IKA RC 2 basic User manual

20000004342
IKA®RC 2 basic
RC 2_112014
Reg.-No. 4343-01
Operating instructions EN

2
Fig. 1
1
4
5
12
12
Fig. 2
2
3
76
8
10
11
12
12
4
9
Item Designation
1
Operator panel and display
2 Mains switch
3 Power socket
4 Venting grid
5Filling lid
6Pump connection IN
7Pump connection OUT
8Overflow
9Backflow
10 RS 232 port
11 USB port
12 Handle

3
Contents
EN
Page
EC Declaration of Conformity....................................................................................................................4
Explication of warning symbols ................................................................................................................4
Safety instructions ....................................................................................................................................4
General information......................................................................................................................................... 4
Fluids ............................................................................................................................................................... 5
Correct use................................................................................................................................................5
Use.................................................................................................................................................................. 5
Range of use (indoor use only) ......................................................................................................................... 5
Unpacking.................................................................................................................................................6
Unpacking ....................................................................................................................................................... 6
Delivery scope.................................................................................................................................................. 6
Preparations..............................................................................................................................................6
Setting up........................................................................................................................................................ 6
Connecting the tubings.................................................................................................................................... 6
Filling and draining........................................................................................................................................... 7
Fluid (Standard information for IKA®fluid) ....................................................................................................... 8
Moving the device ........................................................................................................................................... 9
Operator panel and display .......................................................................................................................9
Commissioning........................................................................................................................................10
Menu settings .........................................................................................................................................11
Menu structure .............................................................................................................................................. 11
Basic guide to setting the menu options......................................................................................................... 11
Operating mode (MODE)................................................................................................................................ 11
Maximum temperature (HI T) ......................................................................................................................... 11
Minimum temperature (LO T) ......................................................................................................................... 11
Maximum speed (HI R)................................................................................................................................... 11
Minimum speed (LO R)................................................................................................................................... 11
Fluid type (FLUI) ............................................................................................................................................. 11
Temperature control type (AUTO) ................................................................................................................... 12
Alarm and key tone (BEEP)
.............................................................................................................................. 12
Calibration (CALI) ........................................................................................................................................... 12
Interface and output ...............................................................................................................................13
Maintenance and cleaning ......................................................................................................................15
Error codes..............................................................................................................................................16
Accessories..............................................................................................................................................16
Technical data .........................................................................................................................................17
Warranty .................................................................................................................................................18
Pump performance curve ........................................................................................................................18

4
Explication of warning symbols
Safety instructions
General information:
• Read the operating instructions completely before
starting up and follow the safety instructions.
• Keep the operating instructions in a place where it can be
accessed by everyone.
• Ensure that only trained staff work with the device.
• Follow the safety instructions, guidelines, occupational
health, safety and accident prevention regulations.
• Set up the device in a spacious area on an even, stable,
clean, non-slip, dry and reproof surface.
• Do not operate the device in explosive atmospheres,
with hazardous substances.
• Protect the device and accessories from bumping and
impacting.
• Check the device and accessories for damage before
each use. Do not use damaged components.
• Safe operation is only guaranteed with the accessories
described in the ”Accessories” section.
• The socket for the mains cord must be easily accessible.
• Socket must be earthed (protective ground contact).
• The voltage stated on the type plate must correspond to
the mains voltage.
• The device can only be disconnected from the mains sup-
ply by pulling out the mains plug or the connector plug.
• Disconnect the mains plug before attaching or changing
any accessories.
• Disconnect the mains plug before cleaning, maintenance
and transportation of the unit.
•
The device must only be opened by trained specialists,
even during repair. The device must be unplugged from
the power supply before opening. Live parts inside the
device may still be live for some time after unplugging
from the power supply.
Coverings or parts that can be re-
moved from the device must be
put back on the device again to
ensure safe operation, for example to keep foreign ob-
jects and liquids, etc. from getting into the device.
• The device may only be used as prescribed and as de-
scribed in these operating instructions. This includes op-
eration by instructed specialist personnel.
• When using critical or hazardous materials in your pro-
cesses, IKA®recommends to use additional appropriate
measures to ensure safety in the experiment. For ex-
ample, users can implement comprehensive monitoring
equipment.
• Process pathogenic material only in closed vessels under
a suitable fume hood. Please contact IKA®application
support if you have any question.
If the OFF switch is not within
reach when device is operating, an
EMERGENCY STOP switch that
can be easily accessed must be installed in the work area.
EC Declaration of Conformity
We declare under our sole responsibility that this product corresponds to the regulations 2006/95/EC, 2004/108/EC,
1999/5/EC and 2011/65/EU and conforms with the standards or standardized documents: EN 61010-1, -2-010, -2-051;
EN 61326-1 and DIN 12876-1.
DANGER
CAUTION
WARNING
NOTICE
indicates an imminently hazardous situation, which, if not avoided, will result in death, serious
injury.
indicates a potentially hazardous situation, which, if not avoided, can result in death, serious
injury.
indicates a potentially hazardous situation, which, if not avoided, can result in injury.
indicates practices which, if not avoided, can result in equipment damage.
NOTICE
WARNING

