IRIS iQ 200 Sprint User manual

Iris Diagnostics, a Division of Iris International, Inc.
iQ
®
200 Sprint ™
Service Manual

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1
Table of Contents
1 - Introduction
Intended Use ......................................................................................................1-2
How to use this Manual ......................................................................................1-2
Precautions and Warnings .................................................................................1-4
Notes ..................................................................................................................1-4
Cautions .............................................................................................................1-4
Warnings ............................................................................................................1-4
Biological Warning..............................................................................................1-4
Warranty.............................................................................................................1-5
Limitation of Liability...........................................................................................1-6
Use of Third Party Computer Products ..............................................................1-6
Iris Diagnostics Contact Information...................................................................1-7
2 - Installation Procedures
Installation Requirements...................................................................................2-2
Space Requirements..........................................................................................2-2
Workstation ........................................................................................................2-2
Location..............................................................................................................2-2
Requirements .....................................................................................................2-2
Installation ..........................................................................................................2-2
Shipping Cartons................................................................................................2-3
Instrument Carton...............................................................................................2-3
Starter Kit A Carton ............................................................................................2-4
Starter Kit B Carton (needs refrigeration)...........................................................2-4
Optional ..............................................................................................................2-4
Installation ..........................................................................................................2-5
Precautions ........................................................................................................2-5
Connecting the Sampler.....................................................................................2-6
Adjust the Height of the Analyzer .......................................................................2-6
Adjust Sampler ...................................................................................................2-7
Installing the Lamina Container..........................................................................2-7
Connecting the Drain Tubing..............................................................................2-7
Computer Connections.......................................................................................2-8
iQ200 Rear Panel SIC Connection Configurations ............................................2-9
iQ200 / AX-4280 Connection Bridge Instructions.............................................2-10
Optional Load/Unload Station Installation ........................................................2-15
Load Station - iQ200 Connection .....................................................................2-15
Load Station - AX-4280 Connection.................................................................2-17
Unload Station..................................................................................................2-19
Startup..............................................................................................................2-21
Run Focus ........................................................................................................2-22

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2
Table of Contents
Run Calibration Rack .......................................................................................2-22
Run Control Rack .............................................................................................2-22
Verify Load/Unload and Bridge Connections ...................................................2-22
Preparation for Starting Operation ...................................................................2-22
Perform Setup ..................................................................................................2-22
3 – Components
Sample Transport Mechanism ...........................................................................3-3
Sampler # 700-3006...........................................................................................3-3
Specimen Tube Detector # 700-3725 ................................................................3-4
Barcode Reader # 700-3315 ..............................................................................3-5
Specimen Presentation Assembly # 700-3700 ..................................................3-6
Crash Detect Function........................................................................................3-6
Waste/Rinse Well # 700-3830............................................................................3-7
Diaphragm Pumps..............................................................................................3-8
Air Pump # 700-3877 .........................................................................................3-8
Waste Pump # 700-3809....................................................................................3-8
Rinse Pump # 700-3874.....................................................................................3-8
Fill Pump # 700-3874 .........................................................................................3-8
Peristaltic Pumps # 700-3820 ..........................................................................3-10
Cannula Pump..................................................................................................3-10
Lamina Pump ...................................................................................................3-10
Evacuation Pump .............................................................................................3-10
Pump Cover Closed and Cassette-In-Place Detection ....................................3-10
Lamina Tank # 700-3784 .................................................................................3-12
Air Tank # 700-3802.........................................................................................3-13
Valves...............................................................................................................3-14
Solenoid Valves................................................................................................3-14
Drain Valve # 700-3844....................................................................................3-14
Evacuation Bypass Valve # 700-3843..............................................................3-14
Lamina Bypass Valve # 700-3842....................................................................3-14
Cannula Bypass Valve # 700-3881 ..................................................................3-14
Pipettor Bypass Valve # 700-3880 ...................................................................3-14
Air Valves .........................................................................................................3-15
Air Mix Valve # 700-3887 .................................................................................3-15
Air Charge Valve # 700-3888 ...........................................................................3-15
Optical Bench Assembly #700-3500 ................................................................3-16
Flow Cell # 525-3077 .......................................................................................3-17
Strobe Light # 065-0059...................................................................................3-17
CCD Camera # 700-3565.................................................................................3-17
20 x Objective # 220-3009 ...............................................................................3-17
Auto Focus System # 700-3505 .......................................................................3-17
Power System Assembly # 230-3000 ..............................................................3-18

