JP Tech 290-AH1 User manual

1
MODELS - 290-AH1 AMP HOUR TOTALIZER
290-AHP3 AMP HOUR CONTROLLER
WITH PRIME FUNCTION FOR
PERISTALTIC PUMP(S)
CONTENTS:
I. OVERVIEW FOR AMP-HOUR BASED METERS: P 2
II. MODEL 290-AHP3 MENU OPERATION: P 3
A. AH TOTAL SCREEN
B. RESETTABLE SCREEN
C. PRIME PUMP SCREEN
1. PRIME FUNCTION SCREEN
D. PRESET1,2 SCREEN
E. PRESET1,2 CNT LEFT SCREEN
F. TIMER1,2 SCREEN
G. TIMER1,2 CNT LEFT SCREEN: P 4
H. CYCLE COUNT 1,2 SCREEN
I. SETUP SCREEN
1. SHUNT SIZE SCREEN
2. RELAY1 SCREEN
3. RELAY2 SCREEN
APPENDIX:
A. CALCULATING THE PRESET AND TIMER VALUES: P 5-6
B. PUMP SETTING WORKSHEET: P 6
C. INSTALLATION / MAINTENANCE / PARTS LIST: P 7-10
D. TROUBLE SHOOTING: P 11
E. WARRANTY & TERMS OF SALE: P 12
REV. 01/27/12
CHECK FOR FREE MANUAL UPDATES AT www.jptech.com
JP Tech, Inc. ▪Box 863 ▪2286A Church Street, Unit C ▪East Troy ▪WI ▪53120
Phone (262) 642-7671 ▪Fax (262) 642-7681

2
I. OVERVIEW FOR AMP-HOUR BASED METERS
JP Tech’s Ampere-Hour meters incorporate several screen options in the Menu designed for ease of operation,
information gathering, and programming. All models incorporate displays showing accumulated Ampere-Hours
or Ampere-Minutes (for Amp Hour meters). A resettable Ampere-Hour/Ampere-Minute display is included. In
addition, there is a “Shunt” display menu item used to select shunt size in hours or minutes and the millivolt input
signal (50mV, 60mV, 75mV, or 100mV).
All models use a two (2) button keypad to cycle through the Menu displays available and to set the parameters of
the meter:
♦
The SELECT key, when pressed and released, cycles through the different screens available to the specific
model. In all screens (except the AH TOTAL, SETUP and SPECIAL FUNCTION screens where applicable),
holding the SELECT key for about 3 seconds will enable the flashing cursor; releasing and pressing again will
position the cursor for a change to be made. The SELECT key will not change any existing information. It only
provides a way to move through the menu or move the flashing cursor.
♦
The CHANGE key is used to change a value or option related to the specific display chosen with the SELECT
key (for example RESETTABLE = VALUE and SHUNT SIZE = OPTION). NOTE: Pressing and releasing, at
any time, BOTH the SELECT and the CHANGE keys together will take you back to the AH TOTAL default
screen.
♦
HOW TO ENTER OR CHANGE DATA:
VALUES: RESETTABLE SCREEN: From the AH TOTAL screen, PRESS/RELEASE the SELECT key.
This will bring you to the RESETTABLE screen. To enter any value or reset existing values, PRESS/HOLD the
SELECT key for about 3 seconds until the LEFT MOST DIGIT begins to flash. When it flashes, release the
SELECT key. If you want the digit that is flashing to stay the same, PRESS/RELEASE the SELECT key
ONCE. This will move the flashing cursor over to the right one digit. [This is the method used to move the
flashing cursor over for all screens.]
If you want to change the value of the flashing digit, PRESS/RELEASE the CHANGE key ONCE. Each time
the CHANGE key is pressed and released, the value of the digit increases by one (1) and cycles from 0 through 9.
[This is the method used to change the value of the flashing cursor for all screens.]
Once you have set the values you want, PRESS/HOLD the SELECT key until the flashing cursor disappears
(about 3 seconds).
