IKA RC 2 lite User manual

RC 2 lite

3
ENGLISH

4 5
Installation.................................................................................................13
EU Declaration of conformity .......................................................................6
Explication of warning symbols ....................................................................6
!
Safety instructions........................................................................................7
Unpacking .................................................................................................11
Operation ..................................................................................................16
Intended use..............................................................................................10
Technical Data .......................................................................................... 31
Maintenance and cleaning........................................................................ 28
Warranty .................................................................................................. 32
Pump performance curve.......................................................................... 32
Error codes ............................................................................................... 29
Accessories ............................................................................................... 30
Operator panel and display ....................................................................... 12
!
!
!
!
!
!
!
!
!
!
Fig. 1
Interfaces and outputs .............................................................................. 25
Menu navigation and structure .................................................................. 19
Transportation and storage ........................................................................ 23
Fluids (Standard information for IKA fluid) ................................................. 24

6 7
EU Declaration of conformity
Source language: German
Explication of warning symbols
/// Warning symbols
Indicates an (extremely) hazardous situation, which, if not avoided, will
result in death, serious injury.
Danger!
Indicates a hazardous situation, which, if not avoided,
can result in
death, serious injury.
Warning!
Indicates a potentially hazardous situation, which, if not avoided, can
result in injury.
Caution!
!
Indicates practices which, if not avoided, can result in equipment damage.
Notice!
/// General Symbols
Position number
A–––
Correct / Result
Shows the correct execution or the result of an action step.
Wrong
Shows the incorrect execution of an action step.
Note
Displays action steps that require particular attention to detail.
/// Device setup
Caution!
› The power switch of the device must be accessed immediately, directly and without risk at
any time. If access to the power switch cannot be ensured, an additional emergency stop
switch that can be easily accessed must be installed in the work area.
!Notice!
›
Set up the device in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface.
› The device has been constructed to the requirements of EU (European Union) and EFTA
(European Free Trade Association) countries.
›
Insucient ventilation may result in the formation of explosive mixtures. Only use the device
in well ventilated areas.
Safety instructions
/// General information
›
Read the operating instructions completely before starting up and follow the safety
instructions.
› Keep the operating instructions in a place where it can be accessed by everyone.
› Ensure that only trained staff work with the device.
› Follow the safety instructions, guidelines, occupational health and safety and accident
prevention regulations.
› The device must only be used in a technically perfect condition.
!Notice!
› Pay attention to the marked sites in Fig. 1.
!
/// Working with the device
Danger!
› Do not use the device in explosive atmospheres, it is not EX-protected.
› With substances capable of forming an explosive mixture, appropriate safety measures must
be applied, e.g. working under a fume hood.
› To avoid body injury and property damage, observe the relevant safety and accident preven-
tion measures when processing hazardous materials.
Warning!
›
Process pathogenic material only in closed vessels under a suitable fume hood. Please contact
IKA application support if you have any question.
› When using critical or hazardous materials in your processes, IKA recommends to use ad-
ditional appropriate measures to ensure safety in the experiment. For example, users can
implement comprehensive monitoring equipment.
Caution!
› Do not start up the device if:
- It is damaged or leaking
- Cable (not only the power cable) is damaged.
We declare under our sole responsibility that this product corresponds to the regulations
2014/35/EU, 2006/42/EC, 2004/108/EC and 2011/65/EU and conforms with the standards or
other normative documents: EN 61010-1, EN 61010-2-010, EN 61326-1, EN 60529, EN ISO 12100
and DIN 12876-1
.
Indicates the risk of fire or explosion!
Warning!

