Iwatsu SC7101 User manual

FREQUENCY
COUNTER
SC7101/SC-7102

INSTRUCTION
MANUAL
FREQUENCY
COUNTER
SC-7I01/SC-7102
F697-816151(L)
F271-816151(L)

SC-7101
/SC-7102
Cautions
on
Operations
When
using
the
Frequency
Counter
SC-7101/SC-
7102,
take
care
on
the
following
items.
Ambient
Temperature
and
Ventilation
The
instrument
operate
normally
in
the
ambient
temperature
of
O°C
to
+40°C.
Be
sure
to
use
the
instrument
within
this
range.
Use
beyond
the
limits
of
this
range
can
result
in
some
trouble.
Do
not
place
anything
near
the
venti-
lating
hole
in
the
cover
to
improve
the
efficiency
of
heat
dissipation.
Check
the
Supply
Voltage
Before
plugging
the
power
cord
to
an
electrical
output,
be
sure
to
check
its
voltage.
The
instru-
ment
can
be
used
on
the
line
voltage
shown
in
Table
1,
which
can
be
selected
with
the
voltage
selector
plug
on
the
rear
panel.
Operating
the
instrument
on
other
than
the
specified
voltages
can
result
in
breakdown.
In
case
of
switching
the
voltage
selector,
remove
the
power
cord.
Then
remove
the
fuse
holder
cap
by
turning
it in
the
direction
shown
by
an
arrow.
Pull
and
remove
the
voltage
selec-
tor,
then
insert
into
the
desired
center
voltage
value.
Table
1
Center
voltage
Voltage
range
AC
100
V
90
to
110
V
AC
120
V
108
to
132
V
AC
220
V
198
to
242
V
AC
240
V
216
to
264
Vv
Do
Not
Apply
Excessive
Voltage
The
maximum
allowable
input
voltage
is
200
V
(DC
+
peak
AC).
Never
apply
any
voltage
beyond
these
voltages.
Connection
of
Input
Signals
Use
a
supplied
coaxial
cable
for
connection
of
input
signals.
Use
of
single
core
wires
may
cause
an
incorrect
measurement
by
electric
noises
due
to
induction.
When
there
is
any
fear
of
mixing
of
noises,
etc.,
check
input
signals
with
the
oscilloscope.
Start
measuring
only
after
confirming
that
there
is
no
mixing
of
noise,
etc.
di
Switch
on
the
Rear
Panel
The
10
MHz STD
OUT-IN
switch
on
the
rear
panel
is
set
to
OUT
side
at
the
time
of
shipment.
When
this
switch
is
set
to
"IN"
side,
no
signal
of
the
internal
reference
oscillator
is
connected.
Be
sure
to
set
this
switch
in
"OUT",
since
setting
in
"IN"
will
not
allow
measurement
without
apply-
ing
external
reference
signal.
Earth
a
Protective
Ground
Terminal
When
supplying
the
line
voltage
from
the
power
consent
using
a
two-conductor
type
power
cord,
ensure
to
connect
the
ground
terminal
on
the
rear
panel
to
the
earth
in
order
to
prevent
a
danger.
When
supplying
the
line
voltage
from
the
power
consent
using
a
supplied
three-conductor
type
power
cord,
grounding
is
provided
with
an
earth-
ing
wire
attached
to
the
power
cord.
Type
of
Input
Circuit
The
input
circuit
of
the
instrument
is
not
of
the
floating
type
but
the
circuit
type
apply-
ing
the
unbalanced
signal
against
the
earth
poten-
tial.
Never
perform
measuring
the
primary
power
source
with
an
unclear
earth
potential,
which
causing
significant
danger.
Ensure
to
Replace
a
Fuse
with
the
Requied
Rated
Current
0.3
A
fuse
is
used
in
the
instrument
in
order
to
prevent
the
damages
to
the
circuit
due
to
excessive
voltage.
If
it is
fused,
make
sure
to
replace
only
fuse
with
the
required
rated
current
after
closely
checking
the
cause
of
fusing
and
repairing
the
defective
portion.
Never
use
those fuses
other
than
the
rated,
as
it
may
result
in
failure
of
the
instrument.