♦DETAILED EXMPLE:
OPTIONS: SHUNT SIZE: To select the appropriate shunt size and millivolt signal, PRESS/RELEASE the
SELECT key until the SHUNT SIZE screen appears (for the TOTALIZER Model) or PRESS/HOLD the SELECT
key at the SETUP screen (for the PUMP CONTROLLER model) for about 3 seconds until the SHUNT SIZE screen
appears.
To change the SHUNT SIZE option, PRESS/HOLD the SELECT key until the flashing cursor appears just to the right
of the “H”. PRESS/RELEASE the CHANGE key to scroll through the amperage size options available until the
correct one for your rectifier rating is shown. IF the millivolt default setting is correct (most rectifiers use a 50mV
signal), PRESS/HOLD the SELECT key for about 3 seconds until the flashing cursor stops. Your new option has
been set. PRESS/RELEASE the SELECT key at this point takes you to the next screen.
If the millivolt default setting is not correct, once you have located the correct amperage rating option, PRESS/
RELEASE the SELECT key once to bring you over to the millivolt option. PRESS/RELEASE the CHANGE key
will toggle between the 50mV, 60mV, and 100mV option. Once your amperage setting and the millivolt setting is
correct, PRESS/HOLD the SELECT key for about 3 seconds until the flashing cursor stops. You have now locked in
the new setttings.

3
A. To the right is the default display screen that shows the CUMULATIVE AMPERE/
HOUR – AMPERE/MINUTE TOTAL. This display will appear when the meter is
first energized and is the screen that will appear when BOTH the SELECT key and
the CHANGE key are pressed and released at the same time when you are in the main
menu. The LmV (Low Millivolt) appears in all screens when there is no millivolt
signal coming to the meter from the rectifier.
NOTE: This value cannot be reset and is good up to one (1) trillion Amp/hours.
II. MODEL 290-AHP3 MENU OPERATION
This model incorporates everything the Model AH-1 (AH Total and AH Total Re-Settable Screens) has plus it can support 2 pumps with
associated Presets and Timers or one (1) pump and one (1) rectifier or two (2) rectifiers. All internal output relays are fused. Also, the status of
the pump(s) will be indicated in each of the main display screens by either a LmV (Low Millivolt signal = no accumaltion of Amp Hours and the
pump(s) is off), P1 (Pump 1 = ON), P2 (Pump 2 = ON), or P12 (both pump 1 and 2 = ON). A blank area here indicates that you are getting a mV
signal but neither pump is on.
B. To the right is the RESETTABLE AMPERE/HOUR screen. This display totals
cumulative ampere/hours/minutes like the AH TOTAL screen but can be reset to zero
at any time.
Programming: See ‘OVERVIEW’
C. To the right is the PRIME PUMP screen. This screen is used to prime the Peristaltic
pumps. To enter the function screen press/hold the SELECT key for about 3
seconds. See next screen for operation of this function.
Programming: See ‘OVERVIEW’
1. To the right is the PRIME PUMP FUNCTION screen. When the cursor is
flashing, push/releasing the SELECT key will toggle between PUMP 1 and PUMP 2. Push/
releasing the CHANGE KEY will engage/disengage the pumps. When a pump(s) is en
gaged, the LmV symbol will change to indicate which pump is on. You can engage
both pumps at one time if you wish. REMEMBER: THERE IS NO AUTO
MATIC SHUT OFF ONCE THE PUMPS ARE ENGAGED. You must either disable
each pump before leaving this screen or when leaving this screen (push/hold the SELECT key for about 3 seconds), both pumps will be disabled
Programming: See ‘OVERVIEW’
D. To the right is the PRESET1 screen. This screen is used to set the interval of
Amp Hours before Pump1 is turned on. The PRESET2 screen operates the same way for
Pump 2. [See “Calculating the Preset and Timer Value Needed” in the appendix to calculate
these values.]
Programming: See ‘OVERVIEW’
NOTE: PRESET 1 VALUES WILL ALSO APPLY TO THE PIGTAIL AS WELL.
