Jorvet J1550 User manual

J1550
JorVet ICU
User Manual
Professional Veterinary Equipment Supplier
Jorgensen Laboratories, Inc
1450 N. Van Buren Ave.
Loveland, CO 80538
(800) 525-5614
www.jorvet.com

Overview
Summary of Important Notes
Product Use
Cleaning and Maintenance
01
02
02
02
05
06
09
14
17
JorVet J1550 ICU Unit
01
Overview
The Instruction Manual is used for JorVet J1550 ICU Unit. Installation, use, cleaning, maintenance
and repair of the ICU, as well as principles and functions of the product, caution and warnings,
common fault analysis and troubleshooting and after-sales services etc. are all in the Overview. Users
should operate the product in accordance with requirements as stated in the Instruction Manual. All
staff who use the equipment shall read the Instruction Manual carefully before turning on the
equipment. If you have difficulty in understanding the Manual, please contact a local JorVet product
distributor or the company for more detailed information.
JorVet ICU Unit is a newly developed pet ICU developed to serve an enormous need and fill a
vacancy in the domestic pet care industry. This moderately sized piece of pet medical equipment will
easily fit into most veterinarians’ practices and pet hospitals. Its characteristics and compact structural
design function perfectly to serve its purpose. The mechanism adopts "dual duct and double
circulation convective thermal regulation" to create an excellent and bacteria-free therapeutic
environment and young pet incubation environment similar to that provided by pet parent
(accompanied by purified air and proper temperature and humidity control).
The machine is equipped with an intelligent alarm device for five faults (sensor, extreme temperature,
excessive moisture, indoor oxygen concentration, lack-of-water in the tank and power supply
Interruption). This alarm system is tested and proven to be safe and reliable. Furthermore, it is
equipped with a unique cabinet structure and double-layer thermostat cover, which can prevent loss
of heat and radiant heat when nursing the pets. In addition, a temperature and humidity display device
is provided, measuring current temperature with the set temperature clearly visible. The heating
control adopts dual duct and double circulation convection heating technology. This prevents rapid
change in temperature minimizing the shock that can be caused. It ensures that temperature of the
ICU can rise evenly and gently. The Thermostat System adopts PID control technology, which can
solve the problem of inertia temperature error comprehensively, thus the temperature control is more
accurate and more energy efficient.
CONTENTS

1. Please read the Instruction Manual carefully before using the machine. Nurses (and other pet care
workers) should be trained to operate the machine under the guidance of qualified medical care personnel
who are familiar with the functions and understand the risks and advantages of the machine.
2. Please count the parts and spare parts when unpacking, and read the Instruction Manual carefully.
3. The equipment shall be placed in a clean place with limited changes in temperature and humidity, protected
from direct sunlight and away from heat radiation.
4. To ensure the proper operation of the Pet ICU, please fill the bottle with distilled water or purified water.
5. After the machine is turned on, please preset the temperature value, and warm it up for 3-5 minutes before
use. Do not use tap water.
6. Please do not load weight on top of the machine, and the maximum load capacity of the machine housing is
1-2 kilograms (2.2 to 4.45lbs).
7.The machine is equipped with an intelligent alarm system sensor for five faults; extreme temperature,
excessive moisture, indoor oxygen concentration, lack-of-water in the tank and power supply Interrup-tion.
This alarm system is tested and proven to be safe and reliable. If the machine fails, please inspect the
systems immediately to find out reasons and eliminate the faults. If the faults cannot be eliminated in a
short time, please stop medical treatment or care for pets, and ask for professionals or personnel from the
JorVet Company to repair the machine.
8. Allow for the machine to have proper ventilation. Do not block the air outlet and return air inlet.
9. After being powered on, if the machine is not used, in order to ensure safety and energy-saving, please
turn the switch to off position and allow for sufficient power down before turning the switch back to on
position.
10. Do not use alcohol or other organic solvents to wipe the thermostat cover of the tank. Keep it moisture
free.
11. The nebulizer should be unplugged as soon as the nebulization is over.
12. When the LAC alarm can be seen on the humidity display panel, the tank inside should be filled to the
maximum level even if the tank has water half of the maximum level.
1. Provides bacteria free and thermostatic incubation cocoon for young pets.
2. Allows for body temperature recovery, fluid infusion, atomization treatment, incubation, health
improving and comfortable, stable stay in the hospital (for observation for sick, dying and weak pets).
Built-in humidifier bottle: some parts of medical atomizer and the tank are separately packed. Please
pay attention not to damage or miss any parts or spare parts when unpacking.
JorVet J1550 ICU Unit cabinet
Nebulizer
Power cord
Waste tray
Water purification filter material (Phosphorus-silicon crystal) User Manual
External Humidifier (optional)
As shown in Figure 4.2.1, the Humidifier Vessel is installed in the right control chamber of the JorVet
J1550 ICU Unit. As shown in Figure 4.2.2, open the outlet cover, take out the built-in Water Vessel, fill
the built-in Water Vessel with water, (using distilled water or purified water)as shown In Figure 4.2.2.
And then put back the vessel and place the cover in the proper position.
Notes:
a) The Humidifier and Water Level Sensor should be placed securely onto the vessel to prevent the
spill of water.
b) Do not fill the vessel with purified or distilled water, which may cause the malfunction of the Water
Level Control of the water tank and may eventually cause water to spill or leak.
c) The water level should not be above the Highest Level Line (HLV).
d) The purpose of the water purification filter material is to remove hard water scale and improve the
water quality. The validity period of each placement is 10-12 months. Colorless, tasteless and non-tox-
ic, the filter material meets the National Sanitary Standard for Drinking Water& GB1499- 1988 and can
effectively improve water quality.
2. Installation of Humidifier Vessel
Outlet Cover
Sensor
of Vessel
Water Vessel
Figure 4.2.1 Figure 4.2.2
02 03
Summary of Important Notes
Product Use
Product Installation
The external Nebulizer can be installed on the left
side of JorVet J1550 ICU Unit. Insert the power plug
into the socket on ICU; insert the nozzle of
Nebulizer into the port on the left side as showed in
Figure 9.1.1.
Before nebulization, switch on the nebulizer by
pressing the switch on the nebulizer and then turn
to the operation panel of ICU.
Figure 9.1.1

