K-Patents PR-43-AC User manual

Sanitary Process Refractometer PR-43-AC/AP
IM-EN-PR43ACAP: Rev. 1.08
Effective: September 20, 2016
Instruction Manual

Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
General safety consideraons
The process medium may be hot or otherwise hazardous. Use shields and
protecve clothing adequate for the process medium - do not rely on avoid-
ance of contact with the process medium.
Precauons when removing a refractometer from the process line :
•Check first that the process line is depressurized and drained.
•Ensure you are clear of any possible spillage and you have a clear emer-
gency escape path.
Disposal
When wishing to dispose of an obsolete refractometer or any parts of a
refractometer, please observe local and naonal regulaons and require-
ments for the disposal of electrical and electronic equipment.

© Copyright K-Patents 2016. All rights reserved.
Symbols and terms used in this manual:
This indicates a warning. It provides safety precauon informaon needed to
avoid injury while operang the refractometer system.
This indicates that something is important for the operaon of the refractome-
ter system.
Note. Notes contain addional informaon and hints.
Warranty
All K-Patents products are guaranteed to be free of material or workmanship defects.
K-Patents provides a limited warranty that covers the repair or replacement, without
charge, of any defected product or part that occurs within two (2) years from the date
of delivery. The repair can only be done by nearest authorized K-Patents repair facility.
K-Patents is not responsible for material selecon of process weed parts.
Warranty does not cover normal wear and tear of the product over me, or any prod-
ucts that are handled, installed or used against manufacturer’s guidance.
Full warranty informaon is available in the Support secon of hp://www.kpatents.com.
Please note that prior to shipping any products or items to K-Patents, a Request for
Return Materials Authorizaon (RMA) must be filled in and submied to K-Patents
Order Handling department. The RMA form is available at hp://www.kpatents.com/
in the Support secon.

Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016

Table of contents
1 Refractometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Mounng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.1 Choosing the mounng locaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Check list for pipe mounng . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.3 Check list for mounng in a tank, a vessel or a large pipe . . . . . . . . . 3
2.4 Electrical connecons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4.1 Wiring opons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Specificaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 Compability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Refractometer rangeability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Sanitary process refractometer PR-43-AC . . . . . . . . . . . . . . . . . . . . . . 9
3.3.1 PR-43-AC model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.2 PR-43-AC mounng hardware model code . . . . . . . . . . . . . . . . . . 12
3.3.3 PR-43-AC specificaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3.4 PR-43-AC parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3.5 PR-43-AC mounng specifics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3.6 Mounng specifics for EHEDG cerfied PR-43-AC configuraon . . 27
3.3.7 3-A Sanitary Standard compliance . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.4 Sanitary process refractometer PR-43-AP . . . . . . . . . . . . . . . . . . . . . 29
3.4.1 PR-43-AP model code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.4.2 PR-43-AP mounng hardware model code . . . . . . . . . . . . . . . . . . 34
3.4.3 PR-43-AP specificaons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.4.4 PR-43-AP parts lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.4.5 PR-43-AP mounng specifics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.4.6 Mounng specifics for EHEDG cerfied PR-43-AP configuraon . . 41
3.4.7 3-A Sanitary Standard compliance . . . . . . . . . . . . . . . . . . . . . . . . . 41
4 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.1 Prism coang . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2 Prism wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2.1 Recommended wash pressures and mes . . . . . . . . . . . . . . . . . . . 44
4.2.2 Prism wash systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.2.3 Prism wash nozzles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
5 Refractometer verificaon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
PR-43 refractometer verificaon form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

1 Refractometer 1
© Copyright K-Patents 2016. All rights reserved.
1 Refractometer
The sanitary refractometers PR-43-AC and PR-43-AP provide a digital interface detec-
on system for alimentary industry.
The PR-43-AC/AP (figure 1.1) measures the refracve index nDand the temperature of
the process medium. The concentraon of the process liquid is calculated from these
values when the composion of the process medium is known.
The output values of the refractometer are transmied through mA output and digitally
through an Ethernet connecon by using a UDP/IP protocol (see the general manual for
specificaons). mA output is available with mA output cable (mA only) or with a split
cable (mA and Ethernet).
PR-43-AC PR-43-AP
Figure 1.1 Refractometer models

2PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
2 Mounng
The refractometer mounng locaon should be chosen with care to ensure reliable
readings from the process.
2.1 Choosing the mounng locaon
The mounng locaon needs to be such that sediments or gas bubbles cannot accumu-
late by the refractometer. Good flow velocity is essenal in keeping the prism clean.
Important: If the process pipe vibrates, support the pipe. A vibrang pipe might dam-
age the in-line refractometer mounted on it.
A K-Patents inline refractometer can be located either indoors or outdoors in most cli-
mates. However, when a refractometer is located outdoors, some basic protecon
against direct exposure to sunlight and rain should be provided. Special care should
be taken if the pipe wall is translucent (e.g. of fiberglass), as light from outside reach-
ing the prism through the pipe wall may disturb the measurement.
The refractometer cover should not be exposed to high temperature radiaon. In most
cases, dra and natural convecon provide sufficient air cooling if the air gets to flow
freely around the refractometer head.
Addional cooling is necessary when the ambient temperature is higher than 45 °C
(113 °F) or when the process temperature is above 110 °C (230 °F) and the ambient
temperature is above 35 °C (95 °F). The air cooling is improved by blowing pressurized
air against the refractometer cover. The pressurized air can be supplied by the ven-
laon system. If no air is available it is also possible to wind a copper coil for cooling
water around the refractometer head cover.
Important: Always mount the refractometer so that the interconnecng cable points
downwards from the refractometer head.
2.2 Check list for pipe mounng
The PR-43-AC and PR-43-AP are mounted in a pipe. K-Patents recommends a flow
velocity between 1 and 3 m/s (3–10 /s). If the flow velocity exceeds 6 m/s (20 /s),

2 Mounng 3
© Copyright K-Patents 2016. All rights reserved.
there is a risk of cavitaon. Cavitaon may damage the refractometer and the piping.
Too slow a flow velocity may cause erraneous measurement readings due to coang of
straficaon (layering) of the sample on the prism.
The diameter and form of the pipe and the process temperature all affect the measure-
ment and need to be taken into account.
−If the process pipe diameter varies, select the posion with the smallest diameter
(and accordingly highest velocity). Then the prism stays clean beer.
−If the refractometer is used in a feed-back control loop, make the me lag short. E.g.
when a diluon valve is controlled, mount the refractometer close to the diluon
point. However, make sure complete mixing has occurred at mounng locaon.
−If the temperature varies along the process pipe, select the posion with the highest
process temperature. That minimizes the risk of coang, because higher tempera-
ture means higher solubility and also lower viscosity.
−Oen the posion with the highest process pressure (= aer pump + before valve)
has favorable flow condions without sedimentaon or air trapping risks.
−The refractometer should be conveniently accessible for service.
2.3 Check list for mounng in a tank, a vessel or a large pipe
A probe refractometer PR-43-AP can be inserted with a flange or clamp into tanks and
vessels which either don’t have a scraper or where the mixer doesn’t touch the vessel
wall. A probe refractometer can also be flush mounted in a cooker where the scraper
touches the wall.
−The inserted probe refractometer is mounted close to a srrer to ensure represen-
tave sample of the process liquid and to keep the prism clean.
−The refractometer should be conveniently accessible for service.

4PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
Prevent outside light
Easy access
Easy access
High pressure
High temperature
High pressure
Small diameter
High temperature
High velocity >1.5m/s
(>5ft/s)
Upper pipe bend
Ambient air
cooling

2 Mounng 5
© Copyright K-Patents 2016. All rights reserved.
2.4 Electrical connecons
The refractometer has an M12 connector in the refractometer for power supply, mA
output and Ethernet connecons.
Figure 2.1 The M12 connector
PR-43 refractometers are powered with 24 VDC. For connecng the refractometer to
Mulchannel user interface MI, see the manual for Mulchannel user interface. For
connecng the refractometer to Compact user interface CI, see the manual for Compact
user interface. See Figure 2.2 for instrucon how to connect the M12 refractometer
cable.
2.4.1 Wiring opons
The PR-43 refractometer provides both analogue (mA) and digital output signals. See
Figure 2.3 for the wiring when only analogue output is used.
Opons for connecng the refractometer with both analogue and digital outputs are
shown in Figure 2.4. Both Compact user interface CI and Mulchannel user interface
MI use the digital output signal.

6PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
Align these three
reference points
Approximately
a quarter turn to
tighten
Figure 2.2 Connecng the refractometer cable to the refractometer
10m (33ft)
90m (295ft)
180m (590ft)
200m (656ft)
P4-001
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
max. 90m (295ft) max. 90m (295ft)
max. 90m (295ft)
10m (33ft)
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8660
Platform 4 cable extender
M12-8 pin, A-code, F+M
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
max. 200m (656ft)
PR-8680
M12-8 pin, A-code, F
Field plug PR-8432
Power supply and
mA output cable
Refractometer only Cables
M
M M
M MM
F
F
F
F
F
F F
Figure 2.3 Wiring with analogue output only

2 Mounng 7
© Copyright K-Patents 2016. All rights reserved.
10m (33ft)
90m (295ft)
180m (590ft)
200m (656ft)
P4-001
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
PR-43 REFRACTOMETER
M-12-8 pin, A-code, M
non-Ex
max. 90m (295ft) max. 90m (295ft)
max. 90m (295ft)
10m (33ft)
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8430
Platform 4 cable
M12-8 pin, A-code, F+M
Ethernet, mA, 24V
PR-8660
Platform 4 cable extender
M12-8 pin, A-code, F+M
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
PR-8431
Split cable M12
M12-8 pin, A-code, F
Ethernet (M12-4 pin, D-code,M)
mA, 24V
1m (3ft)
max. 200m (656ft)
PR-8680
M12-8 pin, A-code, F
Field plug PR-8432
Power supply and
mA output cable
Refractometer only Cables
M
M M
M MM
F
F
F
F
F
F F
Figure 2.4 Wiring opons when both analogue and digital outputs are used

8PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
When a refractometer is connected to a computer or another device with RJ45 port,
an ethernet converter cable is needed as well as the split cable.
Figure 2.5 Connecng the refractometer with network converter

3 Specificaons 9
© Copyright K-Patents 2016. All rights reserved.
3 Specificaons
3.1 Compability
Electrically: The PR-43 refractometers are not interchangeable with any other refrac-
tometer model. All PR-43-AC/AP refractometers are however interchangeable with
each other as long as they have the same prism. The PR-43-AC/AP refractometers are
not compable with the K-Patents indicang transmiers DTR, STR or IT-R.
Mechanically: The sanitary process refractometer PR-43-AC/AP fits the same sanitary
process connecons as PR-23-AC/AP.
3.2 Refractometer rangeability
The refracve index standard range of a PR-43-AC/AP is 1.320-1.530 (corresponds to
0-100 Brix), Figure 3.1, with a Sapphire prism. A PR-43-AC can also be equipped with a
Sapphire prism with a refracve index range 1.260-1.470.
PR-43-AC-73
PR-43-AC-74
1.30
1.25 1.35 1.40 1.45 1.50 1.55
nD
Figure 3.1 PR-43-AC rangeability
3.3 Sanitary process refractometer PR-43-AC
The refractometer PR-43-AC is a 3-A and EHEDG cerfied sanitary process refractome-
ter for measuring concentraons in a pipeline. It is easy to install in any pipe size di-
rectly or using a flow cell. The Sanitary process refractometer is suitable for all food
and beverage processing applicaons where real-me monitoring and control can help
to improve product quality and reduce costs.

10 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
Figure 3.2 PR-43-AC dimensions

3 Specificaons 11
© Copyright K-Patents 2016. All rights reserved.
3.3.1 PR-43-AC model code
SANITARY COMPACT REFRACTOMETER for pipelines
Model and descripon Model
PR-43 = Refractometer PR-43
Refractometer model
-AC = Compact sanitary cerfied process refractometer, inseron length 14 mm -AC
Prism material and Refracve Index range limit
-73 = R.I. 1.320-1.530 nD(0-100 Brix) Sapphire prism -73
-74 = R.I. 1.260-1.470 nDSapphire prism -74
Connecon type and size
-E25-P15 = Varivent DN65, 15 bar (A) -E25-P15
-H25-P15 = Sanitary 3A-clamp, 2½inch, 15 bar (A) -H25-P15
-H25-P40 = Sanitary 3A-clamp, 2½inch, High Pressure, 40 bar @ 20°C (A) -H25-P40
-Z25-P15 = 2½inch I-clamp, 15 bar (B) -Z25-P15
Refractometer weed parts material
-SS = AISI 316 L -SS
-HC = Hastelloy C / ASTM C267 -HC
Electrical classificaon
-UN = Unclassified area, general purpose, ordinary locaon -UN
-AX = EX and IECEx cerfied Ex II 3G, Ex nA IIC T4 Gc
(up to zone 2) (Tamb -40...+65°C)
-AX
-IA = ATEX and IECEx cerfied Ex II 1G, Ex ia IIC T4 Ga
(up to zone 0) (Tamb -40 ... +65°C)
-IA
High accuracy opon
-HAC = High accuracy version in range 0-30 Brix 4-30°C -HAC
EHEDG opon
-EH = EHEDG Type EL class I Cerfied Model -EH
Polishing opon
-EP = Electropolished refractometer weed parts (Ra 0.4 µm, 15 µ inch) -EP
(A) EHEDG cerfied version available
(B) Sanitary 3-A cerfied