E. To the right is the PRESET1 CNT LEFT screen. This screen shows the
amount of Amp Hour preset that is remaining before it activates the timer (see below). This
screen is generally for information purposes only and does not need to be edited. The
PRESET2 CNT LEFT screen operates the same way for Pump 2. [See “Calculating the
Preset and Timer Value Needed” in the appendix to calculate these values.]
Programming: See ‘OVERVIEW’
F. To the right is the TIMER1 screen. This screen is used to set the running time
in seconds for Pump 1. When the PRESET1 CNT LEFT screen counts down to zero, the
TIMER1 value is activated and turns ON Pump 1. Values up to 999999 seconds (16,666
minutes or 11.6 days) are available. TIMER2 for Pump 2 works the same way. [See
“Calculating the Preset and Timer Value Needed” in the appendix to calcuate your values.]
Programming: See ‘OVERVIEW’

4
III MODEL 290-AHP3 MENU OPERATION (Cont.)
G. To the right is the TIMER1 CNT LEFT screen. This screen shows the amount
of seconds remaining during the “ON” pump condition. This screen can also be used to add
time to the current pump cycle for a one time add without effecting the TIMER1 setting.
TIMER2 CNT LEFT for Pump 2 works the same way.
Programming: See ‘OVERVIEW’
H. To the right is the CYCLE COUNT1 screen. This screen shows the number of
times the PRESET1 and TIMER1 have cycled through their counts. CYCLE COUNT2
shows a similar value for PRESET2 and TIMER2.
Programming: See ‘OVERVIEW’
I. To the right is the SETUP screen/menu. This screen provides access to menu
items that usually need to be set only once or changed rarely. The SHUNT SIZE, RELAY1,
AND RELAY2 are Submenus of this screen.
Programming: See ‘OVERVIEW’
SETUP
1. To the right is the SHUNT SIZE menu that is used to select the correct maximum
amperage value for the rectifier, the choice of amp/hours or amp/minutes, and the millivolt
output of the shunt. Selecting an amperage size with an ‘H’ causes the meter to accumulate
Amp/Hours. Selecting a size with an ‘M’ causes the meter to accumulate Amp/Minutes.
The millivolt option toggles between 50mV, 60mV, 75mV, and 100 mV. The default setting
is as shown above.
Programming: See ‘OVERVIEW’
2. To the right is the RELAY1 submenu screen. This display allows you to DISABLE the
output of relay #1, have it control a PUMP, or have it control a rectifier. When first
energized, the default setting is “PUMP” indicating that Pump 1 relay is enabled.
Programming: See ‘OVERVIEW’
NOTE: RELAY 1 WILL ALSO ACTIVE THE PIGTAIL AT THE
SAME TIME. REMEMBER THAT RELAY 1 IS ONLY FUSED FOR 5AMPS.
RELAY1
PUMP
3. To the right is the RELAY2 screen. This screen operates the same as RELAY1. The
default setting is “PUMP”.
Programming: See ‘OVERVIEW’
RELAY2
PUMP
REMEMBER: PRESSING, AT ANY TIME, BOTH THE SELECT KEY AND THE
CHANGE KEY TOGETHER BRINGS YOU BACK TO THE AH TOTAL SCREEN.

5
APPENDIX
A. CALCULATING THE PRESET AND TIMER VALUES NEEDED:
The values you may need to run your metering system correctly can vary significantly. In most cases, the calculations are straightforward and only require a
ratio of need chemical additions per amp/hour units. The following is a step-by-step example of how to derive a series of options for an actual situation. The
user’s chemical representative should have supplied the feed ratio for chemical per amp/hour.
Step 1. Determine Desired Feed Ratio:
Nominal Feed Ratio Desired: 1 gallon per 18,000 amp/hours
(supplied by chemical representative)
Metric: 3784 ml (1 gallon) per 18,000 amp/hours
Step 2. Determine the ratio of feed rate per amp/hour count. Divide the ml by amp/hours.