This External humidifier is helping the internal humidifier in our ICU to reach desired RH level more
easily, especially in extremely dry environment or under high temperature. The power plug should be
plugged into two flat pins receptacle at the rear left corner on the back of the ICU unit, it will be
regulated by the ICU motherboard with special programs installed. (Note: Do not confuse this socket
with another socket for nebulizer, which is next to this socket for humidifier.) Remove the lid of the air
inlet at the back of the ICU, expose the portal, and connect the flexible hose from this humidifier to
the ICU air inlet. The end with large diameter is to the back portal of ICU and the end with small
diameter it to the top portal of the humidifier.
After the unit is plugged into either ICU power supply outlet or wall mains, you can start the machine.
Rotate clockwise the controller to switch on the unit and the blue indicator lamp will be on.
The unit starts working. Rotate the controller knob to adjust the volume of the vapor. Fog can be seen
from the outlet port.
If a warm vapor is needed, press the button below the power button to switch between the default cool
mist mode and warm mist mode.
Remarks: the “warm mist mode” will be needed when the ICU unit is set at “high temperature + high
humidity”, like greater than 28℃+ 60%RH, the warm mist will definitely help humidifying more quickly.
Note:When LAC alarm can be seen on the humidity display panel, the external and internal
humidifiers will both stop working. Fill water in the internal water tank to the maximum level and both of
the humidifiers will start to work.
In certain cases, such as avian or other exotics, a desired environment of higher humidity and
temperature is preferred. In order to achieve these levels you must have increased temperature and
increased humidity. For instance, at 100F the humidity can not go beyond 50% relative humidity
without the external humidifier.
The external humidifier is not necessary when using the unit on standard mammalian patients like
canine and feline.
(optional J1550D1)
04 05
The patented Energy Saving Low Noise Compressor (less than 25 decibels) generates high speed air
jet from filtered clean air (by anti-bacterial cotton), the air stream disperses the medicine solution from
m minus which could reach deep respiratory system like alveolus and fine bronchus. The medicine is
nebulized evenly, no diluted needed, so there is no residues left almost and could achieve very high
medicine utilization rate. This device is preciously made medical equipment with very low faulty rate
and very good durability, could serve for 1000 -1500 hours as minimum lifespan.
Operation Instructions:
The nebulizer is easy to operate, please follow the simple steps as follows.
1. Install the nebulizer on the left side of ICU. Insert the power plug into the socket on the ICU.
2. Screw off the cup lid and nozzle; pour medicine liquid into the nebulizer cup.
3. Screw on the cup lid and nozzle, and put the nebulizer back.
4. Insert the nozzle of Nebulizer into the port. Turn on the nebulizer by pressing the switch on the
nebulizer. And then press Nebulize button on control panel of ICU to switch on nebulization, the
medical fog will be sprayed through grid and get into the ICU chamber. There are two stages of
nebulization treatment timing, in which the basic time is 10 minutes. Once the nebulization process
started, it is 10 minutes countdown automatically; pressing the "-" button on the nebulize adjustment
button will turn it into 20 minutes mode, pressing the "+" will turn back to 10 minutes mode. Pressing
Nebulize button again on the main control panel when the nebulizing process is running will switch the
nebulizer off.
a) The treatment process should not exceed 20 minutes for every nebulizing, misuse would possibly
cause Pulmonary edema.
b) To prevent superinfection, please pay close attention to the sterilization of nebulizer, ambient air and
all the related equipment, also disinfect animal mouth, nose and throat.
Mechanism of action
Animal respiratory system is an open system, where alveolus is the main air-exchanging place. The
object sprayed out of the nebulizer is mainly the mixture of oxygen and medicine – oxygen-medicine
aerosol. This kind of aerosol could improve oxygen metabolism and anoxic metabolism significantly;
increase the oxygen reserve in organisms. The cavitation effect of compressed air could disperse the
liquid medicine into gas phase and turn the liquid into fog form. The suspended particles will get deep
into animal respiratory system and treat the focus of infection directly. The feature of air compressing
even fog could go very deep into bronchus and the particles could settle down in deep
respiratory system, achieves painless cure effect.
Tips:
a) The smaller the particles are, the better curative effect the nebulization could achieve.
b) Normally the fog with bigger particles generated from traditional ultrasonic nebulizer would be
easier to be observed, as the fog seems to be thicker.
Air Compressor Nebulizer uses high pressure gas jet knockout the medicine solution into tiny
particles and carry them into deep respiratory system. The size of droplet will affect the settlement site
of medical fog.
roplet Sie Selement Site
15m Fine roncuslveolus
520m roncus
2040m Upper espiratory Tract mainlyNoseTroatetc
Unscrew the tank cover Add water
Take the water tank out of the base, take away the Valve-cup, and fill the water tank with clear water.
Close the Valve-cup, and put the water tank on to the base carefully.
Nebulization
NOTE: NEVER RUN THE UNIT WHEN THERE IS NO WATER IN WATER
BASIN. Usage Warning:

06 07
Without being heated, high-frequency oscillation technology is adopted to product mist. Compared to
heating atomization, it can save 90% of energy. During atomization, large amounts of negative ions are
released, and have static electricity reaction with smoke and dust in the air, and cause precipitation.
According to the data released by World Health Organization, negative-ion is called “air vitamin”, and
an animal needs about 10 billion ecological negative-Ions every day, while our living environment can
only provide about 0.1-2 billion ones; when the negative-ion concentration in the air is up to 400
ones/cm³, it can satisfy our basic needs for survival; when the negative-ion concentration in the air is
up to 1000-2000 ones/cm³', it can satisfy our basic needs of health; when the negative-ion concentra-
tion in the air is up to 5000-10,000 ones/cm³, it can Improve our immunity and antibacterial ability; and
when the negative-ion concentration in the air is up to 300,000-500,000 ones/cm³, it can Improve our
natural healing ability, and many diseases can be healed without medication. Negative-ion plays a
significant role in improving the air quality of treatment room and improving and maintaining the health
of animals.
JorVet ICU Unit provides a high concentration of negative-ion generator, and the following is the
negative-ion test report:
Heating control adopts dual duct and double circulation convection heating technology to
allow temperature rising quickly, and temperature shock is small to ensure the temperature of the
ICU can raise evenly.
JorVet J1550 ICU Unit
MODEL NO Project Number Applicable specification and model
DC9-12V 1403140013
I Electrical Parameters
Input voltage/frequency DC12V
Output voltage At DC12V -0.5--4.5KVd.c.(-3.5KV DC ± 0.5KV DC )
Anion concentration 6 PCS/cm3 ±10%
Insulation resistance
Dielectric streangth Power line external box AC 1800V
60HZ 5mA 3S No breakdown
Functions of the Product
Tips:
a) There is hissing noise when nebulization starts, that’s generated from the high speed air flow. The
patented low noise air compressing technology ensures no vibration and no electronic noise when
nebulizer runs. The noise is much lower than normal level of most air compressor nebulizer, which is
around 85 decibels normally.
b) When nebulization is ON, the ventilation fans will be turned off to ensure good nebulizing effects,
and humidity and temperature control will be turned off too. The humidity alarm will Not be triggered in
nebulization mode, which will reset in 3 minutes after nebulization function turned off.
1. Mechanism of ICU
The air circulation system as shown in Figure 5.1.1 is driven by the circulating air of After being filtered
by the purifier, the air from the outside enters into the ICU, and "dual duct and double circulation
convection heating technology” is adopted to create an excellent and bacteria-free ICU therapeutic
environment and young pet incuba-tion environment. It provides clean air and proper temperature
and humidity.
2. Main Technical Indicators
1) Power supply voltage: 110V /60Hz
2)
3) Ambient temperature: 10°C ~ 35°C
4) Temperature control range
Cabinet temperature: 15°C ~ 38°C (can be up to 39°C by special operations)
5)
6)
7)
8) Temperature rise time: 5min ~ 20min
9)
10)
11) Withstand voltage: 1500V/50Hz, lasted one minute without breakdown and flashover.
12) Ambient conditions:
①Transportation and storage:
a. Ambient temperature: -10°C~ 40°C
②Operating conditions:
a. Ambient temperature: 18°C ~ 30°C
b. Relative humidity: 40% ~ 65%
Figure 5.1.1
Product Principles and Functions