5
• A laboratory circulator refrigerates and circulates uid
according to specied parameters. This involves hazards
due to low temperatures and general hazards due to the
device of electrical energy. The user is largely protect-
ed by the device of relevant standards. Further hazard
sources may arise due to the type of tempering uid,
e.g. by exceeding or undercutting certain temperature
thresholds or by the breakage of the container and reac-
tion with the refrigeration carrier fluid. It is not possible
to consider all eventualities. They remain largely subject
to the judgment and responsibility of the operator. For
this reason, it may become necessary for user to take
other precautionary safety measures.
• Insucient ventilation may result in the formation of ex-
plosive mixtures. Only use the device in well ventilated
areas.
• Use suitable hoses for connection.
• Secure hoses and tubes against slippage and avoid kinks.
• Check hoses, tubes and bath at regular intervals for pos-
sible material fatigue (cracks/leaks).
• When device is used for external circulation, extra pre-
caution must be taken for cold fluid leakage due to
damaged hose.
Do not start up the device if:
- It is damaged or leaking
-
Cable (not only supply cable) is damaged.
• After a power failure during operation, the device may
start automatically (depending on operating mode).
• Transport the device with care.
• Do not transport or empty the bath while it is still cold.
Check the temperature of the bath fluid before draining.
• Always empty the bath before moving the unit.
Always empty the bath if the unit
is unused for long time.
Fluids:
Only use the uids, which fulll the
requirements for safety, health and
device compatibility. Be aware of
the chemical hazards that may be associated with the
bath fluid used. Observe all safety warning for the fluids.
• Depending on the bath uid used and the type of opera-
tion or toxic can arise. Ensure suitable extraction.
• Do not use any uid which may cause dangerous reac-
tions during processing.
• Only use recommended bath uid. Only use non-acid
and non corroding fluid.
Never operate the device without
sucient uid! Check the uid lev-
el detection at a regular basis (see
the section "Filling and draining").
• Continuous monitoring of the lling level of the bath uid
is required.
• For optimum temperature stability, the uids viscosity
should be 50 mm2/s or less at its lowest operating tem-
perature. This permits good fluid circulation and mini-
mizes heating from the pump.
• Untreated tap water is not recommended. It is recom-
mended to use distilled water or high purity water (ion
exchangers) and add 0.1 g soda (sodium carbonate
Na2CO3) /liter, to reduce corrosive properties.
Don't use following uids:
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, uorides, bromides,
iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.
DANGER
NOTICE
CAUTION
NOTICE
NOTICE
Correct use
• Use:
Use RC (Refrigerated Circulators) for cooling and circulat-
ing fluids.
Intended Use: Tabletop device
• Range of use (indoor use only):
- Laboratories -Schools
- Pharmacies - Universities
This instrument is suitable for use in all areas except:
- Residential areas
-
Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not
supplied or recommended by IKA®.
- If the device is operated improperly or in contrary to the
IKA® specications.
- If the device or the printed circuit board are modied by
third parties.

