Labnet I-4001-HCS User manual

Labnet Microtube Shaking Incubator
Instruction Manual
Catalog Numbers:
I-4001-HCS
I-4002-HCS
This manual is available in additional languages at www.labnetlink.com.

Table of Contents
1.0 Introduction ...................................................................................1
2.0 Symbols and Conventions.......................................................................1
3.0 Safety Information .............................................................................1
4.0 Product ID Labels...............................................................................3
5.0 Package Contents ..............................................................................3
6.0 Specifications ..................................................................................4
7.0 Equipment Overview ...........................................................................4
8.0 Operation Guide ...............................................................................5
8.1 Before turning on power ....................................................................5
8.2 Set up of temperature, vibration time, and vibration speed ...................................5
8.3 Pulse vibration..............................................................................6
8.4 Correct temperature discrepancy ............................................................6
8.5 Buzzer set up ...............................................................................7
9.0 Troubleshooting and Repair .....................................................................8
10.0 Spare Parts and Accessories .....................................................................9
11.0 Limited Warranty............................................................................. 10
12.0 Equipment Disposal .......................................................................... 10

Instruction Manual | 1
1.0 Introduction
The Labnet Microtube Shaking Incubator has a microprocessor controlled cooling and heating
capabilities. The incubator can be equipped with various block sizes to accommodate several tube sizes
and microplates. It can perform a wide range of applications including, sample storage, storage and
reaction of various kinds of enzyme, DNA amplification, pre-denaturation of electrophoresis, and serum
solidification. The Labnet Microtube Shaking Incubator includes the following features:
A vacuum fluorescent display (VFD) display with high-luminosity
Temperature setting value and practical value displayed simultaneously
Display setting time and practical time simultaneously
Aluminum block protecting the sample from contamination
Aluminum block that is easy to replace, clean, sterilize and suitable for different type of tubes
Internal over-temperature protection improving reliability
Warning function after running time ends
Temperature calibration
2.0 Symbols and Conventions
CAUTION: This symbol refers you to important operating and maintenance (servicing)
instructions within the product Instruction Manual. Failure to heed this information may
present a risk of damage or injury to persons or equipment.
WARNING: To avoid accidental bodily harming or burning be very careful touching the metal
parts of the unit. It can be very hot after it is used at high temperatures. Allow the metal
parts to cool down before handling.
3.0 Safety Information
During the operation, maintenance, or repair of the Labnet Microtube Shaking Incubator, the following
safety measures should be taken. Otherwise, the safeguards provided by the Labnet Microtube Shaking
Incubator are likely to be damaged, the rated safety level to be reduced, and the rated operation
conditions to be affected. Labnet shall not be in any way responsible for the consequences resulting
from operator’s not observing the following requirements.
Grounding
AC power’s grounding should be reliable to safeguard against an electric shock. The 3-pin plug supplied
with the Labnet Microtube Shaking Incubator’s power cable is a safety device that should be matched
with a suitable grounded socket. Never allow the third ground pin to be floating. If the 3-pin plug cannot
be inserted, it is recommended to ask an electrician to install an appropriate power socket.
Keeping Away from Electric Circuits
The operator should not open the Labnet Microtube Shaking Incubator without first consulting Service
Department. Changing components or adjusting certain parameters inside the device must be performed by
certificated professional maintenance personnel only. Do not change elements while the power is still on.
AC Power Considerations
Before turning on the power, always make sure that the mains voltage is within the range of required
power supply (±10% difference is allowed) and the rating current of the power socket meets the required
specification as shown in below contents.

