John Bunn Neb-u-Tyke Bella the Butterfly JB0112-071 User manual

Read this manual before operating the nebulizer.
Save this manual for future reference.
Info: The most current version of this manual
.
Federal Law (USA) restricts this device to sale by or
on the order of a physician. This device should not
be used unless the operator has been instructed by
Neb-u-Tyke®
™
Nebulizer Compressor
Item JB0112-071
User Manual
JB0112-071-INS-LAB-RevC17

JB0112-071-INS-LAB-RevC17 2
CONTENTS
INTRODUCTION................................................................................................... 3
INTENDEDUSE............................................................................................ 3
SYMBOLSKEY............................................................................................. 3
IMPORTANTSAFETYPRECAUTIONS............................................................... 4
PRODUCTDESCRIPTION................................................................................... 7
FEATURESANDCOMPONENTS................................................................. 7
SETUP................................................................................................................... 8
OPERATION....................................................................................................... 10
NEBULIZEROPERATION........................................................................... 10
SAFETYOVERLOAD.................................................................................. 11
MAINTENANCE.................................................................................................. 12
CLEANINGTHECOMPRESSOR’SOUTERCASE................................... 12
CLEANINGTHENEBULIZERCUP............................................................ 12
REPLACINGTHEFILTER........................................................................... 13
STORAGE.................................................................................................... 13
TROUBLESHOOTING........................................................................................ 14
SPECIFICATIONS............................................................................................... 15
LIMITEDWARRANTY........................................................................................ 16
COMPRESORDENEBULIZADORMANUALDEUSUARIO(ENESPAÑOL)
..19
This product has been tested and successfully approved with the
following standards:
UL1431
EN 60601-1

JB0112-071-INS-LAB-RevC17 3
INTRODUCTION
This user manual contains important information
and safety precautions for the John Bunn Neb-u-Tyke
Bella the Buttery Nebulizer Compressor.
Before using your Nebulizer Compressor, please read and
understand this entire user manual. Take special note of all
safety precautions marked DANGER and WARNING. Save
this user manual for future reference.
Intended Use
The Nebulizer Compressor System is designed to provide
a source of compressed air for aerosol therapy. It is used
in conjunction with a jet (pneumatic) nebulizer to produce
medicated aerosols for inhalation by pediatric and adult
patients with respiratory symptoms.
Indications for therapy include asthma, chronic bronchitis,
infection of the upper respiratory tract, chronic
obstructive pulmonary disease, and other respiratory
disorders in accordance with a physician’s prescription.
Do not use this product for any other purpose than the
preceding. This device can be used with pediatric or adult
patients under physician’s prescription.
Symbols Key
O Off
I On
~ Alternating current
IPX0 Enclosed equipment without protection against ingress of water
Attention (read the instructions)
II Class II (not applicable for UL1431)
BF: This product is in accordance with the degree of protection against electric
shock for type BF equipment. (not applicable for UL1431)
Disposal of Electrical & Electronic Equipment (WEEE): This product should
not be treated as household waste, but submitted to an appropriate collection
point for the recycling of electrical and electronic equipment. For more detailed
information about the recycling of this product, please contact your local city
ofce, household waste disposal service, or the Graham-Field authorized
distributor from whom you purchased this product.

JB0112-071-INS-LAB-RevC17 4
IMPORTANT SAFETY PRECAUTIONS
When using electrical products, especially when children
are present, always follow basic safety precautions,
including the following:
DANGER:Indicatesapotentialhazardsituationorunsafe
practicethat,ifnotavoided,willresultindeathorserious
personalinjury.
WARNING:Indicatesapotentialhazardsituationorunsafe
practicethat,ifnotavoided,couldresultindeathor
seriouspersonalinjury.
CAUTION:Indicatesapotentialhazardorunsafepractice
that,ifnotavoided,couldresultinminorormoderate
personalinjury.
s NOTICE:Indicatesapotentialhazardorunsafepractice
that,ifnotavoided,couldresultinproductorproperty
damage.
Info: Provides application recommendations or other useful
information to ensure that you get the most from your
product.
READALL INSTRUCTIONSBEFOREUSING
DANGER:Toreducetheriskofelectrocution:
1. Alwaysunplugthisproductimmediatelyafteruse.
2. Donotusethisproductwhilebathing,showering,
washingdishes,orclosetowatersourcesofanykind.
3. Donotplaceorstoredevicewhereitcanfallorbepulled
intoatuborsink.
4. Donotplacedeviceinordropintowaterorotherliquid.
5. Donotreachforadevicethathasfallenintowater.
Unplugitimmediately.