TABLE
OF
CONTENTS
Cautions
on
Operations
..
SECTION
1
SPECIFICATIONS
.
1-1
GENERAL
1-2
SPECIFICATION
.
1-2-1
Frequency
Measurement
.
1-2-2
General
csccescceccceee
1-2-3
Physical
Characteristics
-3
ENVIRONMENTAL
CHARACTERISTICS
-4
ACCESSORIES
..eeeeesseseeee
1
1
SECTION
2
OPERATING
INFORMATION
..cccccscccccccccccscvccccssccsccscccsssssssecsssssssssssssceees
2-1
FRONT
PANEL
.....
2-1-1
Power
Supply
aa
2-1-2
Display
of
Measured
Results
2-1-3
Input
ceccceee
2-1-4
Measuring
Condition
Setting
2-2
REAR
PANEL
wesccccccccescccesece
.
2-3
DISPLAY
UNIT
sescsscseccccccccscscseces
2-4
HOW
TO
USE
THE
CARRYING
HANOLE
(STAND)
NNNNNNNNN
SECTION
3
MEASURING
PROCEOURES
3-1
POWER-ON
.
3-2
ON
RESET
.
3-3
MEASUREMENT
.
3-3-1
Frequency
Measurement
.
3-3-2
DIFF
Calculation
.
3-4
MEASUREMENT
USING
AN
EXTERNAL
REFERENCE
OSCILLATOR
3-5 USE
AS
THE
SECONDARY
FREQUENCY
STANDARD
UUUwnWwaww
SECTION
4
ADJUSTMENT
OF
REFERENCE
OSCILLATOR
sesseeeseeccccccccenccccecccencesssceassssssesessees
4-1
PREPARATION
FOR
ADJUSTMENT
...
4-1-1
Removing
the
Upper
Cover
.
4-1-2
External
Reference
Oscillator
.
4-2
ADJUSTMENT
PROCEDURE
«.sseeeeeeeeee
SECTION
5
SCHEMATIC
DIAGRAM
.
'
PEEUNNA=
'
OUFENNZA245
'
BE
ENN=AS5
'
Nese

MEMO
iv

SC-7101
/SC-7102
Section
1
Specifications
1-1
GENERAL
The
frequency
counter
SC-7101/SC-7102
are
compact
and
light
weight
electronic
measuring
instrument
which
can
measure
frequency
of
wide
range
from
10
Hz
to
200
MHz.
Use
of
large
and
high
luminousity
LED
in
the
display
makes
reading
easier.
This
instrument
is
provided
with
follow-
ing
features.
Use
of
Easy-to-read
Large
Light
Emitting
Diode
The LED
display
is
decimal
9
digits
+
1
digit
of
exponent
and
employs
Zero-blanking
methode.
Capability
of
Measurement
of
Low
Frequency
Signals
with
High
Resolution
This
instrument
allows
measurement
of
signals
of
low
frequencies
with
high
resolution
(normally
7-digits
per
1
s)
in
a
very
limited
time
by
the
reciprocal
methode
frequency
measurement.
Automatic
change
to
direct
frequency
measure-
ment
for
input
signals
of 10
MHz
or
more
enables
this
instrument
to
perform
measurement
of
input
signals
with
the
heightest
resolution.
Provision
of
DIFF
Calculation
Function
This
instrument
can
calculate
and
display
differences
between
the
values
of
input
signals
and
the
reference
value
by
a
simple
operation.
Allows
Easy
Systemization
of
Measuring
Since
mounting
of
an
optional
GP-IB
interface
or
a
centronix
type
parallel
interface
enables
output
of
measured
results,
external
reset,
remote
control
of
the
time
base, etc.,
this
instrument
can
be
built
into
a
systemized
configuration.
Built-in
Highly
Stabilized
Reference
Oscillator
and
Capability
for
External
Connection
The
SC-7101
and
SC-7102
have
built-in
high
stability
reference
oscillator
of
+3
x
10~’/month
and
#2
x
1078/day
respectively.
In
addition,
external
input
of
reference
sig-
nals
of
high
accuracy
is
possible.
4-Step
Setting
of
Sample
Rate.
(SC-7102)
Sample
rates
can
be
set
in
4-steps
of
200
msec
x1,
x2,
x
4
and
x
8
Improved
Shielding
Effect
The
specially
treated
cases
provide
better
shielding
effect
than
that
of
metal
cases
though
these
cases
are
made
of
light
weight
plastics.
Easy
Mounting
of
Optional
Supplies
Following
optional supplies
can
be
easily
mount-
ed.
(For
the
SC-0108
and
the
SC-0110,
select
either
one
only.
No
mounting
of
both
of
them
is
permitted).
*
Parallel
interface
SC-0108.
*
GP-IB8
interface
SC-0110.
On
SC-7102,
there
are
further
additional
op-
tions.
*
High
stability
reference
oscillator
the
SC-
0114.
(10
MHz
+
5
x
10-°/day).
*
High
stability
reference
oscillator
the
SC-
0115.
(10
MHz
+
2
x
10-$/day).
<Reference>
Out
of
the
above
options,
when
SC-0114
is
mount-
ed,
on
the
Name
plate
on
the
rear
panel
showing
the
Model
No., "B"
is
printed,
while
when
SC-
0115
is
mounted,
"C"
is
printed.

