HTI HT-18 User manual

Instruction Manual
Thermal Imaging Camera
MENU SELECTENTER
User Manual Version 1.0. Revised 6/11/2019.

Contents
6
7
8
10
12
16
19
20
8
9
12
17
19
19
20
21
21
21
15
19
1
Safety instruction
Preliminary acquaintance with infrared thermal imager
Overview
Cleaning of the products
Lens maintenance
Charging of the battery and description
Performance index
Product description
1.Instruction to structure
2. Display description
3. Key description
Initial operation
Introduction to the menu
1.“Image overlapping” submenu
1.1 Description of image overlapping
1.2 Application of Image overlapping
2. Introduction to “image” sub-menu
2.1 View image
2.2 Delete images
3. Introduction to “color palette” sub-menu
3.1 Color palette description
22
23
24
25
26
27
27
29
27
27
27
29
28
30
23
3.2 Application of color palette
4. Introduction to “emissivity” sub-menu
4.1 Emissivity description
4.2 Emissivity setting
4.3 The emissivity value of common materials
5. Introduction to “setting” sub-menu
5.1 Auto shutdown setting
5.2 Intensity settings
5.3 Language settings
5.4 Unit setting
5.5 Time format setting
5.6 Time setting
5.7 Enable/disable of the highest and lowest temperature cursor
Troubleshooting
Standard configuration list

Safety instruction
Considerations
Please ensure that you have read and understood the safety precautions
described in the following text before using the product so that you can
operate it correctly.
The safety precautions described in the following text guides you to operate
the product and its accessories correctly and safely to avoid damage and
loss to you,other people and the device.
The warning describes the acts that may cause harm to users. Please
observe the following operation in order to avoid electric shock or personal
injury.
If the product case is damaged, don’t use it continuously.
Please contact the local distributors or agents of Xintai under such
condition.
Warning: Don’t aim the laser pointer at human or
animal's eyes. The laser given out by the laser
pointer may harm the vision.
Please observe the following guide in order not to damage the product:
Don’t assemble or dismantle the product without permission.
The product is a type of super precise equipment. Don’t try to dismantle,
assemble or remodel any part of the product. Repair of the product should
be made by technical personnel designated by Xintai Company.
Avoid damage to probe of the product
Note: don’t put the product close to strong heat source directly (such as
electric iron). Otherwise, it may damage the probe of the product.
Product hum
When the product works, there is light sound of clicking every several
seconds. This is normal phenomenon that the lens captures images.
Warning
During the process of use, if the product is found with smoke,
spark and the burnt smell, please stop using at once.
In case of such condition, the product power supply should be powered off
first.After the smoke and peculiar smell disappears completely, please
contact the local distributors or agents of Xintai.
Don’t remodel the adapter and the data line.
Otherwise, such remodeling may cause short circuit or fire.
Don’t use abradant, isopropanol or solvent to clean the enclosure
of the device.
Such operation may cause the product case to be damaged.
1 2

Don’t weld the battery without permission.
Such operation may damage the battery and result in leakage and
explosion of the battery.
Avoid that the battery is impacted (such as collusion and
falling, etc).
Such condition may damage the battery case or result in that the battery
leaks or explode.
Please pull off the adapter on the power socket when charging is
not made.
The adapter may get overheated if it is connected with power supply for a
long time. This may cause overheating, deformation and fire.
When the plug of the adapter or wire is damaged, stop using
immediately.
Don’t change the battery unless the plug of the charger is inserted into the
socket completely.
Please ensure that the plug of the adapter is inserted into the
specified power socket.
The adapter plug may be different due to the regions. Please make
confirmation whether the specification of the adapter is consist with the
specification of electric appliances in your region. Otherwise, this may
cause overheating of the equipment, electric shock, fire, chemical leakage
inside the battery, explosion and other serious consequences.
Do not touch the electric wire with wet hands.
It is possible to cause electric shock when touching the electric wire with
wet hands. When pulling out the electric wire, hold the electric wire head
rightly to pull out the wire. Don’t pull off the electric wire directly. Otherwise,
the electric wire may be broken, causing electric shock and fire.
Forbid dip the product into water caught in rain.
If the case contact with any liquid, please wipe it dry immediately. If water
or other liquid enters the inside of the instrument, please power off the
power supply at once. Continuous use may result in product damage.
Clean the dust on the adapter plug and the data line.
When it is exposed to dusty and dump environment for a long time, the dirt
surrounding the electric equipment will accumulate moisture. This may
cause short circuit and fire.
Please use the original adapter of the company to charge the
product.
The use of non-original power supply accessories may cause equipment
overheating, electric shot, fire, and leakage of chemicals inside the battery,
explosion and other serious consequences.
3 4