JB0112-071-INS-LAB-RevC17 5
WARNING:Toreducetheriskofburns,electrocution,re
or personalinjury:
1. Electricalshockhazard.Donotremovetheoutercaseof
thedevice.Alldisassemblyandmaintenanceofthedevice
mustbeperformedbyaqualiedservicetechnician.Refer
servicetoqualiedservicepersonnel.
2. Neverleavethisdeviceunattendedwhenpluggedin.
3. Thisproductcontainssmallpartsthatmaypresent
achokinghazardtosmallchildren,andthetubing
maypresentastrangulationhazard.Always use close
supervision when this device is used by or near children
or those who require close supervision.
4. Childrenundertheageof3,oranypatientwhoisunable
touseamouthpieceproperlyundersupervision,should
useamask.
5. Usethisdeviceonlyasintendedanddescribedinthis
manual.Donotuseattachmentsoraccessoriesnot
recommendedbyGFHealthProducts,Inc.
6. Neveroperatethisproductif:
a. ithasadamagedcordorplug,
b. itisnotworkingproperly,
c. ithasbeendroppedordamaged,or
d. ithasbeendroppedintowater.
ReturntheproducttoyourGraham-Fieldauthorized
distributorforexaminationandrepair.
7. Keepthecordawayfromheatedsurfaces.
8. Neverblocktheairopeningsofthisproductorplaceit
onasoftsurface,suchasabedorcouch,wheretheair
openingsmaybeblocked.Keeptheairopeningsfreeof
lint,hairandthelike.
9. Neverdroporinsertanyobjectintoanyopeningsorhose.
10.Neverusethisdevicewhiledrowsyorsleeping.
11.Donotusethisdeviceoutdoorsoroperatewhereaerosol
(spray)productsarebeingused.

JB0112-071-INS-LAB-RevC17 6
12.Donotusethisdeviceinthepresenceofammable
anesthesiaorwhereoxygenornitrousoxideisbeing
administered.ThisdevicehasnoAPorAPGprotection.
13.Connectthisdevicetoaproperlygroundedoutletonly.
14.Thisdevicehasapolarizedplug(onebladeiswider
thantheother).Asasafetyfeature,thisplugwilltinto
apolarizedoutletonlyoneway.Iftheplugdoesnott
fullyintotheoutlet,reversetheplug.Ifitstilldoesnott,
contactaqualiedelectrician.Donotmodifytheplugin
anyway.Donotattempttodefeatthissafetyfeature.
15.Unplugthedevicebeforellingorcleaningthenebulizer.
16.Thisdeviceisoil-free.Donotlubricate.
17.Donotremovethejetnebulizerwhilethedeviceisoperating.
18.Donotleavemedicationinsidethejetnebulizerwhenitis
notinuse.
19.Usethenebulizerintermittently;themaximumoperating
periodis20minutes.Thenebulizermuststandunusedfor
40minutesbeforeuse.
20.Themouthpieceandmasksareforusebyoneperson
only;inordertoavoidinfection,donotsharethem.
21.NoticeforCaliforniaCustomers-CaliforniaProposition65
WARNING:Thisproductcontainsachemicalknowntothe
StateofCaliforniatocausecancerandreproductiveor
developmentalharm.

JB0112-071-INS-LAB-RevC17 7
PRODUCT DESCRIPTION
Features and Components
Please note nebulizer features and components, labeled in
the illustration below.
ON/OFF
switch
stickers
nebulizer
filters
PVCtubing
nebulizercup
mouthpiece
pediatric
mask
Neb-u-Tyke Bella the Butterfly
Nebulizer Compressor
Info: Bella the Butterfly replacement parts are available at
www.grahamfield.com or from your Graham-Field authorized
distributor.