3784ml / 18,000 amp/hours = 0.21 ml per 1 amp/hour
Step 3. Determine the hourly maximum desired feed ratio. Use the shunt size to calculate
this.
0.21 ml per amp/hour count X 6000 amp shunt = 1260 ml per 6000 amp/hours
Step 4. Determine actual pump flow rate. The flow rate should be calculated for one
minute. Use a measured container to gather this information. If using a pulse type
pump, try to remain at 100% stroke and a 100% rate for maximum accuracy.
Pump rate measured: 32 ml per minute
Multiply by 60 minutes: 32 ml X 60 = 1920 ml per hour
Step 5. Determine the feed rate multiplier. Divide the desired feed rate (from Step 3) by
The actual feed rate (from Step 4) to obtain the feed rate multiplier.
1260 ml per 6000 amp/hours (Step 3) / 1920 ml per hour (Step 4) = 0.66
Note: If your multiplier is greater than one, your pump is too small for the
Desired feed rate!
Step 6. Determine pump ON time. Multiply the feed rate multiplier (Step 5) by 3600
seconds.
0.66 (Feed Rate Multiplier) X 3600 seconds = 2376 seconds
Note: This is the duration the pump must remain ON during one hour to pump
the desired amount of chemical.
Step 7. Calculate and select the final settings.
Every 6000 amp/hours the pump will turn on for 2376 seconds (step 6)
Note: THIS IS YOUR FINAL RATIO.
*** Break Down the Ratio to a Usable Level ***
When rounding seconds, always go down to the next lowest whole second (e.g.; 50.9 sec. = 50 sec; 50.1 sec. = 50 sec.) When rounding amp/hours, always
round up to the next whole amp/hour (e.g.; 50.9 = 51; 50.1). This practice will ensure that your meter is adding less rather than too much chemical.
Find the smallest seconds settting:
Ratio: 6000 amp/hours : 2376 Seconds
Round Down: 6000 : 2376
Difference: 0 : 0
% Error: 0% : 0%
Dividing by 10 =
Ratio: 600.0 amp/hours : 237.6 Seconds
Round Down: 600.0 : 237
Difference: 0 : -0.6 Seconds
% Error: 0 : -0.25%
Dividing by 10 =
Ratio: 60.00 amp/hours : 23.76 Seconds
DifferenceRound Down: 60.00 : 23
: 0 : -0.76
% Error: 0 : -3.30%

6
Keep Seconds and Correct for Error =
Ratio: 60.0 amp/hours X [1-0.033(% error)]: 23 Seconds
New Ratio: 58.02 : 23
Round Up: 59 : 23
Difference +0.98 : 23
% Error 1.67% : 0.0
Our final setting will be for every 59 amp/hours the pump must run for 23 seconds. Every 18,000 amp/hours (from step 1) you will need to add 63 ml (3784 ml
X 0.0167) to correct for the 1.67% error.
B. PUMP SETTING WORKSHEET
You will need the following information to use the Pump Settings Worksheet:
Nominal Feed Ratio: (A)_____________ Gal. Per (B)_____________ Amp/Hours
(As recommended by your chemical representative)
Shunt Size: (E) _________________ Amps
Actual Pump Vol: (G) _______________ ml. Per minute (Measured by You)
Step 1. Nominal Feed Ratio:
(A)_____________GAL. Per (B)_____________Amp/Hours
OR (C)_____________ml. Per (B)_______________Amp/Hours
[3784ml = 1 gallon]
Step 2. Feed Ratio per Amp/Hour:
___________ml/__________Amp/Hour = (D)__________ml/1 Amp/Hour
(C) (B) (C)/(B)
Step 3. Desired Feed Rate:
__________ml per 1 Amp/Hour X (E)_________Amps = (F)_________ml/Hour
(D) (Shunt Size) (D) X (E)
Step 4. Measure Actual Pump Rate Per Minute:
(G)__________ml per Minute X 60 Minutes = (H)_____________ml per Hour
(Pump Volume) (G) X 60
Step 5. Feed Rate Multiplier (Desired Feed Rate / Actual Feed Rate):
___________ml per Maximum Amp/Hour /_________ml per Hour = (I)
(F) (H) (F/H)
Step 6. Pump ON Time per Hour:
_____________ X 3600 Seconds = (J)____________ Seconds
( I) (I) X 3600
Step 7. Final Settings:
Every ____________Amp/Hour, the pump must turn ON for______Seconds
(E) (J)
Ratio: _______________:______________
Round _______________:______________
Diff: _______________:______________
%Error _______________:______________

7
PUMP OUTPUT
(Only in Pump Control Versions)
The AH-PMP-3 models come with
either one or two peristaltic pumps.