Nebulizer
The above test report shows that: the nearby negative-ion concentration of JorVet J1550 ICU Unit is up
to 6 million ones/cm
3
standard, which has reached the medical-level Negative-ion emission standard.
It is truly air purifying, and is effectively helpful for the rehabilitation and recovery of pets.
A UVC ultraviolet ray is the one that truly has germicidal effect, and C ultraviolet of which is easily
absorbed by DNA of organisms, especially, ultraviolet ray of 253.7nm is the best. UV sterilizer is a
purely physical method of disinfection, which is characterized by simple and convenient operation, high
efficiency, no secondary pollution, easy for management and automation etc.
JorVet ICU Unit uses ultraviolet sterilizer, which is characterized by strong radiation
sterilization
reaction chamber design is adopted. Compared with similar products from abroad, the sterilization
strength increases by 18% - 27%.
---- Generally, the bactericidal action of ultraviolet ray on bacteria and viruses can achieve sterilization
rate of 99% - 99.9% in one to two seconds.
---- The broad-spectrum of ultraviolet sterilization is the highest, and it can efficiently kill almost all
bacteria and viruses.
---- The ultraviolet sterilizer has no secondary pollution. No chemicals will be added in ultraviolet
sterilization, thus, water and its surrounding environment will not be polluted secondly. Furthermore, it
will not change any ingredients in the water.
---- The operation of ultraviolet sterilizer is safe and reliable: traditional disinfection technologies such
as chloride or ozone, the disinfectants themselves are highly toxic and flammable substances, while
there is no such potential safety hazard in ultraviolet sterilization system.
JorVet ICU Unit is equipped with an external silent medical atomizer, and biomedical engineering
research shows: when the atomized particles are about 10 microns, medicines can only deposit in
the mouth and throat, when the atomized particles are about 3-6 microns, medicines can enter into the
bronchus, and when the atomized particles are less than 2 microns, medicines can enter into the finest
bronchus and alveoli. The JorVet J1550 ICU Unit can produce 0.5-4 micron atomized particles of
medicine. Meanwhile, intelligent control quantity of atomization is provided for your selection,
and the maximum atomizing volume is up to 0.5ml/min. The atomization in the ICU is even,
and the inhalation effect of atomized medicines is obvious. The medical atomizer is
characterized by low noise, which complies with pet ICU noise standard.
An increase in carbon dioxide concentration is accompanied by a decrease in oxygen concentration. Being
anoxic under low oxygen and high concentration of carbon dioxide for a long time will cause pulmonary
edema, cerebral edema, metabolic acidosis, electrolyte imbalance, shock, hypoxic encephalopathy, etc.
When the carbon dioxide concentration reaches 1000PPM above, it will cause dizziness, drowsiness and
reduced resistance to pets; when the carbon dioxide concentration reaches 2000-5000PPM, it will cause
mild nausea, poor breathing, rapid heartbeat and electrolyte imbalance to pets, and seriously affect the
health of pets and their postoperative rehabilitation; and when the carbon dioxide concentration is over
5000PPM, it will cause severe hypoxia, coma and even death.
08 09
The negative ion generator boosts low voltage to negative high voltage DC through the voltage
booster circuit. DC high voltage from the tip of carbon brush is used to produce high corona and
lots of high-speed electrons are released. Since electronics cannot exist in the air for a long time
(the life of existing electrons is n S level), they will be captured by oxygen molecules (O2) in the air
immediately, and then negative ions are formed; the working principle of which is the same as
natural phenomenon (negative ions are produced when there is thunder and lightning).
Test equipment: KEC-990 air ion concentration tester, HG3830 high voltage test bench
In case of no electricity, no abnormalities occur after
being placed for 72 hours at ambient temperature of
+25°C and relative humidity of 95%.
In case of no electricity, no abnormalities occur after
being placed for 72 hours at ambient temperature of
+70°C
In case of no electricity, no abnormalities occur after
being placed for 72 hours at ambient temperature
of-20°C.
Starting at 100cm (39.37”) from the ground, free fall onto
the board, it will still function properly.
Drop test
II Principle of Operation
Low-temperature
characteristics
High-temperature
characteristics
Humidity resistance
Storage temperature
Operating humidity range
Operating temperature range
-20°C~+70°C
40%~85%
-10°C~+40°C
Rated power
Input current
<1W
<80mA