Preliminary acquaintance with infrared
thermal imager
The product temperature may increase after a long time of
charging.
You may feel scorching heat when the hands touch the sensors.
Avoid that the product is impacted (such as collision and falling,
etc).
Such operation may cause the product to be damaged. Please avoid such
operation.
Long-time storage and regular charging.
The product should be placed at a cool and dry environment if it not used
for a long time. If the product installed with battery is stored for a long time,
charging should be made regularly. Otherwise, the battery will run out and
the service life will be shortened.
Problem caused by water condensation
Don’t bring the instrument into low temperature environment from high
tempera-ture environment in a short time or from low temperature
environment to high temperature environment. This may result in that the
inside of the instrument and appearance produce water condensation. In
such condition, the instrument should be placed into portable box or plastic
bag. Before use, make it restore to the environmental temperature and take
it out to use.
If the product has water condensation inside, please power it off
immediately.Otherwise, the instrument may be damaged. Operation is not
allowed unless the water condensation disappears.
For a long time, infrared thermal imaging detection technology has become
an important means to ensure Industrial safety in developed countries. The
application field includes electric power, metallurgy, petro-chemistry,
machinery,coal, traffic, fire control and national defense, etc. It can not only
carries out real-time detection at high voltage, high current and high speed
operation to perform real-time detection for production and instrument but
also need not power off the power supply, stop the machine or stop the
production to find the potential problems and prevent occurrence of
malfunction. The “non contact” modern detection technology is safe,
reliable and quick. It is a type of technical revolution compared with
traditional contact detection method.
The infrared thermal imaging technology is widely used in the following
fields:
The inspection of electric equipment, lines of transmission and
transformers;
Search of concealed fire source in fire control;
Personnel search and rescue and commanding at fire site;
Make analysis of the leakage position and heat loss of thermal pipeline
and heating equipment;
Determine the position of heating failure of the operation train;
Analysis of rationality of wire of microelectronic industry;
Night monitoring of security departments.
5 6

Overview
This product is an infrared camera that integrates surface temperature
measurement and real-time thermal image. The traditional inferred
thermometer needs to measure every component one by one while it is not
necessary for infrared imaging camera, thus saving time. The potential
problems may be displayed on color display screen clearly. Moreover, the
central point measurement cursor is used to locate quickly and accurately
to measure the temperature of the target object.
In order to increase the differentiation, the product is provided with a visible
light camera. The thermal images and visible images are stored in the
device and can be read through USB or stored in a computer to generate a
report or for print.
With small volume, the product is easy to operate and has strong function.
it is the ideal selection for electric power, electronic manufacturing,
Industrial inspection and other fields.
Cleaning of the products
Please use damp cloth or weak soap to clean the enclosure of the device.
Don’t use abradant, isopropanol or solvent to clean. The lens and screen
should be cleaned with use of cleaning agents for professional optical
glasses.
Lens maintenance
Prevent damage of the infrared lens:
Clean the lens:
The following major functions increase the product’s accuracy and usability:
The radiation coefficient may be adjusted to increase the measurement
accuracy of objects with half reflection surface.
The highest temperature and lowest temperature cursor may guide the
users to the areas with highest and lowest temperature of the thermal
images.
The selectable color palette.
Clean the infrared lens carefully. The lens is provided with refined
anti-reflection coating.
Don’t clean with force to prevent damage of the anti-reflection
coating.The selectable color palette.
Please use cleaning solution for lens maintenance such as alcohol-con-
taining commercial lens cleaning solution, alcohol or isopropanol as well
as a piece of lint free cloth or paper tissue. The compressed air tank
may be used to remove the loose particles.
The compressed air tank or dry nitrogen ion gun (if applicable) may be
used to blow the loose particles on the lens surface.
Dip the lint free cloth in alcohol.
Squeeze the excessive alcohol in the cloth or apply the lint free cloth on
dry cloth lightly.
Wipe the lens surface by making circular motion. Then discard the cloth.
If it is necessary to repeat above step, please use new cloth to dip with
the cleaning solution to wipe.
7 8

