Kaise KF-20 User manual

FOR SAFETY MEASUREMENTS!!
WARNING
INTRODUCTION
2. SPECIFICATIONS
3. SAFETY PRECAUTIONS
4. NAME ILLUSTRATION
WARNING 1. Checks of Body and Test Lead
WARNING 3. Warning for High Voltage Measurements
WARNING 4. Dangerous Voltage Measurement Procedures
WARNING 2. High Power Line Measurements Prohibited
WARNING 7. Test Lead Detachment
WARNING 6. Maximum Measurement Range Observance
WARNING 5. Correct Selection of RANGE Switch
1234
To prevent an electrical shock hazard to the operator and/or damage to the
instruments, read this instruction manual carefully before using the
Multitester. WARNINGS with the symbol on the Multitester and this
instruction manual are highly important.
Important Symbols
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution
(refer to instruction manual)".
WARNING : The symbol in this manual advises the user of an
electrical shock hazard that could result in serious injury or even
death.
CAUTION : The symbol in this manual advises the user of an
electrical shock hazard that could cause injury or material damages.
Do not measure High Power Line that might exceed 2kVA. Measurement on
High Power Line is very dangerous. It sometimes includes High Surge
Voltage that could cause dangerous arcs of explosive short in the instrument
and could result in serious injury to the operator. When measuring high
voltage, even for Low Power Line (Low Energy Circuit), take extra care to
avoid electrical shock hazard and/or damage to the instrument.
Thank you for purchasing KAISE "MODEL KF-20 ANALOG MULTITESTER". To
obtain the maximum performance of this instrument, read this Instruction Manual
carefully, and take safe measurement.
Inspect the instrument and acessories for transport damage. If there is any
damage or missing items, ask your local dealer for replacement.
Confirm that the following items are contained in the package.
2-1. MEASUREMENT SPECIFICATIONS
2-2. GENERAL SPECIFICATIONS
3-1. WARNINGS
3-2. GENERAL WARNINGS AND CAUTIONS
1. Analog Multitester
2. Test Lead (100-51)
3. Holster
4. Spare Battery (1.5V R6P)
5. Instruction Manual
Allligator
Clip
Red
Test Lead
Black
Test Lead
Allligator
Clip Red
Test Lead
Black
Test Lead
Children and the persons who do not have enough knowledge
about electric measurements must not use this instrument.
Do not measure the electricity naked or barefooted to protect
yourself from electrical shock hazard.
Be careful not to get hurt with the sharp test lead pins.
Do not polish the case or attempt to clean it with any cleansing
fluid like gasoline or benzine. If necessary, use silicon oil or
antistatic fluid.
Avoid the multitester from hard mechanical shock or vibration,
high temperature and strong magnetic field.
Remove the batteries when the multitester is out of use for a
long time. The exhausted batteries might leak electrolyte and
corrode the inside.
WARNING 1.
WARNING 2.
WARNING 3.
CAUTION 1.
CAUTION 2.
CAUTION 3.
Measurement exceeding the specified maximum values of each measurement
range is prohibited.
Always confirm that RANGE Switch is set to the correct position. Do not measure
voltage in DC-mA, Ωand BAT. ranges.
Detach test leads from the measuring circuit before rotating RANGE Switch or
opening rear case for battery or fuse replacement.
WARNING
Detach test leads from the circuit before rotating RANGE Switch to avoid electric
shock hazard and serious damage to the instrument.
Select suitable scale from "0 - 6", "0 - 12", or "0 - 30" applying certain multiple.
Reading examples :
Multiply the scale value depending on the measurement range.
Reading examples :
Read the scale directly.
For accurate measurement, read following notes carefully.
DC 0.3V range : Read "0 - 30" scale multiplying by 1/100.
DC 120V range : Read "0 - 12" scale multiplying by 10.
DC 60uA range : Read "0 - 6" scale multiplying by 10.
DC 300mA range : Read "0 - 30" scale multiplying by 10.
"x 1" range : Read the scale directly.
"x 100" range : Multiply the scale value by 100.
"x 1k" range : Multiply the scale value by 1,000.
Correct knowledge of electric measurements is essential to avoid unexpected
danger such as operator's injury or damage to the instrument. Read carefully and
observe the following precautions for safety measurements.
Before measurement, confirm the body of this instrument and handle insulators of
the Test Lead have no cracks or any other damages. Dust, grease and moisture
must be removed.