2
Fahrenheit/centigrade can be switched and
displayed with a toggle key. The default tempera-
ture is shown in fahrenheit and in two-digit integer
and one decimal place after which is a letter
showing special indication at the time of setting
the temperature parameter. Switch the
Fahrenheit / centigrade indicator lamp. When the
indicator lamp displays "C", it means centigrade;
when it displays "F", it means Fahrenheit. See fig.
7.2.1.
Humidity / O2concentration display can be
switched with the humidity / O2 concentration key.
The default is humidity display in two-digit integer
and one decimal place. O2concentration is
displayed in four-digit integer. Switch the humidity /
O2concentration indicator lamp. When the
Indicator lamp is "HU", it means humidity; when the
indicator lamp is“O2”, it means O2 concentration.
See fig. 7.3.1.
Key control: press "SET" key plus temperature " + / - " key at the same time to adjust the temperature
set point, and regulate it by ± 0.1 stepping. See fig. 7.4.1.The factory setting is the set point of 68°F.
It can be adjusted down to 59°Fand up to 104°F;
Smart alarm function:
Alarm for beyond the maximum set point: the temperature and the upper dash are displayed alternate-
ly every three seconds with alarm and flashing temperature indicator lamp.
Alarm for beyond the minimum set point: the
temperature and the lower dash are displayed
alternately every three seconds with alarm and
flashing temperature Indicator lamp.
Damaged temperature sensor: flashing "ERR"
display and flashing temperature indicator lamp in
the temperature display.
Figure 7.4.1
Figure 7.3.1
Figure 7.2.1
Hold
10 11
The JorVet J1550 ICU Unit is equipped with a precise oxygen concentration monitor, which
can monitor changes in oxygen concentration in the ICU. When being hypoxic state that
concentration falls beyond the standard value (exceeds the set value), and cannot exceed the set
value for three minutes, It will trigger the alarm system and start the intelligent ventilation system at
the same time, being ventilated with the outside air and thus improve the air quality of ICU,
which becomes the most important life defense line for pets.
JorVet J1550 ICU Unit uses ten-level LED lights adjustment, which can simulate various scenes
similar to parent, forest oxygen bar, warm home, forest dawn and dusk twilight, etc., providing a cozy
and conformable therapeutic environment for pets.
JorVet J1550 ICU Unit adopts intelligent alarm system, and is equipped with
corresponding fault display function, such as extreme temperature and excessive moisture
alarm, lack-of-water alarm and main components fault warning lights.
JorVet J1550 ICU Unit is equipped with intelligent switch power keys, dividing into three controls, i.e.
power-on, start and power-off. This design is energy saving and safer to use.
All equipment components of JorVet J1550 ICU Unit adopt integrated design. All compo-nents are
integrated in a control frame, and completely separated from the control circuit. Only by
disassembling the control rack, the whole machine can be disassembled to clean various
compo-nents, including fan, medical devices, heat sink and filter etc. The operation is extremely
convenient, easy for troubleshooting, maintenance and equipment cleaning.
JorVet J1550 ICU Unit is equipped with two medical device import Interface channels, thus
external medical equipment (such as oxygen machine) can be connected into the ICU,
meanwhile, the side access port adopts medical valve importing design, being convenient for
the importing of pets injection and infusion needles and relative sealing and fixing.
a) One end of the power cord is plugged into the
terminal board at the bottom support of the right side of
the machine, and the other end into the socket of
1100V / 60Hz with reliable grounding. See fig. 7.1.2.
b) The power switch is located at the batter brace of
the left side of the machine. See fig.7.1.2. It is the
on/off switch of JorVet ICU Unit. When it is locked,
all other keys are void; when it is unlocked, all other
keys are valid.
c) The lock indicator lamp is at the on/off switch.
When it is on, it means unlocked; when it is off, it
means locked.
Figure 7.1.2
Power ON/OFF
Power Interface
Product Operation Instructions