6
Unpacking
• Unpacking:
- Please unpack the device carefully
- In the case of any damage a detailed report must be
sent immediately (post, rail or forwarder).
Preparations
• Setting up:
- Place the unit on an even, stable, clean, nonslip, dry
and reproof surface.
- Keep at least 20 cm of open space on the front and
rear side.
- The place for installation should be large enough and
provide sucient air ventilation to ensure the room
does not warm up excessively because of the heat
from device radiates to the environment.
- Do not set up the device in the immediate vicinity of
heat sources and do not expose to sun light.
- Cooling machine, pump motor and electronics pro-
duce intrinsic heat that is dissipated via the venting
grids! Never cover these venting grids!
After setting up the device, wait at
least one hour before starting the
operation.
• Connecting the tubings:
- Unscrew the union nuts and stoppers using a wrench
from the pump connection IN (6,see Fig. 1) and OUT
(7,see Fig. 1).
External
system
IKA®
Calorimeter
Fig. 5
- Connect the hoses for circulating the external system to
the pump connectors M 16 x 1 for IN and OUT directly
or with the olives.
- Screw the hose olive to the pump connection IN and
OUT with union nuts. Slide the hoses (NW 12) onto the
olives. The hoses must be secured with suitable clamps.
- Via a hose to the “Overflow” connection, overowing
fluid can be directed into a suitable vessel. The vessel
should be positioned lower than the “Overflow” con-
nection.
- Connect the “Backflow” connector to IKA® calorim-
eter with a suitable hose.
Note: When the external system is not necessary, please
seal the pump connection IN and OUT with the existing
union nuts and stoppers.
Fig. 3
Union nut
Stopper
Fig. 4
Union nut
Olives
• Delivery scope:
- RC 2basic Refrigerated Circulators
- Mains cables
- Hose olive NW 12 (2 pieces)
- Operating instructions
- Warranty card.
NOTICE

7
Fig. 6
• Filling and draining:
- Before lling the uid into the bath, open the venting
grid as indicated in following gure.
- Check and make sure that the drain valve is closed (the
stop position in clockwise direction, see Fig. 7).
Note:
Please note information in section “Commis-
sioning”.
- Connect the mains plug and turn on the device with
mains switch (2
,see Fig. 1
).
- The low level warning message appear on the display.
Meanwhile, the bottom LED segment of the lights up
in red color to indicate the low fluid level
(see Fig. 8)
.
- Open the lling lid (5
,see Fig. 1
) and ll uid to the
bath.
Stop position
Drain valve
Drain port
Fig. 7
Note: Pay attention to the fluid
level information on the display and
fluid level indicator (see Fig. 8).
Bracket
- To drain the uid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver.
Fig. 9
Red (Low Level)
Red (High Level)
Low Level
High Level
Fig. 8
NOTICE

8
• Fluids (Standard information for IKA® uid):
No.
IKA®Designation
Operating temperature
range for open bath
application
(°C)
Operating temperature
range for closed bath
applications
(°C)
Safety temperature
(°C)
Flash point
(°C)
0
CF.EG28.N10.80.8
-10 ... 80 -10 … 80 80 115
1
CF.EG39.N20.80.16
-20 … 80 -20 … 80 80 115
2
CF.EG44.N25.80.19
-25 … 80 -25 … 80 80 115
3
CF.EG48.N30.80.22
-30 … 80 -30 … 80 80 115
4
UF.Si.N30.150.10LV
-30 … 130 -30 … 150 145>170
5
HF.Si.20.200.50
20 … 200 20 … 200 255 >280
6
HF.Si.20.250.50A
20 … 200 20 … 250 255 >280
7
Water
5 … 95 5 … 95 - -
8
Customized
The listed fluid temperature range may deviate from the device operating temperature ranges and settings.
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0.1 g soda (Na2CO3,
sodium carbonate) per liter of water.
Nomenclaturefor IKA® fluids:
CF.EG28.N10.80.8 --
(1) Classication:
HF: Heating Fluid
CF: Cooling Fluid
UF: Universal Fluid
(2) Chemical composition:
Si:Silicone oil
EG:Ethylene Glycol
(3) Temperature range: (
Minimum temperature. Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa.s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the uid divided by 1000.
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
(5) Addition information
(4) Viscosity (8
mm²/s
)
(3) Temperature range (-10 – 80 °C)
(2) Chemical composition (Ethylene Glycol 28%)
(1) Classication (Cooling fluid)