This instrument is applicable for Low Energy Circuit measurements only.
Do not use for High Power Line (High Energy Circuit) measurements such as
Distribution Transformers, Bus Bars and Large Motors.
Even for Low Energy Circuits of electric/electronic appliances, such as heating
elements, small motors, line cords and plugs, High Voltage Measurements are
very dangerous. Do not touch Multitester, Test Leads, and any part of the circuit.
Generally, shock hazard could occur when the current between the circuit, that
involves more than 33V rms or 46.7V DC or peak, and ground goes up to 0.5mA
or more.
For dangerous voltage measurement, strictly observe the warnings below.
(refer to fig.1)
ANALOG MULTITESTER
KF-20
INSTRUCTION MANUAL
In case of live-line measurement, strictly observe the warnings
below (refer to fig 2.)
fig. 1
fig. 5
fig. 2
fig. 3
fig. 4
1. UNPACKING AND INSPECTIONS
1 pce.
1 set
1 pce.
1 pce.
1 pce.
Turn off the circuit to be measured.
Plug black test lead to "- (COM)" terminal and red test lead into "+" terminal.
Attach black and red alligator clips to test lead pins.
Set RANGE Switch to desired DC or AC voltage range.
Connect black alligator clip to - (earth) side and red alligator clip to + (positive)
side of the measuring circuit.
Do not hold Multitester and test leads in your hands. Keep safety distance
from power source or circuit to be measured not to touch the dangerous
voltage.
Turn on the circuit to be measured, and read the meter.
Turn off the circuit again. Detach alligator clips after confirming the meter
returns into "0".
1.
2.
3.
4.
5.
6.
7.
8.
Set RANGE Switch to desired DC or AC voltage range.
Attach alligator clip to black test lead pin, and connect to - (earth) side of the
measuring circuit.
Do not hold Multitester in your hands. Keep safety distance from power source
or circuit to be measured not to touch the dangerous voltage.
Connect red test lead to + (positive) side, and read the meter.
Detach red test lead from the circuit first, then detach black alligator clip.
1.
2.
3.
4.
5.
Before measurement, confirm that meter points to "0". (zero-adjusted)
Select proper range referring to 「5-1. 6-(1) Range Selection」.
Read the meter from directly over the pointer.
Measurement in strong magnetic field or on iron plate causes reading error or
meter sensitivity wrong.
Once-a-year periodical calibration is necessary to maintain tolerance accuracy
and to take safety measurements.
1.
2.
3.
4.
5.
Set the switch to a desired measurement range.
To measure uncertain voltage or current, firstly measure the approximate
value at the highest range, then change the switch into the suitable range.
For correct voltage and current measurements, select the range to point the
right side of the meter scale. (between center scale and maximum scale)
For correct resistance measurement, select the range to point around the
center of OHMS scale.
a.
b.
c.
DC Voltage ( DC-V )
AC Voltage ( AC-V )
DC Current ( DC-mA )
Resistance ( Ω)
1.5V Battery Test
Hi Range
Lo Range
Temperature ( ℃)
DC : ±3%f.s., AC : ±4%f.s.
Ω / TEMP : ±3%f.s.length
1.5V R6P (AA) battery ×1
136(H)×90(W)×30 (D)mm, 215g
100-51 Test Lead ×1 set
Holster ×1
1.5V R6P (AA) Battery (built-in and spare) ×1 each
Spare Fuse (0.3A/250V)×1 (inside the rear case)
Instruction Manual
660 Clamp Adapter
810-02 Temperature Probe
940 Alligator Clip
0.3 / 3 / 12 / 30 / 120 / 300 / 1200V
12 / 30 / 120 / 300 / 1200V
60 A / 3mA / 30mA / 300mA
5KΩ/ 500KΩ/ 5MΩ
1.5V R20P / R14P / R6P
1.5V 250mA
1.5V 50mA
−50℃〜+250℃
μ
Accuracy
Overload Protection
Power Supply
Dimensions & Weight
Accessories
Optional Accessories
×1 and 300mA range : protected by internal diode
and 0.3A / 250V fuse
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×1×100 ×1k AC-VDC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
1.5V VOLT
BATT
m
A
MADE IN JAPAN
NO.
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
‑50
1.5
1.0
0.5
0
℃
CAL
OHMS
KF-20
C O M
Ω
R
E
P
L
A
C
G
O
O
D
Meter Case
Pointer
Zero Adjustment
Screw
RANGE Switch
Rear Case
Meter Scale
0ΩADJ. Knob
+Terminal
