CONTENTS
PAGE NUMBER
INTRODUCTION
2
WARRANTY
2
DELIVERY
2
1. SAFETY INSTRUCTIONS
2
2. PROPER USE
3
3. INSPECTION
4
4. CONTROL
4
5. TRIAL RUN
6
6. USING THE EXTERNAL TEMPERATURE SENSOR
6
7. RESIDUAL HEAT WARNING ‘HOT’
7
8. LOGGING FUNCTION
7
9. FAULTS
7
10. MAINTENANCE AND CLEANING
8
11. ASSOCIATED STANDARDS AND REGULATIONS
9
12. SPECIFICATIONS
10
13. ORDERING INFORMATION
11

2
INTRODUCTION
Welcome to the Magnetic Hotplate Stirrer User Manual. Users should read this Manual carefully, follow the instructions and
procedures, and be aware of all the cautions when using this instrument.
WARRANTY
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a
period of 12 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to
any product or parts which have been damaged on account of improper installation, improper connections, misuse,
accident or abnormal conditions of operation. For claims under the warranty please contact your local supplier.
DELIVERY
This unit is supplied with one main unit, power cable, stirrer bar and user manual.
1. SAFETY INSTRUCTIONS
Warning!
• Read the operating instructions carefully before use.
• Ensure that only trained staff works with the instrument.
Risk of burn! (hotplate model)
• Caution when touching the housing parts and the hotplate they can reach temperature of 500°C.
• Pay attention to the residual heat after switching off.
Protective ground contact!
• Make sure that socket is grounded (protective ground contact) before use.
• When working wear personal safety guards avoid the risk from:
o Splashing and evaporation of liquids
o Release of toxic or combustible gases
• Set up the instrument in a spacious area that’s stable, clean, non-slip, dry and fireproof surface. Do not operate
the instrument in explosive atmospheres, with hazardous substances or under water.
• Gradually increase the speed but reduce the speed if:
o Stirring bar breaks away due to high speed
o The instrument is not running smoothly, or container moves on the base plate
• When using the hotplate model temperature must always be set to at least 50℃lower than the fire point of the
media used.
• Be aware of hazards due to:
PRODUCT CODE
DESCRIPTIONS
400.100.130
Stirrer Magnetic Hotplate up to 30L

3
o Flammable materials or media with a low boiling temperature (Hotplate model)
o Overfilling of media
o Unsafe container
• Process pathogenic materials only in closed vessels.
• In the case of the stirrer bar is PTFE, please note:
o Elemental fluorine, three fluoride and alkali metals will corrode the PTFE and Halogen alkanes makes it
expand at room temperature
o Molten alkali, alkaline earth metals or their solution, as well as the powder in second and third ethnic of
the Periodic Table of Elements will have chemical reaction with PTFE when temperature reaches 300 ~
400 ℃.
• Check the instrument and accessories prior to each use. Do not use damaged components. Safe operation is only
guaranteed with the accessories described in the user manual. Accessories must be securely attached to the
device. Always disconnect the plug before fitting accessories.
• When using the external temperature sensor, the tip of the measuring sensor must be at least 5-10mm from vessel
bottom and wall.
• The instrument can only be disconnected from the main power supply by pulling out the main or the connector
plug.
• The voltage stated on the label must correspond to the main power supply.
• Ensure that the main power supply cable does not touch the hotplate. Do not cover the device.
• The instrument may only be opened by a qualified technician.
• Keep away from high magnetic field.
2. PROPER USE
The instrument is designed for mixing and/or heating liquids in schools, laboratories or factories.
Observe the minimum distances between the devices and the wall and above the assembly (min. 100 mm)
Figure 1
This device is not suitable for using in residential areas or other constraints.