Charging of the battery and description
Performance index
Use USB data line to charge:
To make the lithium ion battery can play the perfect performance:
The product has built-in chargeable 18650 lithium batteries.
Pull off the USB line after fully charging.
Don’t place the battery on the charger exceeding 24 hours.
The thermal imaging device should be charged for two hours at least
every three months so as to extend the battery service life to the
greatest extent.
Don’t try to charge the battery in extremely cold environment.
The icon “ ” displays when the battery is fully charged.
The top right of the screen will display “ ” at the time of charging.
When the battery level is low, the top right of the screen will display
“ ”.Please charge in time through Micro USB interface.
9 10
Infrared image resolution
visible image resolution
LCD resolution
Field angle
shortest focus length
Thermal sensitivity
Temperature
measurement range
Measurement accuracy
Emissivity
Frame rate of
thermal images
Wavelength coverage
Focus mode
Color palette
Storage capacity
300,000 pixel
-20℃to 300℃(-4°F to 572°F)
Adjustable from 0.01 to 1.00
8-14um
Fixed
Rainbow, iron oxide red, cold color,
black & white, white & black
Built-in 3G (above 20 thousand
image stored)
Display screen 3.2 ″ full angle TFT
display screen
9Hz
±2℃/±2%
56°×42°
0.07℃
320×240
320×240
0.15M

Product description
1. Instruction to structure
English、Chinese、Italian、German
Display screen
keypad
handheld
File format
Automatic power-off time
Power supply
Product size
Product weight
Work temperature
Storage temperature
Relative humidity
Working time
Built-in chargeable 18650 battery
Selectable: 5 minutes/20 minutes/
not power off automatically
90mm×105mm×223mm
389g
0℃to 45℃
-20℃to 60℃
< 85%RH
2-3 hours
JPG
Setting command
Language,
Unit, language, date, time, information
USB
Micro USB 2.0
1
11 12

Visible light camera
Infrared imaging sensor
Image capture key
battery
cap
Micro USB
Please open the cap in use
Please open the cap in use
2 3
13 14

2. Display description
Maximum value/
minimum value of
field temperature
3. Key Description
On/off key Menu key Entry keySelection key
Color code
Current emissivity
Battery power
The highest
temperature
cursor
Temperature
cursor at
central point
Time
Central point
temperature
The lowest
temperature
cursor
1. Color code: used to mark the color corresponding to the relative tempera-
ture from low to high in the field of version.
2. The central point temperature cursor: used to indicate the central position
in the screen area. The cursor color displays white. The temperature value is
displayed top left corner of the screen.
4. The lowest temperature cursor: used to indicate the lowest temperature
position in the screen area. It will move with the movement of the lowest
temperature. The cursor displays red. The temperature value is displayed at
the central position of the screen.
MENU SELECTENTER
Up key
right key
left key
down key
39.1℃
16.5℃
52.3℃
4
15 16

Initial operation
Product boot/shutdown
Press and hold the “ ” button for more than 3 seconds to turn
the Thermal Imaging on or off.
The function to hide highest/lowest temperature column
at the screen bottom
under the operation after normal start up, press “ ” key and
the screen bottom will display highest/lowest temperature column.
Press “ ” can also hide it.
LCD screen display
After turning on the power, open the lens protection cover, as
shown in , the screen shows the thermal imaging status.
Witching between infrared thermal image and visible image
press “ ” or “ ” key to switch the degree of fusion between
inferred thermal images and visible images (the degree of fusion is
0%, 25%, 50%, 75% and 100%).
Note:
Time adjustment may be required when you move the camera
between environments with widely varying ambient temperatures.
Image capture
press the image capture key. When the capture is successful, the
screen will display “store photo?” prompt. If “yes” is selected,
please press “ ” key to determine to save the image. If “no” is
selected, press “ ” key to determine not to save the image.
Image output
The saved images through capture may be checked and output
by connecting with a computer through Micro USB.
Read images
Open the USB protective cover as shown in . Use USB line to
connect the USB port and then computer to read the images or
save it into computers.images or save it into computers.
The supported operating system through verification includes:
winxp, win7, win 8, win10, Apple system.
It is suggested to use the attached USB line or USB line with
higher quality.
Note:
When connecting with a computer, pull off the data line after
selecting “pop out device safely” to avoid causing file system
damage and other problems. If “unable to save” and other
problems occur, you may find the hard disc in the computer and
fix it.
MENU
SELECT
2
3
17 18