−(COM)
Terminal
100-51 Test Lead
KF-20 contains 1 x fast-acting 0.3A/250V (φ5x20mm) fuse to protect Resistance
and Current measurement ranges. For replacement, refer to 「6-2. FUSE
REPLACEMENT」.
Insert black test lead into - (COM) terminal, and red test lead into + terminal.
Always confirm that RANGE Switch is set to the proper measurement range. To
avoid serious damage to the instrument, do not measure voltage in DC-mA, Ω
and BAT ranges.
3. FUSE
Before measurement, confirm that meter points to "0". If not, adjust the pointer by
Zero Adjustment Screw to take accurate measurement.
4. ZERO ADJUSTMENT
5. TEST LEAD
7. Meter Scale Reading
8. Measurement Accuracy
DC/AC scale (for AC/DC Voltage, DC Current measurements : V, μA, mA)
OHM scale (for Resistance measurement : Ω)
TEMP scale (for Temperature measurement : ℃)
(1) Range Selection
(2) Confirmation of the Range
6. RANGE SWITCH
Read INSTRUCTION MANUAL carefully to understand the specification and
functions correctly. 「3. SAFETY PRECAUTIONS」is higly important for safety
measurement.
KF-20 contains 1 x 1.5V R6P (AA) battery. Replace it referring to 「6-1.
BATTERY REPLACEMENT」when exhausted.
5-1. PREPARATION FOR USE
1. INSTRUCTION MANUAL
2. BATTERY
5. MEASUREMENT PROCEDURES
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
1.5V VOLT
BATT
0599060356-1
MADE IN JAPAN
NO.
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
R
E
P
L
A
C
E
G
O
O
D
UCHIDA
BATT scale
TEMP scale
KF-20
DC/AC scale
OHMS scale
TAUT BAND
DROP
PROOF
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
KF-20
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
KF-20
outlet
when measuring DC. mA /Ωrange
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
KF-20
×
DROP
PROOF
70-1201-KF20-1 0802
Do not measure high power line that might exceed 2kVA.
Do not exceed the maximum input value of each range.
Confirm that RANGE Switch is set to the correct position.
Read 「3. SAFETY PRECAUTIONS」carefully for safety measurement.
KF-20 is warranted in its entirety against any defects of material or workmanship
under normal use and service within a period of one year from the date of
purchase of the original purchaser. Warranty service is available at KAISE
AUTHORIZED SERVICE AGENCY through your local dealer. Their obligation
under this warranty is limited to repairing or replacing KF-20 returned intact or in
warrantable defect with proof of purchase and transport charges prepaid. KAISE
AUTHORIZED DEALER and the manufacturer, KAISE CORPORATION, shall
not be liable for any consequential damages, loss or otherwise. The foregoing
warranty is exclusive and in lieu of all other warranties including any warranty of
merchantability, whether expressed or implied.
This warranty shall not apply to any instrument or other article of equipment which
shall have been repaired or altered outside of KAISE AUTHORIZED SERVICE
AGENCY, nor which have been subject to misuse, negligence, accident, incorrect
repair by users, or any installation or use not in accordance with instructions
provided by the manufacturer.
Check the battery connection, polarity, and capacity.
Confirm that RANGE Switch is set to the correct position.
Confirm that the body of this instrument and handle insulators of the Test
Leads have no cracks or any other damages.
Check if any noise affects the instrument. This instrument is fully shielded
against noise, but possibly to be affected by very strong noise.
Repair service is available at KAISE AUTHORIZED SERVICE AGENCY through
your local dealer. Pack the instrument securely with your name, address,
telephone number and problem details, and ship prepaid to your local dealer.
KAISE AUTHORIZED DEALER
Product specifications and appearance are subject to change without
notice due to continual improvements.
5-3. AC VOLTAGE MEASUREMENTS (AC-V)
WARNINGS
●
●
●
●
6-1. BATTERY REPLACEMENT
6-4. REPAIR
WARNINGS
WARRANTY
1.
2.
3.
4.
5.
1.
2.
3.
4.
Replace the batteries when it is exhausted and "0ΩAdjustment" does not work.
Check the following items before asking repair service.