Section
1
Specifications
SC-7101
/SC-7102
1-2
SPECIFICATION
1-2-1
Input
Frequency
Measurement
Measuring
range
10
Hz
to
200
MHz
1
kHz
to
200 MHz
with
time
base
of
1
ms
100
Hz to
200 MHz
with
time base
of
10
ms
Input
voltage
At
ATTEN
1:
20
mrms
to
1
Vrms
(1
kHz
to
100
MHz)
50
mUrms
to
1
Vrms
(10
Hz
to
200
MHz)
At
ATTEN
1/20
:
0.4.Vrms
to
20
Vrms
(1
kHz
to
100
MHz)
1
Vrms
to 20
Vrms
(10
Hz
to
200
MHz)
Maximum
input
voltage
200
V
(OC
+
peak
AC)
Input
RC
Approximately
1
M2//25
pF
or
less
Attenuator
1
or
1/20
Input
coupling
AC
Sc-7101
SC-7102
Time
base
10
ms,
0.1
s,
1s,
10s
|
ims,
10
ms,
0.1
s,
1s,
105
Resolution
10
MHz
or
more
100
Hz/10
ms,
10
Hz/O.1
s,
|
1
kHz/1
ms,
100
Hz/10
ms,
1
Hz/1
s,
0.1
Hz/10
s
10
Hz/0.1
s,
1
Hz/1
s,
0.1
Hz/10
s
10
MHz
or
less
S
digits/10
ms
4
digits/1
ms
6
digits/0.1
s
5
digits/10
ms
7
digits/
1s
6
digits/0.1
s
8
digits/10
s
7
digits/1
s
8
digits/10
s
Reading
unit
EXPONENT
(0,
#3,
+6,
#9)
Hz
Error
10
MHz
or
more
Accuracy
of
reference
oscillator
+
1
count
10
MHz
or
less
Accuracy
of
reference
‘
Trigger
error
+
1
reference
time
oscillator
Measured
frequency
x
Time base

SC-7101/SC-7102
1-2-2
General
Calculating
function
Display
method
Display
digits
Display
time
Sample
rate
Reference
oscillator
Oscillation
frequency
Aging
rate
Rise
time
Temperature
Vibration
Impact.
Output
of
reference
frequency
Frequency
Released
voltage
50
terminal
Input
of
reference
frequency
Frequency
Input
voltage
Maximum
input
voltage
Input
resistance
Input
coupling
Power
supply
Supply
voltage
Section
1
Specifications
OIFF
$X-Xref
X
:
Measured
value
X-ref
Storage
display
method
At
mantissa
DIFF
Calculation
mal
8
digits
Reference
value
using
the
light
emitting
diode,
zero
blanking
decimal
9
digits
when
the
result
is
negative,
-
sign
and
deci-
At
exponent
21
digit
(0,
+3,
+6,
+9)
Time
base
+
sample
rate
or
infinite
display
SC-7101
Sc-7102
Approximately
200
ms
10
MHz
+3
x
10°7/month,
#10
x
10°?/year
(On
the
basis
of
1
hour
after
Power
on)
+3
x
10%
(On
the
basis
of
1
hour
after
Power
on,
from
immediately
after
power
on)
#3
x
1078/0
°C
to
+40
°C
#1
x
107
or
less
(For
30
minutes
each
in
three
direction
at
frequency
of
10 Hz
to
55
Hz
and
with
am-
plitude
of
1.5
mm)
#1
x
1077
or
less
(Dropping
from
height
of SO
mm
above
a
hard
wood
three
times
each
in
three
directions)
Approximately
200
ms,
400
ms,
800
ms
or
1.6
s
10
MHz
+2
x
10°8/day,
#10
x
10°°/year
(On
the
basis
of 24
hour
after
power
on)
+5
x
107°
(On
the
basis
of
1
hour
after
power
on,
from
10
minutes
after
power
on)
+5
x
10°8/0
°C
to
+40
°C
+5
x
1078
or
less
(For
30
minutes
each
in
three
direction
at
frequency
of
10 Hz
to
55
Hz
and
with
am-
plitude
of
1.5
mm)
+5
x
1078
or
less
(Dropping
from
height
of
50
mm
above
a
hard wood
three times
each
in
three
directions)
10
MHz
2
Vp-p
or
more
350
mVp-p
or
more
10
MHz
1
Vrms
to
5
Vrms
10
Vrms
Approximately
10
k®
AC
Center
voltage
Voltage
range
AC
100
V
90
to
110
V
AC
120
V
108
to
132
V
AC
220
V
198
to
242
V
AC
240
V
216
to
264
V
Selected
with
the
voltage
selector
plug

Section
1
Specifications
SC-7101
/SC-7102
Frequency
range
50
Hz
to
400
Hz
Power
consumption
Approximately
11
w
(SC-7101)
Approximately
15
w
(SC-7102)
1-2-3.
Physical
Characteristics
Weight
Approximately
2.5
kg
Dimension
(210#2)W
x
(1002)H
x
(310+2)L
[mm]
1-3.
ENVIRONMENTAL
CHARACTERISTICS
Operating
temperature
0
°C
to
+40
°C
Operating
humidity
85
$RH
or
less
(0
°C
to
+40
°C)
Storage
temperature
-20
°C
to
+70
°C
Storage
humidity
90
#RH
or
less
(0
°C
to
+70 °C)
1-4
ACCESSORIES
Power
cord
+s.
Coaxial
cable
Fuse
(FSA-0.3A)
Accessory
bag
.-
Instruction
manual
Figure
1-2-4.
Dimension
(SC-7102)
r
230
-
|
21022
=
10H
|
oe
aaa
1
:
31082
+
[=
gesgeesee
6
<|'
|
el
(SC-7101
and
SC-7102
are
of
the
same
dimensions)