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 3
Table of Contents
Cooling Fans # 700-3892 & 700-3412 .............................................................3-18
Fuses and Visible LEDs ...................................................................................3-18
Backplane Fuses and LEDs .............................................................................3-19
PCBA Summary ...............................................................................................3-21
Cardcage..........................................................................................................3-22
High Level Control Board (HLCB), # 101-5000 ................................................3-22
Motor/Scanner Control Board (MSCB), # 101-5001.........................................3-22
DC Valve Board (DCVB), # 101-5043 ..............................................................3-22
Stepper/Scanner/Motor Board (SSMB), # 101-5007........................................3-23
Backplane # 101-5003 .....................................................................................3-23
Signal Interface Concentrators (SICs)..............................................................3-24
Power SIC # 101-5020 .....................................................................................3-24
Pump SIC #101-5020.......................................................................................3-25
FSV SIC # 101-5045 ........................................................................................3-25
SPA SIC # 101-5004 ........................................................................................3-27
SPA SIC Visible LEDs......................................................................................3-27
SPA SIC Push-Button Switches .......................................................................3-27
OBA SIC #101-5053.........................................................................................3-28
Do-All SIC # 101-5046 .....................................................................................3-29
Rear Panel SIC # 101-5030 .............................................................................3-30
STM SIC # 101-5014........................................................................................3-30
Other PCBAs....................................................................................................3-30
ESN Board # 101-5049 ....................................................................................3-30
Lamina ESN # 101-5048 ..................................................................................3-30
FITD PCBA # 101-5041 ...................................................................................3-30
SM Status Board PCBA # 101-5032 ................................................................3-31
STD PCBA # 101-5034 ....................................................................................3-31
Pressure Transducer Board #101-5016 ...........................................................3-31
SPA Optical Board # 101-5052 ........................................................................3-31
LED Visual Indicators .......................................................................................3-31
4 – Adjustments
Accessing the Service Screen............................................................................4-3
Calibration Table ................................................................................................4-4
Calibration Auto-Sets .........................................................................................4-4
Sampler Calibration............................................................................................4-6
Material required ................................................................................................4-6
Procedure...........................................................................................................4-6
Specimen Tube Detector Alignment...................................................................4-7
Material required ................................................................................................4-7
Procedure...........................................................................................................4-7
Specimen Tube Detector Calibration .................................................................4-8
Material required ................................................................................................4-8

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 4
Table of Contents
Procedure...........................................................................................................4-8
STD Verification Procedure................................................................................4-9
Specimen Presentation Assembly (SPA) and Waste/Rinse Well Calibration ..4-10
Material required ..............................................................................................4-10
Procedure.........................................................................................................4-10
Waste Well Positioning.....................................................................................4-10
Test Tube Centering Position...........................................................................4-11
Test Tube Depth Positioning ............................................................................4-11
Waste Well Depth Positioning ..........................................................................4-12
Lamina Consumption Adjustment ....................................................................4-12
Air Pressure Calibration ...................................................................................4-13
Material required ..............................................................................................4-13
Procedure.........................................................................................................4-13
Barcode Reader Alignment ..............................................................................4-14
Material Required .............................................................................................4-14
Procedure.........................................................................................................4-14
Prepare Barcode Reader Connections ............................................................4-14
Load the Alignment Tool ..................................................................................4-15
Perform barcode alignment ..............................................................................4-15
Validation..........................................................................................................4-18
Collimation........................................................................................................4-19
What is a Collimation?......................................................................................4-19
Condenser Alignment.......................................................................................4-19
Run Auto Focus................................................................................................4-20
Flow Cell Replacement ....................................................................................4-21
Flow Cell Removal ...........................................................................................4-21
Flow Cell Installation ........................................................................................4-21
Flow Cell Lateral Adjustment............................................................................4-22
Flow Cell Tilt Adjustment..................................................................................4-23
5 - Service Application Software
Accessing the Service Screen............................................................................5-3
Startup................................................................................................................5-3
Establishing Link ................................................................................................5-4
Live Operation ....................................................................................................5-5
Tree View Pane..................................................................................................5-5
Commands .........................................................................................................5-6
Preset Moves .....................................................................................................5-6
Sensors ..............................................................................................................5-6
Register 0x0A.....................................................................................................5-7
Calibration Table ................................................................................................5-7
Calibration Auto-Sets .........................................................................................5-7
Sequences .........................................................................................................5-7