2| LABNET MICROTUBE SHAKING INCUBATOR
AC Power Line Considerations
As an accessory of the Labnet Microtube Shaking Incubator, the AC power supply should be the default
device. Once it is damaged, the AC power line may not be repaired, but must be replaced by a new one.
The power supply should be free of heavy objects during the Labnet Microtube Shaking Incubator’s
operation. Keep the power supply away from high traffic areas.
Connect the A.C. Power Line
While connecting or disconnecting the power line, user should insert the plug firmly to ensure good
contact between the plug and socket. Pull the plug, but not the cable, when the plug needs to be
disconnected from the mains.
Design Environments
The Labnet Microtube Shaking Incubator should be placed in a low-humidity, dustfree, well-ventilated
room without caustic gas or powerful magnetic interference. The Labnet Microtube Shaking Incubator
should not be operated in close proximity to water sources, such as pools and water pipes. Never cover
or obstruct the openings of the Labnet Microtube Shaking Incubator, which are designed for ventilation
and to prevent the device’s interior from becoming too hot. When a single device is running, the shortest
distance between its openings and the nearest object should be at least 20 inches (50 cm); when two
devices or more are running simultaneously, the shortest distance is 40 inches (100 cm) among these
machines. Do not place the device on a soft surface. Doing so will result in adverse ventilation near
the device’s bottom openings. Operating conditions that cause a high temperature environment will
result in degraded performance or failure of Labnet Microtube Shaking Incubator. Additionally, the
device should be protected against any kind of heat sources such as sunlight, ovens, or central heating
equipment. If the Labnet Microtube Shaking Incubator will sit idle for extended periods of time, it is
recommended to disconnect the power line from the mains and cover the device with a piece of soft
cloth or plastic to prevent against dust.
Once one of the following events occurs, the operator is advised to disconnect the power
cable from the mains, and contact the distributor or ask a certified maintenance engineer
for assistance.
Liquid enters into the device
The device malfunctions, giving off an abnormal sound or odor
The device falls onto the floor or the housing is damaged
Significant changes in the device’s performance

Instruction Manual | 3
4.0 Product ID Labels
For model I-4001-HCS
LABNET
Model˖I-4001-HCS
Power˖AC100-120V50-60Hz 150W
Name˖The MicrotubeShakingIncubator
Fuse˖125V 2.5A (5×20mm) SN
˖*******-*** Date˖****.**.**
LABNET
Model˖I-4002-HCS
Power˖220-240V~50/60Hz 150W
Name˖The Microtube ShakingIncubator
Fuse˖250V1.5A (5×20mm)
SN˖BYQ6008E-*** Date˖****.**.**
For model I-4002-HCS
LABNET
Model˖I-4001-HCS
Power˖AC100-120V50-60Hz 150W
Name˖The MicrotubeShakingIncubator
Fuse˖125V 2.5A (5×20mm) SN
˖*******-*** Date˖****.**.**
LABNET
Model˖I-4002-HCS
Power˖220-240V~50/60Hz 150W
Name˖The Microtube ShakingIncubator
Fuse˖250V1.5A (5×20mm)
SN˖BYQ6008E-*** Date˖****.**.**
5.0 Package Contents
Description Quantity
Labnet Microtube Shaking Incubator 1 each
Instruction Manual 1 each
Inner-Hexagon Round Head Screws 4 each
Spring Washer 4 each
T-Wrench 1 each
Rubber Gasket 1 each

4| LABNET MICROTUBE SHAKING INCUBATOR
6.0 Specications
Operating conditions 10°C to 30°C, relative humidity ≤ 70%
Transportation and storage
conditions
-20°C to 55°C, relative humidity ≤ 80%
Electrical I-4001-HCS: AC100-120V, 50/60Hz, 150W
I-4002-HCS: AC220-240V, 50/60Hz, 150W
Dimension (L × W × H) 12.9 x 6.5 x 9.8 in. (328 × 166 × 249 mm)
Weight 18.75 lbs. (8.5 kg)
Temperature control range 0°C to 105°C
Temperature range Room temperature -14°C to 100°C
Timing range 1 min. to 99h. 59 min.
Cooling time From room temperature to room temperature - 10°C: ≤ 8min
From 100°C to room temperature +10°C: ≤ 15min
Heating time ≤ 12 min (from 20°C to 100°C)
Temperature accuracy ≤ ±0.5°C
Temperature uncertainty ≤ ±0.5°C
Heating rate About 6°C/min. (from 20°C to 100°C)
Block temperature uniformity ≤ ±0.5°C
Mixing rate 300 to 1,500 rpm
Amplitude 3 mm
7.0 Equipment Overview
Block
Case (main body)
Display panel
and keyboard
Sticker
Power connector, fuse
and power switch