To avoid electrical shock, detach test leads from circuit when when
opening rear case for battery or fuse replacement.
Be sure to tighten the screw of rear case after the battery replacement
Detach test leads from the circuit.
Unscrew rear case.
Remove rear case opening from meter side and slide to the upward.
Replace an exhausted battery into a new 1.5V R6P (AA) battery in correct
polarity.
Close rear case fitting from bottom to up, and tighten the screw.
422 Hayashinogo, Ueda City, Nagano Pref., 386-0156 Japan
TEL : +81-268-35-1600 (REP.) / FAX : +81-268-35-1603
E-mail : [email protected]
http://www.kaise.com
5678
fig. 10
6.MAINTENANCE
Do not measure over 300mA DC.
Do not exceed the maximum input value of each range.
Confirm that RANGE Switch is set to the correct position.
Read 「3. SAFETY PRECAUTIONS」carefully for safety measurement.
5-4. DC CURRENT MEASUREMENTS (DC-mA)
WARNINGS
●
●
●
●
Do not measure high power line that might exceed 2kVA.
Do not exceed the maximum input value of each range.
Confirm that RANGE Switch is set to the correct position.
Read 「3. SAFETY PRECAUTIONS」carefully for safety measurement.
WARNINGS
●
●
●
●
The internal protection circuit against over current for meter movement and the
internal circuit is installed. But, the instrument should be defective by such as over
input measurement, or applying voltage to resistance or current range.
9. Protection Circuit
5-2. DC VOLTAGE MEASUREMENTS (DC-V)
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to a desired DC-V range.
NOTE :
Connect black test lead to −(earth) side and red test lead to +(positive) side
of the measuring circuit.
NOTE : Connect the multitester IN PARALLEL to the circuit.
NOTE : Use alligator clips for dangerous voltage measurement.
Read the meter on DC/AC scale referring to 「5-1. 7. Meter Scale Reading」.
Detach test leads from the circuit.
To measure uncertain voltage, firstly measure the approximate value
at 1200V range, then change the switch into the suitable range.
Detach test leads from the circuit before rotating RANGE Switch.
1.
2.
3.
4.
5.
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to a desired AC-V range.
NOTE :
Connect black test lead to −(earth) side and red test lead to +(positive) side
of the measuring circuit.
NOTE : Connect the multitester IN PARALLEL to the circuit.
NOTE : Use alligator clips for dangerous voltage measurement.
Read the meter on DC/AC scale referring to 「5-1. 7. Meter Scale Reading」.
Detach test leads from the circuit.
To measure uncertain voltage, firstly measure the approximate value
at 1200V range, then change the switch into the suitable range.
Detach test leads from the circuit before rotating RANGE Switch.
Installed battery is for inspection purpose and might be exhausted
earlier than the new battery.
300mA range is protected by 0.3A/250V fuse, but 60uA, 3mA, and 30mA
ranges are not.
Do not measure voltage in DC-mA range.
Do not measure wall outlet or car battery by mistake.
1.
2.
3.
4.
5.
Do not measure voltage in Ωrange.
When measuring in-circuit resistor, turn off the circuit to be measured and
discharge the capacitors.
5-5. RESISTANCE MEASUREMENTS (Ω)
WARNINGS
●
●
●
●
6-2. FUSE REPLACEMENT
6-3. PERIODICAL CHECK AND CALIBRATION
WARNINGS
Always use the specified fuse. Do not use the instrument shorting fuse
holder or without using the fuse.
Be sure to tighten the screw of rear case after the fuse replacement.
●
●
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to a desired Ωrange.
When measuring in-circuit resistor, turn off the circuit to be measured and
discharge the capacitors.
0ΩAdjustment
Short-circuit metal pins of the test leads. Turn 0ΩADJ. Knob until the meter
points to "0Ω".
NOTE : 0ΩAdjustment is necessary for each time when Ωrange is selected.
NOTE : 0ΩAdjustment does not work if the battery is exhausted. Replace it
referring to「6-1. BATTERY REPLACEMENT」.
Connect test leads to resistor to be measured detaching its one side from the
circuit.
Read the meter on OHMS scale referring to 「5-1. 7. Meter Scale Reading」.
