CONDUCTIVITY CALIBRATION
It is recommended that the KS-EC-10L-E should be calibrated regularly to ensure accurate measurements. The KS-
EC-10L-E allows 1 point calibration and the instrument’s acceptable calibration ranges from 100 to 180 µS/cm. These
solutions are sold separately. Ensure that you will not reuse spent calibration solutions.
1. Rinse the conductivity electrode with distilled water then rinse with a small amount of calibration solution.
2. Dip the electrode in your chosen calibration solution (146.5 µS/cm is required), stir the electrode gently.
3. Make sure there are no air bubbles trapped in the slot of the sensor.
4. Press and hold the CAL key. The tester will now enter into calibration mode.
5. Press or to set the calibration value
6. Ensure the displayed value matches your calibration standard. Press MEAS to confirm.
7. The calibration value will then automatically flash three (3) times and then go back to measurement mode to
complete the calibration.
USING THE TESTER
CONDUCTIVITY MEASUREMENT
1. Press the MEAS button to initiate measurement mode.
2. Rinse the conductivity electrode with distilled water.
3. Dip the conductivity electrode into the sample solution. Gently stir the tester.
4. Wait for the reading to stabilize and record the measurement.
TEMPERATURE MEASUREMENT
To measure the temperature, press the while in measurement mode. Pressing it again returns to conductivity
mode.