8 9
/// Disposal instructions
›
The device, accessories and packaging must be disposed of in accordance with local and
national regulations.
/// Maintenance
› The device must only be opened by trained specialists, even during repair. The device must
be unplugged from the power supply before opening. Live parts inside the device may still
be live for some time after unplugging from the power supply.
/// Power supply / switching off the device
›
The voltage stated on the type plate must correspond to the power voltage.
› The device can only be disconnected from the power supply by pulling out the power plug
or the connector plug.
› The device must only be operated with the original power cord set.
› The socket for the power cord must be easily accessible.
› Socket must be earthed (protective ground contact).
› After a power failure during operation, the device may start automatically (depending on
operating mode).
› Disconnect the power plug before attaching or changing any accessories.
› Disconnect the power plug before cleaning, maintenance and transportation of the unit.
/// Fluid
Warning!
› Only use the fluids, which fulfill the requirements for safety, health and device compatibility.
Be aware of the chemical hazards that may be associated with the bath fluid used. Observe
all safety warning for the fluids.
›
Depending on the bath fluid used and the type of operation or toxic can arise. Ensure suitable
extraction.
› Do not use any fluid which may cause dangerous reactions during processing.
!Notice!
› Only use recommended bath fluid. Only use non-acid and non corroding fluid.
›
Never operate the device without sucient fluid! Check the fluid level detection at a regular
basis.
› Continuous monitoring of the filling level of the bath fluid is required.
›
To ensure a sucient fluid circulation, the viscosity of the bath fluid must not exceed of 50 mm/s
at the lowest operating temperature.
› Untreated tap water is not recommended. It is recommended to use distilled water or high
purity water (ion exchangers) and add 0.1 g soda (sodium carbonate NaCO) /liter, to reduce
corrosive properties.
› Don’t use following fluids:
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides, iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.
/// Accessories
› Protect the device and accessories from bumping and impacting.
› Check the device and accessories beforehand for damage each time when you use them. Do
not use damaged components.
›
Safe operation is only guaranteed with the accessories described in the ”Accessories” section.
!Notice!
› Covers or parts that can be removed from the device without tools must later be refitted to
ensure safe operation. This will prevent the infiltration of foreign objects, liquids and other
contaminants.
› The feet of the device must be clean and undamaged.
› Always empty the bath if the unit is unused for long time.
› A laboratory circulator refrigerates and circulates fluid according to specified parameters. This
involves hazards due to high and low temperatures and general hazards due to the device
of electrical energy. The user safety can not be ensured simply with design requirements on
the part of the device. Further hazard sources may arise due to the type of tempering fluid,
e.g. by exceeding or undercutting certain temperature thresholds or by the breakage of the
container and reaction with the carrier fluid. It is not possible to consider all eventualities.
They remain largely subject to the judgment and responsibility of the operator. For this
reason, it become necessary for user to take other precautionary safety measures.
› The device may only be used as prescribed and as described in these operating instructions.
This includes operation by instructed specialist personnel.
› When device is used for external circulation, extra precaution must be taken for cold fluid
leakage due to damaged hose.
- Use suitable hoses for connection.
- Secure hoses and tubes against slippage and avoid kinks.
- Check hoses, tubes and bath at regular intervals for possible material fatigue (cracks/leaks).
› Do not transport or empty the bath while it is still cold.
› Always empty the bath before moving the unit.

10 11
Intended use
///
Use
Use RC 2 lite for tempering and circulating fluids.
Intended use: Tabletop device
Unpacking
›
Please unpack the device carefully.
› Any damage should be notified immediately to the shipping agent (post oce, railway
network or logistics company).
1 RC 2 lite
2Power cord set
3USB cable
4Hexagon socket offset screw key
5Hose olives NW 8 (2 x)
6User guide
7Warranty card
IKA RC 2 lite
Garanty
1+2+3
12
6
7
5
4
3
/// Refrigerant
Warning!
› The device is not allowed to be used in an ATEX (Atmosphere Exlosible) zone.
› Flammable refrigerant used. Do not use mechanical devices or other means to accelerate the
defrosting process, other than those recommended by the manufacturer.
› Do not damage the refrigerant circuit.
›
Insufficient ventilation may result in the formation of explosive mixtures. Only use the device
in well ventilated areas and keep all air vents open.
›
Flammable refrigerant used. Only the manufacturer may open and repair the closed refrigeration
system.
› The minimum room size is 9 m³, so that in the event of an error (exit of the complete refrigerant
filling quantity) no flammable refrigerant-air mixture can arise (R 290 max. 8 g/m³ ambient air).
/// Range of use
Indoor environments similar to that a laboratory of research, teaching, trade or industry.
The safety of the user cannot be guaranteed:
›
if the device is operated with accessories that are not supplied or recommended by the
manufacturer.
›
if the device is operated improperly or contrary to the manufacture’s specifications.
›
if the device or the printed circuit board are modified by third parties.

