Labino Apollo 1.0 User manual

Light Meter
Apollo1.0 and Apollo 2.0
OPERATING INSTRUCTIONS

Rev. 02 May 2018
2
CONTENTS
1. INTRODUCTION 3
2. OPERATION 4
2.1 Turning on Apollo 4
2.2 The main menu 6
2.3 Switching off Apollo 6
2.4 Average reading function 7
2.5 Change of measuring units 7
2.6 The peak function 8
2.7 The hold function 8
3. PAIRING SENSOR UNITS 9
4. TECHNICAL INFORMATION 10
4.1 Batteries 10
4.2 Operating conditions 10
4.3 Relative spectral response 10
5. CALIBRATION 10

3
1. INTRODUCTION
Apollo 1.0 and Apollo 2.0 are instruments for accurate measurements of
UV-A irradiation and visible illuminance. Extra engineering eort is taken to
make an accurate measurement of visible light emission from a UV-A lamp
by incorporating a superior bandpass lter containing only non-uorescent
lters. e instrument provides fast measurements as it oers autoranging and
concurrent measuring of visible light and UV-A irradiation. It is ergonomic and
easy to use due to its lightweight chassis, a wireless sensor unit and compact
size. Apollo 1.0 and Apollo 2.0 are traceable to NIST (USA’s National Institute of
Standards and Technology). In addition, Apollo 2.0 complies with the ISO 3059
standard.
Sensor measurements and transmission of the data is done via Bluetooth. e
wireless sensor enables the user to measure from a distance of up to ve meters
(16feet). is feature ensures that the sensor unit is stable, and no movement
occurs from connecting cables during measurement. Each sensor unit has both
the UV and white light sensors incorporated.
Apollo 1.0 and 2.0 come as a single kit or as a double kit. e single kit includes
one reader unit and one wireless sensor unit. A double kit includes one reader
unit and two wireless sensor units. As only the sensor needs to be sent for
calibration, the double kit is a convenient tool so that operations are never
disrupted. Please note that Aerospace companies that are audited by Nadcap
must send in both the reader and the sensor for calibration.

4
2. OPERATION
2.1 Turning on Apollo
a) Turn on the Hand Unit by pressing the “Power On” button (see Figure 1).
Figure 1. Hand Unit
Hold
Enter
Power On
Peak
Arrow Up
Arrow Down
LCD
Screen

5
b) Turn on the Sensor Unit by pressing the “Power On” button. e LED will
ash once to conrm (see Figure2). Note that the Sensor Unit will turn o
automatically aer 30 s if no connection to the Hand Unit was established or
if the Hand Unit was switched o. In these cases, the LED ashes twice.
Progress bar –
Bluetooth connection
Figure 3
Figure 2. Sensor Unit
LED Power On
UV-A
Sensor
VIS
Sensor
c) e Hand Unit and Sensor Unit will now connect. A progress bar indicates
the connection process (see Figure 3).
d) Once the connection is established, you will see the measuring screen that
shows the UV-A intensity in µW/cm² as measured by the UV-A sensor of the
Sensor Unit and the white light illuminance in lux as measured by the VIS
sensor.

6
2.2 The main menu
e main menu is reached when pressing “Enter” (Figure 1).
Here, you have the following options:
POWER OFF
SETTINGS
SENSOR UNITS
<<<
Use the “Arrow Up” and “Arrow Down” function to navigate through the
options. Press “Enter” to choose a submenu. Press <<< to return to the
measuring screen. e dierent options will be discussed in the following
sections.
2.3 Switching off Apollo
a) Turn o the Hand Unit by selecting POWER OFF in the main menu.
b) e Sensor Unit is automatically switched o 30 s aer the Hand Unit is
switched o. In this case, the LED ashes twice.
In addition, there is an automatic function which turns o the Hand Unit aer 5
min if no button has been pressed.

7
2.4 Average reading function
e average reading function allows integrating over the last 10 or 25 values that
are measured by the sensor.
a) Select SETTINGS in the main menu. e following submenu opens
UV-A: μW/cm2
VIS: lx
AVER AGE: OFF
<<<
b) Navigate to AVERAGE: OFF. Repeated application of the “Enter” button
allows you to switch between three values (10, 25, and OFF).
c) Return to the measuring screen by either pressing <<< twice or by pressing
the “Peak” button once. If you chose AVERAGE: 10 or AVERAGE: 25, you
will see the averaging symbol in the display (see Figure 4).
d) If no key has been pressed for 1 min, the Hand Unit will automatically return
to the main menu.
2.5 Change of measuring units
Measuring units for the UV intensity can be chosen between µW/cm² (default)
and W/m² and units for the visible illuminance can be chosen between lux
(default) and foot candles (fc).
a) Select SETTINGS in the main menu.
b) To change the settings for the UV-intensity, navigate to UV-A: µW/cm².
Repeated application of the “Enter” button allows you to alternate between
the measuring units.
c) To change the settings for the visible illuminance, navigate to VIS:lx.
Repeated application of the “Enter” button allows you to alternate between
the measuring units.
Average activated
Visible reading
UV-A reading
Figure 4