SC-7101
/SC-7102
Section
2
Operating
Information
CAUTION
©
Check
mode
and
measurement
mode
In
this
manual,
the
status,
where
LED
located
above
the
|CHECK|
key
is
turned
on,
is
called
a
“check”
mode
and
the
status,
where
such
LED
turns
out,
is
called
a
“measurement”
mode.
With
the
“check’””
mode
selected,
performs
self-check
and
displays
the
frequency
(10MHz)
of
the
internal
refer-
ence
oscillator.
With
the
“measurement”
mode
selected,
measures
a
signal
applied
to
INPUT
A
or
B.
©
Hold
mode
and
sample
mode
In
this_manual,
the
status,
where
LED
located
above
the
|HOLD|
key
is
turned
on,
is
called
a
“hold”
mode,
and
the
status,
where
such
LED
is
turned
out,
is
called
a
“sample”
mode.
With
the
“hold”
mode
selected,
measurement
is
stopped,
while
such
measurement
already
started
be-
fore
the
|HOLD]
key
is
pressed,
is
continued
until
completion.
With
the
“sample”
mode
selected,
the
SC-7101
repeats
measuring
at
the
rate
of
approximately
200
msand
the
SC-7102,
at
the
preset
sample
rate.
2-1
FRONT
PANEL
This
paragraph
describes
the
switches
and
keys
on
the
front
panel
and
the
display
which
indi-
cates
measured
results.
See
to
Figure
2-1-1-(1)*(2)
2-1-1
Power
Supply
@
POWER
ON-OFF
Selects
line
voltage
on
or
off.
2-1-2
Display
of
Measured
Results
@
Numerical
display
LED
Indicates
the
measured
results
in
9
digits
of
the
mantissa
(or
8
digits
with
a
negative
sign)
and
one
digit
of
the
exponent
with
-
sign.
@
STATUS
LED
Indicates
operating
status,
which
are
REMOTE,
OVER,
GATE
and
OVEN
(for
SC-7102)
statuses.
REMOTE
:
Turns
on
in
only
GP-IB
remote
status.
OVER
Turns
on
when
the
measured
results
causes
overflow.
GATE
+
Turns
on
at
the
measuring
gate
open.
QVEN
+
Indicates
to
use
a
high
stability
crystal
oscillator
with
thermostat
(for
SC-7102
only).

Section
2
Operating
Information
2-1-3.
Input
@®
BNC
connector
Signal
input
connector
to
measure
frequency
from
10
Hz to
200
MHz.
Input
Voltage
is
200
V
(DC
+
AC
peak).
©
FILT
NORM-10
kHz
Selects
the
low
pass
filter
of 10
kHz
on
or
off.
In
such
a
case
where
noises
of
high
frequencies
are
contained
in
low
frequency
signals,
attenuate
frequency
higher
than
10
kHz.
©
ATTEN
1/20-1
Selects
attenuator
of
1/1
or
1/20.
X1
position
connects
input
signal
directly
to
input
amplifiers,
X20
position
attenuates
input
signal
by
a
factor
of
20
(nominal).
SC-7101/SC-7102
2-1-4
Measuring
Condition
Setting
@
TIME
BASE
Selects
a
TIME
BASE.
The
TIME
BASE
are
displayed
by
1
ms
(for
SC-7102
only),
10
ms,
0.1
s,
1
s
and
10
s
(im
black
color)
which
locate
above
the
keys.
Mounting
of
the
GP-IB
interface
(optional
supply)
allows
the
external
remote
control.
CHECK
Selects
the
"check"
mode
or
"measure"
mode.
When
the
LED
located
above
the
[CHECK]
key
turns
on,
the
"check"
mode
is
selected
and
the
instrument
performs
self-checks
with
frequency
(10
MHz)
of
the
internal
reference
oscillator.
When
the
LED
located
above
the
[CHECK]
key
turns
out the
"measure"
mode
is
selected
and
the
instrument
measures
input
signals
applied
to
BNC
connector
of
INPUT.
Figure
2-1-1-(1).
Front
Panel
(SC-7101)

