another
country for repair, Dongguan Xintai Instrument Co.,Ltd. reserves the right
charge the import fees for the parts and components to be repaired or
to replaced from the purchasers.
When warranty service is repaired, please contact the service center
authorized by Dongguan Xintai Instrument Co.,Ltd. to obtain the
authorization information for return. Then post the product together with
problem description to the service center with postage and insurance
expenses prepaid. Our company will not undertake any responsibility for
any damage occurred during the process of transportation. The product will
be posted to the purchaser after repair with the transportation expenses
paid by the purchasers.
If the product failure is verified by negligence, misuse, pollution, change,
accidents or improper operation or handling after inspection, including the
over voltage failure caused by use without following the rated value
stipulated by the product or daily service wear the of the parts, our company
will estimate the repair expenses and carry out repair after obtaining the
consent of the purchasers. The product will be posted back to the purchaser
after repair with the transportation expenses paid by the purchasers.
The guarantee is the only compensation that users can be obtained and
does not include other express or implied guarantees. Dongguan Xintai
Instrument Co.,Ltd. will not undertake any responsibility for any special,
indirect, accidental or subsequent damage or loss, including any data loss
caused by any reasons or inference.
Because some countries or states do not allow implied warranties and
incidental or consequential damages.Restrictions, so the above limitation of
liability and regulations may not apply to every buyer.
The users shall make promise that they familiarize with the application field
and occasi on. The purchase indicates that the users are deemed to know
the suitability for the products.

43
Safety Instruction
Considerations
Please ensure that you have read and understood the safety precautions
described in the following text before using the product so that you can
operate it correctly.
The safety precautions described in the following text guides you to operate
the product and its accessories correctly and safely to avoid damage and
loss to you,other people and the device.
The warning describes the acts that may cause harm to users. Please
observe the following operation in order to avoid electric shock or personal
injury.
If the product case is damaged, don’t use it continuously.
Please contact the local distributors or agents of Xintai under such
condition.
Please observe the following guide in order not to damage the product:
Don’t assemble or dismantle the product without permission.
The product is a type of super precise equipment. Don’t try to dismantle,
assemble or remodel any part of the product. Repair of the product should
be made by technical personnel designated by Xintai Company.
Avoid damage to probe of the product.
Note: don’t put the product close to strong heat source directly (such as
electric iron). Otherwise, it may damage the probe of the product.
Product hum.
When the product works, there is light sound of clicking every several
seconds. This is normal phenomenon that the lens captures images.
Warning
During the process of use, if the product is found with smoke,
spark and the burnt smell, please stop using at once.
In case of such condition, the product power supply should be powered
off first.After the smoke and peculiar smell disappears completely,
please contact the local distributors or agents of Xintai.
Don’t remodel the adapter and the data line.
Otherwise, such remodeling may cause short circuit or fire.
Don’t weld the battery without permission.
Such operation may damage the battery and result in leakage and
explosion of the battery.
Avoid that the battery is impacted (such as collusion and falling,
etc).
Such condition may damage the battery case or result in that the
battery leaks or explode.

5 6
Please pull off the adapter on the power socket when charging is
not made.
The adapter may get overheated if it is connected with power supply for
a long time. This may cause overheating, deformation and fire.
When the plug of the adapter or wire is damaged, stop using
immediately.
Don’t change the battery unless the plug of the charger is inserted into
the socket completely.
Please ensure that the plug of the adapter is inserted into the
specified power socket.
The adapter plug may be different due to the regions. Please make
confirmation whether the specification of the adapter is consist with the
specification of electric appliances in your region. Otherwise, this may
cause overheating of the equipment, electric shock, fire, chemical
leakage inside the battery, explosion and other serious consequences.
Do not touch the electric wire with wet hands.
It is possible to cause electric shock when touching the electric wire
with wet hands. When pulling out the electric wire, hold the electric wire
head rightly to pull out the wire. Don’t pull off the electric wire directly.
Otherwise, the electric wire may be broken, causing electric shock and
fire.
Forbid dip the product into water caught in rain.
If the case contact with any liquid, please wipe it dry immediately. If
water or other liquid enters the inside of the instrument, please power
off the power supply at once. Continuous use may result in product
damage.
The product temperature may increase after a long time of
charging.
You may feel scorching heat when the hands touch the sensors.
Problem caused by water condensation.
Don’t bring the instrument into low temperature environment from high
tempera-ture environment in a short time or from low temperature
environment to high temperature environment. This may result in that
the inside of the instrument and appearance produce water
condensation. In such condition, the instrument should be placed into
portable box or plastic bag. Before use, make it restore to the
environmental temperature and take it out to use.
If the product has water condensation inside, please power it off
immediately.Otherwise, the instrument may be damaged. Operation is
not allowed unless the water condensation disappears.
Clean the dust on the adapter plug and the data line.
When it is exposed to dusty and dump environment for a long time, the
dirt surrounding the electric equipment will accumulate moisture. This
may cause short circuit and fire.
Please use the original adapter of the company to charge the
product.
The use of non-original power supply accessories may cause
equipment overheating, electric shot, fire, and leakage of chemicals
inside the battery,explosion and other serious consequences.
Such operation may cause the product case to be damaged.
Don’t use abradant, isopropanol or solvent to clean the enclosure
of the device.

