Kambic OBM User manual

User Manual
Calibration Bath
OBM

User Manual
OBM
2
WARRANTY
Kambic d.o.o. guaranties that the product
delivered has been thoroughly tested to ensure
that it meets its published specifications. The
warranty is valid only if the product has been
installed and used in accordance to the
instructions supplied by Kambic d.o.o.
Kambic d.o.o. shall in no event be liable for
incidental or consequential damages, including
without limitations, lost profits, loss of income, loss
of business opportunities, loss of use, and other
related exposures, caused by incorrect use of
product.
Warranty is valid for 13 months from date of
shipping and includes spare parts and labor.
Type: OBM
Serial number:
Date of shipping:
Signature/stamp:
Manufacturer:
Kambic d.o.o.
Metliska cesta 16
8333 Semic
Slovenia
Tel: +386 7 35 65 220
Fax: +386 7 35 65 232
Email: info@kambicmetrology.com
Web: www.kambicmetrology.com
Thank you for your trust and for buying this
device. We hope it will successfully serve the
purpose for many years.
CAREFULLY READ THESE INSTRUCTIONS
BEFORE USING THE DEVICE!

User Manual
OBM
3
CONTENT
1. WARNINGS AND SAFETY PRECAUTIONS........... 4
2. INTRODUCTIONS ................................................ 6
3. TECHNICAL DATA ................................................ 7
3.1 Front view............................................................. 8
3.2 Rear view ............................................................. 8
3.3 Top view............................................................... 9
4. INSTRUCTIONS FOR USE.................................. 10
4.1 Basic Setup ......................................................... 10
4.2 Control Panel with OLED display and keyboard...... 10
4.2.1 Main Screen (running) ...................................... 11
4.2.2 Main Screen (not running)................................. 11
4.2.3 Program selection and start............................... 12
4.2.4 Menu and navigation ........................................ 15
4.3 Program Setup ................................................... 16
4.4 Instrument Setup ............................................... 18
4.4.1 Display Parameters ........................................... 18
4.4.2 Time and Date ................................................. 19
4.4.3 System Parameters........................................... 19
4.4.4 Bath Calibration ............................................... 20
4.4.5 PID parameters ............................................... 20
4.4.6 Change Password ............................................. 21
4.5 Service .............................................................. 22
5. SERVICE AND MAINTENANCE .......................... 23
5.1 Draining the Bath ................................................ 24
6. REQUIRES SPACE AND CONECTIONS .............. 25
7. SAFE USE AND SAFETY ELEMENTS ................... 26
8. DISMANTLING AND DISPOSAL ........................ 27
9. ABBRIVATIONS ................................................. 28
10. TROUBLESHOOTING ....................................... 29
11. ATTACHMENTS ................................................ 30
11.1 Wiring diagram
11.2 Declarations
11.3 RS232 Communication protocol

User Manual
OBM
4
1. WARNINGS
AND SAFETY
PRECAUTIONS
Always keep this manual in close reach for reference. This product should only be used
and operated by skilled and trained personnel familiar with technical capabilities and
potential safety hazards.
Always follow and obey the national regulations related to technology and labor laws.
To avoid injuries and equipment damage study and follow the safety instructions and
warnings.
Legal:
The content of the manual is subject to change without notice.
Understanding and following this operating manual is first step to hazard-free and safe
use of the equipment.
This user manual can not reach out to all application that might even develop in the
future. If you find information unclear or you are not sure about safe use of the product in
further information or opinions.
In no event shall Kambic d.o.o. be held liable for any damages, direct or incidental related
to the use of this product and manual.
Product can be equipped with variety of accessories and options. This warning list might
include elements that are applicable only with certain models and accessories.
READ ALL INSTRUCTIONS BEFORE USE!
Reduce the possibility of injury and failure.
WARNING
Before you perform
any kind of maintenance, disconnect all electrical power
to the unit! DO NOT operate this unite without a properly grounded,
properly polarized power cord. DO NOT connect this unite to a non-
grounded, non-polarized outlet.
DANGER
Appliance must not be installed in a room where the possibility of explosive
atmosphere is present!
DANGER OF BREAKDOWN
Service only allowed to authorized personnel!

User Manual
OBM
5
WARNING
Never use with toxic, corrosive, flammable or organic materials unless special
precautions are in place to prevent injury to personnel or damage to equipment.
WARNING
Never operate the appliance with the housing removed!
RISK OF FAILURE
Due to exceptional volumes of fluid expansion at elevated temperatures,
regularly check the fluid level in the bath!
DANGER
Temperature fluid expansion may cause overspill!
Adjust fluid level!
RISK OF FAILURE
Do not use rubber cap at elevated temperatures
DANGER OF BURNS
Use safety gloves when operating with high/low temperatures!
ELECTRIC HAZARD
Do not operate the equipment in wet and condensing environment!
STABILITY HAZARD
Do not load or affect product stability with heavy loads or climbing the product!
RISK OF FAILURE
Ensure correct flow. Insuficient flow will resault poor stability and uniformity.
Check stir magnet-bar position and rotation to ensure stirring / flow.
✓