12 13
Operator panel and display
AOn / Off button Switch on / off the device.
B“Set” LED The LED lights up at the same time as the set value is displayed.
CCooling LED The light LED indicated the cooling function is activated.
DLED display
Display the settings and actual temperature values.
E“Menu” button Press it once: menu option is display.
Press it once again: back to working screen.
FMinus (-) button
Decrease the temperature setting.
Navigation, selecting the settings in the menu.
Change the menu value setting.
GPlus (+) button
Increase the temperature setting.
Navigation, selecting the settings in the menu.
Change the menu value setting.
H“OK / Temp” button Start / Stop the tempering function.
Confirm the menu options.
BC D
FG
H
AE
Installation
›
Place the unit on an even, stable, clean, nonslip, dry and fireproof surface.
›
Keep at least 20 cm of open space on the front and rear side.
›
The place for installation should be large enough and provide sucient air ventilation to
ensure the room does not warm up excessively because of the heat from device radiates
to the environment.
›
Do not set up the device in the immediate vicinity of heat sources and do not expose to
sun light.
›
Cooling machine, pump motor and electronics produce intrinsic heat that is dissipated via
the venting grids! Never cover these venting grids!
2x
2x
19 mm
2x
2x
1
2
/// Connecting the tubings and external device

14 15
Example
/// Filling the device
1
Fluid level
Low
High
2
2x
3
/// Connecting to power
1.
2.
220 - 240 V ~
50/60 Hz
250 W
IP 21
1

16 17
/// Setting temperature
/// Starting tempering function
/// Switching on
Operation
!Notice!
Before commissioning, make sure that the device has not been moved in one hour.
The socket used must be earthed (fitted with earth contact).
Observe the ambient conditions (temperature, humidity etc.) listed under “Technical Data”.

18 19
///
Menu navigation
›
Enter the menu by pressing ”Menu” button (E).
›
Press the plus (+) button (G) or minus (-) button (F) to change the menu options.
›
Press “OK / Temp“ button (H) to enter the select menu option.
›
Press the plus (+) button (G) or minus (-) button (F) to change the menu value settings.
›
Confirm the menu settings by pressing the “OK / Temp“ button (H).
Menu navigation and structure
The type of temperature
PID control arithmetic (A)
Menu
/// Menu structure
Default settings
AUTO 1
Minimum temperature (L)
MODE (M)A
C
B
Fluid type (F)
AUTO 0
Activated
Activated
Maximum temperature (H)40 °C
-10 °C
8
-
-
-
Kp 1(P)40
Ti 1 (i)10.0
Td 1 (d)3.0
Reset to factory settings (r)0
3
2
Calibration (C)0Activated
-
-
Alarm and key tone (b) 0
1Activated
-
/// Switching off
/// Stopping tempering function

20 21
AUTO 0:
For special requirements, AUTO 0 can be used with manual adjustment of the control
parameters.
If the PID control (AUTO 0) option is selected, the following items including Kp 1, Ti 1 and Td 1
could be set. Otherwise they are not displayed in the menu list.
°C
t
Note: Inappropriate settings may produce the following cooling down curves:
Proportional coecient of PID (Kp 1)
Integral time of PID (Ti 1)
Differential time of PID (Td 1)
The proportional coecient Kp is the controller amplification and determines how strongly
the control deviation (the difference between the target temperature and actual temperature)
directly affects the control variable (cooling down time). Kp values that are too large can lead
to the controller overshooting.
The integral time Ti (s) is the correction time and determines how strongly the duration of the
control deviation affects the control variable. Ti compensates an existing control deviation. A
high Ti means a smaller and slower effect on the control variable. Ti values that are too small
can lead to instability of the controller.
The differential time Td (s) is the derivative time and determines how strongly the rate of
change of the control deviation affects the control variable. Td compensates for rapid control
deviations. A high Td means a smaller and slower effect on the control variable. Td values that
are too small can lead to instability of the controller.
Kp too high
°C
t
Kp too low
Ti too high
°C
t
Td too high
°C
t
Ti too low
Minimum temperature (L):
The minimum adjustable value: -10 °C.
Fluid type (F):
The chosen fluid (No.) limits the maximum and minimum temperature.
Customized fluid (No.: 8) enables the whole working temperature range. See Section “Fluids (Stand-
ard information for IKA fluid)”.
Maximum temperature (H):
The maximum adjustable value: 40 °C.
Temperature control type (A 1):
AUTO 1:
AUTO 1 is the default setting. The default PID settings are used automatically.
°C
t
Cooling down curve in AUTO 1 control (AUTO 1):
Operating mode (M):
Mode A: After power on/power failure no automatic restart of functions.
Mode B: After power on/power failure automatic restart of functions, depending on previous
settings.
Mode C: Set values (set in mode A or B) cannot be changed. After power on/power failure
automatic restart of functions, depending on previous settings.
///
Menu details