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 5
Table of Contents
Fluid Schematic Pane ........................................................................................5-8
Diaphragm Pumps..............................................................................................5-8
Peristaltic Pumps................................................................................................5-8
3-Way Valves .....................................................................................................5-9
2-Way Valves .....................................................................................................5-9
Motors ................................................................................................................5-9
Tube Connectors..............................................................................................5-10
Other Regions of Interest .................................................................................5-10
Menu Bar..........................................................................................................5-10
File Menu..........................................................................................................5-10
View Event Log ................................................................................................5-10
Show Versions Report......................................................................................5-11
Exit ...................................................................................................................5-11
Edit Menu .........................................................................................................5-11
View Menu........................................................................................................5-11
Toolbar .............................................................................................................5-11
Status Bar.........................................................................................................5-12
Polling...............................................................................................................5-12
Video ................................................................................................................5-12
Record Video Statistics ....................................................................................5-12
Terminal ...........................................................................................................5-12
Jog Menu..........................................................................................................5-13
Status Bar.........................................................................................................5-13
Current Message Status...................................................................................5-13
Link Status........................................................................................................5-14
“Starting ASTM”................................................................................................5-14
“Establishing link (may take a minute)” ............................................................5-14
“Could not establish link”..................................................................................5-15
Troubleshooting Failure to Establish Link ........................................................5-15
“Connection established”..................................................................................5-15
“Sending initialization messages”.....................................................................5-16
“Sending initial query messages” .....................................................................5-16
“Sending query messages” ..............................................................................5-16
“Idle” .................................................................................................................5-16
Caps Lock Status .............................................................................................5-16
Num Lock Status ..............................................................................................5-16
Scroll Lock Status.............................................................................................5-16
Sequences & Sensor Identifications for Troubleshooting.................................5-17
Master (Master Controller) ...............................................................................5-17
Sequences .......................................................................................................5-17
Level sensors ...................................................................................................5-17
SPA ..................................................................................................................5-17
Pumps/Evacuation pump/Sensors ...................................................................5-18
STM..................................................................................................................5-18

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 6
Table of Contents
Sensors ............................................................................................................5-18
FBA ..................................................................................................................5-18
Pumps/Sheath pump/Sensors..........................................................................5-18
Pumps/Cannula pump/Sensors........................................................................5-18
Program Operation...........................................................................................5-19
Frequently Asked Questions (FAQ) .................................................................5-19
Why Does a Sensor Display the Wrong Value?...............................................5-19
Which Controller Does Component X (Pump, Valve, etc.) Belong To? ...........5-19
What Are the Communication Addresses of the Major System Components? 5-20
What Should I Do If the Service Application Appears to Hang?.......................5-20
Service Tree .....................................................................................................5-21
Master ..............................................................................................................5-21
SPA Motors ......................................................................................................5-22
SPA Pumps and Valves ...................................................................................5-23
FBA Pumps and Sequences ............................................................................5-25
STM Motors 1...................................................................................................5-26
STM Motors 2, Sensors, Calibration Table and Sequences ............................5-27
OBA..................................................................................................................5-28
Versions 1 ........................................................................................................5-29
Versions 2 ........................................................................................................5-30
6 - Troubleshooting
Bypassing Error Condition..................................................................................6-2
Decoding the Valves field in Valve Error Reports ..............................................6-2
Iris Diagnostics Contact Information...................................................................6-2
Troubleshooting Guide .......................................................................................6-3
System locked up ...............................................................................................6-3
Autofocus failed..................................................................................................6-3
IQ 200 fails to go to Standby ..............................................................................6-4
ID errors .............................................................................................................6-4
Illumination/Sequential errors.............................................................................6-5
Valve failure to open or close .............................................................................6-5
System does not go into Standby mode.............................................................6-5
Samples take more than 60 secs to process and complete...............................6-5
Positive control failure ........................................................................................6-5
Calibration rack failed.........................................................................................6-5
Non uniform flow in Runfile ................................................................................6-6
Inconsistent cell count from run to run in Runfile ...............................................6-6
SP-FLI (spatial flow index) is high in all Runfiles................................................6-6
SP-KSP are high in all Runfiles..........................................................................6-6
Bubbles are present in fluid system ...................................................................6-6
Waste well overflow............................................................................................6-6
Event Log Error Messages.................................................................................6-7