User Manual
OBM
6
2. INTRODUCTIONS
Calibration bath OBM was developed and manufactured specifically for the calibration of
temperature instruments, probes and liquid thermometers at temperatures from 35°C to
220°C. Fluid calibration baths ensure high temperature stability and uniformity across
working volume.
To ensure such specifications all calibration baths of our brand are designed with a special
double cylinder construction with uniform vertical flow in the working volume.
Of course, the construction is not enough that is why we are using high capacity
microprocessor controllers and different levels of heating.
Control panel with display is used to set the individual temperatures. All controllers are
also equipped with communication modules, so the appliance can be operated and
controlled via a PC.
Bath top, bath bottom and working chamber is completely made of stainless steel AISI 304,
which guarantees easy maintenance and long life time. Exterior housing is made of
aluminum sheets coated with high quality varnish.
WARNING
Temperature range must be limited / adjusted to reflect the properties of
the selected fluid!

User Manual
OBM
7
3. TECHNICAL DATA
OBM
Width
225 mm
Height
638mm
Depth
330 mm
Working chamber dimensions
50 x 210 mm
Volume
~ 1 L
Temperature range
+ 35 °C …+ 220 °C
Temperature Stability
from ± 0.005 °C to ± 0.05 °C
Temperature homogeneity
Better than ± 0.05 °C
Display resolution
0.001 °C
Set-up resolution
0,01 °C
Control
MPC –PID
Power supply (±10%)
230 V / 50 Hz
Wattage
300 W
Communication port RS 232
Built-in
Temperature probe
PT 100
Construction materials
SS vessel, exterior housing & cover
Width
225 mm

User Manual
OBM
8
3.1 Front view
3.2 Rear view
Control panel
Bath working chamber
Rubber cap
Refrigeration system
compartment –air intake
Perforated back
Main switch
Power cord connection
RS 232 PC port
USB port
Rubber legs
Front housing
Back housing
Carrying handle
Tank drain plug

User Manual
OBM
9
3.2 Rear view
Bath working chamber
Removable inner tank wall
Openings on inner tank wall

User Manual
OBM
10
4. INSTRUCTIONS
FOR USE
4.1 Basic setup
Step 1: Remove rubber cap and fill the bath with appropriate fluid (up to 80%)
Step 2: Connect the power cord into the power supply 230V±10%; 50Hz
Step 3: Put the main switch in position 1 (on the back of the device)
Step 4: Press button and in Program Start Menu select the desired program. After selection press
button to start cycle
Step 5: Check the fluid level and add if necessary
Step 6: Press button to stop the cycle
4.2 Control panel with OLED display and keyboard
START button
ESC button
STOP button
OLED display
Increase value button
Decrease value button
Navigate right button
Navigate left button
Navigate down button
Navigate up button
ENTER button
MENU button

User Manual
OBM
11
4.2.1 Main screen (running)
Main screen while cycle is running:
4.2.2 Main screen (not running)
After button is pressed you see this screen:
Setpoint temperature
Heating power
Stirrer speed in %
Bath temperature
Current program number
Temperature status
Bath temperature
Heating power 0%
Bath status (not running)
Current date
Current time

User Manual
OBM
12
4.2.3 Program selection and start
The bath must be in Main screen (not running):
Press button to enter Program Start Menu:
With and buttons select desired program. Select from 0 to 10. Program 0 is reserved for manual
program and 1 to 10 is automatic program.
After selection press or button to start the cycle.
To skip start press or button to go back to Main screen.
Program number

User Manual
OBM
13
When cycle has started you see this screen for manual program :
Bath temperature: current measured bath temperature is 19.85°C
Heating power: current heater duty cycle 14% (from 0% to 100%)
Temperature status: Stabilizing = temperature is stabilizing and it is not yet in setpoint tolerances
Stable = temperature is stable and it is in setpoint tolerances
Current program: the current running program number is 0(0- manual program, 1÷10 for automatic
program)
Setpoint temperature: current setpoint temperature 50.00°C (setpoint range from 35.00°C to 220.00°C)
Stirrer speed: the set speed for stirrer is 100% ( set range from 30% to 100%)
WARNING
When operating at higher temperatures set the stirring speed at 80%!
Setpoint temperature
Heating power
Stirrer speed
Bath temperature
Current program
Temperature status