When the preset number of amp hours is
attained, the pump(s) will be energized
for the length of time that the operator
has designated in TIMER1 or TIMER2.
Facing the controller, Pump 1 is to the
left and Pump 2 is to the right.
120 VAC POWER
The meter operates
off 120 VAC or
220VAC. The AH-
PMP-3 meters use a
three wire grounded
plug for 110VAC or
a terminal strip for
220VAC hardwire..
WARNING:
All board components and
circuitry use a “floating
ground” and must remain
isolated from all other cir-
cuits and grounds. This is
only an issue if board is
removed outside its protec-
tive casing.
PUMP 1 PUMP 2
INSTALLATION
110VAC
MILLIVOLT INPUT
All models of the meter come with 15
feet of shielded twisted pair wire for
millivolt input. This wire can be
connected either to the shunt or to the
back of the ammeter. The red wire
must be connected to the positive
terminal and the black wire to the
negative terminal.
DO NOT CONNECT THE
BRAIDED WIRE OR SHIELD TO
GROUND!
NOTE:
CHECK
SQUEEZE
TUBES
REGU-
LARLY FOR
INTEGRITY.
INSTALLATION
220VAC
230 VAC
POWER
The meter operates
off 230 VAC. No
ground wire is
needed. The elec-
tronics work on a
“floating” ground.
Connect to Termi-
nal strip inside
PUMP
PUMP OUTPUT
(Only in Pump Control Versions)
The AH-PMP-3 models come with
either one or two peristaltic pumps.
When the preset number of amp
hours is attained, the pump(s) will be
energized for the length of time that
the operator has designated
in
TIMER1
or
TIMER2. Facing the
controller, Pump 1 is to the left and
Pump 2 is to the right.
WARNING:
All board components and
circuitry use a “floating
ground” and must remain
isolated from all other cir-
cuits and grounds. This is
only an issue if board is
removed outside its protec-
tive casing.
PUMP 2
MILLIVOLT INPUT
All models require a shielded
twisted pair wire for millivolt
input. This wire can be connected
either to the shunt or to the back of
the ammeter. The red wire must be
connected to the positive terminal
and the black wire to the negative
terminal.
DO NOT CONNECT THE
BRAIDED WIRE OR SHIELD
TO GROUND!
IMPORTANT
PARAMETERS:
10’ HEAD MAX
LATERAL RUN:
75’ MAX

8
INSTALLATION GUIDE
NOTE: CHECK SQUEEZE TUBE REGULARLY FOR INTEGRITY
1. Read Manual to familiarize yourself with the operation of the meter.
2. Mount Unit. Attach feed tubes to the Squeeze tubes using wire ties or stainless
hose clamps. NOTE: Tote nozzle may float so you might need to weight it down.
3. Connect the 15’ of 2 wire twisted pair mV wire to the rectifier’s shunt (red to the
more positive side of shunt) or connect to the amperage meter’s plus/minus leads.
IMPORTANT: If you shorten this wire, remove all shielding from the ends of
the twisted pair and put tape around the ends. At no time must the shielding be
grounded.
4. Connect the 110VAC power source plug into the 110VAC receptacle.
5. Apply power to the unit
6. IMPORTANT: Set Shunt size in the “shunt size” screen. Remember, you want to
set the shunt size to the amperage rating of the rectifier (i.e., 15,000A rating,
3000A rating, etc.) not your operating amperage setting.
