HWM COMLog 2 IS User manual

1
Quick Start Guide: COMLog 2 IS
Installation (Using HWM IDT).
(Automatic Meter Reading models).
Warning: Always read this manual in conjunction with the Safety Supplement
which contains important safety and operating information.
Please read, understand and follow the instructions in the manuals.
Additionally, carefully read and follow the information in the
“Safety Warnings and Approvals Information” document.
MAN-157-0005-A November 2020

2
TABLE OF CONTENTS
1Prepare...............................................................................................................................3
2Choose Logger Model .......................................................................................................4
3Installation .........................................................................................................................5
4Activate Logger / Wireless Communications ..................................................................6
5Checking and adjusting logger configuration. ................................................................6
6Set Meter Reading of Flow Channel(s)...........................................................................11
7Add any additional Combos ...........................................................................................13
8Call settings and tests .....................................................................................................13
9Complete installation......................................................................................................14
Note(s):
1. Loggers will be set to call in with meter reading data and other messages to a
server. This could be one of the HWM central servers or to your company’s own
server (with HWM logger communication software installed).
2. Your logger may be pre-configured with a security setting requiring the use of
the HWM DataGate or FCS OmniColl central servers; these are referred to as
“secure loggers” and notes 5, 6 and 7 apply (see below).
Without this security setting, the loggers are referred to as “insecure loggers” and
notes 5, 6 and 7 do not apply (insecure loggers can be used without DataGate).
Alternatively, notes 5, 6 and 7 can be optionally applied (insecure loggers can be
optionally used with DataGate).
3. IDT will allow use without the app first logging in to the DataGate server (see link
at the app login page) but will only work with insecure loggers.
4. For additional information, refer to main product manual,
MAN-157-0002 “User Guide: COMLog2-IS / ISLog - Installation and Setup”.
5. Loggers must be pre-registered on DataGate.
6. User must be pre-registered on DataGate.
Contact your System Administrator (or your HWM Sales representative) to obtain
your server URL, username, and password information, etc.
7. Assumes DMAs are pre-registered on DataGate.

3
1PREPARE
1. Install required apps
Install “IDT” onto
a mobile phone
Links: IDT:
Google Play and the Google Play logo are trademarks of Google LLC.
2. Configure IDT app and Login
Launch IDT.
To edit or enter information on any screen, tap on a line.
Then view or change the editable information.
Tap on the “Save” or “Cancel” button when finished.
The “Devices” page appears.
(When IDT used with DataGate login):
Contact your supervisor or
administrator for the Server URL,
your username and your password.
Add Server URL.
Add Username.
Add Password.
Login.
✓
(When IDT used without DataGate login):
Tap on the link
“Log in without DataGate”.

4
2CHOOSE LOGGER MODEL
Several logger types are available. Examine the meter at the installation site and choose
the correct logger model (and adaptor cables) for the job.
Common Models and examples of use: (Note: ‘ * ’ indicates unspecified option)
HIS/F0K0Q/*/*/C - Flow (Uni directional, 1 channel) + Tamper alarm option.
1 x Flow pulse input channel (2-wire);
RJ11 plug (male).
(Configured as 1-wire Unidirectional Flow
+ option for Tamper Alarm).
1 x Flow pulse (replicated) output channels (2-wire);
RJ11 socket (female).
(Can be configured to give replicated output of input pulses).
Example of use: (1) Monitor meter flow pulses (with tamper-alarm option).
Signal and wire colour
details are shown
opposite for both
RJ11 plug (male) and
RJ11 Socket (female).
HIS/F084Q/*/*/C - Flow (Uni directional, 2 channels).
2 x Flow pulse input channels (1-wire);
RJ11 plugs (male).
2 x Flow pulse (replicated) output channels (1-wire);
RJ11 sockets (female).
Example of use: (1) Monitor of regular meter pulses and pulses from a volume
corrector (with replicated pulse outputs).
Signal and wire colour
details are shown
opposite for both
RJ11 plug (male) and
RJ11 Socket (female).
Key:
“LF Pulse” =
Pulse ch1 / Flow1
“Tamper” =
Pulse ch2 / Flow2
Key: (2 channels)
“LF Pulse” =
Pulse ch1 / Flow1
“LF Pulse” =
Pulse ch2 / Flow2