Figure
2-1-1-(2).
Front
Panel
(SC-7102)
SC-7101
/SC-7102
©
DIFF
Performs
DIFF
calculation
(X-Xref)
while
LED
located
above
the
[DIFF]
key
turns
on,
indicates
that
the
instrument
is
put
in
DIFF
calculating
status,
while
this
LED
turns
out,
the
DIFF
calculation
is
released.
To
perform
calculation,
set
in
DIFF
calcula-
tion
status
and
then
register
the
reference
value
(Xref)
with
the
key.
For
further
details,
refer
to
"3-3-2
DIFF
Cal-
culation".
HOLD
Selects
the
hold
or
sample
mode.
Whi
lighting
of
LED
located
above
the
key,
the
instrument
is
put
in
the
hold
status
to
stop
measurement
and
keep
displaying
the_
results
measured
before
pressing
the
[HOLO}
key.
However,
such
measurement,
which
has
already
started
before
the
[HOLD]
key
is
pressed,
is
per-
formed
until
completion.
While
this
LED
turns
out,
the
instrument
is
put
in
the
sample
status
to
repeat
mea-
surement
in
accordance
with
the
preset
sample
rate
(SC-7101
is
fixed
at
200ms).
Section
2
Operating
Information
RESET
(LOCAL)
Performs
the
following
operation
by
pressing
key.
Interrupts
all
process
being
executed
when
the
key
is
pressed.
*
Clears
the
splay
of
the
measured
re-
sults.
*
Updates
measuring.
*
If
DIFF
calculation
is
selected,
resisters
internally
the
restarted
measuring
re-
sults
as
Xref
by
pressing
the
key.
*
In
GP-IB
mode,
it
is
recognized
as
key.
SAMPLE
RATE
x2,
x4,
x8
(sc-7102
only)
Selects
SAMPLE
RATE
time.
Each
pressing
the
[SAMPLE
RATE]
key
allows
selection
of
sample
rates
in
the
following
sequence.
All LED
LED
LED LED
turns
out
turns
on
turns
on
turns
on
ba
be
*
Gd
x1
>
x2
—>
x4
>
x8
Approxi- Approxi- Approxi-
Approxi-
mately
mately
mately
mately
200
ms
400
ms
800
ms
1.6s

Section
2
Operating
Information
2-2
REAR
PANEL
Describes
switch
and
connectors
on
the
rear
panel.
Refer
to
Figure
2-2-1.
®
10
MHz
STD
IN-OUT
BNC
connector
Either
provides
10
MHz
of
the
internal
ref-
erence
oscillator
or
apply
external
signals
of
the
reference
frequency
of
high
accuracy.
OUT:
Provides
from
the
10
MHz STD
connector
immediately
when
10
MHz
internal
reference
oscillator
is
internally
connected.
IN:
Disconnects
internal
connection
of 10
MHz
internal
reference
oscillator.
In
this
case,
applies
10
MHz
of
an
external
refer-
ence
oscillator
to
the
10
MHz
STD
connec-
tor.
@
FUSE,
VOLTAGE
SELECTOR
Both
the
fuse
holder
and
the
power
supply
voltage
selector.
To
select
the
power
supply
voltage,
remove
the
fuse
holder
cap
and
then
select
it.
Figure
2-2-1.
Rear
Panel
SC-7101
/SC-7102
©
AC
LINE
INPUT
Connect
the
power
cord
to
it.
+
(GND)
Protective
grounding
terminal.
DATA
OUT
connector
installed
location
(option)
Installs
the
data
output
connector
for
the
Parallel
interface
SC-0108.
REMOTE
connector
installed
location
(option)
Installs
the
input/output
connector
for
the
GP-IB
interface
SC-0110.
:
Switch
installed
location
(option)
Installs
the
switch
for the
parallel
inter-
face
SC-0108
or
the
GP-IB
interface
SC-0110.

SC-7101
/SC-7102
2-3.
DISPLAY
UNIT
Section
2
Operating
Information
11
digits
LED
to
shows
measured
results
of
input
signals.
Turns
on
while
GP-IB
remote
status.
Turns
on
while
the
|
measurement
overflows.
Gate
Turns
on
while
the
=
4
oogc000
Indicates
the
high
stability
crystal
oscillator
with
a
thermostat
(SC-7102)
a
ETS
EI
TTT
>
oven
6
z
EXPONENT
gate
is
open.
Display
of
Hysteresis
in
a
Mantissa
In
measurement
of
the
signal
of 10
MHz
or
less,
resolution
is
determined
according
to
the
mea-
surement
time.
Thus,
if
MSD
value
of
the
mea-
sured
results
changes
from
9
to 10
or
vice
versa,
read-out
of
the
results
becomes
difficult,
since
the
position
of
decimal
point
and
exponent
change
while
the
number
of
digit
displayed
remains
same.
To
solve
this
problem,
when
MSD
value
changes
from
9
to
10,
the
display
of
the
mantissa
is
shifted
to
left
by
1
digit
to
blank
the
lowest
digit.
(For
the
external
output
data,
0
is
added
to
the
lowest
digit).
When
MSD
value
changes
from
10
to
9,
the
display
of
the
mantissa
is
shifted
to
right
by
omitting
1
digit.
Through
the
above
operations,
hysteresis
is
provided
with
display
for
easy
read-out.
If
this
operations
are
performed
8
times
continu-
outly
or
if
the
MSD
value
becomes
any
one
of
2
through
8,
results
will
be
displayed
without
such
operations.
In
the
RECIPROCAL
mode
measurement,
where
the
result
is
obtained
by
calculating
the
count
data,
there
may
be
1
digit
difference
in
the
effective
digit
due
to
calculated
error
even
in
measurement
in
the
same
range.
Also,
in
such
a
case,
the
same
processes
as
above
are
performed
to
ensure
easy
read-out.
—
Mantissa
Exponent
On
Displays
*
The
highest
9
digits
indicate
the
mantissa.
However,
in
the
case
when
the
measured
result
is
negative,
the
first
digit
(the
left-end
of
the
mantissa
display)
takes
"-"
sign
and
the
display
of
the
measured
result
is
maximum
8
digits.
*
The
lowest
two
digits
(10th
and
11th
digits
from
the
left)
display
the
sign
of
the
exponent
and
1
digit
of
the
exponent.
The
displays
of
the
exponent
are
by
such
multi-
ples
of
3,
as,
-9, -6,
-3,
0,
3, 6,
9.
When
the
sign
of
the
exponent
is
+,
this
position
is
left
blank.
*
For
display
of
the
mantissa,
the
leading
zero
blanking
occurs.
°
When
the
measurement
overflows,
normally
the
OVER
LED and
the
date
display
indicates
overflow
mark
"Uuueee".
However,
though
counting
has
been
10
or
more
digits,
the
data
display
shows
the
mantissa
in
9
digits
from
the
lowest
digits
and
indicates
the
overflow
by
turning
on
the
OVER
LED.