4
3. INSPECTION
Receiving Inspection
Unpack the equipment carefully and check for any damages which may have arisen during transport. Please contact
manufacturer/supplier for technical support.
Note:
If there is any apparent damage to the system, please do not plug it into the power line.
4. CONTROL
Control elements
ITEM
DESCRIPTION
TEMPERATURE CONTROL KNOB
The heating function is switched
SPEED CONTROL KNOB
Set the rated rotary speed. The stirring function is switched ON or OFF by
LED HEAT
When the heating function is switched ON, the Led Heat is lit
LED STIR
When the stirring function is switched ON, the LED Stir is lit
TIMER
Press the know and then press “-“ or “+” to set the heating time value
LCD DISPLAY
If rotate the heating knob, LCD displays the temperature setting value (right
side) and real value (left side)
SD CARD
The Sd Card can save the data of test
POWER SWITCH
Switch ON or OFF the instrument

5
DISPLAY
CHARACTERS
DESCRIPTIONS
TEMP AND °C
(HOTPLATE MODEL ONLY)
Displays temperature when the heating function is switched ON
STIR
Displays stirring state when the stirring function is switched ON
HOT
Displays hot warning if the heating plate temperature is above 50°C after switching OFF
the heating function.
PROBE
Displays when using external probe.
ERR
Display when error is occurring
SD
Displays when using SD cards
RATED VALUE/REAL VALUE
Displays value in case of heating or stirring function switching ON.
Note:
If both heating and stirring functions have been started at the same time, heating function always has higher
priority. If in this case speed is changed via the stirring knob, it displays stirring speed and reverses to
temperature in the duration of 5 seconds.
Temp Stir
℃
Remote Probe Hot
Err
Real value
Rated value
Temperature
( Hotplate model) Stir
Error
Hotplate waning
(Hotplate model)
External probe
( Hotplate model )
Remote control

6
5. TRIAL RUN
• Make sure the required operating voltage and power supply voltage match.
• Ensure the socket is properly grounded.
• Plug in the power cable and ensure the power is on and begin initialising.
• Add the medium into the vessel with an appropriate stirring bar.
• Place vessel on the work plate.
• Set the target stirring speed and begin.
• Observe the stirring bar and LCD display (digital model).
• Set the target temperature and start heating.
• Observe the real temperature on LCD display (digital model).
• Stop the heating and stirring functions.
• If these operations above are normal, the device is ready to operate. If these operations are not normal, the
device may be damaged during transportation, please contact supplier for technical support.
Warning!
Do not transfer the vessel when the instrument is powered on.
6. USING THE EXTERNAL TEMPERATURE SENSOR
The external temperature sensor PT1000 is a standard accessory. If the sensor is plugged in “Probe” will be displayed on
the LCD digital display to indicate the sensor is operating. The setting value of the external temperature sensor and actual
temperature are displayed.
The safety circuit controls the hotplate temperature. Comparing with the temperature control of the hotplate, the external
temperature sensor can control the medium’s temperature more precisely. The heating function will be stopped
automatically under abnormal conditions. Please operate following the instructions below:
• Switch OFF the instrument
• Plug in the external PT1000
• Switch ON the instrument and run heating function
• Set the temperature via rotating the temperature control knob slowly to the target value.
If the heating function does not work, please contact the supplier for support.

7
7. RESIDUAL HEAT WARNING ‘HOT’
In order to prevent the risk of burns from a hotplate, digital hotplate has a residual heat warning function. When the heating
function is switched off and the heating plate temperature is still above 50°C, “Hot” will flash to warn that there is a hazard of
burns from the hotplate. When the hotplate temperature drops to below 50°C, the unit will automatically switch off. If users
want to turn off the LCD immediately, just pull out the plug directly. When the plug is pulled out, the residual heat warning
function will not work.
8. LOGGING FUNCTION
The data can be saved by SD card. The process is as follows:
• With the power switch turned off, insert the SD card into the slot on the left side of the panel.
• After inserting it, turn on the power switch and operate the heating as described above. Set temperature at 1
minute intervals
• The current temperature will be logged automatically.
• The logged data can be checked in Excel format by connecting the SD card to a personal computer
SD Card
9. FAULTS
• Instruments can’t be power ON
o Check whether the power line is unplugged.
o Check whether the fuse is broken or loose.
• Fault in power on self-test
o Switch OFF the unit, then switch ON and reset the instruments to factory default setting.
• Stir speed cannot reach set point
o Excessive medium viscosity may cause abnormal speed reduction of the motor.
• Unit cannot be powered off when switched off
o Check if the residual heat warning function is still ON and hotplate temperature is above 50℃(the LCD
still work and “Hot” flash).
If these faults are not resolved, please contact supplier.