9
max. 10 ˚
• Moving the device:
Empty all fluid in the bath before moving device from
one place to other place.
The device can be lifted up and moved by using the top
handles or the bottom handles (12, see Fig. 1).
It can also be moved on flat surface by lifting and push-
ing the front of the device. It is easy to move the device
with the help of the wheels.
The angle of inclination should never be more than 10 º
at any direction when move the device!
The device must not be moved
during operation. After moving
the device, you must wait at least
one hour before restarting the unit.
Fig. 10
NOTICE
Operator panel and display
Item Designation Function
A ON/OFF button: Switch on/off the circulator.
B “Temp” button: Start/stop the
refrigerating
function.
C “Temp (+)” button: Increase the temperature setting.
Navigation, selecting the settings in the menu
.
D “Temp (-)” button: Decrease the temperature setting.
Navigation, selecting the settings in the menu.
E “OK / Pump” button: Start/stop the pump function.
Conrm the menu options.
F “Pump (+)” button: Increase the pump speed setting.
Change the menu value setting.
G “Pump (-)” button: Decrease the pump speed setting.
Change the menu value setting.
H “Menu” button: Press it once: menu option is display.
Press it once again: back to working screen.
I Key button:
Lock/unlock keys.
J LED, Key button: Indicate the keys are
deactivated.
K LED display:
Display the settings and actual values.
L Fluid level indicator: Indicate the uid level (when the uid level is too low or too high, the bottom or the
top LED segment change into red color).
M LED, cooling: The light LED indicate the
refrigerating
function is activated.
O LED, set: The LED lights up at the same time as the set value is displayed.
A
B FG
H I
ECD
K
L
JO M
Fig. 11

10
Before commissioning, make sure
that the device has not been moved
for one hour.
Check whether the voltage specied on the type plate
matches the mains voltage available.
The power socket used must be
earthed (protective earthing).
If above conditions are met, the instrument is ready for
operation after plugging in the mains plug.
If these conditions are not met, safe operation is not guar-
anteed and the device could be damaged.
Observe the ambient conditions (temperature, humidity,
etc.) listed under “Technical Data”.
After switching on the mains switch (2) at the back of the
device or pressing the ON/OFF key (A), all LED segments light
up during the self test.
Commissioning
After that, following two screens appear alternately.
Display field
Software version (ver)
Operating mode (mode)
Working setting
Then the device enters standby status and is ready for
operation.
Change the temperature setting with “Temp (+)“ button
(C) or “Temp (-)“ button (D).
Change the pump speed setting with “Pump (+)“ button
(F) or “Pump (-)“ button (G).
In standby status, activate the refrigerating function by
pressing the “Temp“ button (B). The pump and fan starts
to run at once. The refrigerating function may delay to
start at maximum 2 minutes.
In working status, press the “OK/Pump“ button (E) to
stop the pump function. The refrigerating function and
pump stops.
Note: In running status, the pump speed adjusts auto-
matically according to the load. When the actual speed
doesn’t reach the setting speed, the actual speed and set-
ting speed will appear on the screen alternately.
In standby status, press the “OK/Pump“ button (E) to
start the pump function. The refrigerating function will
not be activated.
In working status, press the “Temp“ button (B) to stop
the refrigerating function, the pump keep running.
The controls of the device allow to be locked by pressing
the key button (I), so no accidental changes during opera-
tion are possible (LED (J) lights up).
By pressing the key button (I) again, the controls are re-
leased (LED (J) no longer lights up).
NOTICE