7. Set your Presets (how often to feed chemicals) and Timers (how much chemical to
feed). Use worksheet to figure these values. These values can be changed any
time to get the correct balance.
8. You can use the “Prime Pump” function to prime these pumps if need be.
CHECK FOR FREE MANUAL UPDATES AT www.jptech.com
JP Tech, Inc. ▪Box 863 ▪2286 Church Street, Unit C ▪East Troy ▪WI ▪53120
Phone (262) 642-7671 ▪Fax (262) 642-7681

9
MAINTENANCE
The PERISTALTIC Series of metering pumps require a minimal amount of
maintenance to achieve optimal performance. Periodically check the squeeze
tube for cracks, deterioration, or swelling. The squeeze tube will typically need
to be replaced about every 6 months (chemical compatibility and duty cycle
can cause this interval to vary). NOTE: squeeze tubes are NOT a warranty
item.
VERY IMPORTANT: When replacing squeeze tubes, DO NOT TWIST
THE TUBES WHEN FITTING THEM AROUND THE ROLERS. Insert
them so they remain flat in the same plane. (The writing on the tubes should be
inline on both sides of the rollers.)
Appling lube to the squeeze tube once a month will extend the life of the tube,
minimize wear on other contacting parts, and promote smoother pump opera-
tion. Use Knight Tube Lube (P/N 245-4-SQZLUBE) or an equivalent silicone-
based lubricant.
1. Remove the faceplate of the pump.
2. Apply a thin bead of Tube Lube to the inner surface (the side that the
rollers contact) of the squeeze tube between the 9 o’clock and 3 o’clock
positions. Avoid getting the lube near the pinch points where the bottom
of the faceplate grips the tube.
3. Put the faceplate back on the pump
Activate the pump under normal conditions—the lubricant will be evenly
distributed as the pump rotates.
CAUTION: To avoid severe or fatal shock, always disconnect main
power when servicing the unit.

10
PARTS LIST FOR 290-AHP3
ITEM PART NUMBER
ELASTIC SQUEEZE TUBE 245-4-ELSQZTUB
( Standard with unit. Good for weak to strong alkalis, weak to medium acids.)
THERMISTANT SQUEEZE TUBE 245-4-TMSQZTUB
(Has superior acid-resistant and alkali-resistant qualities.)
FLUORO-VITON SQUEEZE TUBE 245-4-FVSQZTUB
( Good for Strong solvents and acids.)
SQUEEZE TUBE LUBRICANT 245-4-SQZLUBE
24 VDC PERISTALTIC GEAR MOTOR 245-4-GEARMOTOR-FV
ROLLER FOR PERISTALTIC UNITS 245-4-ROLLER
FACEPLATE FOR PERISTALTIC UNITS 245-4-FACEPLATE
ROLLER / TUBE PUMP HOUSING 245-4-PMPBODY
RED/BLACK TWISTED PAIR SIGNAL WIRE 210-1-TWPR22RD
DELIVERY TUBING—500’ ROLL 245-4-PLYTB500
(Polyethylene 1/4’ OD, rigid wall tubing)
DELIVERY TUBING—100’ ROLL 245-4-PLYTB100
PERISTALTIC HINGE 245-4-HINGE
NOTE: CHECK YOUR SQUEEZE TUBES PERIODICALLY FOR INTEGRITY (ABLE
T0 EXPAND AFTER BEING SQUEEZED BY THE ROLLERS).
ORDER BY:
PHONE: 1-262-642-7671 POST: JP TECH, INC.
FAX: 1-262-642-7681 P.O. BOX 863
USA

11
TROUBLESHOOTING FOR AMP HOUR BASED METERS
AMP HOUR ISSUES:
NO AMP HOURS RECORDING WHEN RECTIFIER IS ENGAGED
.