Section
2
Operating
Information
2-4
HOW
TO
USE
THE
CARRYING
HANDLE
(STAND)
The
lock
of
the
carrying
handle
is
released
by
pulling
both ends
of it
outward,
allows
turning
of
the
handle
approximately
every
20
degree.
When
fixed
at
the
bottom
of
the
instrument,
the
handle
serves
as
the
stand.
Figure
2-4-1.
How
to
Use
the
Carrying
Handle
SC-7101/SC-7102
When
this
carrying
handle
is
not
used
as
the
stand
or
the
instrument
is
stored,
turn
the
handle
to
the
instrument
rear
side,
via the
bottom
and
store
the
handle
as
shown
in
Figure
2-4-2.
Figure
2-4-2.
Storing
of
the
Carrying
Handle

SC-7101
/SC-7102
Section
3
Measuring
Procedures
3-1
POWER-ON
When
the
POWER
switch
is
turned
on,
display
tests
of
LEDs
are
performed
in
the
following
sequence.
1.
All
LEDs turn
on
and off
three
times
in
suc-
cession
except
GATE
and
OVEN
(for
SC-7102)
LEDs.
2.
Mumeral
display
LEDs
display
each
number
from
0
to
9
in
this
order
once
for
each
number,
except
the LED
which
indicates
-
sign
of
the
exponent.
3.
Set
the
keys
as
follows:
*
TIME
BASE
key
1s
(LED
turns
on)
*
CHECK
key
CHECK
MODE
(LED
turns
on)
4.
The
data
display
indicates
the
frequency
of
the
internal
reference
oscillator
as:
"10,000,000
6Hz"
(Refer
to
Figure
3-1-1.)
Mounting
of
the
parallel
interface,
the
SC-
0108
(optional
supply)
enables
any
selection
of
initial
values
of
TIME
BASE
and
CHECK
ON-OFF.
Any
selection
of
INPUT
SELECT
initial
value
is
also
possible.
Figure
3-1-1.
The
Display
of
Initial
Setting
3-2
ON
RESET
following
operations
by
pressing
of
key.
*
Interrupts
all
process
being
executed
when
the
RESET]
key
is
pressed.
Clears
the
display
of
the
measured
results.
Updates
measuring.
S
°
If
OIFF
calculation
is
selected,
resistors
internally
the
restarted
measuring
results
as
Xref
by
pressing
the
[RESET]
key.
In
the
GP-IB
mode,
it
is
recognized
as
a
[LOCAL]
key.
Bee
sc-7102
rreouency
counter
=
(@0000000
6
ow

Section
3
Measuring
Procedures
3-3.
MEASUREMENT
3-3-1
Frequency
Measurement
Internal
Self-check
With
turning
on
the
power
switch,
after
perform
ing
"power
on
self-check",
measuring
the
10
MHz
of
the
internal
reference
oscillator
with
TIME
BASE
1
s,
displays
"10.000
000
6
Hz".
Then,
the
display
is
changed
by
TIME
BASE
as
listed
in
Table
3-3-1.
Table
3-3-1.
TIME
BASE
Count
display Display
unit
Mantissa
1
ms
10.000
Hz
10
ms
10.0000
Hz
O1s
10.00000
Hz
1s
10.000000
Hz
10s
10.0000000)
Hz
<Note>
1
ms
of
TIME
BASE,
available
for
the
SC-7102
only.
Measurement
Procedure
*
Press
the
[CHECK]
key
and set
to
the
"MEASURE"
mode.
(LED
turns
out).
Operate
ATTEN
and
FILT
switches.
Connect
the
signals
to be
measure:d
to
the
input
connector.
Connect
the
signals
to
be
measured
to
the
input
connector
selected
with
the
INPUT
SELECT
key.
Select
the
most
suitable
TIME
BASE.
In
case
of
holding
the
measured
results,
press
the
key.
Then
the
LED
located
above
the
key
turns
on
and
holds
the
results.
.
SC-7101
/SC-7102
Reference
for
Frequency
Measurement
When
the
signals
to be
measured
are
higher
than
10
MHz;
The
measuring
gate
opens
for
the
time
selected
as
the
TIME
BASE,
and
counts
and
displays
frequencies
of
input
signals
during
this
period.
When
signals
to
be
measured
are
lower
than
10
MHz;
Measure
by
the
RECIPROCAL
method.
REFERENCE
RECIPROCAL
Method
.
The
reference
gate
opens
for
the
time
selected
as
the
TIME
BASE,
then
the
measuring
gate
opens
in
triggering
the
input
signals,
After
closing
of
the
reference
gate,
the
measuring
gate
closes
triggering
with
the
input
signals.
The
instrument,
counting
the
measuring
gate
open
term
“t”
with
the
reference
clock
(10
MHz)
also
counts
a
number
of
period
“n”
of
the
signals
to
be
measured
within
this
term
and
obtains
the
frequency
by
executing
calculation
of
f
=
n/t
and
displays
the
result.
(reciprocal
of
the
average
period
measurement).
When
this
method
is
used,
irrespective
of
the
input
frequency,
resolution
of
the
results
is
determined
by
the
measuring
time
(displays
7
digits
for
1s
gate,
normally).