NOTE : Read the scale directly when measuring in "x 1" range. In "x 100"
range, multiply the scale by 100, and in "x 1k" range, multiply by 1,000.
NOTE : For correct resistance measurement, select the range to point around
the center of OHMS scale.
Detach test leads from the resistor.
1.
2.
3.
4.
5.
6.
7.
Do not test non-specified batteries to avoid electric shock hazard and
serious damage to the instrument.
Do not test car battery.
Do not measure voltage or current in BAT range.
Battery exhausted level can be tested in Hi (1.5V 250mA), or Lo (1.5V 50mA)
ranges.
Testable batteries : 1.5V R20P (D), 1.5V R14P (C), 1.5V R6P (AA), 1.5V R03 (AAA)
5-6. 1.5V BATTERY TEST (BATT)
WARNINGS
●
●
●
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to "Hi" or "Lo" of BAT range.
NOTE : "Lo" range applies 50mA, and "Hi" range applies 250mA to the battery.
Confirm consumption current of the instruments that the battery will be used to
select the proper testing range.
Popular range selection : 1.5V R20P (D), 1.5V R14P (C) : "Hi" range
1.5V R6P (AA), 1.5V R03 (AAA) : "Lo" range
Connect black test lead to −(earth) side and red test lead to +(positive) side.
Read the meter on BATT scale.
Meter points "GOOD (blue)" : The tested battery is good.
Meter points "REPLACE (red)" : The tested battery is exhausted and needs to
be replaced.
NOTE : Even if meter points "REPLACE (red)" for 1.5V R6P (AA) battery tested
in "Lo" range, it may be used with low consumption current instruments.
1.
2.
3.
4.
810-02 temperature probe (option) is necessary. Measurement range is -50℃to
+250℃.
5-7. TEMPERATURE MEASUREMENTS (℃)
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to TEMP (x100) range.
0ΩAdjustment
Short-circuit metal pins of the test leads. Turn 0ΩADJ. Knob until the meter
points to "CAL" line on TEMP scale.
Remove test leads from the terminals, and insert 810-02 temperature probe.
Connect temperature probe to the object to be measured.
Read the meter on TEMP scale.
1.
2.
3.
4.
5.
6.
Plug Black Test Lead into "- (COM)" terminal and Red Test Lead into "+"
terminal.
Set RANGE Switch to a desired DC-mA range.
WARNINGS
Turn off the circuit to be measured and discharge the capacitors.
Connect black test lead to −(earth) side and red test lead to +(positive) side
of the measuring circuit.
NOTE : Connect the multitester IN SERIES to the circuit.
NOTE : Using alligator clips is recommended for safety measurement.
Turn on the circuit to be measured.
Read the meter on DC/AC scale referring to 「5-1. 7. Meter Scale Reading」.
Turn off the measured circuit again and discharge the capacitors.
Detach test leads from the circuit.
1.
2.
3.
4.
5.
6.
7.
8.
●
●
●
NOTE :
1.
2.
3.
DC Current and Resistance measurement cannot be made if fuse blows out.
This instrument contains one spare fuse inside the rear case.
Periodical check and calibration is necessary to make safety measurements and
to maintain the specified accuracy. The recommended check and calibration term
is once a year and after the repair service. This service is available at KAISE
AUTHORIZED SERVICE AGENCY through your local dealer.
Open rear case referring to 「6-1. BATTERY REPLACEMENT」.
Replace a blowout fuse into a new one.
Fuse type : Fast-acting 0.3A/250V (φ5x20mm)
Close rear case fitting from bottom to up, and tighten the screw.
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
mA
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
UCHIDA
battery
KF-20
Red
Test Lead
Black
Test Lead
Red
Test Lead
Black
Test Lead Red
Test Lead
Black
Test Lead
Red
Test Lead
Black Test Lead
fig. 6
fig. 7
fig. 9
fig. 8
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
UCHIDA
outlet
KF-20
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
UCHIDA
+−
+−
Switch
KF-20
1200 1200
300
120
30
12
300
120
30
12
3
3
60 30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC DC
AC
BATT
TEMP
TAUT BAND
1.5VVOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000/V AC
20,000/V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CAL
OHMS
C O M
Ω
R
E
P
L
A
C
E
G
O
O
D
Resistor
KF-20
Rear Case 1.5V R6P Battery
Screw