22 23
Reset to factory settings (r):
r 0: Without resetting to factory settings.
r 1: Resetting to factory settings.
Calibration of the other points is performed in the same way.
Transportation and storage
› Remove the power cord before transporting the device.
› The device must be protected against mechanical impact, vibrations, dust deposits and corrosive ambient
air during transportation and storage.
In addition, take care to ensure that the relative humidity does not exceed 80 %.
› The device must be completely emptied before transportation and storage.
1.
2.
1 2
› Beware of the high dead weight of the device when transporting.
› Ensure that your fingers do not get crushed when setting down the device.
› The device can be transported:
- by hand / unaided (2 to 4 persons).
- with a pallet truck.
Caution!
Beware of tipping over because of the high center of gravity when lift and move the device!
Confirm the value by pressing the “OK / Temp“ button. The first point calibration is finished now.
The screen displays the set value for the first point calibration and flashes.
Alarm and button tone (b):
b 0: no alarm and button tone
b 1: alarm and button tone is activated
Calibration and adjustment (C):
C 0: Reset calibration
C 2: 2-point calibration
C 3: 3-point calibration.
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring instrument into the bath fluid.
Select 2-point calibration with plus (+) button / minus (-) button and “OK / Temp“ button in
the menu.
Set the first point temperature (step 1, e.g. 10 °C) with plus (+) button / minus (-) button.
Confirm the setting by pressing “OK / Temp“ button. The unit now starts and controls to the
set value.
When the temperature has reached the set value and has become constant, the value in the
screen changes from the set temperature value to "0.0" and flashes.
Input the calibration value from the reference measuring instrument (e.g. 10.3 °C) with plus (+)
button / minus (-) button.
Note: the difference between the input calibration value from reference measuring instrument
and set value is limited to less than +/-3 K.
Press the “OK / Temp“ button to start the 2-point calibration. The last calibration temperature
value flashes.

24 25
/// USB interface
The Universal Serial Bus (USB) is a serial bus for connecting the device to the PC. Equipped with
USB devices can be connected to a PC during operation (hot plugging).
Connected devices and their properties are automatically recognized.
USB 2.0
type B
or
/// USB device drivers
First, download the latest driver for IKA devices with USB interface from:
www.ika.com/ika/lws/download/usb-driver.zip.
Install the driver by running the setup file. Connect the IKA device through the USB data cable
to the PC. The data communication is via a virtual COM port. Configuration, command syntax
and commands of the virtual COM ports are as described in RS 232 interface.
Note: the USB driver is uncessary for Windows 10 system. Don't install it for Windows 10
system!
Fluids (Standard information for IKA fluid)
No.
IKA Designation
Operating
temperature range
for open bath
application
(°C)
Operating
temperature range
for closed bath
applications
(°C)
Safety
temperature
(°C)
Flash point
(°C)
0
CF.EG28.N10.80.8
-10 ... 80 -10 … 80 90 115
1
CF.EG39.N20.80.16
-20 … 80 -20 … 80 90 115
2
CF.EG44.N25.80.19
-25 … 80 -25 … 80 90 115
3
CF.EG48.N30.80.22
-30 … 80 -30 … 80 90 115
4
UF.Si.N30.150.10LV
-30 … 130 -30 … 150 145 >170
5
HF.Si.20.200.50
20 … 200 20 … 200 255 >280
6
HF.Si.20.250.50A
20 … 200 20 … 250 255 >280
7
Water
5 … 95 5 … 95 - -
8
Customized
(1) Classification:
HF: Heating Fluid
CF: Cooling Fluid
UF: Universal Fluid
(2) Chemical composition:
Si: Silicone oil
EG: Ethylene Glycol
(3) Temperature range: (
Minimum temperature. Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa
•
s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the
fluid divided by 1000.
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
(5) Addition information
(4) Viscosity (8
mm²/s
)
(3) Temperature range (-10 – 80 °C)
(2) Chemical composition (Ethylene Glycol 28%)
(1) Classification (Cooling fluid)
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content
may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for
operation due to corrosive properties of these media. High purity water and distillates are
suitable as a medium after adding 0.1 g soda (Na2CO3, sodium carbonate) per liter of water.
Note: The temperature limit values are adjustable in accordance with the fluid used.
Check the suitability of the liquid according to your application.
Nomenclature for IKA fluids:
CF.EG28.N10.80.8 --
Interfaces and outputs
The device can be connected to a PC and operated with the laboratory software labworldsoft
®
through USB or RS 232 interface.
The device software can also be updated with a PC via the RS 232 or USB port.
Note: Please comply with the system requirements together with the operating instructions
and help section included with the software.