87
Overview
Preliminary Acquaintance with Infrared
Thermal Imager
This product is an infrared camera that integrates surface temperature
measurement and real-time thermal image. The traditional inferred
thermometer needs to measure every component one by one while it is not
necessary for infrared imaging camera, thus saving time. The potential
problems may be displayed on color display screen clearly. Moreover, the
central point measurement cursor is used to locate quickly and accurately
to measure the temperature of the target object.
In order to increase the differentiation, the product is provided with a visible
light camera. The thermal images and visible images are stored in the
device and can be read through USB or stored in a computer to generate a
report or for print.
With small volume, the product is easy to operate and has strong function.
it is the ideal selection for electric power, electronic manufacturing,
Industrial inspection and other fields.
The following major functions increase the product’s accuracy and usability:
The radiation coefficient may be adjusted to increase the measurement
accuracy of objects with half reflection surface.
The highest temperature and lowest temperature cursor may guide the
users to the areas with highest and lowest temperature of the thermal
images.
The selectable color palette.
Avoid that the product is impacted (such as collision and falling,
etc).
Such operation may cause the product to be damaged. Please avoid
such operation.
Long-time storage and regular charging.
The product should be placed at a cool and dry environment if it not
used for a long time. If the product installed with battery is stored for a
long time, charging should be made regularly. Otherwise, the battery
will run out and the service life will be shortened.
For a long time, infrared thermal imaging detection technology has become
an important means to ensure Industrial safety in developed countries. The
application field includes electric power, metallurgy, petro-chemistry,
machinery,coal, traffic, fire control and national defense, etc. It can not only
carries out real-time detection at high voltage, high current and high speed
operation to perform real-time detection for production and instrument but
also need not power off the power supply, stop the machine or stop the
production to find the potential problems and prevent occurrence of
malfunction. The “non contact” modern detection technology is safe,
reliable and quick. It is a type of technical revolution compared with
traditional contact detection method.
Personnel search and rescue in the fire, and fire command;
Analyze the location and heat loss of leak points in heat pipes and
heating equipment;
Determining the location of the heating fault of the operating train;
Security department night monitoring.
The infrared thermal imaging technology is widely used in the following
fields:
Power equipment, transmission, transformer line inspection;
Hidden fire source search in fire protection;

9 10
Cleaning of the Products Charging of the Battery and Description
The shell of this product is cleaned with a damp cloth or diluted soapy
water. Do not use abrasives, isopropyl alcohol or solvents to clean the lens
and screen. Use professional optical lens cleaners.
Lens Maintenance
Prevent damage of the infrared lens:
Use USB data line to charge:
To make the lithium ion battery can play the perfect performance:
Clean the lens:
Clean the infrared lens carefully. The lens is provided with refined
anti-reflection coating.
Don’t clean with force to prevent damage of the anti-reflection
coating.The selectable color palette.
Use a cleaning solution for lens maintenance, such as alcohol-based
commercial lens cleaners, alcohol, and a lint-free cloth or paper towel.
Compressed air tanks can be used to remove loose particles.
The product has built-in chargeable 18650 lithium batteries.
The compressed air tank or dry nitrogen ion gun (if applicable) may be
used to blow the loose particles on the lens surface.
Dip the lint free cloth in alcohol.
Squeeze the excessive alcohol in the cloth or apply the lint free cloth on
dry cloth lightly.
Wipe the lens surface by making circular motion. Then discard the cloth.
If it is necessary to repeat above step, please use new cloth to dip with
the cleaning solution to wipe.
Pull off the USB line after fully charging.
Don’t place the battery on the charger exceeding 24 hours.
The thermal imaging device should be charged for two hours at least
every three months so as to extend the battery service life to the
greatest extent.
Don’t try to charge the battery in extremely cold environment.
When the battery level is low, the top right of the screen will display
“ ”.Please charge in time through Micro USB interface(When
the product is off, you can charge).