May have the leads from the twisted pair mV wire connected wrong at the shunt or analog
amp meter in the remote. Reverse the leads of the Twisted Pair mV wire on the shunt or ana-
log amp meter. [Red = positive, Black = negative}
AMP HOURS BEING RECORDED AT A VERY FAST RATE
.
May have the Twisted Pair mV wire connected to Voltmeter rather than the analog amp meter.
Disconnect and reconnect to the amp meter.
AMP HOURS COMING IN TOO SLOW OR TOO FAST FOR WHAT WE ARE USE TOO.
May have the wrong “SHUNT SIZE” setting for the size of the rectifier / shunt the meter is connected
too. Go to the “SHUNT SIZE” screen (see Manual) and make sure the correct shunt size is entered.
PUMP ISSUES:
PUMP(S) STAY ON ALL THE TIME.
May be due to the Preset amp hours coming in quicker than the Timer settings have a chance
to time out (the timer will re-set to start value each time preset reaches it’s set mark). Make
sure calculations for feed rate is correct. May need a bigger pump or, if using our double peri-
staltic model, use both pumps together.
PUMP(S) WON’T ENGAGE WHEN TIMER TIMES OUT OR WHEN ACTIVATED IN “PRIME
PUMP” SCREEN.
May have blown the 5A fuse protecting the relays. Open lid of controller and check the fuse for conti-
nuity. (Unplug unit while doing this step.)
If fuse is OK, engage the relay in “Prime Pump” screen and check for 110VAC out of the receptacle
with a DMM meter. If there is 110VAC, the pump controller is OK and you need to check the pump
and it’s connection
.
PERISTALTIC PUMPS DON’T PUMP AS MUCH AS THEY DID IN THE SAME AMOUNT OF
TIME.
May be due to squeeze tubes that have deteriorated to point that they can’t re-expand after being com-
pressed. They need to be replaced.
STROKES (AH-PMP-2 MODELS) ARE NOT BEING RECORDED WHEN PUMP IS ENGAGED.
First you need to determine where the problem is located. Place the controller in the “MANUAL
PUMP” mode and carefully take a jumper and, intermittently, go across terminal #9 & #10 for Stroke 1
and terminals #9 k& #11 for Stroke 2. If strokes are recorded in the Manual Pump screen for each
stroke pump, the controller is working find. The problem may be with the signal coming from the
pump to the controller. Please call your source to get suggestions on how to fix this problem. If NO
strokes are recorded in the Manual Pump screen, than the controller is not recording the signal. Please
call your source of this meter for directions for repair.

12
Revised 10/3/01
JP TECH, INC.
LIMITED PRODUCT WARRANTY
JP Tech warrants to first user of each new JP Tech product or component that it is free from defect in material and workmanship. The obligations of JP Tech
under this warranty are expressly limited to the following:
• JP Tech will repair or replace, at its option, any defective components for a period of twelve (12) months from date of shipment. No charges are
covered for the removal or replacement of defective components.
• This warranty applies only if the product is defective under normal use. It does not apply to breakage or defect from accident, alteration, misuse, or
abuse of the product or component. In addition, this warranty is effective only if the product or component is installed in a location and manner
prescribed by JP Tech’s instructions and only if it is so maintained. This warranty becomes null and void if the product or component is altered by
anyone other than JP Tech, it’s authorized representatives, or by expressed written authorization for a specific situation.
• If JP Tech elects to send a service technician to a customer site to repair a defect, the cost of transportation and/or living expenses will be paid for by
the customer. Should the defect turn out to be the result of the customer’s misuse, improper installation, or maintenance of the product or compo-
nent, the customer will be responsible for the full cost of the service call including labor charges plus the aforementioned travel and living expenses.
• JP Tech will repair or replace any defective part within a product at the sole discretion of JP Tech. If JP Tech should choose to supply a part to the
customer as a no-charge warranty replacement, the customer assumes all cost of installation associated with the replacement part. If the product
needs to be returned for warranty service, a Returned Material Authorization (RMA) must be issued by JP Tech prior to such return. All returned
material must be sent freight prepaid or it will not be accepted by JP Tech irrespective of warranty issues.