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 7
Table of Contents
7 - Electronics
Electronics..........................................................................................................7-2
General Functions ..............................................................................................7-2
Cabling Diagram Overview.................................................................................7-3
View of Backplane inside Cardcage...................................................................7-4
Rear view of Backplane -- Fuses and Connections to Internal parts of Microscopy
Module................................................................................................................7-5
Addresses and Summary ...................................................................................7-6
Hardware Bit Assignments and Details ............................................................7-11
Hardware Assignment Cross Reference ..........................................................7-21
ESN I2C Addresses .........................................................................................7-27
8 - Schematics
998-3006 SM SMCS Assemblies Block Diagram...............................................8-3
700-3800 System Fluid Assembly......................................................................8-4
Master Controller State records .........................................................................8-5
SPA State records............................................................................................8-21
FBA State records ............................................................................................8-25
STM State records ...........................................................................................8-27
9 - Preventive Maintenance
Material...............................................................................................................9-2
Semi-Annual PM kit Part Number 799-3050 ......................................................9-2
Annual PM Kit Part Number 799-3051 ..............................................................9-2
Cleaning .............................................................................................................9-3
Instrument Clean Up ..........................................................................................9-3
Fluidics Clean Up ...............................................................................................9-3
Rinse/Waste Well Clean Up ...............................................................................9-3
Air blow...............................................................................................................9-3
Computers Clean Up..........................................................................................9-3
Laser Printer.......................................................................................................9-3
Fluidics ...............................................................................................................9-4
Liquid Waste.......................................................................................................9-4
Peristaltic pump cassette replacement...............................................................9-4
Tubing replacements..........................................................................................9-4
Adjustment verification .......................................................................................9-5
Sampler Rack alignment ....................................................................................9-5
Specimen Tube Detector verification .................................................................9-5
Barcode Reader alignment verification ..............................................................9-5
SPA alignment verification .................................................................................9-5

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 8
Table of Contents
Fluidic inspection................................................................................................9-6
Autofocus run .....................................................................................................9-6
Optical Bench Verification ..................................................................................9-6
Replenish Consumables ....................................................................................9-6
Performance verification.....................................................................................9-7
Software .............................................................................................................9-7
Calibration run ....................................................................................................9-7
Control run..........................................................................................................9-7
Documenting the Preventive Maintenance.........................................................9-7

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-1
1. Introduction
1
Introduction
Intended Use..................................................................................................................1-2
How to use this Manual..................................................................................................1-2
Precautions and Warnings .............................................................................................1-4
Notes ..........................................................................................................................1-4
Cautions .....................................................................................................................1-4
Warnings ....................................................................................................................1-4
Biological Warning......................................................................................................1-4
Warranty.........................................................................................................................1-5
Limitation of Liability.......................................................................................................1-6
Use of Third Party Computer Products ..........................................................................1-6
Iris Diagnostics Contact Information ..............................................................................1-7