1211
English、Chinese、Italian、German
Performance Index
Product size
Product weight
Rainbow, iron, cold color, white hot,
black hot
Field angle
Infrared detector type
Focus mode
Temperature
measurement range
Measurement accuracy
Temperature measurement
mode
Vanadium Oxide Uncooled Infrared
Focal Plane
35°×27°
Non-heating fixed focus lens
-20℃to +550℃(-4 ℉to 1022 ℉)
±2℃or ±2%
Center point/hot and cold spot tracking
Cell size
NETD
Lens focal length
Color palette
12μm
<50mK @ F#1.0
3.2mm
Infrared resolution
Infrared response band
256x192
Emissivity setting
Thermal imaging frame rate
Adjustable from 0.01 to 1.00
≤25Hz
Display size
2.8-inch(240×320)
Image display mode
Infrared/visible light/dual light fusion
8 to14μm
Device storage
Storage Image/Video Format
Image/video export method
Built-in 4G eMMC (user available
storage space is about 3G)
JPG/MP4
USB connection to computer export
Lighting system
LED fill light
Battery Type
Battery capacity
Working time
Removable Rechargeable Lithium
Battery (18650)
2000mAh
2 to 3 hours
Power interface
Micro USB
Power configuration
5 minutes / 20 minutes / no automatic
shutdown
Work temperature
0℃to +45℃
(32 ℉to 113 ℉)
Storage temperature
-20℃to +60℃
(-4℉to 140℉)
Relative humidity
10% to 85%RH (non-condensing)
389g
96mm×72mm×226mm
Image analysis function
Menu language
Support offline analysis on PC

13 14
Product Description
1. Instruction to Structure
Visible light camera
LED lighting lamp
Infrared imaging sensor
Image capture key
Display screen
Keypad
Handheld
Battery
Micro USB
Please open the cap in use
12

1615
2. Display Description
Initial Operation
Product boot/shutdown
Color code: Used to mark the color corresponding to the relative temperature
from low to high in the field of version.
Press and hold the “ MENU/ ” button for more than 3 seconds to
turn the Thermal Imaging on or off.
The central point temperature cursor: Used to indicate the central position in
the screen area. The cursor color displays white. The temperature value is
displayed top left corner of the screen.
The lowest temperature cursor: Used to indicate the lowest temperature
position in the screen area. It will move with the movement of the lowest
temperature. The cursor displays green. The temperature value is displayed
at the central position of the screen.
LCD screen display
After turning on the power, the screen shows the thermal imaging
status.
Note:
Time adjustment may be required when you move the camera
between environments with widely varying ambient temperatures.
The highest temperature cursor: Used to indicate the highest temperature
position in the screen area. It will move with the movement of the highest
temperature. The cursor displays red. The temperature value is displayed
at bottom left corner of the screen.
Maximum value/
minimum value of
field temperature
Color code
Current emissivity
Battery power
The highest
temperature
cursor Temperature
cursor at
central point
Time
Central point
temperature
The lowest
temperature
cursor
3. Key Description
Menu key/On/off key
Navigation key: Up, down, left and right
Selection key/Entry key
Temperature range Wifi
L
H
Temperature range: The range of temperature measurement.

17 18
The function to hide highest/lowest temperature column
at the screen bottom
under the operation after normal start up, press “ ” key and the
screen bottom will display highest/lowest temperature column.
Press “ ” can also hide it.
Witching between infrared thermal image and visible image
press “ ” or “ ” key to switch the degree of fusion between
inferred thermal images and visible images (the degree of fusion is
0%, 25%, 50%, 75% and 100%).
Image capture
press the image capture key. When the capture is successful, the
screen will display “store photo?” prompt. If “yes” is selected,
please press “MENU/ ” key to determine to save the image. If
“no” is selected, press “ SELECT/ENTER ” key to determine not to
save the image.
Image output
The saved images through capture may be checked and output
by connecting with a computer through Micro USB.
Read images
Open the USB protective cover as shown in . Use USB line to
connect the USB port and then computer to read the images or
save it into computers.images or save it into computers.
The supported operating system through verification includes:
winxp, win7, win 8, win10, Apple system.
It is suggested to use the attached USB line or USB line with
higher quality.
Note:
When connecting with a computer, pull off the data line after
selecting “pop out device safely” to avoid causing file system
damage and other problems. If “unable to save” and other
problems occur, you may find the hard disc in the computer and
fix it.
2
Video recording
In the normal startup and running interface, long press the capture
button, the screen will display the prompt “record video?”, if you
want to select “Yes” at this time, please press the “MENU/ ”
button or the capture button to start recording. If you want to select
“No”, press the “SELECT/ENTER” key to cancel the recording.
When the recording is finished, press and hold the capture button
again to end the recording.
On/off of LED lighting lamp
Hold down the “image capture” key for about 3 seconds to power
on/off the LED lighting lamp.