Instruction Manual | 5
Temperature increasing
indicator
Constant temperature
indicator
Temperature decreasing
indicator
Timer indicator
Vibration speed indicator
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
Button explanation
Temp Temp
Temperature set up buttons
Press or to adjust the value ant set the requested temperature.
Keep pressing or may help to speed up the setting time.
Time Time
Vibration time set up buttons
Press or to adjust the value and set the requested vibration time.
Keep pressing or may help to speed up the setting time.
Speed Speed Vibration speed set up buttons
Press or to adjust the value and set the requested vibration speed.
Short
Prog
Instant spot vibration button
Press it to start spot vibration. The running speed is the highest vibration speed.
It is also a function button for temperature calibration.
Start
Stop
Start/Stop button
Press this button to start or stop the mixing of the device (valid for only mixing).
8.0 Operation Guide
8.1 Before turning on power
Check and confirm the following items first
Unit is on stable and level surface
The voltage is comply with the specification (see Section 5)
The power plug is tightly plugged into the outlet
The ground and module connection are reliable
8.2 Set up of temperature, vibration time, and vibration speed
Approximately 5 seconds after the power is turned on, the
monitor will display a temperature of 28.5°C, which is the instant
temperature of the module.
The display of TEMP 100.0 is the set up temperature of last run, the
display of TIME 80:00 is the set up vibration time of last run, and the
display of SPEED 1000 is the vibration speed of last run.
At same time, the device will automatically heat or cool, to achieve
the setting temp. point.
If the display is abnormal when the power is turned on, please shut down immediately and contact
the supplier.
Set temperature
Set vibration time
Set vibration speed
Actual temperature
Actual vibration speed
Actual vibration time

6| LABNET MICROTUBE SHAKING INCUBATOR
The digits of the TEMP display may be reduced or increased by
pressing Temp or Temp respectively. The digits of the TIME
display may be reduced or increased by pressing Time or Time
respectively.
Press the above mentioned buttons for more than 3 seconds to set
up the exact value.
If the temperature should be set at 40°C and the time should be
set up at 60:00, then press the Temp and not release to see the
number reducing. When the number shows 40.0, release the
button, and it will be automatically saved in 3 seconds. Next
press the Time and not release to see the number reducing.
When the number shows 60:00, release the button, and it will be
automatically saved in 3 seconds.
Press Speed or Speed to set up for the speed. The number will
increase or reduce at a rate of 50 rpm, release the button when the
number reaches exact value, and it will be automatically saved in 3
seconds. When set up is done, press Start/Stop to start mixing.
After the setting value of temperature is saved successfully, the device
will automatically heat or cool, to achieve the setting temp. point.
8.3 Pulse vibration
Press Short/Prog to start instant vibration or short term vibration. Press the button to start it and release
the button to stop it. The running speed of pulse vibration is the highest vibration speed.
8.4 Correct temperature discrepancy
The temperature of the instrument has been calibrated before shipment. However, due to various
conditions, there might be discrepancy between the actual temperature and displayed temperature.
On such case, you may use the calibration button to correct the discrepancy.
To ensure the accuracy of the temperature, please wait for at least 30 minutes after the instrument
is set at a constant temperature, and then perform the calibration.
Please use certified standard class two mercury thermometers to calibrate this instrument.
Calibration point: Center hole of the module. Please fill the hole with paraffin oil and immerge
the thermometer bulb in it.
The method of the temperature calibration of this instrument is two-point linear calibration.
The calibration point may be set freely, and by setting the second calibration point the same as the
first calibration point, you can adjust the calibration point to one temperature point. For two-point
calibration, if set the two points at 40°C and 100°C, then other temperature points will be automatically
adjusted per the linear relationship of the two calibration points.
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
Press both Temp and
Temp at the same time,
and the display will show
the set up value of last
time (for example 20°C)
and the cursor of the first
calibration point, C_1.
Press the temperature
set up button Temp or
Temp to set up the first
calibration temperature
such as 40°C.
Perform the same
operation again to set up
the second calibration
temperature such as
100°C.
NOTE: When setting up the temperature points, an order of low temperature point first and high
temperature after is recommended. If on the case high temperature point is set first and low
temperature is set after, then when performing the following actual calibration, please follow the order
of low temperature first and high temperature after.
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
Temp
Time