26 27
Commands Function
IN_PV_2 Read the internal actual temperature
IN_SP_1 Read the internal setting temperature
OUT_SP_1 xxx Set the internal setting temperature XXX
OUT_SP_12@n Set the WD safety temperature with echo of the set (defined) value.
OUT_WD1@n Start the watchdog mode 1 and set the time for the watchdog to n (20...1500)
seconds.
Echo of the Watchdog time.
During a WD1-event, the tempering and pump functions are switched off.
This command needs to be sent within the watchdog time.
OUT_WD2@n Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety
temperature and the pump set speed is set to the WD safety speed.
This command needs to be sent within the watchdog time.
RESET Reset the PC control and stop the device functions.
START_1 Start the tempering function.
STOP_1 Stop the tempering function.
/// Connections between device and external devices
PC 1.1 Cable:
This cable is required to connect RS 232 interface to a PC.
1
2 RxD
3 TxD
4
5 GND
6
7 RTS
8 CTS
9
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
PC
1
2
3
4
5
6
7
8
9
9
8
7
6
5
4
3
2
1
USB 2.0 Cable (A - B):
This cable is required to connect USB interface to a PC.
The above details correspond as far as possible to the recommendations of the NAMUR working
party (NAMUR recommendations for the design of electrical plug connections for analogue and
digital signal transmission on individual items of laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specific IKA commands serve only as low level
commands for communication between the device and the PC. With a suitable terminal or
communications program these commands can be transmitted directly to the device. The IKA
software labworldsoft®, provides a convenient tool for controlling the device and collecting
data under MS Windows.
/// Command syntax and format
The following applies to the command set:
› Commands are generally sent from the computer (Master) to the lab device (Slave).
› The lab device only sends at the computer’s request. Even fault indications cannot be sent
spontaneously from the lab device to the computer (automation system).
› Commands are transmitted in capital letters.
› Commands and parameters including successive parameters are separated by at least one
space (Code: hex 0x20).
› Each individual command (incl. parameters and data) and each response are terminated with
CR LF (Code: hex 0x0d hex 0x0A) and have a maximum length of 80 characters.
› The decimal separator in a number is a dot (Code: hex 0x2E).
/// Device software update
Keep your device up-to-date with the IKA Firmware update tool.
The firmware update can be done with a computer connected through USB-Interface.
For this, you need register on our website MyIKA first.
After registering your device IKA will inform you about available updates for your devices.
Please download the software "FWUToolSetup.zip" from our IKA service website www.ika.com.
/// RS 232 interface
Configuration:
›
The functions of the interface connections between the device and the automation system are
chosen from the signals specified in EIA standard RS 232 in accordance with DIN 66 020 Part 1.
› For the electrical characteristics of the interface and the allocation of signal status, standard
RS 232 applies in accordance with DIN 66 259 Part 1.
› Transmission procedure: asynchronous character transmission in start-stop mode.
› Type of transmission: full duplex.
› Character format: character representation in accordance with data format in DIN 66 022 for
start-stop mode. 1 start bit; 7 character bits; 1 parity bit (even); 1 stop bit.
› Transmission speed: 9600 bit/s.
› Data flow control: none
› Access procedure: data transfer from the device to the computer takes place only at the
computer’s request.