Control program:
a) When the setting temperature is higher than the measured temperature, start the PID precise
heating system and left/right fans. When the measured temperature is higher than the setting tempera-
ture, stop heating step by step.
b) When the measured temperature is higher than the setting temperature, start the PID precise
cooling system (need additionally purchased cooler) and the left fan to realize ventilation and air
exchange inside and outside. When the measured temperature is lower than the setting temperature,
stop cooling step by step and shut down fans after one-minute delay.
c) When there is no additionally connected cooler, try to place an ice tube inside the atomizing
chamber or at the sink as auxiliary cooling. The system will start the right fan and the left fan to realize
ventilation and cooling.
Indicator lamp control:
Heating---the temperature indicator lamp is red
Cooling---the temperature indicator lamp is blue
Delay---the temperature indicator lamp is yellow
Before humidification, check if the water is enough
in the internal water tank.
If you have an external humidifier and need high
humidity in high temperature or dry environment,
just install the external humidifier and start it before
setting.
Key control: press "SET" key plus humidity "+/-" key
at the same time to adjust the humidity set point,
and regulate it by±0.1 stepping. See fig. 7.4.1. The
factory setting is the humidity set point of 55%. It
can be adjusted down to 40% and up to 65%.
Alarm function:
a) Alarm for beyond the maximum set point: the humidity and the upper dash are displayed
alternately every three seconds with alarm and flashing temperature indicator lamp.
b) Alarm for beyond the minimum set point: the humidity and the lower dash are displayed alternately
every three seconds with alarm and flashing temperature indicator lamp.
c) Damaged humidity sensor: flashing "ERR" display and flashing humidity indicator lamp in the humid-
ity display.
d) Internal humidifier water shortage alarm (the water level sensor of the humidifier is disconnected):
in this case, if the humidifier is started, the humidity Indicator lamp is off with water shortage alarm, and
the humidity display shows "LAC". Meanwhile, it starts the smart water shortage protection system.
Control program:
a) When the setting humidity is higher the measured humidity, start the humidifier and the right fan.
b) When the measured humidity reaches the setting humidity, stop humidifying.
c) When the measured humidity is higher than the setting humidity, start dehumidifying and the left fan;
reversely start the right fan for air exhaust as auxiliary dehumidifying. If an additionally purchased
cooler is connected, start the cooling function for dehumidifying.
Indicator lamp control:
The humidity indicator lamp is always on at the time of humidifying.
The humidity indicator lamp is flashing at the time of dehumidifying.
Figure 7.5.1
Hold
Press "anion" key to switch it ON; and press it
again to switch it OFF.
When the anion is open, the anion indicator lamp
is on; when the anion is closed, the anion indicator
lamp is off. See fig. 7.6.1.
The atomizing volume of the medical atomizer has two grades, measuring by time and quantity. Press
"medical atomization" key to start the atomizing function (it is the higher grade with the atomizing
volume of 0.375ml/min and the auto-timing of 10 minutes). Press "atomizing volume regulation" key to
switch between the higher grade and the lower grade of atomizing volume. The lower grade is
0.5ml/min with the auto-timing of 20 minutes. When the medical atomization is started, the atomization
indicator lamp is flashing quickly (correspondent to the higher grade). Press "regulation" key to adjust
it to the lower grade at the time of which the indicator lamp is slowly flashing. Press "medical atomiza-
tion" key again to stop atomization, and the atomization indicator lamp is off.
a) Press "atomization" key to start the higher grade; press atomizing volume regulation key "+". To
adjust it to the lower grade, press atomizing volume regulation key "-". See fig. 7.7.1.
b) Press "atomization" key and get into the medical atomization status (higher grade) during which the
humidity display and control are shielded. The humidity display shows accumulated atomization time
counting (in seconds). After atomization and 10 sec delay, it automatically exits the atomization mode
and reverts to the humidity control mode.
c) During atomization, press "atomization" key to manually stop the atomization mode and revert to the
humidity control mode.
d) In the atomization mode, the humidity control is not effective. Therefore, when the humidity exceeds
the maximum set point, there is no alarm. If the humidity still exceeds the maximum set point in three
minutes after it exits (closes) the medical atomization mode, the alarm is started.
Atomizer water shortage alarm: when the water tank or the medicine cup of the atomizer is short of
water, stop atomization. The humidity display shows "ELAC" and the atomization lamp is off with water
shortage alarm. Press "atomization" key again to enter the "atomizer off" status and recover humidity
display.
Figure 7.6.1
Figure 7.7.1 Figure 7.8.1
Hold press once message come out
Press once for start
12 13

Alarm and Processing Control: when the Oconcentration is higher than the set point and remains
constant for some time (to avoid false alarm for an abrupt increase of partial concentration), the O
Concentration Indicator Lamp is flashing with the alarm. Start both the left fan and the right fan to
enhance ventilation inside and outside. If the Oconcentration remains unchanged, open the ICU door
to enhance ventilation. When the Oconcentration is equal to or lower than the set point, switch off the
Oindicator lamp and close the alarm. If fans are not needed when other functions are working the
fans automatically stop.
1. Disinfection & Sterilization: after incubation or caring of one pet finishes treatment and care, it is
suggested to disinfect and sterilize this machine. Do this once every week when not in use.
2. Separate cleaning of the Humidifier
If the Humidifier fails to produce moisture, or the water spills out, take the following measures:
a) As shown in Figure 8.2.1-a,open the cover plate of the right Control Rack.
b) As shown in Figure 8.2.1-b and Figure 8.2.1-c,remove Water Level Sensors and Humidifier from the
Water Vessel. Take out the Water Vessel.
c) As shown in Figure 8.2.1-d, empty the Vessel. Brush off the mineral substance such as water scale
or verdigris on the Vessel, Water Level Sensors and Humidifier. If the filter of the Humidifier fails to
work, replace with a new filter.
d) Install Water Level Sensor and Humidifier, place the water vessel back and secure the cover plate.
Press "SET" key to remove disengage alarms. See fig. 7.11.1.
Sensor
of Vessel
Water Vessel
Figure 8.2.1-bFigure 8.2.1-a
10. O
Press illumination adjustment key "+/-" to adjust
ten grades of illumination intensity. See fig. 7.9.1.
The highest is grade nine and the lowest is
closed. Under the Initial start-up status, it is
closed.
Key control: press "SET" key for five seconds to enter the parameters setting status. The humidity / O
display shows flashing digits of Oconcentration set value. When the indicator lamp "P" (Oconcentra-
tion) is flashing, it is able to adjust the set value of Oconcentration alarm. Now press humidity "+/-"
key to adjust Oset value by ±1 stepping. The default factory setting is the alarm set point of 21%.
a) On: before starting the sterilization function, make sure that there is no pet in the ICU and the ICU
door is closed. Meanwhile, press "SET" key plus "sterilization" key to enter the disinfection mode. See
fig. 7.8.1.
b) Operation: under this mode, all other function keys are void except "SET" key and "sterilization" key.
Both temperature display and humidity display show "DISI" (abbreviated disinfection); under the
disinfection mode, switch on the UV disinfection lamp. There is violet light inside the ICU (with weak
acrid scent). Meanwhile, start the left fan and the right fan to effect ventilation inside.
c) Off: press "SET" key plus "disinfection" key at the same time again to switch off the UV disinfection
lamp, the disinfection indication lamp and the left fan to exit the disinfection mode and recover the
normal work mode. At this time all function keys are valid except the disinfection key.
Note: when the ICU is under the normal work mode, it is invalid to press "disinfection" key individually.
Figure 7.9.1
Figure 7.10.1 Figure 7.11.1
Hold for 5 seconds
valne for o2alarm
14 15
Cleaning and Maintenance