11
Menu settings
• Maximum temperature (HIT):
The maximum adjustable value: 40 °C.
• Minimum temperature (LO T):
The minimum adjustable value: -20 °C.
•
Maximum speed (HI R):
The maximum adjustable value: 3200 rpm.
•
Minimum speed (LO R):
The minimum adjustable value: 1000 rpm.
•
Operating mode (MODE):
Mode A: After power on/power failure no automatic
restart of functions.
Mode B: After power on/power failure automatic re-
start of functions, depending on previous settings.
Mode
C:
Set values (set in Aor B) cannot be changed.
After power on/power failure automatic restart of func-
tions, depending on previous settings.
• Basic guide to setting the menu options:
F
Enter the menu by pressing ”Menu” key (H).
F
Press the
“Temp (+)“ button (C) or “Temp (-)“ button
(D)
to change the menu options.
F
Press the
“Pump (+)“ button (F) or “Pump (-)“ button
(G)
to change the menu value settings.
F
Conrm the menu settings by pressing the
“OK/Pump“
button (E)
.
MODE
Maximum temperature (HI T)
Minimum temperature (LO T)
Maximum speed (HI R)
Minimum speed (LO R)
Fluid type (FLUI)
The type of temperature PID control arithmetic (AUTO)
Alarm and key tone (BEEP)
Calibration (CALI)
Menu
• Menu structure:
Default settings
A
B
C
Proportional coecient of PID (Kp 1)
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)
AUTO 1
AUTO 0
BEEP 0
BEEP 1
CALI 0
CALI 2
CALI 3
Activated
-
-
40 °C
-20 °C
3200 rpm
1000 rpm
8
Activated
3.0
20.0
2.0
-
Activated
Activated
-
-
•Fluid type (FLUI):
The chosen fluid (No.) of the temperature setting is lim-
ited. See Section “Fluids (Standard information for
IKA® uid)”.

12
• Temperature control type (AUTO):
AUTO 1: AUTO 1 is the default setting. The default PID
settings are used automatically.
Cooling down curve in AUTO 1 control (AUTO 1):
Kp too high
°C t
Ti too high
°C t
Kp too low
Ti too low
°C t
Td too high
°C t
Fig. 13
optimum setting
°C
t
Proportional coecient of PID (Kp 1)
Integral time of PID (Ti 1)
Differential time of PID (Td 1)
Fig. 12
The proportional coecient Kp is the controller ampli-
cation and determines how strongly the control devia-
tion (the difference between the target temperature and
actual temperature) directly affects the control variable
(cooling down time). Kp values that are too large can
lead to the controller overshooting.
The integral time Ti (s) is the correction time and deter-
mines how strongly the duration of the control devia-
tion affects the control variable. Ti compensates an ex-
isting control deviation. A high Ti means a smaller and
slower effect on the control variable. Ti values that are
too small can lead to instability of the controller.
The differential time Td (s) is the derivative time and de-
termines how strongly the rate of change of the control
deviation affects the control variable. Td compensates
for rapid control deviations. A high Td means a smaller
and slower effect on the control variable. Td values that
are too small can lead to instability of the controller.
AUTO 0: For special requirements, AUTO 0 can be used
with manual adjustment of the control parameters.
If the PID control (AUTO 0) option is selected, the fol-
lowing items including Kp 1, Ti 1 and Td 1 could be
set. Otherwise they are not displayed in the menu list.
Note: Inappropriate settings may produce the follow-
ing cooling down curves:
•
Alarm and key tone (BEEP):
BEEP 0: no alarm and key tone
BEEP 1: alarm and key tone is activated
•
Calibration and adjustment (CALI):
CALI 0: Reset calibration
CALI 2: 2-point calibration
CALI 3: 3-point calibration