5
3INSTALLATION
1. Un-pack logger
Dispose of packaging
responsibly.
2. Install logger.
This product is only to be installed and connected by a fully ATEX trained installer.
Check on available space in the meter cabinet and decide where to position the
logger, antenna, and cables. Ensure there is access to the magnetic sensor (see 4).
Securely mount the logger with suitable fixings; keyhole mounting holes
are provided.
Note: The logger should be installed upright (pictured) to give best battery life.
If required, add anti-tamper seals between lid and base using holes provided.
3. Connect antenna to logger
Place the antenna in a suitable location.
Connect antenna to the logger.
Ensure antenna cable has no sharp bends.
4. Connect logger to meter output(s).
Connect the logger Flow1 input to the
meter output.
Use a suitable adapter cable if required.
Alternatively, where a volume corrector
is in place, Flow 2 can be used for
corrected meter flow.
Use a suitable adapter cable if required.
If required, activate pulse repeat outputs of the logger, and connect to any downstream
equipment (not shown).
Locate the cables so they cannot be damaged or cause a trip-hazard, and secure in position.

6
4ACTIVATE LOGGER /WIRELESS COMMUNICATIONS
Logger must be activated during installation.
Hold a strong magnet in the position shown.
Wait 12s, then remove the magnet.
•If logger Bluetooth-compatible communications
is in standby, it becomes active.
•Logger communications stays active for 120s,
waiting for IDT to communicate.
•Communication with phone remains open
for 600s after last use.
5CHECKING AND ADJUSTING LOGGER CONFIGURATION.
1. Read barcode / Bluetooth scan
Tap “Barcode Scan” icon.
Scan barcode of logger.
IDT connects to Logger.
If fail, try (4) again.
Alternative:
Tap “Bluetooth Scan”.
Select logger from list; IDT connects to logger.
Note: When using IDT app with DataGate login, insecure loggers may be hidden from lists.
There is a menu option to view both secure and insecure loggers together.

7
2. Setup Logger ID and time-zone
Tap
“Configure Device”.
Tap
“Device Information”.
Set local time-zone.
3. Display existing Channel settings (summary)
The required logger settings will vary according to the meter and other equipment used at
the site.
Tap the Channels icon.
(A “Combo” is a way of
setting a combination of
conditions that, together, can generate an alarm).
A summary of number of channels and combos
already configured is shown.
(The illustrations opposite show an example of a single
flow channel with Tamper alarm set).
Flow channels appear as Logging channels. A tamper alarm
uses a status input which appear as a Trigger channel.

8
4 Check / Edit logger Flow Channel 1 settings
Tap the Flow channel line to show its details.
This screen is to set the volume measurement
characteristics of the logger to match the meter’s pulse
outputs.
Note: These details can normally found within the
data-sheet for the meter, but typically are found
on the meter or pulse-head, as shown below.
The meter unit above is metric, so the logger
is required to be set up to use metric units (m3).
Check the “BASIC” tab channel settings match the meter, and if not adjust the settings.
e.g. Using the gas meter pulse characteristic shown above…
•The sensor type is set to Gas.
This matches the meter. ✓
•The unit of measurement shown is set to ft3, which is an imperial measurement.
- The channel requires adjusting to a metric measurement unit.
Tap on “Recording Unit” and set to m3.
•The “Units / Pulse” field now requires editing. The volume represented by the
meter-pulse needs to be scaled to the setting of the Recording Unit field.
“1 PULSE = 100dm3 “ corresponds to 0.1 m3of gas per
pulse.
The channel therefore requires adjustment of the
“Units/Pulse” to be 0.1 to match.
•For a Flow reading, the channel must be set to “Spot”
(the result is the pulse count since last measurement).
The edited screen is shown opposite.
(Settings are correct for this example).
If any settings were changed, tap on the “ACCEPT” button.