3. Complete Cleaning and Maintenance: remove and clean parts and assemblies periodically. Desired
results are achieved only by cleaning after removal.
Steps for Disassembling Components:
Remove four screws from the master control rack on the right to take down the right master control
rack, (remove the left master control rack in the same way). See fig. 8.3.1.
Remove power connection plugs on power cords on one side of the master control rack one by one.
Note that plugs have different colors for each electrical component.
Warning: do not remove control racks or electrical components from control racks immediately after
shut down; give it some time to cool down. The heaters in control racks reach high temperature.
Remove control racks or contact heaters after a 30 minutes respite after shutdown.
Remove four screws in the fan box and take out the whole heater. Clean the fan. Try not to remove heat
sinks from the fan, especially PTC heater bands that cannot be removed from heat sinks. See fig.
8.3.2.
Remove and clean filters behind the square plastic section inside the chamber.
Take out the humidifier. If it is working normally, it is not necessary to clean the humidifier. For separate
cleaning of the humidifier, it is not necessary to remove control racks. See the chapter of separate
cleaning of the humidifier above. If the humidifier is definitely damaged or does not work after separate
cleaning, remove the humidifier for cleaning and maintenance. (See fig. 8.3.3):
Sensor
Figure 8.3.1-b
Figure 8.3.1-a
Figure 8.2.1-c Figure 8.2.1-d
a) As shown In Figure 8.3.3-a, Open the cover plate of the Water Vessel of the humidifier (on the
control rack).
b) As shown In Figure 8.3.3-b remove Humidifier from the Water Vessel and as shown In Figure
8.3.3-c, remove Water Level Sensors.
c) Take out the Water Vessel of the humidifier as shown In Figure 8.3.3-d.
Temperature
Humidity sensor
Fan
Fan
PTC Heater
PTC Heater
screws of fan
PTC Heater
Sensor
of Vessel
Water Vessel
Sensor
Figure 8.3.3-a
Figure 8.3.2-a Figure 8.3.2-b
Figure 8.3.3-b
Figure 8.3.3-c Figure 8.3.3-d
4. Cleaning
The ICU must be empty at the time of cleaning. Cleaning is feasible only after removal.
5. Maintenance
It is suggested to maintain the machine every two months during normal operation, which shall be
carried out in accordance with removal, installation and maintenance procedures above.
16 17

After the power switch is pressed,
there is no reaction or the power
indicator lamp is still off
Phenomenon
18 19
1.Bad power connection and
plug connection
2.Damaged switch
Cause analysis Solution
1. Check power connection
2.Replace the power switch
Little atomizing volume form
the medical atomizer
1.Too big atmizing particles of
solution causes little atomizing
volume
2.Solution contains syrup or
colloform
3.Tablets are used for
dissolution
4.Oily impurities on the surface
of solution
1.Too big particles are not
suitable for operation of this
atomizer
2.When solution contains
syrup or colloform,it requires
with dry filter paper
3.When tablets are used for
dissolution,dissolve sediments
to remove impurites first
4.Remove oily impurities on
the surface of solution with
clean tissue firstivvv
No atomizing from the atomizer
1.Water shortage of the water
tank the atomizer
2.Damaged ultrasonic chip of
the atomizer
3.Water scale in the tank of the
atomizer
1.Add water into the water
tank until it is at the same
level of the cup bottom
2.Replace the ultrasonic chip
3.Gently scrub the inside of
the tank with lint
Remaining solution not
atomized in the cup after
atomization of the medical
atomizer
Atomizing time and volume
depend on concentration of
atomizing liquid
Repeat the atomization
operation procedure if not
completely atomized once
The humidity is not increased
after raising the set point of
humidity
1.Water shortage in the bottle
of the humidifier
2.Damaged chip of the
humidifier
3.Water scale in the metal
water level sensor inside the
humidifier tank
1.Add water into the bottle
2.Replace the humidifier
3.Remove water scale inside
according to the chapter of
“separate cleaning of the
humidifier”, or try to dip some
vinegar into the water bottle of
the humidifier
Operational keys on the
panel do not work
Bad contact of panel
plugins
Damaged panel keys
1.Install plug-ins properly
2.Replace the panel
No heating
1.Failed or damaged fan
2.Heater (heat sink) with
bad contact or damaged
3.ICU door opened
1.Check and replace fan
2.Replace heater band or
heat sink
3.Close ICU door
No cooling
1.No external cooler
connected
2.Cooling pipe of the cooler
not connected to the tank
3.Signal line of the cooler
not connected to the
machine
1.Connect it to an externally
purchased cooler
2.Check the cooling pipe
3.Plug the signal line into the
colling socket of the tank
Continuous alarm for O2
concentration
1.Failed or damaged fan
2.Not enough instant ventila-
tion inside the chamber
3.damaged O2sensor
1.Chech fans
2.Open the ICU door
3.Replace the O2concentra-
tion sensor
Troubleshooting