1.“Image overlapping” submenu
2. Introduction to “image” sub-menu
1.1. Description of image overlapping
2.1. View image
Introduction to the menu
As shown in the figure, Then press “ ” key to enter image list.
Press “ ”or “ ” key in navigation to select the image. Then
press “ ” key to view image.
When viewing the images, press “ ” key to view the precious
image, Press “ ” to view the next image.
Press “ ” key to return. Press “ ” key to exit from the menu.
Press the left of “ ” key and the menu bar appears. They are “image
overlapping”, “image”, “color palette”, “emissivity” and “setting” submenus.
Press the “ ” button to enter the main menu, and select “ ”
(image) in the main menu.
Image overlapping makes it easier for users to understand the infrared
images by using aligned visible images and infrared images. The use
of image overlapping can capture the visible image of every infrared
image so as to display the temperature distribution in the target region
correctly and share with other people more effectively.
1.2. Application of Image overlapping
Press the “ ” button to enter the main menu, and select “ ”
(Image registration) in the main menu.
Press the “ ” button to enter the image overlap adjustment
mode.Press the navigation keys (up, down, left and right buttons) to
perform the visible image shift operation.
Press the “ ” button to exit the image blending mode (Note: If there
is no operation for more than 6 seconds, the image blending mode will
be automatically exited).
171229-105030
171229-105031
171229-105032
171229-105033
MENU
MENU
MENU
ENTER
ENTER
SELECT
MENU
SELECT
19 20

3. Introduction to “color palette” sub-menu
2.2. Delete images
3.1. Color palette description 3.2. Application of color palette
When viewing the images, the screen will show the prompt of “Delete
photo?” by pressing “ ” key. If “yes” is selected at the moment,
press “ ” key to determine to delete the image. If “no” is
selected, press “ ” key to determine not to delete the image.
The following is the image of the same object with selection of different
color palettes.
As shown in the figure, press “ ” key to enter the main menu and
select “ ” (color palette) option and press “ ” key to enter the
color palette list. Press “ ” and “ ” keys in navigation to
select the color palette. Then press “ ” key to select the color
palette. Press “ ” to return. Press “ ” key to exit from the
menu.
Spectra
Iron
Cool
White
Black
Color palette
The palette can be used to change the pseudo-color display of the
infrared image on the display. Some tune
Swatches are better suited for specific applications and can be set
up as needed. The palette is divided into: rainbow, iron red, cool,
white hot, black hot, five palettes. These palettes work best with
high thermal contrast and provide additional color contrast between
high and low temperatures.
Suitable selection of color palette displays the details of the target
objective better. Rainbow, iron oxide red and cold color palettes
focus on display of color. Such color palettes are very suitable for
high heat contrast and are used to improve the color contrast
between high temperature and low temperature. But the black &
white and white & black color palettes provide even linear color.
MENU
MENU
SELECT
MENU
SELECT
Iron oxide redRainbow Cold color White heat Black heat
21 22

As shown in the figure, press “ ” key to enter the main menu
and select “ ”(emissivity) option and press “ ” key to enter
the emissivity list.
Press “ ” and “ ” keys in navigation key to select the
emissivity. Then press “ ” key to determine selection of the
emissivity. Press “ ” key again to return.
The operating step is as the following:
4.2. Emissivity setting
4. Introduction to “emissivity” sub-menu
4.1. Emissivity description
The emissivity of the product can be adjusted from 0.01 to 1.00 with the
default value of 0.95. Many common objects and materials (such as
timber, water, skin and textile fabric) can reflect the heat energy
effectively. So it is easy to obtain relatively correct measurement value.
The emissivity is usually set as 0.95 when the coarse objects that are
easy to give out energy. For semi-matte objects that give out less energy,
the emissivity is usually about 0.85 and the emissivity of semi-gloss
objects is 0.6. The shiny objects are divided into materials with low
radiation coefficient. The emissivity is usually set as 0.3 at the time of
measurement. Correct setting of the value of emissivity is very important
for you to carry out the most correct temperature measurement. The
surface emissivity will produce giant impact on surface temperature
measured by the product. Understanding the surface emissivity will
enable you to obtain correct temperature measurement result.
The product is provided with four types of object measurement modes:
Coarse object (0.95)
Semi-matte object (0.85)
Semi-shiny object (0.60)
Shiny object (0.30)
According to the characteristics of the measured objects, users may set
the emissivity value through the “self-define” option (please refer to the
table of “emissivity of common materials”).
Matt (e=0.95)
Emissivity
Semi-matt (e=0.85)
Semi-glossy (e=0.60)
Glossy (e=0.30)
Custom
0.01
MENU
SELECT
23 24

