Icel Manaus TR-5300 User manual

TR-5300
CLAMP-ON GROUND
RESISTANCE TESTER
MANUAL

Table of Contents
Ⅰ.Attention....................................................................................1
Ⅱ.Brief Introduction.........................................................................1
Ⅲ.Specification...............................................................................2
1. Model of Series............................................................................2
2. Ranges and Accuracy of Measurement...............................................2
3. Technical Specifications.................................................................3
4. Reference conditions.....................................................................3
5. Variations in the nominal working range.............................................4
Ⅳ.Structure of Meter.........................................................................4
Ⅴ.Crystal Display.............................................................................4
1. LCD Screen.................................................................................4
2. Description of Special Symbols........................................................5
3. Examples of displays.....................................................................5
Ⅵ.Operate the Meter.........................................................................6
1.Boot Up .....................................................................................6
2.Shutdown .................................................................................7
3.Resistance Measurement..............................................................7
4.Data Lock/Release/Store..............................................................7
5.Data review.................................................................................8
6.Configure alarm value....................................................................9
7.Review alarm setting......................................................................9
8.Delete Stored data.......................................................................10
Ⅶ.On site application......................................................................12
1. Application in telecommunication building........................................12
2. Application in transmitter tower......................................................12
3. Application in thunder-prevention grounding system of
4. Application in grounding system in gas station...................................12
Ⅷ. Important notes for earth resistance measuring................................13
1. Comparison of clamp method and tranditional voltage-current
measurement of earthing.................................................................13
2. Selection for measuring point........................................................14
Ⅸ. Bill of Loading...........................................................................15
I. Attention
Thank you for purchasing clamp on ground resistance tester TR-5300
In order to have better experience with this product, please
be certain:
---To read this user manual carefully.
---To comply with the operating cautions presented in this manual.
¨Under any circumstances, pay special attention to safety in the use of the Meter.
¨Do not exceed the measuring range and endenviroment provided.
¨Pay attention to text labeled on front panel and back plane of the meter.
¨Before booting up, trigger should be pressed for a couple of times to ensure jaws are
well closed.
¨Measurement can be performed o ' ' l inspenction innly when LCD shows OL after se
¨DO NOT OPERATE the trigger or clamp wires during boot-up stage.
¨DO NOT OPERATE the trigger or clamp wires during boot-up stage.
1
buildings...........12
method in

Model
TR-5300
Jaw specication(mm)
55x32
boot-up stage.
¨Keep jaw and jaw contact clean.
¨Pollution will effect accuracy of measurement.
¨Keep meter away from any impact, especially jaw contact planes.
¨It is normal that the meter has some sound buzzing during measurement.
Please remove batteries if long time without operation.
¨Disassembling, calibration and maintenance on the Meter shall be operated by the
authorized engineer.
¨In case of hazards generated itself caused by continuing use, stop using immediately,
send to authorized agencies for further operation.
II. Brief Introduction
Clamp on ground resistance tester is a major breakthrough in traditionalTR-5300
grounding resistance measurement. It is widely used in the grounding resistance
measurement of the power, telecommunications, meteorology, oilfield, construction
and the industrial and electrical equipment.
Clamp on ground resistance tester , in the measurement of a groundingTR-5300
system with loop, does not require breaking down the grounding wire, and need no
auxiliary electrode. It is safe, fast and simple.
Clamp on ground resistance tester is equipped with a long jaw sized 55mmXTR-5300
32mm, particularly suitable for the grounding with the flat steel.
III. Specification
1. Model of Series
2. Ranges and Accuracy of Measurement
Earth resistance
2
Range Resolution Accuracy
0.010-0.099Ω
0.10-0.99Ω
1.0-49.9Ω
50.0-99.5Ω
100-199Ω
0.001
0.01Ω
0.1Ω
0.5Ω
1Ω
( 1%+0.01 )±Ω
( 1%+0.01 )±Ω
( 1.5%+0.1±)Ω
( 2%+0.5±)Ω
(3%+1 )Ω

Conditions Reference
Ambient
Relative humidity
Battery voltage
External magnetic eld
External electric eld
Operating position
(20 3 )±℃
50%RH 10%±
6V 0.5V±
<40A/m
<1V/m
Clamp horizontal
3 . Technical Specifications
Power Source: 6VDC (4 X 5# alkaline battery)
Working Temperature: -10 to 55℃℃
Relative Humidity: 10%-90%
LCD: 4-digital LCD, 47 X 28.5mm in length
Span of Jaw: 28mm
Meter Quality (including batteries): 1320g
Meter Size: 286mm(L) X 84mm(W) X 56mm(H)
Protection Level: double insulation
Featured: by clamp
Range Shift: Automatic
External Magnetic Field: <40A/m
External Electric Field: <1V/m
Single Measuring Time: 1 second
Maximum resolution of resistance Measurement Resolution٠٫٠٠١ :Ω
Range of resistance Measurement Range١٠٠٠-٠٫٠١ :Ω
Number of record sets stored: 99 sets
Confingurable resistance alarm value: 1-199Ω
4. Reference conditions
3
200-395Ω5Ω( 6%+5±)Ω
400-590Ω
600-1000Ω
10Ω
20Ω
( 10%+10 )±Ω
( 20%+20±)Ω

Distirtion quantity Limit of operating range
Temperature
Relative humidity
Battery voltage
Conductor position
Clamp position
Proximity of
magnetic mass
Magnetic eld 50-60Hz
Electric eld 50-60Hz
-10 to 55℃
10%RH to 90%RH
5.5V to 6.5V
From edge to centre
±°180
1mm steel plate
against jaw face
400A/m
0-10KV/m
Distortion
1.5 class of
accuracy per 10°
1.5 class
0.25 class
0.1 class
0.5 class
0.25 class
0.25 class
0.25 class
5. Variations in the nominal working range
IV. Structure of Meter
1. Liquid Crystal Display (LCD)
2. Trigger: to open and close the jaw
3. Jaw: sized 55mm(L) x 32mm(W)
4. POWER Key: Boot Up / Shutdown /*Quit
5. HOLD Key: lock / Release display
6. MEN/READ:data store and review
V. Crystal Display
1. LCD Screen
(1) Alarm: Shows when the measured value exceed the configured alarm value.
4
Position of conductor in the clamp
Proximity to metallic the clamp
Loop resistance
Rate of distortion
Interferance current on measurement of
loop resistance
Centred
>10cm
Non choke resistance
<0.5%
Nil
Table of contents
Other Icel Manaus Test Equipment manuals
Popular Test Equipment manuals by other brands

Redtech
Redtech TRAILERteck T05 user manual

Venmar
Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions
Test Instrument Solutions SafetyPAT operating manual

Hanna Instruments
Hanna Instruments HI 38078 instruction manual

Kistler
Kistler 5495C Series instruction manual

Waygate Technologies
Waygate Technologies DM5E Basic quick start guide

StoneL
StoneL DeviceNet CK464002A manual

Seica
Seica RAPID 220 Site preparation guide

Kingfisher
Kingfisher KI7400 Series Training manual

Kurth Electronic
Kurth Electronic CCTS-03 operating manual

SMART
SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies
Agilent Technologies BERT Serial Getting started