JorVet ICU Unit
Dear user:
Thank you for using the JorVet ICU Unit. We provide following services as dictated in this
warranty manual (and the invoice or other certificates of product purchase according to "Law
of the People's Republic of China on Protection of Consumers' Rights and Interests", as well
as "Provisions on the Liability for the Repair, Replacement and Return of Some
Commodities" issued by the CSBTS and the Bureau of Finance.
1. One-year warranty for the whole machine.
2. Two-year warranty for key components (including medical atomizer, humidifier, anion
generator, ultrasonic sterilizer, fan, tank, door, master control panel, PTC heating chip).
3. Six-month warranty (from date of purchase) for replacement of key integral parts (including
medical atomizer, humidifier)
4. Warranty services are not provided under the following situations, (but paid repair services
are available for the life time of the equipment).
[A] No warranty card or any other certificate of product purchase presented
[B] Altered certificate of product purchase
[C] Damage caused by accident, abuse or misuse
[D] Repair privately without authorization of our company
[E] Products can be used after repair beyond the valid period of the three warrantees
5. The valid period of the three warrantees is from the date of product purchase, not including
time of repair and waiting for components if out of stock. During the valid period of three
warrantees, consumers can have products repaired, replaced or returned with invoice or
certificates of product purchase.
6. Within 15 days after purchase, consumers can return, replace or repair products if there is
any performance failure found.
7. Within 30 days after purchase, consumers can replace or repair products if there is any
performance failure found.
8. "Performance failure" indicates that products cannot meet safety and hygiene requirements
with unreasonable danger threatening safety of human body or properties; or products do not
have complete functional performance as required, or cannot meet explicitly expressed quality
conditions.
Advance replacement warranty service:
Your warranty service may involve advance replacement warranty service. If the hardware is
found to be defective after diagnosis, we will send replacement parts directly to you. After
receiving the replacement, you must return the defective part to us within the specified period
(normally five days). The transportation and insurance fees rising from return of the defective
part needs to be paid by you. If you cannot return the defective part, we will ask you for
replacement fees. We provide replacement services during normal working hours,
8:00 a.m. to 5: 00 p.m. Monday to Friday. The office time may be different according to local
time zones.
20
Mail Warranty Service
Your warranty service may involve mail or courier. According to mail service provisions, you must
send defective hardware to an authorized maintenance facility via freight collect. We will repair and
return it to you.
Door-to-door pickup and return warranty service:
Your warranty service may involve door-to-door pickup and return services. The company or a
designated repair point will send technical employees to pick up defective hardware at your
location and send it back after repair. For such services, the company pays all maintenance,
transportation and insurance fees.
User Instructions:
For products beyond the warranty period, either the company sends service employees to pick up
or you can mail the defective part to the maintenance point. We provide our repair or replacement
services and parts (fees for maintenance after warranty) covering shipment costs, labor costs for
repair and component cost. Users can refuse to pay if maintainers cannot comply with unified
charging standards of the company.
Contact Customer Support and Service:
If information provided in these user instructions cannot solve your problem, contact the customer
support of JorVet Company at this website:
www.jorvet.com
Here you can:
1. Communicate with our technical employees online.
2. Use English if technical support is not available in some specific language.
3. Send an email to the customer support of JorVet;
4. Find the global/domestic customer support telephone number of JorVet
5. Find JorVet service center.
21
Table of contents
Other Jorvet Pet Care Product manuals
Popular Pet Care Product manuals by other brands

Wahl
Wahl ZX657 manual

Braecrest Design
Braecrest Design Skye View Post Mount Bird Feeder Kit Assembly instructions

Home Pet Spa
Home Pet Spa RA-07X quick start guide

GEOHOME
GEOHOME I602-40 Assembly instruction

Assisi Animal Health
Assisi Animal Health Assisi Loop 150 Instructions for use

PAPIFEED
PAPIFEED PF005 manual

Patpet
Patpet T600 user manual

METROPOLITAN VACUUM CLEANER CO
METROPOLITAN VACUUM CLEANER CO Master Blaster MB-3 instruction manual

Petsafe
Petsafe ST-70 Operating and training guide

Petsafe
Petsafe Innotek Ultrasmart IUT-300 quick start guide

Petwant
Petwant F9-L user manual

Master equipment
Master equipment PolyPro TP4180 operating instructions