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-2
1. Introduction
Intended Use
This Service Manual is intended for the iQ®200 Sprint and the iQ®200 2nd
generation Automated Urine Microscopy Analyzer – HVL (High Volume
Laboratory).
There are two HVL configurations:
• The iQ®200 Sprint can process 101 samples per hour
• The iQ200 2nd generation can process 60 samples per hour
These instruments have a new fluidic design, the capability of connecting
Load and Unload stations, and improved error reporting.
The iQ®200 Sprint™ Automated Urinalysis System (iQ®200 System) is an in-
vitro diagnostic use device composed of the iQ®200 Automated Urine
Microscopy Analyzer (iQ®200), connected physically and electronically to
the AUTION MAX™ AX-4280 Automated Urine Chemistry Analyzer (AX-
4280), and a workstation. The iQ®200 System is used to automate the
complete routine urinalysis profile, including urine test strip chemistry
panel, specific gravity, color, clarity, and microscopic analysis, providing
quantitative or qualitative counts of formed elements, such as cells, casts,
crystals, and organisms.
Optionally, the iQ®200 Automated Urine Microscopy Analyzer can be used
as a stand-alone device for microscopic analysis, and the results from the
iQ®200 can be combined with other urine chemistry results.
This manual is intended to be used by Iris Diagnostics trained personnel.
How to use this Manual
References are made to other sections within the Manual. In order to
avoid searching for a specific document or section, electronic hyperlinks
have been created to let you jump to other locations and to give you
immediate access to related information.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-3
1. Introduction
To follow a link:
1. Select the hand tool
2. Links are indicated in the following manner:
• The text is colored blue and underlined. Example: Iris Diagnostics
Contact Information.
3. Position the pointer over the linked area on the page until the pointer
changes to the hand with a pointing finger . Then click the link.
4. To return to the
previous section, click
on the Go to Previous
View button.
5. The Bookmark pane,
located on the left side
of the screen, can be
used as a “linked” table
of contents. Click on
the +or -sign located
next to the title to
expand or collapse the
selection for a
bookmark. Clicking on
a bookmark will take you automatically to the starting page of the
selected section.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-4
1. Introduction
Precautions and Warnings
The Manual includes information and warnings that must be observed by
the operator to ensure safe operation of the system. Important messages
are highlighted with borders and special icons identifying the type of
message enclosed.
There are four types of messages.
Notes
NOTE: Highlights important facts, gives helpful information and tips, and
clarifies procedures.
Cautions
CAUTION: Electrical caution! Unplug before handling.
CAUTION: If any liquid is spilled onto the Sampler, wipe the area clean
before testing. Leaving the spill may cause crystals to form and block the
movement of the sample racks.
Warnings
WARNING: Identifies potentially hazardous situations that could result in
serious injury to laboratory personnel.
Biological Warning
WARNING: Wear protective gloves to prevent exposure to pathogens.
Discard contaminated materials according to applicable regulations.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-5
1. Introduction
Warranty
Iris Diagnostics, a subsidiary of Iris International, (Iris) warrants that the
products manufactured by it, its divisions or subsidiaries and sold
hereunder shall be free from defects in material and/or workmanship,
under normal use and service, for the period expiring twelve (12) months
from the completion of installation under standard procedure by Iris
Diagnostics or an authorized Iris Diagnostics distributor, or upon
Purchaser’s signature on Iris Warranty/Acceptance form, or eighteen (18)
months from shipment, whichever occurs first. Iris Diagnostics makes no
warranty whatsoever regarding products manufactured by persons other
than Iris Diagnostics, its Divisions or Subsidiaries and Purchaser’s sole
source of warranty therefore, if any, is the original manufacturer’s
warranty.
No warranty extended by Iris Diagnostics shall apply to any products
which have been modified, altered, or repaired by persons others than
those authorized or approved by Iris or to products sold as “used.”
Iris Diagnostics’ obligation under this warranty is limited SOLELY to the
repair or replacement, at Iris Diagnostics’ option, of defective parts,
F.O.B. warehouse or local Iris office, or as otherwise specified by Iris.
Repairs or replacement deliveries shall not interrupt or prolong the term of
this warranty. Iris Diagnostics warranty does not apply to consumable
materials, except as specially stated in writing, not to products or parts
thereof manufactured by Purchaser.
This limited warranty is made on condition that immediate written notice
of any defect be given to Iris Diagnostics and that Iris inspection reveals
that the Purchaser’s claim is valid under the terms of this warranty.
Iris Diagnostics makes no warranty other than the one set forth herein or
that which may be provided in a separate warranty covering the
applicable product category. Such limited warranty is in lieu of all other
warranties, expressed or implied, including but not limited to any
expressed or implied warranty of merchantability or fitness for particular
purposes and such constitutes the only warranty made with respect to the
products.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-6
1. Introduction
Limitation of Liability
Iris Diagnostics shall not be liable for any loss of use, revenue or
anticipated profits, or for any consequential or incidental damages
resulting from the sale or use of the products.
Use of Third Party Computer Products
Iris Diagnostics does not recommend that the workstation provided as a
functional part of the iQ®200 Automated Urine Microscopy Analyzer or the
iQ®200 System be employed for performing any software or hardware-
based applications other than those specifically furnished to operate and
support the Iris instrument system, or those recommended and offered by
Iris Diagnostics specifically as accessories or enhancements for the Iris
instrument system. No other third party application software should be
installed in these microcomputers in addition to those provided or
recommended by Iris Diagnostics, without the expressed approval of Iris
Diagnostics Technical Services, in order to avoid potential performance
and reliability problems which can result from incompatibility factors,
errors in use of such software, or software-based “viruses.”
Installation of such third party software, or non-approved electronic cards
or other devices, without advance Iris Diagnostics approval may affect the
terms of or void any Iris Diagnostics warranty otherwise in effect, covering
Iris Diagnostics supplied software and hardware on the microcomputers
and the overall performance and reliability of the entire Iris Diagnostics
instrument system.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 1-7
1. Introduction
Iris Diagnostics Contact Information
Iris Diagnostics
A Division of Iris International, Inc.
9172 Eton Avenue
Chatsworth, CA 91311
USA
Telephone:
From U.S. locations (800) PRO-IRIS (776-4747)
From outside the U.S. +1-818-709-1244
Fax: +1-(818) 700-9661

