10.2 Possible measuring errors and their remedy
If Err2 appears, the reagent being measured is "lighter" than the liquid used for calibration. Please check
whether calibration has been carried out with the correct liquid; if necessary, recalibrate and subsequently
repeat the measurement. If Err2 is still displayed, check your reagent and, if necessary, renew it.
Subsequently carry out a new measurement.
The measuring range is exceeded if the water sample briefly turns red or if Err4 appears on the display after
the measurement. In this case, the water sample should be diluted with distilled water in advance.
To achieve this, multiply the measured value displayed by the instrument with the dilution factor.
11. Overview of Available Parameters
Parameter Measuring range Instrument
article no.
Reagent article
no.
Aluminium 0 - 1 mg/l 320181 410651
Ammonium 0 - 1 mg/l 320182 410681
Bromine 0 – 5 mg/l 320183 410521
Chlorine dioxide 0,01 – 1 mg/l 320171 410525
Chloride 0 - 200 mg/ 320172 410527
Chromate 0 - 1 mg/l 320173 410533
Iron L 0 - 1 mg/l 320178 410548
Iron H 0 - 10 mg/l 320179 410545
Hydrazine 0 - 1 mg/l 320184 410557
Copper 0 - 2 mg/l 320185 410563
Manganese L 0 - 0,25 mg/l 320186 410661
Manganese H 0 - 20 mg/l 320180 410569
Nitrate 0 - 10 mg/l 320187 410671
Nitrite 0 - 1 mg/l 320188 410581
Ozone 0 - 1,5 mg/l 320189 410517
Phosphate 0 - 10 mg/l 320191 410593
Silicate 0 - 2 mg/l 320192 410623
Sulphite 0 - 10 mg/l 320193 410635
If you require new reagents, please order the corresponding parameter refill packs under the
specified reagent article number!