9
5 Check / Edit logger Flow Channel 1 Pulse Sample rate and Pulse replication settings
Tap on the “Advanced” tab.
Check the channel settings…
•The “Global pulse sample rate” should be set to
“Low Power” for most meters, for best battery life.
(Pulse width minimum 500ms, 50% duty cycle).
“High Speed” must be set for meters with high pulse
rates. (Pulse width maximum 10ms, 50% duty cycle).
Refer to the datasheet of the meter.
•If required, the “Replicate Channel” control should be active, as pictured.
(This setting ensures pulses from the Flow input are duplicated on the corresponding
output, allowing down-stream equipment to use the meter pulses).
If any settings were changed, tap on the “ACCEPT” button.
6. Check / Edit Flow Channel 2 settings
If the logger is dual channel, repeat the checks
and adjustments (steps 4 and 5)
for the second meter measurement
(e.g. from the volume corrector).
(Only required on dual-channel models.
For single-channel models, the option is
replaced by a Tamper Alarm option; see below).
7. Check / Edit logger Tamper Alarm settings
Some logger models can support a Tamper alarm to detect un-plugging of the logger from
the meter.
If the logger is single channel, and the meter supports a tamper alarm, then confirm this is
set up (see end of description). If not, follow the instructions below:
Tap on Combo:
No combos have been
set up.
Tap on the “ + “ line.
Ch1✓ Ch2 ?

10
Tap on “Tamper Alarm Combo” to set up the tamper alarm.
(This activates IDT to make several changes to logger settings in
order to produce a Tamper Alarm).
The tamper alarm settings are shown below (a) and (b):
(a)
A new channel is
created measuring the
open-circuit time of
input Status2 (/ Flow2).
The channel input is
tested every 2 seconds.
The channel is not
logged.
The tamper input is
not replicated to any
output channel due to
the “Replicate channel”
control being switched off (see diagram).
(With this setting is not available for use by other equipment).
(b)
There is an alarm condition set up
(via a “Combo” condition test).
The test checks the last 20 samples.
The OK condition (non-tamper) is for a closed-circuit to be
present; the time of an open-circuit condition is to be “0”.
The settings opposite create a trigger condition as soon as
a single measurement shows an open-circuit condition
exists.
The Action taken is to immediately report the alarm (to the
server). It is to report it as a tamper condition.
(Functionality and settings in the server determine if the
alarm is just recorded, or if it is actively forwarded to alert a
user).
When the triggering condition no longer exists, the logger will also send an “alarm cleared”
message. However, this will be sent when the unit next calls in to the server, not
immediately.

11
The channels summary shows
the presence of the newly
added channel and a combo
being set to trigger in the new
channel.
A summary of the
tamper alarm combo
is shown opposite
and can be used to
check if the combo
has already been set.
6SET METER READING OF FLOW CHANNEL(S)
1. Set logger with current Meter Reading.
The logger tracks the volume of gas passing through the meter (from the pulses) and sends
a meter reading into DataGate when it calls in with the flow measurements data.
Initially, the logger has not had the meter value set.
Tap on “Metering Settings”.
Find the units the meter is displaying. For the meter in the example we are
using, the dial clearly states “m3“(see diagram opposite).
The “Meter Factor” must next be determined and entered.
Meter factor =
In our example: The flow channel was set up in m3and the meter similarly measures in m3.
The ratio would be 1 m3/ 1 m3;
Set “Meter Factor” equal to “1”.
Example 2: If the flow channel measured in ft3, whilst
the meter measures in 100s of ft3,
the ratio would be 100 ft3/ 1 ft3;
Set “Meter Factor” equal to “100”.
Unit of volume measurement used on the Meter for each digit
Unit of volume measurement used on the Flow channel

12
Read the main meter, noting its format.
Enter the meter reading into the “Initial reading” field.
When entering the meter
reading, include any
leading zeros, the decimal
point and any other
displayed digits.
For the example shown,
enter “020339 . 82”.
Tap the back-arrow and the entered reading for Meter 1
is now shown. (This is not a live value).
You can confirm logger is set up correctly by doing the
following:
After waiting for the meter to change by a few digits,
note the meter reading and tap the “Meter Settings”
control.
Confirm the “Current value” reading matches the meter
reading (noted earlier).
If the reading is not correct, re-check:
The cables go to the correct meter.
The meter factor is correct.
The channel settings (units/pulse and recording unit) are correct.
… and then repeat the test.
Where two flow channels are used, the meter reading entry
page will be as shown opposite.
Make the settings (as above) for each meter.
Add the initial reading and Meter Factor for the second meter.
Then tap the back-arrow.
A meter reading is now
shown for both meters.
Confirm the readings of both “Current value” fields are
operating as expected, using the method described earlier.