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-1
2. Installation Procedures
2
Installation
Procedures
Installation Requirements ..............................................................................................2-2
Space Requirements ..................................................................................................2-2
Workstation.................................................................................................................2-2
Location......................................................................................................................2-2
Requirements .............................................................................................................2-2
Installation ..................................................................................................................2-2
Shipping Cartons ...........................................................................................................2-3
Instrument Carton.......................................................................................................2-3
Starter Kit A Carton ....................................................................................................2-4
Starter Kit B Carton (needs refrigeration) ...................................................................2-4
Optional ......................................................................................................................2-4
Installation......................................................................................................................2-5
Precautions.................................................................................................................2-5
Connecting the Sampler .............................................................................................2-6
Adjust the Height of the Analyzer ...............................................................................2-6
Adjust Sampler ...........................................................................................................2-7
Installing the Lamina Container..................................................................................2-7
Connecting the Drain Tubing......................................................................................2-7
Computer Connections...............................................................................................2-8
iQ200 Rear Panel SIC Connection Configurations.....................................................2-9
iQ200 / AX-4280 Connection Bridge Instructions.....................................................2-10
Optional Load/Unload Station Installation ................................................................2-16
Load Station - iQ200 Connection..........................................................................2-16
Load Station - AX-4280 Connection .....................................................................2-20
Unload Station ......................................................................................................2-24
Startup .........................................................................................................................2-27
Run Focus ................................................................................................................2-28
Run Calibration Rack................................................................................................2-28
Run Control Rack .....................................................................................................2-28
Verify Load/Unload and Bridge Connections............................................................2-28
Preparation for Starting Operation ............................................................................... 2-29
Perform Setup.......................................................................................................2-29