12 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
3.3.2 PR-43-AC mounng hardware model code
Mounng hardware (without wash nozzles)
MODEL AND DESCRIPTION MODEL
AFC = Elbow flow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
10 = 25 mm (1 inch) (A) 10
15 = 40 mm (1 1/2 inch) 15
20 = 50 mm (2 inch) 20
25 = 65 mm (2 1/2 inch) (A) 25
30 = 80 mm (3 inch) (A) 30
40 = 100 mm (4 inch) (A) 20
Flow cell inlet type
-SI = Straight pipe -SI
-RI = Reduced pipe (cone) -RI
Opons
-EP = Electropolished process weed parts (RA 0.4µm, 15 µ inch) -EP
(A) with -SI opon only
EHEDG cerfied mounng hardware
MODEL AND DESCRIPTION MODEL
AFC = Elbow flow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
20 = 50 mm (2 inch) 20
Flow cell inlet type
-SI = Straight pipe -SI
EHEDG
-EH = EHEDG Type EL Class I Cerfied model -EH
Polishing opon
-EP = Electropolished process weed parts (Ra 0.4µm, 15 µ inch) -EP
For EHEDG cerfied refractometers use ISO2852 type EHEDG cerfied gasket PR-9202-
EH (2½inch) or PR-9272-EH (4 inch).

3 Specificaons 13
© Copyright K-Patents 2016. All rights reserved.
Mounng hardware with wash nozzles
MODEL AND DESCRIPTION MODEL
AFC = Elbow flow cell AFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
10 = 25 mm (1 inch) 10
15 = 40 mm (1 1/2 inch) 15
20 = 50 mm (2 inch) 20
25 = 65 mm (2 1/2 inch) (A) 25
30 = 80 mm (3 inch) (A) 30
40 = 100 mm (4 inch) (A) 40
Flow cell inlet type
-SI = Straight pipe -SI
-RI = Reduced pipe (cone) -RI
Wash nozzle connecon
-NC = Nozzle connecon -NC
Wash nozzles for 10/15 flow cells
-SN = Steam nozzle, threads G ¼ inch female -SN
-WN = Water nozzle, threads G ¼ inch female -WN
-WP = Pressurized water nozzle, threads G ¼ inch female -WP
-PG = Plug for nozzle connecon -PG
(A) with -SI opon only
Example: Refractometer:PR-43-AC-73-H25-P15
Flow cell: AFC-HSS-H20-SI
Note: Prism wash control and diagnoscs require the use of a Mulchannel user in-
terface MI (or customer’s own control system).

14 PR-43-AC/AP instrucon manual
Document/Revision No. IM-EN-PR43ACAP/1.08 Effecve: September 20, 2016
Mounng hardware, mini flow cell
MODEL AND DESCRIPTION MODEL
MFC = Mini flow cell MFC
Refractometer connecon
-H = Sanitary 3A clamp, 2 1/2 inch -H
Construcon material
SS = AISI 316 SS
Process connecon
-H = Sanitary 3A clamp -H
Pipe secon diameter
05 = 15 mm (½ inch) 05
Mounng hardware, Varivent DN65 connecon
MODEL AND DESCRIPTION MODEL
TDN = Varivent®in-line access unit clamp DN65 Type N (A) TDN
Pipe secon diameter
40 = 100 mm (4 inch) -40
50 = 50 mm (2 inch) -50
65 = 65 mm (2 1/2 inch) -65
80 = 80 mm (3 inch) -80
100 = 100 mm (4 inch) -100
125 = 125 mm (5 inch) -125
150 = 150 mm (6 inch) -150
Counter flange opons
-SN = Steam nozzle, threads G ¼ inch female -SN
-WN = Water nozzle, threads G ¼ inch female -WN
-WP = Pressurized water nozzle, threads G ¼ inch female -WP
-PG = Varivent blind flange type N -PG
(A) Includes one 2½inch Type N blind flange with 2½inch EPDM gasket
and 2½ inch Varivent clamp Type N
This manual suits for next models
1
Table of contents
Other K-Patents Measuring Instrument manuals

K-Patents
K-Patents PR-14038 User manual

K-Patents
K-Patents PR-43 Series User manual

K-Patents
K-Patents PR-23 User manual

K-Patents
K-Patents PR-33-S User manual

K-Patents
K-Patents PR-33-AC User manual

K-Patents
K-Patents PR-21-S User manual

K-Patents
K-Patents SAFE-DRIVE PR-23-SD User manual

K-Patents
K-Patents PR-33-S User manual

K-Patents
K-Patents PR-43 Series User manual

K-Patents
K-Patents SAFE-DRIVE PR-23-SD User manual