8
10. MAINTENANCE AND CLEANING
• Proper maintenance can keep instruments working properly and lengthen its lifetime.
• Do not spray cleanser into the instrument when cleaning.
• Unplug the power line when cleaning.
• Only use recommended cleansers:
DYES
Isopropyl alcohol
CONSTRUCTION MATERIALS
Water containing Isopropyl alcohol
COSMETICS
Water containing Isopropyl alcohol
FOODSTUFFS
Water containing tenside
FUELS
Water containing tenside
• Before using other method for cleaning or decontamination, the user must ascertain with the
manufacturer/supplier that this method will not harm the instrument.
• Wear proper protective equipment during cleaning of the instrument.
• Send in the case of service the instrument back in the packaging carton. Storage packing is not sufficient for
despatch. Please use additional suitable packaging for transportation.
• The enamel makes the hotplate easier to care for and more resistant to acids and bases. However the hotplate is
also more susceptible to extreme fluctuations in temperature and the force of impact. This can result in cracks
forming or the coating flaking off.
Note:
• Electronic device cannot clean with cleanser.
• If you require maintenance service, must be cleaned the instrument in advance to avoid pollution of
hazardous substances, and to send back into original packing.
• If the instrument will not use for a long time, please switch off and place in a dry, clean, room
temperature and stable location.

9
11. ASSOCIATED STANDARDS AND REGULATIONS
Construction in accordance with the following safety standards:
EN 61010-1
UL 3101-1
CAN/CSA C22.2(1010-1)
EN 61010-2-10
Construction in accordance with the following EMC standards:
EN 61326-1
Associated EU guidelines:
EMC-guidelines: 89/336/EWG
Instrument guidelines: 73/023/EWG
Changes or modifications not approved by the party responsible for compliance could void the user's authority to operate
the equipment. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at your expense.

10
12. SPECIFICATIONS
ITEMS
SPECIFICATIONS
VOLTAGE [VAC]
200-240
FREQUENCY [HZ]
50/60
POWER [W]
*550 50
STIRRING POINT POSITION QUANTITY
1
MAX. STIRRING QUANTITY (H2O) [L]
30
MAX. MAGNETIC BAR [Lר, MM]
80x10
MOTOR TYPE
DC brushless motor
MAX. POWER INPUT OF MOTOR [W]
18
MAX. POWER OUTPUT OF MOTOR [W]
10
SPEED RANGE [RPM]
100-1500, increment: 10
ROTARY SPEED DISPLAY
LCD
PLATE MATERIAL
Stainless steel
DIMENSIONS OF WORKPLATE (W X D)
254 x 254mm
*HEATING POWER [W]
500
*TEMPERATURE RANGE [°C]
RT-340, increment: 1
*TEMPERATURE DISPLAY [°C]
LCD
*TEMPERATURE DISPLAY ACCURACY [°C]
±0.1
*THE SAFE TEMPERATURE OF HOTPLATE [
℃
]
360
*TEMPERATURE SENSOR IN MEDIUM
PT1000
*CONTROL ACCURACY OF HEATING TEMP WITH
TEMP SENSOR [
℃
]
±0.2
*RESIDUAL HEAT WARNING
50℃
DIMENSIONS (MM)
280×160×85
WEIGHT [KG]
2.8
PERMITTED AMBIENT TEMPERATURE [°C]
5-40
PERMITTED RELATIVE HUMIDITY
80%
PROTECTION CLASS ACC. TO DIN 60529
IP21

11
13. ORDERING INFORMATION
PRODUCT CODE
DESCRIPTIONS
400.100.130
Stirrer Magnetic Hotplate up to 30L
ACCESSORIES
400.100.215
Sensor Temperature PT1000 100mm
400.100.218
Sensor Temperature PT1000 230mm
400.100.219
Sensor Temperature PT1000 230mm
400.100.225
Clamp Support for PT1000 Sensor

12
Page Left Intentionally Blank\

13
Page Left Intentionally Blank

14
Page Left Intentionally Blank

15
Table of contents
Other LabCo Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Sartorius
Sartorius Centrisart G-16 operating manual

Fritsch
Fritsch ANALYSETTE 28 IMAGESIZER operating manual

Molecular Devices
Molecular Devices SpectraMax M3 user guide

MGI
MGI Complete Genomics DNBSEQ-G800RS user manual

Metrohm
Metrohm Titrando 851 manual

Steris
Steris AMSCO 600 Medium 639V-3 Operator's manual