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-2
2. Installation Procedures
Installation Requirements
Space Requirements
The iQ®200 is a tabletop unit requiring free space on a standard (36 inch:
0.9m) height countertop of approximately 36 inches wide by 24 inches
deep (91cm by 61cm), with vertical clearance of at least 24 inches
(61cm).
With optional Load and Unload stations, the width requirement is 60
inches (147 cm.)
Workstation
The Workstation is composed of a desktop computer, video monitor,
keyboard and mouse requiring a suitable counter or desktop, providing
comfortable access and a good viewing angle. It is suggested that, if
possible, the WorkStation be placed in a work area, which can be semi-
darkened to increase screen display visibility and reduce glare.
The data connections are made between the iQ®200 and the Workstation
(and the AX-4280 and the WorkStation, if applicable) via the special
cables provided.
Location
For most laboratories, the iQ®200 is placed near an open benchtop work
area which can be used for preparation of urine tubes and sample racks.
All data transmissions from the iQ®200 (and the Ax-4280, if used) are
routed to the Workstation, which manages and controls all
communications to the Laboratory Information System (LIS) via serial
connections.
Requirements
The iQ®200 uses 90 to 240V AC, 50-60Hz input source. Neither voltage
nor frequency parameters are customer configurable. Uninterruptible
power supplies are recommended and available from Iris Diagnostics for
the iQ®200 and the Workstation (and the AX-4280 if applicable) to maintain
system operation during short power outages and brownouts. This allows
for an orderly shutdown of instruments without the loss of data.
Installation
The iQ®200 will be installed by a factory-trained representative from Iris
Diagnostics.

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-3
2. Installation Procedures
Shipping Cartons
The iQ®200 is shipped in three cartons.
Instrument Carton
Item number Description Qty
800-3320
800-3322
iQ®200 60/hour or
iQ®200 Sprint TM 101/hour 1
700-3103 Printer 1
250-3082 Printer Cable 1
700-3120 Computer Version 3.0 1
210-3004 Flat Panel Monitor 1
700-3101 Keyboard and Mouse 1
700-3022 Sample Racks (1 Set of 9) 1
700-3008 Calibrator Rack 1
700-3007 Control/Focus Rack 1
700-3781 iQ Lamina Cap & Filter 1
700-3701 Drain Tubing 1
700-3702 Lamina Tubing 1
250-3501 Serial Cable, PC to Microscopy Module 1
700-3011 Operators Manual CD 1

Iris Diagnostics, a Division of Iris International, Inc.
iQ®
200 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005 2-4
2. Installation Procedures
Starter Kit A Carton
Item number Description Qty
475-0047 iQ®200 Lamina 1
475-0021 Iris Diluent 1
475-0003 Iris System Cleanser 1
660-0036 Culture Tube, 16x100 250 pack 1
800-3211 iQ®200 Dilution Labels Package 1
Starter Kit B Carton (needs refrigeration)
Item number Description Qty
800-3104 iQ®200 Control/Focus Set (four bottles) 1
800-3103 iQ®200 Calibrator (four bottles) 1
Optional
Item number Description Qty
800-3810 Load and Unload Set 1
800-3514 Load Station (only) 1
800-3515 Unload Station (only) 1
800-3501 Ax-4280 Bridge Connection Kit 1
Table of contents
Other IRIS Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific Nicolet iS5 Safety guide

Julabo
Julabo MAGIO MS-Z Original operating manual

Agilent Technologies
Agilent Technologies 1260 Infinity Operator's quick guide

Shodex
Shodex Silica C18M 4D Operation manual

Excelitas Technologies
Excelitas Technologies X-Cite XYLIS XT700 Series user guide

REITEL
REITEL ROTOGRIND NT operating instructions

OHAUS
OHAUS Frontier FC5706 instruction manual

Eisco
Eisco CH0814 instruction manual

Viso Systems
Viso Systems LabRail user manual

Linkam Scientific Instruments
Linkam Scientific Instruments TST350 user guide

VERDER
VERDER Carbolite Gero HTMA 7/95 Installation, operation and maintenance instructions

Sartorius
Sartorius Biotrate user manual