2019
1.“Image Overlapping” Submenu
2. Introduction to “image” Sub-menu
1.1. Description of Image Overlapping
2.1. View Image
2.2. Delete Images
Introduction to the Menu
As shown in the figure, Then press “ ” key to enter image list. Press
“ ”or “ ” key in navigation to select the image. Then press
“
SELECT/ENTER
” key to view image.
When viewing the images, press “ ” key to view the precious image,
Press “ ” to view the next image.
Press “
SELECT/ENTER
” key to return. Press “
MENU/
” key to exit
from the menu.
When viewing the images, the screen will show the prompt of “Delete
photo?” by pressing “ ” key. If “yes” is selected at the moment,
press “
MENU/
” key to determine to delete the image. If “no” is
selected, press “
SELECT/ENTER
” key to determine not to delete the
image.
Press the left of “
MENU/
” key and the menu bar appears. They are
“image overlapping”, “image”, “video”,“color palette”, “emissivity” and
“setting” submenus. Press the “
MENU/
” button to enter the main menu, and select
“ ” (image) in the main menu.
Image overlapping makes it easier for users to understand the infrared
images by using aligned visible images and infrared images. The use of
image overlapping can capture the visible image of every infrared image
so as to display the temperature distribution in the target region correctly
and share with other people more effectively.
1.2. Application of Image overlapping
Press the “
MENU/
” button to enter the main menu, and select “ ”
(Image registration) in the main menu.
Press the “
SELECT/ENTER
” button to enter the image overlap
adjustment mode.Press the navigation keys (up, down, left and right
buttons) to perform the visible image shift operation.
Press the “
SELECT/ENTER
” button to exit the image blending mode
(Note: If there is no operation for more than 6 seconds, the image blending
mode will be automatically exited).
171229-105030
171229-105031
171229-105032
171229-105033

21 22
4. Introduction to “Color Palette” Sub-menu
4.1. Color Palette Description
The following is the image of the same object with selection of different
color palettes.
The palette can be used to change the pseudo-color display of the
infrared image on the display. Some tune Swatches are better
suited for specific applications and can be set up as needed. The
palette is divided into: rainbow, iron,
cold color
, white hot, black hot,
five palettes. These palettes work best with high thermal contrast
and provide additional color contrast between high and low
temperatures.
Suitable selection of color palette displays the details of the target
objective better. Rainbow, iron and cold color palettes focus on
display of color. Such color palettes are very suitable for high heat
contrast and are used to improve the color contrast between high
temperature and low temperature. But the
White hot
and
Black hot
color palettes provide even linear color.
IronRainbow Cold color White hot Black hot
3. Introduction to “Video” Submenu
3.1.View Video
Press “
MENU/
” key to enter the main menu, select “ ” in the
main menu(video), as shown below.
3.2. Delete Video
Press the “ ” key to enter the video list, press the “ ” and “ ”
keys in the navigation keys to select the video, and then press the
“ SELECT/ENTER ” key to view the video. When viewing a video (not
playing), press the “ ” key to view the previous video, and press the
“ ” key to view the next video.
Press the “
MENU
/
” key to exit from the menu.
When viewing the video (if it is not playing), press the “ ”key, and
the prompt “Delete video?” will appear on the screen. If you want to
select “Yes” at this time, please press the “
MENU/
” key to confirm
the deletion of the video. If you want to select “No”, press the
“SELECT/ENTER” key to make sure the video is not deleted.
20220211-154307
20220211-124307
20220210-174307
20220210-164307

