Kensington Scientific KS-EC-10L-E User manual

User Manual
KS-EC-10L-E | Pocket Conductivity Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
OVERVIEW
Below are the steps you need to follow to begin using your Kensington KS-EC-10L-E pocket conductivity tester. This
manual will cover how insert the batteries, using the function keys, making measurements, calibration and sensor
replacement.
INSERTING THE BATTERIES
1. Your KS-EC-10L-E pocket pH tester comes with one G13a button battery. Remove it from the package.
2. Remove the button cap found on the bottom of the tester as shown in the above image.
3. Place the battery into its designated slot. Take note of the proper position of the battery to ensure the polarities
placed correctly.
4. Replace the battery cap.
KEYPAD INDEX
Key
Function
1. Serves as the power button of the tester
2. Initiates Data Hold. Press once to hold the current reading, press again to resume making
measurements.
3. Exit button during calibration
1. Press once to conduct calibration of the instrument
2. Press the button again to choose the type of buffer used in the calibration.
1. Press to show the current temperature during measurement mode
2. Decreases the setting value in calibration mode.
User Manual
KS-EC-10L-E | Pocket Conductivity Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
CONDUCTIVITY CALIBRATION
It is recommended that the KS-EC-10L-E should be calibrated regularly to ensure accurate measurements. The KS-
EC-10L-E allows 1 point calibration and the instrument’s acceptable calibration ranges from 100 to 180 µS/cm. These
solutions are sold separately. Ensure that you will not reuse spent calibration solutions.
1. Rinse the conductivity electrode with distilled water then rinse with a small amount of calibration solution.
2. Dip the electrode in your chosen calibration solution (146.5 µS/cm is required), stir the electrode gently.
3. Make sure there are no air bubbles trapped in the slot of the sensor.
4. Press and hold the CAL key. The tester will now enter into calibration mode.
5. Press or to set the calibration value
6. Ensure the displayed value matches your calibration standard. Press MEAS to confirm.
7. The calibration value will then automatically flash three (3) times and then go back to measurement mode to
complete the calibration.
USING THE TESTER
CONDUCTIVITY MEASUREMENT
1. Press the MEAS button to initiate measurement mode.
2. Rinse the conductivity electrode with distilled water.
3. Dip the conductivity electrode into the sample solution. Gently stir the tester.
4. Wait for the reading to stabilize and record the measurement.
TEMPERATURE MEASUREMENT
To measure the temperature, press the while in measurement mode. Pressing it again returns to conductivity
mode.
User Manual
KS-EC-10L-E | Pocket Conductivity Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
HOLD FUNCTION
1. Pressing the HOLD key while in measurement mode will activate the hold function. The HOLD indicator will appear
on the display and further measurement readings will not be possible.
2. Press the HOLD key again to disable the hold function.
ELECTRODE SENSOR MAINTENANCE
1. It is recommended to keep the sensitive membrane of the sensor wet. This is accomplished by periodically dipping
the sensor in distilled water or by using an electrode storage solution when the tester is not in use.
2. If the instrument will be kept in storage for a long time, place the sensor in an electrode storage solution.
3. Make sure to wash the sensor with distilled water before and after use.
4. Do not reuse prepared buffer solutions.
ELECTRODE REPLACEMENT
One of the advantages of using the Kensington KS-EC-10L-E is that the sensor can be replaced instead of buying a
new tester when the sensor fails during calibration. Removing and installing the sensor is easy and convenient and
also increases the utility of the instrument as additional sensors can be purchased for different applications. To replace
the sensor:
1. Twist the electrode collar counter-clockwise and pull the electrode away from the instrument
2. Align the new electrode into the same spot, push, and turn clockwise to keep it secure.
User Manual
KS-EC-10L-E | Pocket Conductivity Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
SPECIFICATIONS
Model
KS-PH-10S-E
Range
1.0~199.9µS/cm
Accuracy
± 0.1% FS
Calibration points
1 point
Calibration solutions
146.5µS/cm
Power off
Manual or Automatic (8 minutes after last key pressed)
Temperature compensation
0~50°C
Temperature coefficient
2%/°C
Cell constant
K=1
Normalization Temperature
25°C
Hold function
Manual
Operating temperature
0~60°C, 32~140°F
Power requirements
3× 1.5V "G13A" Batteries
Dimensions
185(L)× 40(Dia.)mm
Weight
100 grams

Popular Test Equipment manuals by other brands

Vaisala AQT400 user guide

Vaisala

Vaisala AQT400 user guide

Fluke ScopeMeter 190 Series III Service manual

Fluke

Fluke ScopeMeter 190 Series III Service manual

Tektronix 561A instruction manual

Tektronix

Tektronix 561A instruction manual

Drummond MTL10 instruction manual

Drummond

Drummond MTL10 instruction manual

Tektronix 465B instruction manual

Tektronix

Tektronix 465B instruction manual

Secure Care Products 135 user guide

Secure Care Products

Secure Care Products 135 user guide

Extech Instruments ET10 user manual

Extech Instruments

Extech Instruments ET10 user manual

VOLTCRAFT IRS-350 Blackbody operating instructions

VOLTCRAFT

VOLTCRAFT IRS-350 Blackbody operating instructions

Micsig SigOFIT MOIP01P SigOFIT MOIP01P quick guide

Micsig

Micsig SigOFIT MOIP01P SigOFIT MOIP01P quick guide

Multi-Amp SR-51 A instruction manual

Multi-Amp

Multi-Amp SR-51 A instruction manual

InstruStar Electronic Technology ISDS205C user manual

InstruStar Electronic Technology

InstruStar Electronic Technology ISDS205C user manual

Tegam 2602A Instruction and service manual

Tegam

Tegam 2602A Instruction and service manual

Mitutoyo SJ-301 user manual

Mitutoyo

Mitutoyo SJ-301 user manual

M.C. MILLER VADER user guide

M.C. MILLER

M.C. MILLER VADER user guide

Micronta 22-204C instruction manual

Micronta

Micronta 22-204C instruction manual

Horotec FLASHTEST Calibration manual

Horotec

Horotec FLASHTEST Calibration manual

BMS HARTIP 2000 Operation manual

BMS

BMS HARTIP 2000 Operation manual

Ebyte MBL Series user manual

Ebyte

Ebyte MBL Series user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.