User Manual
OBM
14
When cycle has started you see this screen for automatic program :
Bath temperature: current measured bath temperature is 26.31°C
Heating power: current heater duty cycle 0% (from 0% to 100%)
Temperature status: Stabilizing = temperature is stabilizing and it is not yet in setpoint tolerances
Stable = temperature is stable and it is in setpoint tolerances
Current program: the current running program number is 2(0- manual program, 1÷10 for automatic
program)
Setpoint temperature: current setpoint temperature 50.00°C (setpoint range from 35.00°C to 220.00°C)
Stirrer speed: the set speed for stirrer is 100% ( set range from 30% to 100%)
Step dwell time: remaining time of temperature stability for current step is 0:30:00 = 30 minutes
(in hh:mm:ss format). Set range from 00:01 to 24:00 (in hh:mm format).
For endless dwell time set program to 00:00.
Step information: 1/ 1 ( current temperature step / all temperature steps in program).
Maximum number of temperature steps is 10.
When cycle has finished you see confirmation screen:
Setpoint temperature
Heating power
Stirrer speed
Bath temperature
Current program
Temperature status
Step information
Step dwell time

User Manual
OBM
15
4.2.4 Menu and navigation
Press button to enter MAIN MENU:
The Main Menu includes three submenus: Program Setup,Instrument Setup and Service:
Use navigation buttons and to select and then press button to enter selected submenu.
Press button to go back one level.
MAIN MENU
Program Setup
Instrument Setup
Service

User Manual
OBM
16
4.3 Program Setup
Use navigation buttons and to select parameter and then press button to edit parameter.
With and buttons change value of parameter.
To move cursor left and right use and button.
Press button to confirm or button cancel changes.
After edit press button to go back to MAIN MENU.
1
2
3
4
5
6
7
8
9
10
11
12
13
Current program number setup
Current program number and step number setup

User Manual
OBM
17
1. Program: choose desire program number to edit settings (0- manual program, 1÷10 for automatic
program).
2. Steps: choose the required number of temperature steps (set range from 1to 10).
3. Stirring: choose speed for stirrer for fluid mixing (set range from 30% to 100%)
4. Auto Start: controller has option to set automatic start at certain hour and date (select YES or NO).
If you would like to start the program right after pressing the start button the auto start should be
set to NO.
5. Cycling: here you activate or deactivate cycling of steps (select YES or NO). If you would like to
work without cycling of steps then select NO.
6. Start Time: set the time for delayed automatic start (set range from 00:00 to 23:59 in hh:mm
format).
7. Start Date: set the date for delayed automatic start (set range ALL and 1to 31). The
number is day in month. If you set to ALL then all days in month are selected.
8. Cycles: set how many times the cycling of steps is repeated (set range from 1to 99).
9. Cycle from step: set the step number from which the cycling will start (set range from 1to
maximum 10 if we have 10 temperature steps). Cycling from step number must always be
bigger than Cycling to step number!
10. Cycle to step: set the step number to which the cycling will go (set range from 1to maximum
10 if we have 10 temperature steps) Cycling to step number must always be smaller than Cycling
from step number!
Example: “number of temperature steps” is set to 5, cycling of steps is set to 3, “cycle from
step” is 5, “cycle to step” is set to 1. So the program shall do 5 steps and then repeat them
for 3 times. Together we have 5 + 3 x 5 = 20. We have 20 steps to execute.
11. SETPT: set the setpoint temperature for current step (set range from 35.00°C to 220.00°C)
12. Dwell time: set the duration time of temperature stability for current step (set range from 00:01 to
24:00 in hh:mm format). For endless dwell time set time to 00:00. The countdown will start only
when temperature is within allowed tolerances from set values!
13. Ramp: set the speed how fast the temperature in bath changes (set range from 1°C/hr to 250°C/hr
(°C degrees per hour)). To activate speed ramp set mark ✓in rectangle.
Fluid mixing speed
After changing this parameter wait for a few
minutes for the system to stabilize!!!
Never change for more than a few % at a time!

User Manual
OBM
18
4.4 Instrument Setup
Use navigation buttons and to select parameter and then press button to edit parameter.
With and buttons change value of parameter.
To move cursor left and right use and button.
Press button to confirm or button cancel changes.
Press button to go back one level.
4.4.1 Display Parameters
Show display language –or select display language if there is more languages installed:
4.4.1
4.4.2
4.4.3
4.4.4
4.4.5
4.4.6

User Manual
OBM
19
4.4.2 Time and Date
Set time and date:
Time format: hh:mm:ss
Date format: yyyy.mm.dd
4.4.3 System Parameters
Setup system parameters. To change parameter you must first enter password:
Default values
Max SETP limit: 220°C
Heat power limit: 100%
USB COM Baudrate: 230400
RS232 Baudrate: 230400

User Manual
OBM
20
4.4.4 Bath Calibration
Setup bath temperature PT100 probe corrections. To change parameter you must first enter password:
Default values
At: 50°C 49.600°C
At: 100°C 99.500°C
At: 150°C 149.500°C
At: 200°C 199.500°C
At: 220°C 219.500°C
4.4.5 PID parameters
Setup bath PID parameters. To change parameter you must first enter password:
Default values
PID Pb: 2.35°C
PID Ti: 180sec
PID Td: 15
WARNING
Changing the PID parameters can seriously affect temperature control, stability,
stabilization time,…
This should only be done by skilled professional staff!!!
Table of contents
Other Kambic Laboratory Equipment manuals