2423
4.2. Application of Color Palette
As shown in the figure, press “
MENU/
” key to enter the main menu
and select “ ” (color palette) option and press “ ” key to enter
the color palette list. Press “ ” and “ ” keys in navigation to
select the color palette. Then press “
SELECT/ENTER
” key to select
the color palette. Press “ ” to return. Press “
MENU/
” key to exit
from the menu.
Spectra
Iron
Cool
White
Black
Color palette
5. Introduction to “Emissivity” Sub-menu
5.1. Emissivity Description
The emissivity of the product can be adjusted from 0.01 to 1.00 with the
default value of 0.95. Many common objects and materials (such as
timber, water, skin and textile fabric) can reflect the heat energy
effectively. So it is easy to obtain relatively correct measurement value.
The emissivity is usually set as 0.95 when the coarse objects that are
easy to give out energy. For semi-matte objects that give out less energy,
the emissivity is usually about 0.85 and the emissivity of semi-gloss
objects is 0.6. The shiny objects are divided into materials with low
radiation coefficient. The emissivity is usually set as 0.3 at the time of
measurement. Correct setting of the value of emissivity is very important
for you to carry out the most correct temperature measurement. The
surface emissivity will produce giant impact on surface temperature
measured by the product. Understanding the surface emissivity will
enable you to obtain correct temperature measurement result.

25 26
As shown in the figure, press “
MENU/
” key to enter the main menu
and select “ ”(emissivity) option and press “ ” key to enter
the emissivity list.
Press “ ” and “ ” keys in navigation key to select the
emissivity. Then press “
SELECT/ENTER
” key to determine selection
of the emissivity. Press “ ” key again to return.
5.3. The Emissivity Value of Common Materials
The operating step is as the following:
5.2. Emissivity Setting
The product is provided with four types of object measurement modes:
Coarse object (0.95)
Semi-matte object (0.85)
Semi-shiny object (0.60)
Shiny object (0.30)
According to the characteristics of the measured objects, users may set
the emissivity value through the “self-define” option (please refer to the
table of “emissivity of common materials”).
Thermal radiation
Black cloth
Human skin
Foam
Charcoal dust
Paint
Matte paint
Black rubber
Plastic
Timber
Paper
Chromium
hemitrioxide
Copper oxide
Ferric oxide
Textile
Substance Thermal radiation
Substance
Bitumen
Concrete
Cement
Sand
Earth
Water
Ice
Snow
Glass
Ceramics
Marble
Gypsum
Mortar
Brick
If you select "
self-defined
" emissivity, press the "
SELECT/ENTER
" button
to enter the editing state.Press “ ” / “ ”
keys
to select the number
to be
changed,
press “ ” “ ”
keys
to
change
the value. After
themodification is completed, press “
SELECT/ENTER
” to confirm, then
press“ ” to return. The “
MENU/
” button exits the menu.
Matte (e=0.95)
Emissivity
Semi-matte (e=0.85)
Semi-glossy (e=0.60)
Glossy (e=0.30)
Custom
0.01

2827
After selecting “ ” (unit), press the “ ” key in the navigation
key to enter the temperature unit setting. Can be set to Celsius or
Fahrenheit.
6.1. Auto Shutdown Setting
6.2. Intensity Settings
After entering the "Settings" sub-menu, select "
MENU/
" (automatic
shutdown), press navigationThe “ ” key in the button enters the auto
power off setting. Can be set to not automatically shut down or 5 points
the clock is turned off or turned off in 20 minutes.
After selecting “ ” (brightness), press the “ ” key in the
navigation key to enter the brightness setting. Can be set to low or medium
or bright.
6.3. Language Settings
After selecting " " (language), press the " " key in the
navigation key to enter the language setting.
Available in 4 languages: English, Chinese, Italian, German.
6.4. Unit Setting
6. Introduction to “Setting” Sub-menu
Press “
MENU/
” key to select the “ ”(setting) option in the main
menu. Press “ ” key again to enter the “setting” sub-menu.
After selecting
“ ”
(temperature range), press the
“ ”
key in the
navigation keys to enter the temperature range setting. Can be set to low
temperature (-20°C to 120°C) or high temperature (120°C to 550°C).
6.5. Tempertature Range Setting
L
H
WiFi
Settings
Auto shutdown Auto
shutdown
Intensity
Language
Unit
Time Format
Year
Month
Day
Hour
Minute
Second
Set time
Spot Off
On
24 hour
AM/PM
English
Chinese
Italian
German
Celsius
Fahrenheit
NO
5min
20min
Low
Medium
Hight
Intensity
Language
Unit
Temperature range Range
Time format
Set time
Spot
2020
10
26
02
52
03
low(-20℃~120℃)
high(120℃~550℃)
24h
L
H
WiFi
Off
On
SN:Thermal-0000