JB0112-071-INS-LAB-RevC17 8
SETUP
Info: Before initial operation, clean the nebulizer
cup assembly following the instructions given in
MAINTENANCE/Cleaning the Nebulizer Cup.
WARNING:Ensurethatthenebulizercompressoris
unpluggedduringsetup.
1. Before each use, inspect the compressor and nebulizer
cup assembly for damage or wear; replace as needed.
2. Place the compressor on a table or other at stable
surface (not on the oor), in an area that is clean and
dust-free. Be sure you can easily reach the controls
when seated.
3. Open the nebulizer medication cup: Twist the cap
counter-clockwise until it stops, then lift cap off.
4. Pour the prescribed medication into the nebulizer
medication cup. Do not ll with more than 5 ml per
treatment.
5. Install cap on cup. Turn clockwise
to tighten.
6. If using the mouthpiece for
treatment, proceed to step 7. If
using a mask, install the bottom
part of the mask directly onto the
top of the nebulizer medication
cup cap as shown at right.
Proceed to step 8.
7. Install mouthpiece onto top of
nebulizer medication cup cap.
install cap
on cup
cap
cup
install mask
on cap
mask
cap
install
mouthpiece
oncap
mouthpiece
cup

JB0112-071-INS-LAB-RevC17 9
8. Connect the tubing to the adapter on the bottom of the
nebulizer medication cup.
9. Connect the tubing's other end to the compressor air
outlet connector.
WARNING:Beforeuse,checkthatallpartsaresecurely
connected.Useofanimproperlyassemblednebulizer
coulddiminishorpreventadequatedeliveryofmedication
andtheeffectivenessofthetreatment.

JB0112-071-INS-LAB-RevC17 10
OPERATION
Nebulizer Operation
s NOTICE:TheNebulizerCompressorisdesignedfor
intermittentuseonly.Donotoperateitcontinuouslyfor
morethan20minutesforasingleusewithoutturningit
offandallowingacoolingperiodofleast40minutes.
1. Ensure that the ON/OFF switch is in the OFF position.
2. Plug the power cord into a properly grounded electrical
wall outlet.
3. Move the ON/OFF switch to the ON position to start the
compressor.
4. Place the nebulizer mouthpiece between teeth and
close mouth (or install the mask on face).
5. As aerosol starts owing, inhale deeply and slowly
through mouthpiece (or mask).
WARNING:Neveruseanebulizerwithacloggednozzle.
Ifthenozzleisclogged,theaerosolmistwillbereduced,
alteredorstopped,whichwilldiminishoreliminatethe
effectivenessofthetreatment.Ifcloggingoccurs,replace
thenebulizer.
Info: If you need to interrupt treatment, move the ON/OFF
switch to the OFF position.
6. When the treatment is complete, move the ON/OFF
switch to the OFF position to turn off the compressor.
7. Unplug the compressor from power outlet.
8. Disconnect nebulizer from tubing.
9. Clean nebulizer as instructed in MAINTENANCE/
Cleaning the Nebulizer Cup.
WARNING:Thisproductcontainssmallpartsthatmay
presentachokinghazardtosmallchildren,andthetubing
maypresentastrangulationhazard.Always use close
supervision when this device is used by or near children
or those who require close supervision.

JB0112-071-INS-LAB-RevC1711
Safety Overload
s NOTICE:Donotexceed20minutesofcontinuous
operation.
1. The motor of this device has a built-in thermal overload
protector. Should the motor overheat, the protector
will automatically shut off the motor. Should this occur,
move the ON/OFF switch to the OFF position and allow
the motor to cool down for approximately 40 minutes
before turning it on again.
2. If the overload protector shuts off the motor frequently,
you may have an unstable voltage situation.
3. If the device shuts down and cannot restart, it may
need to be replaced. Call your Graham-Field authorized
distributor immediately.

JB0112-071-INS-LAB-RevC17 12
MAINTENANCE
Cleaning the Compressor’s Outer Case
WARNING:Electricalshockhazard.Donotremove
theoutercaseofthedevice.Alldisassemblyand
maintenanceofthedevicemustbeperformedbya
qualiedservicetechnician.Referservicetoqualied
servicepersonnel.
WARNING:Thisdevicedoesnotrequireoil.Donot
attempttolubricateanyinternalparts.
WARNING:Unplugdevicebeforecleaning.Donot
submergeinwaterforcleaning.
Wipe the compressor's outer case with a clean, damp
cloth weekly to keep it dust-free. Do not use any powdered
cleaners or soap.
Cleaning the Nebulizer Cup
All parts of the disposable nebulizer, except tubing, should
be thoroughly cleaned after each use. If your physician
species a specic cleaning procedure, follow that
procedure.
WARNING:Topreventpossibleriskofinfectionfrom
contaminatedmedication,cleanthenebulizeraftereach
treatment.
1. After each use, disassemble the nebulizer cup and
mouthpiece completely, wash all parts in warm soapy
water, and rinse thoroughly. Allow to air dry. Store in a
clean, dry location.
2. Once daily, after washing as above, soak nebulizer
components in a fresh solution consisting of one part
warm water and one part white vinegar for 30 minutes
followed by a warm water rinse. Allow to air dry. Store
in a clean, dry location.