13
Set the rst point temperature (step 1, e.g. 10 °C) with
“Pump (+)” button (F) or “Pump (-)” button (G). Con-
rm the setting by pressing “OK/Pump“ button (E).
Left area of the display appears the set value (10 °C) and
the right area of the display appears the actual tempera-
ture value measured. The “Set” LED (O) flashes briefly.
The unit now starts and controls to the set value. When
the temperature has reached the set value and has be-
come constant, the “Set” LED (O) no longer flashes and
the following display appears. The temperature mea-
sured by the unit appears on the left hand side.
Input the calibration value from the reference measuring
instrument (e.g. 10.2 °C) with “Pump (+)” button (F) or
“Pump (-)” button (G).
Conrm the value by pressing the “OK/Pump“ button (E).
Back to previous screen for input a new value by press-
ing “Temp” button (B).
Back OK
The rst point calibration is nished now.
Calibration of the other points is performed in the same
way.
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring
instrument into the bath fluid.
Select 2-point calibration in the menu. Press the “OK/
Pump“ button (E) to start the 2-point calibration.
Interface and output
The device can be operated in “Remote” mode via the RS
232 interface (10, see Fig. 1) or the USB interface (11, see
Fig. 1) connected to a PC and with the laboratory soft-
ware Labworldsoft®.
Note: Please comply with the system requirements to-
gether with the operating instructions and help section
included with the software.
USB interface:
The Universal Serial Bus (USB) is a serial bus for connect-
ing the device to the PC. Equipped with USB devices can
be connected to a PC during operation (hot plugging).
Connected devices and their properties are automatically
recognized. The USB interface can also be used to update
rmware.
USB device drivers:
First, download the latest driver for IKA®devices with USB
interface from http://www.ika.com/ika/lws/download/
usb-driver.zip and install the driver by running the setup
le. Then connect the IKA®device through the USB data
cable to the PC.
The data communication is via a virtual COM port. Con-
guration, command syntax and commands of the virtual
COM ports are as described in RS 232 interface.
RS 232 interface:
Conguration
- The functions of the interface connections between the
stirrer machine and the automation system are chosen
from the signals specied in EIA standard RS 232 in ac-
cordance with DIN 66 020 Part 1.
-
For the electrical characteristics of the interface and the
allocation of signal status, standard RS 232 applies in ac-
cordance with DIN 66 259 Part 1.
- Transmission procedure: asynchronous character trans-
mission in start-stop mode.
- Type of transmission: full duplex.
-
Character format: character representation in accordance
with data format in DIN 66 022 for start-stop mode. 1
start bit; 7 character bits; 1 parity bit (even); 1 stop bit.
- Transmission speed: 9600 bit/s.
- Data ow control: none
- Access procedure: data transfer from the stirrer machine to
the computer takes place only at the computer’s request.

14
Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master)
to the device (Slave).
- The device sends only at the computer’s request. Even
fault indications cannot be sent spontaneously from the
device to the computer (automation system).
- Commands are transmitted in capital letters.
- Commands and parameters including successive parame-
ters are separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data)
and each response are terminated with Blank CR LF
(Code: hex 0x20 hex 0x0d hex 0x20 hex 0x0A) and have
a maximum length of 80 characters.
- The decimal separator in a number is a dot (Code: hex
0x2E).
The above details correspond as far as possible to the rec-
ommendations of the NAMUR working party (NAMUR rec-
ommendations for the design of electrical plug connections
for analogue and digital signal transmission on individual
items of laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specic IKA®
commands serve only as low level commands for commu-
nication between the device and the PC. With a suitable
terminal or communications programme these commands
can be transmitted directly to the circulator equipment. The
IKA® software package, Labworldsoft®, provides a conve-
nient tool for controlling circulating equipment and collect-
ing data under MS Windows, and includes graphical entry
features, for pump motor speed ramps for example.
PC 1.1 Cable:
This cable is required to connect RS 232 port (10) to a PC.
1
2 RxD
3 TxD
4
5 GND
6
7 RTS
8 CTS
9
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
PC
1
2
3
4
5
6
7
8
9
9
8
7
6
5
4
3
2
1
Fig. 14
Commands:
NAMUR
Commands
Function
IN_PV_2 Read the internal actual temperature
IN_PV_4 Read the pump actual speed
IN_SP_1 Read the internal setting temperature
IN_SP_4 Read the pump setting speed
IN_TMODE Read temperature control
0: internal regulation
OUT_SP_1 xxx Set the internal setting temperature XXX
OUT_SP_12@n Set the WD safety temperature with echo of the set (dened) value.
OUT_SP_4 xxx Set the pump speed XXX
OUT_SP_42@n Set the WD-safety speed with echo of the set (dened) value.
OUT_WD1@n Start the watchdog mode 1 and set the time for the watchdog to n (20...1500) seconds.
Echo of the Watchdog time.
During a WD1-event, the heating and pump functions are switched off and message PC 1 is displayed
This command needs to be send within the watchdog time.
OUT_WD2@n Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump
set speed is set to the WD safety speed. The PC 2 warning is displayed.
This command needs to be send within the watchdog time.
RESET Reset the PC control and stop the device functions.
START_1 Start the tempering function
START_4 Start the pump function
STOP_1 Stop the tempering function
STOP_4 Stop the pump function