29 30
After you select " "(time format), press the “ ” key in the
navigation key to enter the time format setting. Can be set to 24 hours
or 12 hours.
6.7. Time Setting
6.8. Enable/Disable of the Highest and Lowest
Temperature Cursor
Press “ ”/“ ” to select year/month/day/hour/minute.
After selecting, press “
SELECT/ENTER
” key to enter the edit state.
Press“ ” and “ ” keys to select the figure to be changed.
Press“ ” /“ ” key to change the value. After completing the
change, press “
SELECT/ENTER
” to enter.
After the time setting is completed, press “ ” key to return. Press
“
MENU/
” exit from the menu.
As shown in the figure, after selecting “ ” (cold hotspot), press the
“ ” button in the navigation key to enter the cold hotspot setting.
As shown in the figure, after selecting “ ” (set time), press “ ”
in the navigation key to enter the setting time.
Press “ ” /“ ” key to select “enable” or “disable” the option.
Then press “
SELECT/ENTER
” key to determine selection.
After the setting is completed,press “ ” key to return. Press
“
MENU/
” key to exit from the menu.
6.6. Time Format Setting
24h
Year
Month
Day
Hour
Minute
2016
9
9
11
33
Second
33
Set time
Spot
Off
On
24h
L
H
24h
L
H

3231
6.9. Wireless WIFI on and off
As shown in the figure, after selecting “ ” (WiFi), press the “ ”
button in the navigation key to enter the WIFI setting.
Press “ ” /“ ” key to select “ on ” or “ off ” the option.
Then press “
SELECT/ENTER
” key to determine selection.
After the setting is completed,press “ ” key to return. Press
“
MENU/
” key to exit from the menu.
WiFi
Off
On
L
H
WIFI
SN:Thermal-0000
After turning on WIFI, use the mobile device to search for WIFI, the WIFI
name can be found in the local menu (as shown above); access the
device through the IP address 192.168.230.1.
In the mobile device, you can set the color palette, fusion, emissivity,
maximum/minimum value, language conversion, and then it can be
synchronized to the machine, and the settings on the machine can also be
synchronized to the mobile device (Note: If not converted, the refresh on
the device).
Click to take a photo on the mobile device, and the image will be saved on
the machine synchronously, as shown in the figure below.
Image view: click "Image" in the mobile device to enter the image list, and
click "Infrared video streaming" to return to the video streaming page.
24h

33 34
Thermal Imaging Image Analysis Software
Operation Guide
1. Connect the product with a USB cable, and then connect to the computer.
Open the removable disk in the computer, find the "INSTALLATION
PACKAGE" folder and open it, and click to install the software.
Software Installation And Operation
2. After installation, open Enter the picture analysis software
interface,click "Open File", and select the picture to be analyzed,as shown
in the figure:
1. Select the language in the upper right corner of the interface, you can
choose Chinese, English, and switch between the two languages.
2. On the right are shown as: fusion rate, color palette, temperature unit,
picture information.
3. "Open File" in the upper left corner of the interface, open the thermal
image to be read. When you need to save the picture after the operation,
click "Save"(Note: the saved picture cannot be opened again with this
software).
Fusion rate: Moving the slider can realize the fusion of visible light image
and infrared thermal image. The left end is a visible light image, and the
right end is an infrared thermal image.
Palette: In the palette, click the “ ” icon to choose from five color
palettes: rainbow, iron red, cold, white hot, black hot.
Temperature unit: In the temperature unit, click the “ ” icon to select
three temperature units: Celsius, Fahrenheit, and Kelvin.
Picture information: display file name, emissivity, capture time, maximum
temperature, minimum temperature, center point temperature, original
thermal image.
Interface Introduction

35
Icon Introduction
1. " " icon, take the temperature of the thermal image.
2. " "icon, select the maximum temperature and minimum temperature
of the thermal image in a straight line.
3. " "icon, select the thermal imaging map to select the maximum
temperature and minimum temperature.
4. " " icon, circle the thermal imaging map to get the maximum
temperature and the minimum temperature.
5. " "icon, arbitrarily circle the thermal imaging map to take the
maximum temperature and the minimum temperature.
6. " "Return icon, when an operation error occurs, you can click this
icon to return to the previous operation.
7. " "Delete icon, when you select multiple temperature values and
need to cancel all of them, click Delete.