8
d) Return to the measuring screen by either pressing <<< twice or by pressing
the “Peak” button once.
e) If no key has been pressed within 1 min, the Hand Unit will automatically
return to the main menu.
2.6 The peak function
Aer the “Peak” button is pressed while you are on the measuring screen, the
Hand Unit shows the largest measured value since the button was pressed. A
small triangle can be seen in the display (see Figure 5). To deactivate the peak
function, press the “Peak” button again.
2.7 The hold function
Aer the “Hold” button is pressed while you are on the measuring screen, the
Hand Unit will show the latest measured value and does not change. A small
“H” can be seen in the display (see Figure 6). To deactivate the hold function,
press the “Hold” button again.
Peak function
activated
Figure 5
Hold function
activated
Figure 6

9
3. PAIRING SENSOR UNITS
e Sensor and the Hand Unit are paired by the factory. Only one Hand Unit
and one Sensor Unit can be paired at one time. In case of a double kit, the Sensor
Unit and the Hand Unit can be paired manually.
a) Select SENSOR UNITS in the main menu.
b) You see a submenu that looks like this
D588647BB843
NEW SENSOR
<<<
e rst line shows the unique ID of the Sensor Unit (12 characters, always
starting with a letter) that is currently paired with the Hand Unit. e ID is
printed on the label on the back side of the Sensor Unit.
c) Select NEW SENSOR. e Hand Unit it now disconnecting from the old
Sensor Unit and the display reads “Scanning …”.
d) Switch on the Sensor Unit you want to pair with.
e) A list of available Hand Units appears on the screen.
f) Select the desired Sensor Unit and press “Enter”.
g) Return to the measuring screen by either pressing <<< twice or by pressing
the “Peak” button once.

10
4. TECHNICAL INFORMATION
4.1 Batteries
e active measuring time of the Hand Unit is estimated to about 100 h. If the
display does not light up aer pressing “Power On”, new batteries (3 x AA 1,5V) are
needed. e batteries used for the Hand Unit can be replaced by the user.
e active measuring time for the Sensor Unit is estimated to approximately 600h.
If the battery is discharged and/or if the Sensor Unit is cold (below -30°C/-22°F)
the LED will ash 5 times and the Sensor Unit will turn o. e battery shall only
be replaced by Labino or a Labino authorized calibration center (see section 5).
Batteries are changed by the factory when the spectrometer is recalibrated.
4.2 Operating conditions
a) Operating temperature: 0˚C to +40˚C
b) Storage temperature: -30˚C to +60˚C
c) Relative humidity: 5% to 95%, noncondensing
4.3 Relative spectral response
e wavelength dependence of the relative spectral responsivities of the Apollo 1.0
and 2.0 UV-radiometers are shown in Figure 7. e relative spectral response of
Apollo2.0 fullls the requirements of ISO 3059 which are:
a) the relative spectral response shall not exceed 105% for any wavelength;
b) the peak relative spectral response shall occur between 355 nm and 375 nm;
c) the relative spectral response at a wavelength of 313 nm shall be less
than 10%;
d) the relative spectral response at a wavelength of 405 nm shall be less than 2%.
5. CALIBRATION
e recommended calibration interval is six months and is done by Labino or a
Labino authorized calibration center. ese centers can be found online at
www.labino.com.
Labino AB
Fågelsångsvägen 16
186 42 Vallentuna
Sweden
Tel.: +46 (0)883 90 70
labino@labino.com

11
Figure 7.
Relative spectral responsivities of the Apollo 1.0 and
the Apollo 2.0 UV-radiometers

DISTRIBUTOR:
Labino AB shall not be held liable for any errors or omissions
resulting from the test procedures that were used in validating
performance of any Labino AB product nor for unforeseen errors
or omissions in digital or printed material.
www.labino.com
Light Meter Apollo 1.0 and Apollo 2.0 Operating Instructions, revision 02, May 2018, Vallentuna
Distributed by:
This manual suits for next models
1
Table of contents
Other Labino Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

SENSIT Technologies
SENSIT Technologies GAS-TRAC LZ-30 instruction manual

Danaher
Danaher Veeder-Root 7990 Series quick start guide

Contrec
Contrec 515 Operation manual

Sper scientific
Sper scientific 850081DOK instruction manual

Sifam Tinsley
Sifam Tinsley N20ZPLUS user manual

Spectroline
Spectroline AccuPRO XP-2000 quick guide