JB0112-071-INS-LAB-RevC17 13
Replacing the Filter
s NOTICE:Changethelteronceevery30daysorwhenthe
lterchangescolor.
Info: Additional filters can be purchased at
www.grahamfield.com or from your Graham-Field authorized
distributor.
1. Ensure that ON/OFF switch is in
the OFF position. Open the lter
cap located on the bottom of the
nebulizer.
2. Remove the dirty lter with a small,
pointed object. Discard the lter.
3. Replace with a clean lter.
4. Replace the lter cap.
s NOTICE:Donotusecottonorothermaterialsasalter.
s NOTICE:Donotwashorcleanthelter.
s NOTICE:Donotreusethelter.Donotoperate
compressorwithoutalter.Compressordamagecan
result.
WARNING:Thedisposablenebulizerkitisforsingle
patientuseonly.
Storage
Keep the device and accessories dry. Avoid exposure to
direct sunlight.
filtercompartment
(shownwithoutfilter)

JB0112-071-INS-LAB-RevC17 14
TROUBLESHOOTING
If your Nebulizer Compressor fails to function, consult the
following Troubleshooting guide. If the problem persists,
consult your Graham-Field authorized distributor.
Problem Cause and Solution
Device doesn’t
operate
Check if plug is properly t into a grounded electrical outlet.
When device has been run continuously for more than 20 minutes,
auto-shut-down may have been activated by the built-in thermal
protector; ensure that device is OFF for at least 40 minutes before
next usage.
Nebulization is
weak
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if lter is dirty; replace lter as needed.

SPECIFICATIONS
Specication Description
Electrical requirements 120 VAC 60Hz
Current 1.1A
Fuse 250V, 5A
Power cord Black, 7 ft (84" / 2,184 mm)
Plug 2-prong
Max pressure (compressor) 42 psi
Max ow rate 8 LPM
Nebulization rate .18 ml/min
Particle dimension .5 - 10 microns
Capacity 5 ml
Filter Sponge
Dimensions (L x W x H) 6.34" x 9.1" x 3.7" (161 mm x 231 mm x 93 mm)
Product weight 2.88 lb (1.31 kg)
Sound level < 60 dBA
Operation mode 20 minutes On / 40 minutes Off
Environment Temperature Operation: 50°F to 104°F (10°C to 40°C)
Storage: -4°F to 140°F (-20°C to 60°C)
Humidity Operation: 10% to 95% RH
Storage: 10% to 95% RH
JB0112-071-INS-LAB-RevC17 15

JB0112-071-INS-LAB-RevC17 16
LIMITED WARRANTY
SCOPE OF WARRANTY
GF Health Products, Inc. (“GF”) warrants to the original purchaser only that it
will replace or repair components, at GF’s sole discretion, that are defective
in material or workmanship under normal use and service. All warranties
are conditioned upon the proper use of the products strictly in accordance
with good commercial practice and applicable GF instructions and manuals,
including proper use and maintenance. To the extent that a component is
warranted by a third party, GF conveys all of its rights under that warranty
to the original purchaser, to the extent permitted. This limited warranty shall
only apply to defects that are reported to GF’s customer service team within
the applicable warranty period and which, upon examination by GF or its
authorized representative, prove to be a warranty item. This limited warranty
is not transferable.
The warranted components and time period are set forth below:
Compressor ...............................ve years
Accessories................................six months
The applicable warranty period shall commence from date of shipment to
the original customer, unless there is an expiration date on the component in
which case the warranty shall expire on the earlier of warranty period or the
expiration date.
OBTAINING WARRANTY SERVICE
This limited warranty shall only apply to defects that are reported to the
Distributor from whom the Customer purchased the product within the
applicable warranty period. If there is not a Distributor, you must contact GF
directly by calling 1-770-368-4700, sending a fax request to 1-770-368-2386, or by
e-mailing a request to cs@grahameld.com. Specic directions will be provided
by the Customer Service Representative. Failure to abide by the specic
directions will result in denial of the warranty claim.
EXCLUSIONS
The warranty does not cover and GF shall not be liable for the following:
1) Defects, damage, or other conditions caused, in whole or in part, by misuse,
abuse, negligence, alteration, accident, freight damage, tampering or failure
to seek and obtain repair or replacement in a timely manner;
2) Products which are not installed, used, or properly cleaned and maintained
as required in the ofcial manual for the applicable product;
3) Products considered to be of a non-durable nature including, but not limited
to: casters, lters, fuses, gaskets, lubricants, and charts;
4) Accessories or parts not provided by GF;
5) Charges by anyone for adjustments, repairs, replacement parts, installation
or other work performed upon or in connection with such products which
are not expressly authorized in writing, in advance, by GF;
6) Any labor or shipping charges incurred in the replacement part installation
or repair;
7) Costs and expenses of regular maintenance and cleaning; and
8) Representations and warranties made by any person or entity other than
GF.

ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL
DAMAGES DISCLAIMER
THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES
OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. IF ANY MODEL OR SAMPLE WAS
SHOWN TO THE CUSTOMER, SUCH MODEL OR SAMPLE WAS USED MERELY
TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF THE PRODUCT AND
NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM
TO THE MODEL OR SAMPLE IN ALL RESPECTS. THIS WARRANTY IS LIMITED
TO THE REPAIR OR REPLACEMENT OF THE DEFECTIVE PARTS. GF SHALL NOT
BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT,
INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT
NOT LIMITED TO: DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE,
DOWNTIME, COVER, OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES,
PAYMENTS AND BENEFITS.
The warranties contained herein contain all the representations and
warranties with respect to the subject matter of this document, and supersede
all prior negotiations, agreements and understandings with respect thereto.
The recipient of this document hereby acknowledges and represents that it
has not relied on any representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this document.
For additional information on this product or this warranty, please contact a
GF Customer Service Representative.
NOTES:
1) Additional terms and conditions may apply.
2) Freight claims must be notated on the appropriate shipping documents and
must be made with immediacy. International, federal and state regulations
govern specic requirements for freight claims. Failure to abide by those
regulations may result in a denial of the freight claim. GF will assist you in
ling the freight claim.
3) Claims for any short shipment must be made within three (3) days of the
invoice date.
JB0112-071-INS-LAB-RevC17 17

1.770.368.4700
Information contained herein is subject to change.
The most current and complete product information can be found on our website.
www.grahameld.com
© 2012, GF Health Products, Inc. All Rights Reserved.
Bella the Buttery, Graham-Field, John Bunn, and Neb-U-Tyke
are trademarks of GF Health Products, Inc.
GF Health Products, Inc. is an ISO 13485 Certied Company.
Manufactured for GF Health Products, Inc.
Made in China

Lea antes de operar el nebulizador. Guarde este
manual para referencia en el futuro.
Info: La más reciente versión de este manual podrá
La Ley Federal (USA) restringe este aparato para
ventas por la orden de su médico. Este aparato no
debe ser utilizado si el operador no ha sido adiestrado
Neb-u-Tyke®
™
Compresor de Nebulizador
Modelo JB0112-071
Manual de Usuario
JB0112-071-INS-LAB-RevC17

JB0112-071-INS-LAB-RevC17 2
CONTENIDOS
INTRODUCCIÓN................................................................................................... 3
USODELCOMPRESORDENEBULIZADOR.............................................. 3
SÍMBOLOS.................................................................................................... 3
DIRECTIVASIMPORTANTESDESEGURIDAD................................................. 4
DESCRIPCIÓNDELPRODUCTO........................................................................ 7
CARACTERÍSTICASYCOMPONENTES.................................................... 7
ENSAMBLAJE...................................................................................................... 8
OPERACIÓN....................................................................................................... 10
OPERACIÓNDELCOMPRESORDENEBULIZADOR.............................. 10
SEGURIDADENCASODESOBRECARGA............................................. 11
MANTENIMIENTO............................................................................................. 12
LIMPIEZADELACARCASADELCOMPRESOR..................................... 12
LIMPIEZADELATAZADELNEBULIZADOR........................................... 12
INSTRUCCIONESPARAELCAMBIODELFILTRODEAIRE.................. 13
ALMACENAJE............................................................................................ 13
DETECTANDOPROBLEMAS............................................................................ 14
ESPECIFICACIONES......................................................................................... 15
GARANTÍALIMITADA........................................................................................ 16
Este producto ha sido examinado y aprobado con los siguientes
estándares:
UL1431
EN 60601-1
Table of contents