28 29
Error codes
Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
› Turn off the device by using the power switch.
› Carry out corrective measures.
› Restart the device.
Causes
›
Safety shutdown of the compressor.
Effect
› Pump off
› Tempering function off
Solutions
› Check operating conditions according to the "Technical data".
› Wait until over-temperatures / -pressures are equilibrated.
› Restart the device.
E 15
Causes
› Compressor suction temperature fault.
Effect
› Pump off
› Tempering function off
Solutions
› Check operating conditions according to the "Technical data".
› Wait until over-temperatures / -pressures are equilibrated.
› Restart the device.
E 16
Causes
› Compressor discharge temperature fault.
Effect
› Pump off
› Tempering function off
Solutions
› Check operating conditions according to the "Technical data".
› Wait until over-temperatures / -pressures are equilibrated.
› Clean the filter.
E 17
Causes
›
Fluid temperature too high.
Effect
› Pump off
› Tempering function off
Solutions
› Check operating conditions according to the "Technical data".
› Wait until over-temperatures / -pressures are equilibrated.
› Switch the device off and let the fluid cool down.
E 19
/// Repairs
Please send in device for repair only after it has been cleaned and is free from any
materials which may constitute a health hazard.
For repair, please request the “Decontamination Certificate” from IKA, or download printout of
it from the IKA website www.ika.com.
If you require servicing, return the device in its original packaging. Storage packaging is not
sucient. Please also use suitable transport packaging.
Maintenance and cleaning
To avoid contamination, it is essential to change the bath fluid regularly.
If water is used for bath fluid, we recommend to add Water bath protective media. The protective
media stops the growth of algae, bacteria and other microorganisms.
To keep the full cooling performance, the dust filter of the cooler must be checked regularly and
cleaned if necessary.
› Switch off the device and disconnect power cable.
› Take out the dust filter.
/// Ordering spare parts
When ordering spare parts, please indicate:
› Device type.
› Serial number, see type plate.
› Item and designation of the spare part, see: www.ika.com, spare parts diagram and spare
parts list.
› Software version (briefly visible in the display when the device is switched on).
/// Cleaning
For cleaning disconnect the mains plug!
Use only cleaning agents which have been approved by IKA to clean the devices:
Water containing surfactant /isopropyl alcohol.
› Wear protective gloves during cleaning the devices.
› Electrical devices may not be placed in the cleansing agent for the purpose of cleaning.
› Do not allow moisture to get into the device when cleaning.
›
Before using another than the recommended method for cleaning or decontamination, the
user must ascertain with that this method does not destroy the device.
› Clean condenser dust protection sheet with a vacuum cleaner or wash the sheet in the water and
dry the sheet before assembly.
Note: Don’t touch the condenser surface with hard parts.

30 31
Technical data
Operating voltage VAC 230 ± 10 %
100 ... 115 ± 10 %
Frequency Hz 50 / 60
Max. input power W 250
Fuse T10A 250 V (2 x)
Working temperature range °C - 10 ... RT
Operating temperature range
(with external heating)
°C - 10 ... + 70
Temperature stability – internal temperature
control, water (according to DIN 12876)
K ± 0.5
Temperature control PID
Temperature measurement, absolute accuracy K ± 0.5
Temperature setting Buttons
Temperature setting resolution K 0.1
Temperature display LED
Temperature display resolution 0.1
Cooling capacity according to DIN 12876:
+ 20 °C
+ 10 °C
0 °C
- 10 °C
W 400
350
250
140
Refrigerant R 290
Refrigerant quantity g 70
Max. refrigerating system pressure bar 21
Max. pump pressure / suction bar 0.35 / 0.15
Max. flow rate (at 0 bar) l/min 18
Bath volume I 1 ... 3.5
Fluid maximum viscosity mm2/s 50
Low fluid level protection Yes
Interface USB, RS 232
Permitted on-time %100
IP code according to EN 60 529
IP 21
Protection class
I
Overvoltage category II
Contamination level 2
Permissible ambient temperature °C +5 ... +32
Permissible relative humidity % 80
Dimensions (W × D × H) mm 225 x 430 x 385
Weight kg 24.5
Operation at a terrestrial altitude m max. 2000
Note: Refrigerant must be disposed of in accordance with local and national regulations.
Subject to technical changes!
If the actions described fails to resolve the fault or another error code is displayed then take
one of the following steps:
›
contact the service department.
›
send the device for repair, including a short description of the fault.
Accessories
/// Tubings, hoses and adapters
LT 5.20 Metal hose (isolated M16 x 1)
LT 5.21 PTFE hose (isolated M16 x 1)
H.PVC.8 PVC tube (nominal width 8 mm)
H.PVC.12 PVC tube (nominal width 12 mm)
H.SI.8 Silicone tube (nominal width 8 mm)
H.SI.12 Silicone tube (nominal width 12 mm)
H.PUR.8 PUR tube (nominal width 8 mm)
H.PUR.12 PUR tube (nominal width 12 mm)
H.FKM.8 FKM tube (nominal width 8 mm)
H.FKM.12 FKM tube (nominal width 12 mm)
Hose olive NW 6 (2 pieces)
Hose olive NW 8 (2 pieces)
Hose olive NW 10 (2 pieces)
Hose olive NW 12 (2 pieces)
/// Tubing insulations
ISO. 8 Insulation (8 mm)
ISO.12 Insulation (12 mm)
/// Additional accessories
PC 1.1 Cable (RS 232)
USB 2.0 cable (A - B)
Labworldsoft®
See more accessories on www.ika.com.
Causes
›
Motor is not connected
Effect
› Pump off
› Tempering function off
Solutions
› Send device for repair.
E 22