SC-7101/SC-7102
Section
3
Measuring
Procedures
Errors
in
Measurement
When
average
values
of
input
signals
during
Fi
:
-
Errors
in
frequency
measurements,
there
are
N
period
are
measured:
igure
3-3-2.
Frequency
and
Error
in
Frequency
Measurement
I
errors
due
to
accuracy
of
the
reference
oscil-
2a
1
En
1
(Sinewave;
when
S/N
is 40
dB)
(SC-7101)
lator",
"+
1
count
error"
and
the
"Trigger
error".
NT
7
*
es
*
Ww
.
Showing
1/N
reduction
in
error.
Consequently,
Error
due
to
accuracy
of
the
reference
oscil-
measurements
of
average
values
of
various
periods
My
lator":
are
advantageous
as
they
can
reduce
influence
Since
the
counting
time
(gate
time)
of
the
of
noise.
+
10
MHz
or
more
frequency
measurement
are
determined
by
dividing
1
kH2/1
ms,
10
H2/O.1
sy
1
Hz/1
5,
O61
Hz/10
s
the
oscillation
frequency
of
the
reference
oscil-
How
do
these
errors
appear
in
the
actual
mea-
10
fie
on
Less:
lator
(crystal
oscillator),
this
error
depends
surement.
Figure
3-3-2
and
3-3-3
shous
relation
‘an
aah
ria
tack
aca
shee
on
the
accuracy
of
the
reference
oscillator.
between
input
signal
and
error
frequency
after
cee
calibration
of
the
reference
oscillator
and
[
"Error
of
+
1
count":
errors.
Although
the
digital
measuring
instrument
is
a
measuring
instrument
of
high
accuracy,
on
in-
Figure
3-3-1.
Trigger
Error
[
struments
which
count
by
opening
and
closing
of
gates,
a
quantization
error
of
+
1
count
oc-
curs.
Limit
due
to
accuracy
of
the
internal
reference
oscillator
Noiseless
Es
signal
"Trigger
Error":
Normally
the
measuring
error
due
to
the
trigger
error
is
proportional
to
the
noise
level.
This
error
is
larger
when
the
slope
of
the
area
set-
———1
period
4
eale
ting
the
trigger
level
of
the
input
signal
wave-
form
is
less
sharper.
This
error
is
the
minimum
when
the
trigger
level
of
a
sine
waveform,
as
shown
in
Figure
3-3-1,
is
set
in
the
center
level
of
the
waveform.
Relation
of
the
input
signal
voltage,
noise
volt-
age and
the
error
can
be
expressed
by
the
follow-
ing
equation.
2aT
1,
£0
(..fn
27s,
Teo
Es
at”
TT
A
i
H
H
H
H
1
.
Signals
with
noise
Figure
3-3-3.
Frequency
and
Error
in
frequency
Measurement
II
(Sinewave;
when
S/N
is
40
dB)
(SC-7102)
Maximum
period
-——|
10
MHz
or
more
|
1
kHz/1
ms,
10
Hz/0.1
s,
1
H2/1
s,
0.1
Hz/10
s
10
Mz
or
less
4
digits/1
ms,
6
digits/0.1
s,
7
digits/
1
s,
8
digits/
10
s
As
it
is
clear
from this
equation,
in
case
of
the
sinewave,
this
error
is
solely
determined
by
the
ratio
of
noiseless
signal
Es
and
noise
En,
irrespective
of
the
frequency.
For
example,
the
error
when
S/N
is 40
dB,
comes
to
0.3
%
when
ghee
60
cB,
0.03
%
from
the
above
equation.
Limit
due
to
accuracy
of
the
internal
reference
oscillator.
OOMHE
OH:
100Hz~=—«sTkHz
~=—«10kHz_-—=«100kHz»=«
MHz
©
10MHe
200MHz