• There are no implied warranties of merchantability or of fitness for a particular purpose. The above warranty is made in lieu of all other guarantees
or warranties, express or implied. JP Tech distributors or OEMs who purchase JP Tech products for resale are not authorized to assume any other
obligation or liability for JP Tech.
• JP Tech will in no case or under any circumstances be liable for special, incidental or consequential damages, loss of profit or commission for any
loss caused by any delay in production or shipment of product, or defect of any kind in any product or component covered by the sale. Without
limitation, JP Tech will not be so liable with respect to furnishing of any product, or component, delay in such furnishing, use, resale, or other cause.
JP Tech’s liability arising out of the supply of any product or component, its use, resale or other disposition, or out of any guarantee or warranty,
express or implied, or any other cause, shall in no way exceed the cost to JP Tech of the product or component which JP Tech agrees to repair or
replace. JP Tech’s liability for any product or component terminates upon expiration of the applicable repair or replacement period.
This implementation of this warranty may, under separate agreement, be subrogated to exclusive distributors or manufacturers who shall assume all or
portions of the liability associated with warranty costs.
This warranty may be modified, wholly or in part, at any time by JP Tech without notice to past or future customers. The warranty revision in effect at the
time of shipment shall prevail in any claims rendered.
Revised 10/3/01
JP TECH, INC.
TERMS AND CONDITIONS OF SALE
The purchase of any products or services supplied by JP Tech shall be governed by the terms of this agreement. Purchaser of these products and services ac-
knowledges and agrees to these terms without modification by any competing document or any agreement not reduced to writing and authorized by an officer of
JP Tech, Inc.
• Pricing is the effective price at the time of the order. If the shipment of product is postponed by buyer, the price may be changed to reflect any price
changes enacted by JP Tech. Prices may be changed by JP Tech at any time for any reason without notice to purchaser except for accepted orders
not affected by a purchaser initiated delay. Prices, unless otherwise stipulated, do not include shipping and handling charges.
• Certain products may require initial and progress payments before the commencement and continuation of design, engineering, component procure-
ment, and manufacture. These products will not be shipped until all progress payments have been made. Cancellation of any orders in progress will
necessitate the forfeitures of any payments received to date as well as payment of any costs accrued in excess of paid amounts.
• Orders must be accepted by JP Tech at their home office. Acceptance of any purchase order, regardless of the method, is conditioned on assent of
buyer to the terms and conditions contained herein.
• Sales are FOB point of shipment. Sales terms are net 30 days from date of shipment. Present or future sales, use, or other taxes on sales, installation
or use shall be paid buy purchaser. Purchaser shall pay 1% interest per month on all outstanding amounts due to JP Tech. Interest accrual shall
begin on the 31st day after shipment for all outstanding amounts.
• All sales are final. Any decisions to accept return of product after shipment and receipt by purchaser shall be at the sole discretion of JP Tech and
not until payment has been made and agreement by purchaser to pay all shipping, cancellation, and restocking charges that may accrue.
• Shipping dates given prior to shipment are estimated, actual delivery will be based on factory and engineering loading at the time of manufacture as
well as the availability of parts required for manufacture. JP Tech shall not be liable for any costs or damages arising out of or related to any delays
in shipment or delivery, including but not limited to liquidated damages, unless otherwise agreed in writing.
• JP Tech may change design or construction of any product or component in any way they see fit. Upgrades for previously purchased products may
be available for certain products for a price that will be determined when appropriate.
• Except as provided herein, any controversy, claim or dispute arising out of or related to any order or sale or breach there of, including but not limited
to any breach of warranty claims, shall be litigated in state court, Walworth or Waukesha Counties, Wisconsin, and shall be governed by the laws of
Wisconsin. If JP Tech is the prevailing party, JP Tech shall be entitled to collect all reasonable fees and costs, including court costs and attorney
fees.