32
Warranty
In accordance with IKA warranty conditions, the warranty period is 24 months. For claims under
the warranty please contact your local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use,
insufficient care or maintenance not carried out in accordance with the instructions in this operating
manual.
Pump performance curve
Pump performance curve measured with water:
(Measurements done according DIN 12876-2 with water at 20 °C; pump in a closed-loop circuit).
0
0
100
-100
-200
-300
-400
-500
200
300
400
246810 12 14 16 18
Q [l/min]
∆P [mbar]

20000031217a_EN_RC 2 lite_072021_web
Technical specifications may be changed without prior notice.
IKAworldwide IKAworldwide /// #lookattheblue @IKAworldwide
Discover and order the fascinating products of IKA online:
www.ika.com
JAPAN
IKA Japan K.K.
Phone: +81 6 6730 6781
eMail: [email protected]
INDIA
IKA India Private Limited
Phone: +91 80 26253 900
eMail: [email protected]
UNITED KINGDOM
IKA England LTD.
Phone: +44 1865 986 162
eMail: [email protected]
MALAYSIA
IKA Works (Asia) Sdn Bhd
Phone: +60 3 6099-5666
eMail: [email protected]
CHINA
IKA Works Guangzhou
Phone: +86 20 8222 6771
eMail: [email protected]
POLAND
IKA Poland Sp. z o.o.
Phone: +48 22 201 99 79
eMail: [email protected]
USA
IKA Works, Inc.
Phone: +1 910 452-7059
eMail: [email protected]
KOREA
IKA Korea Ltd.
Phone: +82 2 2136 6800
eMail: [email protected]
BRAZIL
IKA Brazil
Phone: +55 19 3772 9600
eMail: [email protected]
IKA-Werke GmbH & Co. KG
Janke & Kunkel-Straße 10, 79219 Staufen, Germany
Phone: +49 7633 831-0, Fax: +49 7633 831-98
eMail: [email protected]
VIETNAM
IKA Vietnam Company Limited
Phone: +84 28 38202142
eMail: [email protected]
Table of contents
Other IKA Chiller manuals
Popular Chiller manuals by other brands

AERMEC
AERMEC NRL Technical manual installation maintenance

ACS
ACS HE Series Operation and instruction manual

Midea
Midea LSBLG MCF Series Technical & service manual

Mueller
Mueller Falling Film Chiller Installation and operation manual

Mueller
Mueller PMC 40/50 Instructions for installation and operation

enware
enware DFSA121-PBBF manual

OHM ELECTRIC
OHM ELECTRIC COOL CABI OCA-S2300AC instruction manual

teca
teca TLC-700 Product information packet

Trane
Trane Manhattan Gen II Installation, operation and maintenance

York International
York International YCAA-B Installation, Commissioning, Operation and Maintenance Manual

Elkay
Elkay ER52Y Installation, care & use manual

Harbour
Harbour HBRBC93B quick start guide