5. Introduction to “setting” sub-menu
Settings
Auto shutdown Auto
shutdown
Intensity
Language
Unit
Time Format
Year
Month
Day
Hour
Minute
Second
2017
12
28
15
15
15
Set time
Spot Off
On
24 hour
AM/PM
English
Chinese
Italian
German
Celsius
Fahrenheit
NO
5min
20min
Low
Medium
Hight
Intensity
Language
Unit
Time format
Set time
Spot
24h
4.3. The emissivity value of common materials
Thermal radiation
Black cloth
Human skin
Foam
Charcoal dust
Paint
Matte paint
Black rubber
Plastic
Timber
Paper
Chromium
hemitrioxide
Copper oxide
Ferric oxide
Textile
Substance Thermal radiation
Substance
Bitumen
Concrete
Cement
Sand
Earth
Water
Ice
Snow
Glass
Ceramics
Marble
Gypsum
Mortar
Brick
Press “ ” key to select the “ ”(setting) option in the main
menu. Press “ ” key again to enter the “setting” sub-menu.
If you select "
self-defined
" emissivity, press the " " button to enter
the editing state.Press “ ” / “ ”
keys
to select the number to be
changed,
press “ ” “ ”
keys
to
change
the value. After
themodification is completed, press “ ” to confirm, then press
“ ” to return. The “ ” button exits the menu.
ENTER
MENU
MENU
SELECT
25 26

After you select " "(time format), press the " " key in the
navigation key to enter the time format setting. Can be set to 24 hours
or 12 hours.
5.1. Auto shutdown setting
5.2. Intensity settings
5.6. Time setting
Press “ ”/“ ” to select year/month/day/hour/minute.
After selecting, press “ ” key to enter the edit state.
Press“ ” and “ ” keys to select the figure to be changed.
Press“ ” /“ ” key to change the value. After completing the
change, press “ ” to enter.
After the time setting is completed, press “ ” key to return.
Press “ ” exit from the menu.
After entering the "Settings" sub-menu, select " " (automatic
shutdown), press navigationThe “ ” button in the button enters the
auto power off setting. Can be set to not automatically shut down or 5
points The clock is turned off or turned off in 20 minutes.
As shown in the figure, after selecting “ ” (set time), press “ ”
in the navigation key to enter the setting time.
After selecting “ ” (brightness), press the “ ” button in the
navigation key to enter the brightness setting. Can be set to low or
medium or bright.
24h
Year
Month
Day
Hour
Minute
2016
9
9
11
33
Second
33
Set time
ENTER
MENU
SELECT
5.3. Language settings
After selecting " " (language), press the " " button in the
navigation key to enter the language setting.
Available in 4 languages: English, Chinese, Italian, German.
5.4. Unit setting
After selecting “ ” (unit), press the “ ” button in the navigation
key to enter the temperature unit setting. Can be set to Celsius or
Fahrenheit.
5.5. Time format setting
24h
27 28

Name Quantity
Thermal imaging device 1 set
USB connection line 1 piece
User manual 1 copy
Warranty card 1 copy
Certificate of Quality 1 copy
Adaptor 1 piece
Instrument bag 1 piece
Color box 1 piece
Troubleshooting
Standard configuration list
5.7. Enable/disable of the highest and lowest
temperature cursor
As shown in the figure, after selecting “ ” (cold hotspot), press the
“ ” button in the navigation key to enter the cold hotspot setting.
If you encounter with any problem when using the thermal image device,
please use the following table for repair.If the problems are not solved,
please cut off the power supply and contact with the manufacturer.
Press “ ” /“ ” key to select “enable” or “disable” the option.
Then press “ ” key to determine selection.
After the setting is completed,press “ ” key to return. Press
“ ” key to exit from the menu.
SELECT
MENU
Spot
Off
On
24h
Failure phenomenon
The thermal imaging
device cannot start
The thermal imaging
device powers off
automatically
No thermal images The lens cover is not
opened
Open the lens cover
Failure cause
The battery is not
installed
The power of the battery
is used up
Solution
Install the battery
Replace with new
battery or charge it
The power of the
battery is used up
Replace with new
battery or charge it
The time set for
automatic power off
is due
Restart or change the
time for automatic power
off after restarting
(refer to 5.1)
29 30
This manual suits for next models
1