13
7ADD ANY ADDITIONAL COMBOS
1. Add any required Combos.
If any additional alarms are required, they should be added by using a custom “Combo”.
(e.g. If flow rate exceeds rated flow of the meter).
8CALL SETTINGS AND TESTS
1. Set Call-in Settings
Tap
“Call Settings”.
Set your required
logger call-in details
Set your DataGate destination
Set network details for your SIM provider.
Note: These settings can (as an option) be
pre-configured by HWM to customer
requirements.
2. Call Test.
Check in final position with
door of meter cabinet (or lid on
chamber) closed.
Tap
“Call Test”.
Wait until complete.
Check “Call successful”.
Check CSQ is a good value.
(If not, check CSQ level with
antenna in various positions,
as described below)
Call-in daily at set times.
Save battery –Do not call if
latest data already sent OK.
Times at which to call with
data.

14
3. Check CSQ level (If required).
Tap
“Signal Test”
Close door of meter cabinet and check network CSQ level.
Summary of CSQ levels is shown below:
0-7 Poor.
(The logger may be able to register with network but will not be able to send
or receive data reliably).
8-14 Acceptable.
(Depending upon the ambient conditions data transmission may be possible.
It is important to select the correct antenna and install it in the most suitable
location).
15+ Good.
(Data transmission should be reliable).
If required, adjust the antenna position or type to get best results.
•Check result with meter cabinet door open, for comparison.
•Move position
•Adjust height
•Change antenna type.
(Where the meter cabinet is constructed of metal, it may be required to cut a hole
and fit a button-type antenna). –Contact HWM for antenna options.
Repeat the call test when optimal antenna position is found.
9COMPLETE INSTALLATION
1. Make a record of the logger deployment location (for office use).
2. Logger Deployment complete.
Leave site and go to the next site.
The logger communications
will go into standby.
3. Logger makes measurements and logs data.
Logger takes measurements at regular sample intervals and saves datapoints at specified
intervals.
Z
Z
Z

15
4. Logger calls in with measurement data (when used with DataGate).
Logger contacts server
at call-in time, and uploads
measurement data.
Logger can call server
immediately if an alarm condition
is detected and the combo settings require it.
(or other server)

16
HWM-Water Ltd
Ty Coch House
Llantarnam Park Way
Cwmbran
NP44 3AW
United Kingdom
+44 (0)1633 489479
www.hwmglobal.com
©HWM-Water Limited. This document is the property of HWM-Water Ltd. and must not be copied or
disclosed to a third party without the permission of the company. All images, text and designs are protected
by international and UK copywrite law and remain the property of HWM-Water. It is against the law to copy
or use any of the content from the HWM website or literature without the written consent of HWM-Water.
HWM-Water Ltd reserve the right to vary the specification.
MAN-157-0005-A November 2020.
Table of contents
Popular Measuring Instrument manuals by other brands

CUBETAPE
CUBETAPE C190POS quick start guide

PASCO
PASCO Pasport PS-2152 instruction manual

MONARCH INSTRUMENT
MONARCH INSTRUMENT Pocket Laser Tach 200 instruction manual

Siemens
Siemens Sicam P855 7KG85 Series Device manual

HT Italia
HT Italia HT950N user manual

OKOndt GROUP
OKOndt GROUP UTG-8 quick start guide

Ossila
Ossila FOUR-POINT PROBE user manual

Essilor Instruments
Essilor Instruments ALM 500 user guide

Emerson
Emerson AMS 2140 Pocket guide

SKC
SKC chek-mate 375 series operating instructions

Labom
Labom BX1008 Series operating instructions

Radians Innova
Radians Innova EnviroMentor Field Finder User instructions