16
Error codes
Accessories
•
Water bath protective media:
AQ.1 Aquacide 1
•
Tubing, hoses, adapters and couplings:
LT 5.20 Metal hose (isolated M16 x 1)
LT 5.21 PTFE hose (isolated M16 x 1)
H.PVC.8 PVC hose (nominal width 8)
H.PVC.12 PVC hose (nominal width 12)
H.SI.8 Silicone hose (nominal width 8)
H.SI.12 Silicone hose (nominal width 12)
See more accessories on www.ika.com.
Any malfunctions during operation will be identied by an error message on the display.
Proceed as follows in such cases:
F
Switch off device using the main switch at the back of the device
F
Carry out corrective measures
F
Restart device
Error code Effect Cause Solution
Err 02 Pump off
Refrigerating off
Motor over current (rate current) - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 04 Pump off
Refrigerating off
Motor hall signal missing - Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Err 06 Pump off
Refrigerating off
Too low liquid level - Check the fluid level and buoyage
Err 15 Pump off
Refrigerating off
Compressor power supply
disconnected
- Restart the device
Err 16 Pump off
Refrigerating off
Suction temperature error - Restart the device
Err 17 Pump off
Refrigerating off
Discharge temperature error - Restart the device
Err 19 Pump off
Refrigerating off
Fluid temperature too high - Switch the device off and let the fluid cool down
Err 20 Pump off
Refrigerating off
Condenser fan error - Restart the device
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
•
Tubing Insulations:
ISO. 8 Insulation (8 mm)
ISO.12 Insulation (12 mm)
•
Additional accessories:
RV 10.5003 Pressure regulating valve
RV 10.5002 Filter
PC 1.1 Cable (RS 232)
USB cabel A - Micro B 2.0
Labworldsoft®

17
Technical data
Nominal voltage VAC 230 ± 10% / 115 ± 10%
Frequency Hz 50 / 60
Max. input power W300
Working temperature range °C -20 ... RT
Operating temperature range (with external heating)
°C -20 ... + 40
Temperature stability – internal temperature
control (according to DIN 12876)
K± 0.1
Temperature control PID
Temperature measurement, absolute accuracy K± 0.5
Temperature setting Key on the front of the device
Temperature setting resolution K0.1
Temperature display LED
Temperature display resolution K0.1
Classication according to DIN 12876-1
Class III (FL) suitable for flammable and non-flammable fluids
Cooling capacity according to DIN 12876
+ 20 °C
+ 10 °C
0 °C
- 10 °C
- 20 °C
W400
370
320
240
130
Pump speed (adjustable) rpm 1000 ... 3200
Max. pump pressure/suction bar 0.3/0.2
Max. flow rate (at 0 bar) l/min 15
Bath volume I1.4 – 4.0
Fluid maximum viscosity mm2/s 50
Sub-level protection Yes
Interface USB, RS 232
Permitted on-time %100
Protection class according to EN 60 529
IP 21
Protection class
I
Excess voltage category II
Contamination level 2
Permitted ambient temperature °C + 5 ... + 32
Permitted ambient humidity %80
Dimension (W x D x H) mm 220 x 525 x 475
Weight kg 28
Operation at a terrestrial altitude mmax. 2000
Subject to technical changes!

18
Warranty
In accordance with IKA® warranty conditions, the warranty
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giv-
ing reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insucient
care or maintenance not carried out in accordance with
the instructions in this operating manual.
Pump performance curve
250
200
150
100
50
0
51
01
5
Q [l/min]
P [mbar]
0
-200
-150
-100
-50
300
Fig. 16
3200 rpm
2000 rpm
1000 rpm
Pump performance curve measured with water

Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details
below.
UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, Engl nd.
Tel: +44 (0)1245 600560
F x: +44 (0)1245 808399
Em il: s [email protected].uk
Please note - Product designs and specifications are sub ect to change without notice. The user is responsible for determining the
suitability of this product.
Table of contents
Other IKA Chiller manuals
Popular Chiller manuals by other brands

Waterworks
Waterworks IC8 Installation

Cold Shot Chillers
Cold Shot Chillers ACWC-024-E-DR-LT-0-2 Installation, operation and maintenance instructions

Daikin
Daikin Pathfinder AWV Series Installation, operation and maintenance manual

Carrier
Carrier 16DF013-050 installation instructions

Daikin
Daikin HYDROCUBE EWWQ014KBW1N installation manual

Randell
Randell BC-18-PUB Operator's manual