Welcome to the RespiPro™ Quick Setup Guide intended for users of the RespiPro™. This guide includes
steps to:
1. Remove the RespiPro™ from the shipping crate
2. Assemble and mount the Instructor Display
3. Assemble and mount the Learner Display
NOTE
It is recommended that the removal and assembly of RespiPro™ be done with at least two (2)
people.
2SAFETY CONSIDERATIONS
For correct and effective use of the product, it is mandatory to read and to observe all instructions,
WARNING, and CAUTION statements in this guide and in the RespiSim® Operating Instructions. If the
product is not used as instructed, the safety protection provided may be impaired.
Moving the RespiPro™ Mobile Cart
The RespiPro™ Mobile Cart has the option to be untethered to a wall outlet in combination with a UPS
battery. Always lock the brakes on all four castors when the RespiPro Mobile Cart is stationary.
Before moving the cart, make sure that:
•the instrument cabinet is in its locked position so the ASL 5000™Breathing Simulator, the
RespiPro™ Interface Box, and other accessories in the slide-out cabinet are secured in place.
•the drawers of the utility cabinet are locked with the provided key lock.
•the bed angle is lowered into the horizontal position.
•the Learner Display and the laptop tray are lowered and swiveled inward.
•the RespiPro™ Manikin is secure on the bed.
When moving the cart from one location to another, have a second person assist with guiding the cart
along, as you would do with a hospital bed. The fully equipped RespiPro™ weighs approximately 375
lbs (170 kg). Therefore, a significant amount of momentum will occur when moving it. It is
recommended that a second person assists in guiding the RespiPro™ Mobile Cart along any sloped
surface.

4
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
WARNING!
Use appropriate caution when moving on ramps. A runaway RespiPro™ Mobile Cart can cause
serious injury and/ or equipment damage!
Using the height adjustment feature:
Before using the height adjustment feature, check that no ancillary cables, lines, or patient circuits are
being stretched, kinked, or pinched.
When lowering the bed height, keep hands, feet, and any objects clear of areas where the cabinetry of
the cart gets close to the cart base.
WARNING!
Risks of pinch injury indicated by this symbol.
Do not obstruct the upwards movement of the electrical column. It is rated for the combined weight of
the bed structure and the equipment in the cabinetry only.
CAUTION!
Obstructing the movement of the electrical column will damage the motor drive!
The maximum combined load for the drawer cabinet (all three drawers) is 25 lbs (11 kg).
Do not exceed the maximum load of the laptop tray (7lbs/ 3kg).
Weight of additional equipment attached to the 10 x 50 mm rail must not exceed 20 lbs (9kg).
CAUTION!
RespiPro™ equipment rail is not designed to support oxygen tanks.
Electrical connection
Always pull the retractable power cord gently out of the reel and allow the cord to settle with sufficient
slack so that a taut cable will not create a tripping hazard.
WARNING!
Connect RespiPro™ only to a properly grounded power outlet supplying the correct voltage!
RespiPro™ is either configured for 100..120V or 220..240V AC.

5
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
WARNING!
Electric shock hazard if connected to an ungrounded outlet!
CAUTION!
Connecting a RespiPro™ Mobile Cart configured for 100..120V AC to a higher voltage will
damage the electrical column and the Uninterruptable Power Supply (UPS).
WARNING!
Electromagnetic Interference: Do not use RespiPro™ in patient rooms or other areas where
life supporting equipment is in use.
WARNING!
Use of RespiPro™ in the presence of flammable anesthetics may present an explosion hazard.
WARNING!
Fire Hazards related to the use of oxygen:
When using RespiPro™ with elevated concentrations of oxygen (ventilators set to FiO2 >
21%), observe all precautions applicable to the use of oxygen indoors.
•Always use extreme caution when using oxygen!
•No smoking, no open fire in areas where oxygen is in use!
•Always provide adequate ventilation in order to maintain ambient O2 concentrations < 24%.
•Always secure O2 cylinders against tipping, do not expose to extreme heat.
•Do not use oil or grease on O2 equipment such as tank valves or pressure regulators.
Do not touch with oily hands. Risk of fire!
•Open and close valves slowly, with smooth turns. Do not use any tools.