Section
3
Measuring
Procedures
3-3-2
DIFF
Calculation
Performs
calculation
of
difference
between
the
measured
value
X
and
the
reference
value
Xref.
Register
the
reference
value
Xref
in
ad-
vance.
Measurement
1.
Apply
the
reference
value
Xref.
2.
Press
the
|
DIFF}
key
(LED
turns
on).
3.
Press
the
[RESET]
key
and
register
the
refer-
ence
value
Xref.
4.
Apply
the
measured
signal
X.
5.
"X-Xref"
is
displayed
on
the
data
display.
<Note>
*
To
re-register
the
reference
value
Xref,
apply
the
signal
of
the
new
reference
value
and
then
press
the
[Rt
key.
Registered
Xref
values
remain
unchanged
even
when
CHECK
mode,
MEASURE
mode,
INPUT
SELECT,
SAMPLE
RATE
and
TIME
BASE
are
changed.
When
DIFF
calculation
is
turned
on
again,
after
once
turned
off,
the
Xref
which
has
been
pre-
viously
registered,
remains
unerased.
(When
using
the
same
value
as
the
Xref,
no
new
re-
gistration
is
required).
Turning
on
line
voltage,
clears
Xref,
resetting
to
0.
SC-7101/SC-7102
3-4
MEASUREMENT
USING
AN
EXTERNAL
REFERENCE
OSCILLATOR
The
instrument
allows
measurement
of
higher
accuracy
by
using
external
reference
oscillator
which
has
higher
accuracy
than
the
internal
reference
oscillator.
In
this
case,
set
the
IN-OUT
switch
on
the
rear
panel
to
"IN"
and
apply
signals
from
the
external
reference
oscil-
lator
into
the
input-output
connector
of
10
MHz
STD.
CAUTION
Make
sure
to
set
the
IN-OUT
switch
to
“OUT”
except
when
applying
signals
of
the
external
reference
oscillator.
3-5
USE
AS
THE
SECONDARY
FREQUENCY
STANDARD
When
the
In-Qut switch
on
the
rear
panel
is
set
to
"QUT",
the
input-output
connector
of
the
10
MHz STD
provides
signals
of
the
10
MHz
reference
oscillator.
This
signal,
because
of
its
very
high
accuracy,
can
be
used
as
the
external
driving
signal
of
pulse
generators
or
time
markers
for
adjusting
the
oscilloscope.

SC-7101/SC-7102
Section
4
Adjustment
of
Reference
Oscillator
Though
the
SC-7101
and
the
SC-7102
counters
use
a
high
accuracy
crystal
oscillator,
when
continuous
measurement
of
high
accuracy
are re-
quired,
adjust
the
crystal
oscillator.
4-1
PREPARATION
FOR
ADJUSTMENT
Perform
adjustment
of
the
reference
oscillator
by
removing
the
upper
case.
4-1-1
Removing
the
Upper
Cover
Remove
the
upper
cover
by
the
following
pro-
cedures.
1.
Face
the
instrument
bottom
up
and
remove
3
screws
on
the
bottom.
(See
to
Figure
4-1-
1).
Hold
the
bottom
and
upper
covers
between
both
hands
securely
and
place
with
the
bottom
down.
Lift
the
upper
cover
rear
end
slightly
up
and
remove
the
upper
case
by
pulling
it
back-
ward.
(The
upper
cover
front
end
is
inserted
into
the
panel
frame).
(See
Figure
4-1-2).
2
3
Figure
4-1-1.
Screw
Location
on
Bottom
|
|
&
HAT
TMT
TE
HEH
HEH
f
o
6
&
o
2.
a
o
1g
WHT
<Note>
To
install
the
upper
cover,
reverse
the
above
procedures
to
remove
the
upper
cover.
4-1-2
External
Reference
Oscillator
Accuracy
of
the
reference
oscillator
is
as
follows:
Sc-7101
10
MHz
reference
signal
of
+1
x
10°’.
SC-7102
10
MHz
reference
signal
of
+1
x
1078.
Figure
4-1-2.
Removing
the
Upper
Cover
oF
ae
=
Right
side!
@
This manual suits for next models
1
Table of contents
Popular Cash Counter manuals by other brands

VISIONARY DESIGNS
VISIONARY DESIGNS MOD-1701 Setup instructions

TrueTime
TrueTime 800 Series General information

Lighthouse
Lighthouse ApexPortable Read me first

HYOSUNG
HYOSUNG NH-2700 Operator's manual

Royal Sovereign
Royal Sovereign RBC-ED350 owner's manual

Capetti Elettronica
Capetti Elettronica WineCap WSD12-2DI user manual

LZR
LZR SIGMA quick guide

TSI Instruments
TSI Instruments AEROTRAK 9306 Operation manual

Volumatic
Volumatic CountEasy Advanced configuration manual

Met One Instruments
Met One Instruments GT-321 Operation manual

ARPEGE SECURITY
ARPEGE SECURITY TC-5500 Operation manual

HYOSUNG
HYOSUNG NH-1800CE Operator's manual