Other Kaise Multimeter manuals

Kaise SK-350 User manual

Kaise

Kaise SK-350 User manual

Kaise KT-2021 User manual

Kaise

Kaise KT-2021 User manual

Kaise SK-4033 User manual

Kaise

Kaise SK-4033 User manual

Kaise 7050 User manual

Kaise

Kaise 7050 User manual

Kaise SK-7820 User manual

Kaise

Kaise SK-7820 User manual

Kaise KU-2600 User manual

Kaise

Kaise KU-2600 User manual

Kaise SK-7720 User manual

Kaise

Kaise SK-7720 User manual

Kaise KT-2022 User manual

Kaise

Kaise KT-2022 User manual

Popular Multimeter manuals by other brands

Tietzsch VARIOSAFE EXM 25 User instructions

Tietzsch

Tietzsch VARIOSAFE EXM 25 User instructions

Gossen MetraWatt METRAHIT IM E-DRIVE FUSO Short-form operating instructions

Gossen MetraWatt

Gossen MetraWatt METRAHIT IM E-DRIVE FUSO Short-form operating instructions

Orbit Controls OC502-t owner's manual

Orbit Controls

Orbit Controls OC502-t owner's manual

Velleman DVM4200 user manual

Velleman

Velleman DVM4200 user manual

Extech Instruments DM4750D manual

Extech Instruments

Extech Instruments DM4750D manual

VOLTCRAFT VC930 operating instructions

VOLTCRAFT

VOLTCRAFT VC930 operating instructions

LIMIT 500 Auto operating manual

LIMIT

LIMIT 500 Auto operating manual

KPS KPS-MT440 instruction manual

KPS

KPS KPS-MT440 instruction manual

Gossen MetraWatt METRAport 32 XS operating instructions

Gossen MetraWatt

Gossen MetraWatt METRAport 32 XS operating instructions

Velleman DVM 890L user manual

Velleman

Velleman DVM 890L user manual

Micronta 22-191 owner's manual

Micronta

Micronta 22-191 owner's manual

Elenco Electronics M-1150K Assembly and instruction manual

Elenco Electronics

Elenco Electronics M-1150K Assembly and instruction manual

Megger M8035 user guide

Megger

Megger M8035 user guide

Bosch MMD 540H instruction manual

Bosch

Bosch MMD 540H instruction manual

Fluke III Series instruction sheet

Fluke

Fluke III Series instruction sheet

Agilent Technologies Agile or Elusive Signals Using Information sheet

Agilent Technologies

Agilent Technologies Agile or Elusive Signals Using Information sheet

Smart Electrician 364-5017 Operator's instruction manual

Smart Electrician

Smart Electrician 364-5017 Operator's instruction manual

Extech Instruments RHT50 user guide

Extech Instruments

Extech Instruments RHT50 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.