6
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
3REMOVING RESPIPRO™
Unlatch the clasps on the front of the crate and lower the door. The door also serves as a ramp for
removing the cart from the crate.
Figure 3-1: Crate Door/ Ramp
Remove the Learner Display, located on the upper right side of the cart, by pulling the Styrofoam
towards yourself.
Figure 3-2: Remove Learner Display

7
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
Pull the pins on either side of the wooden beam and remove it by lifting upwards.
Figure 3-3: Remove Pins
Unlock the 2 caster brakes closest to you by lifting up on the brake lever.
Figure 3-4: Unlock Caster
Loosen and remove both tie-downs that are attached to the D-shaped anchors. Remove the cart from
the crate. It is recommended that a second person assists in guiding the RespiPro™ Mobile Cart along
any sloped surface, including the crate ramp.

8
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
4CONNECTIONS
4.1 Learner Display
Remove the packaging from the monitor pole, and remove the plastic cap from the top of the monitor
pole.
Figure 4-1: Learner Display Monitor Pole
Unlock the cabinet with the keys taped to the top drawer and remove the monitor swivel arm and
prepared power cable.
Figure 4-2: Learner Display Power Cable

9
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
Drop the long cable through the notched side of the monitor pole. Pull the remaining length so that
one cable ‘sheath’ (circled in red) slides through the monitor pole.
Figure 4-3: Learner Display Power Cable
Drop the power cable through the hole in the bottom of the monitor pole bracket and place the
monitor pole into the bracket. Secure the pole by tightening the plastic thumb screws.
Figure 4-4: Learner Display Power Cable
Plug the female power cable connector into the only loose power cable attached to the power strip.
Slide the cable sheath up into the hole in the bottom bracket, then attach the sheath to the plastic clip.

10
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
Figure 4-5: Connect Learner Display to Swivel Arm
Hold the monitor up to the monitor swivel arm and, using the 4 bolts provided in the top drawer of the
cabinet, hand tighten the bolts through the VESA mounting plate and into the monitor. Remove the
plastic panel covering the video and power connections and plug the power cable into the monitor.
Place the plastic covering back onto the monitor.
4.2 Instructor Display
Remove the 2 pieces of the tablet stand from the bottom drawer of the cabinet. Attach the base to the
swivel arm by snapping them together.
Figure 4-6: Instructor Computer Swivel Arm
Place the swivel arm into the other open bracket and tighten the plastic knobs to secure the tablet
stand.

11
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
Figure 4-7: Instructor Swivel Arm
The tablet stand is intended to work with a Microsoft Surface tablet as the Instructor Display
computer. The stand can be configured with or without the detachable keyboard/ touchpad. When
using the Instructor Display computer as a touchscreen only, place the stand with the lip down and fit
the computer between the clamps.
Figure 4-8: Instructor Computer in Touchscreen Only Configuration

12
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
Spin the stand 180 degrees where the lip is facing up. Connect the keyboard/ touchpad. Open the
Surface stand and place it on the lip. Place the clamps around the top of the keyboard/ touchpad.
Figure 4-9: Instructor Computer in Keyboard/Touchpad Configuration
5POWER
To power the system, pull the retractable extension cord from the center of the cart and plug the
power cord into a wall outlet.

13
31 30 170 Rev. 1 RespiPro™ Quick Setup Guide, © IngMar Medical, LLC. 2021
6GET STARTED
With the RespiPro™ setup complete and the system connected to a wall outlet, your simulation
environment is ready. Go to www.ingmarmed.com/resipsimdownloads to find the User’s Manual and
other helpful quick start guides.
Figure 6-1: RespiPro™ with ICU Ventilator (not included)