Instruction Manual | 7
Press Short/Prog for once
and change ON to OFF.
Then press both Speed
and Speed at the same
time to save the setting,
and keyboard beep will
be disabled. If you want
to exit during set up, you
may press Start/Stop
and the setting will be
ineffective.
TEMP
TIME
TIME
TEMP
TIME
TEMP
TIME
TEMP
Press both Speed and
Speed at the same time,
and the display will show
BEEP.
Press Speed and
Speed at the same time
to save the setting, and
buzz will be disabled upon
failure’s occurrence or
ending. The display will
then show the keyboard
beep status. If the display
will show ON K_S. Press
Short/Prog once and
change ON to OFF.
TEMP
TIME
TIME
TEMP
TIME
TEMP
TIME
TEMP
TEMP
TIME
TIME
TEMP
TIME
TEMP
TIME
TEMP
TEMP
TIME
TIME
TEMP
TIME
TEMP
TIME
TEMP
Press both Temp and
Temp at the same time
to start the calibration. The
display will show 40.0 and
the temperature of the
instrument will automatically
increase to 40°C. Once the
temperature is kept constant,
timing will start.
30 minutes later, press
Short/Prog for one time,
and the display will show the
blink of TEMP. At this time,
the display will show the
actual temperature reading
of the thermometer. For
example, if the thermometer
reads 40.5°C, you can adjust
the temperature display
to 40.5°C by pressing the
temperature set up button.
Next press Short/Prog for
one time, and the display
will show 100.0°C. Following
the same instruction, when
the temperature of the
instrument increases to 100°C,
wait for 30 minutes, input
the calibrated value and press
Short/Prog to save it.
When the calibration
of 40°C is done, press
Start/Stop to exit. At this
time the calibration of
40°C is ineffective. When
the calibration is done
for once at both 40°C
and 100°C , check for
the discrepancy between
the two calibration
points and the actual
temperature points.
It should be within
0.5°C. If it’s more than
0.5°C, please follow the
above instruction and
recalibrate until the
request is met.
TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP TEMP
TIME
TIME
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TIME
SPEED
TEMP
TEMP
TIME
TIME
TEMP
TIME
TEMP
TIME
TEMP
8.5 Buzzer set up
Upon failure’s occurrence or ending, the instrument will buzz. You may disable it if you do not need
this function. However, the default setting is to have it on.
This instrument has keyboard beep. Once the button is pressed, it will beep once. You may disable it if
you do not need this function. However, the default setting is to have it on.

8| LABNET MICROTUBE SHAKING INCUBATOR
9.0 Troubleshooting and Repair
Problem Probable Cause Remedy
No display shown in
the monitor when the
power is on.
Power is not connected. Check the power and connect it.
Fuse has burned out. Change the fuse:
BYQ15412A000020: 125V, 2.5A Φ5x20
BYQ15411A500170: 250V, 1.5A Φ5x20
Switch is broken. Change the switch.
Others. Contact the supplier of manufacturer.
During run, the housing
shakes abnormally and
makes abnormal noise.
Use improper block Change the block. (Contact supplier or
manufacturer to make sure whether the block is
suitable.)
Unit is placed on
uneven surface
Place unit on even, stable surface
Round feet badly worn Change round feet.
Others Contact supplier or manufacturer.
Bad temperature
uniformity of block.
Improper installation of
block.
Re-install the block.
Bad cooling
components.
Contact supplier or manufacturer.
The system no longer
operates after changing
block.
Bad grounding of
machine.
Check power cable to insure reliable grounding;
Do remember to turn off the power when
changing block.
Serious discrepancy
between the actual
temperature and
displayed temperature.
Broken sensor or bad
contact.
Contact the supplier or manufacturer.
The cooling of the
module drastically
slows down or the
temperature cannot
reach below room
temperature.
Thermoelectric cooling
module is broken.
Contact the supplier or manufacturer.
Fan is broken or does
not work.
The module is neither
able to heat nor cool.
Temperature sensor is
broken.
Contact the supplier or manufacturer.
Thermoelectric cooling
module is broken.
Vibration stops. Drive is broken. Contact the supplier or manufacturer.
Motor is broken.
Fuse has burned out. Change the fuse.
BYQ15412A000020: 125V, 2.5A Φ5x20
BYQ15411A500170: 250V, 1.5A Φ5x20
User shall not open the cover of this instrument when under warranty. If housing does need to be
opened to perform certain repair per the above instruction, please contact the supplier or manufacturer.
Should you have a question about the operation of the Labnet Microtube Shaking Incubator or if service
is required, contact Customer Service. Do not send in a unit for service without first calling to obtain a
repair authorization number. Should the unit require return for service, it should be properly packed to
avoid damage. Any damage resulting from improper packaging shall be the responsibility of the user.

Instruction Manual | 9
10.0 Spare Parts and Accessories
Spare Parts
Cat. No. Description Quantity
BYQ121L00000020 L socket head wrench 1
BYQ900800000070 Cushioning balls for foot 1
BYQ15411A500170 Fuse used for 230V model (250V, 1.5A Φ5x20 mm) 1
BYQ15412A000020 Fuse used for 120V model (125V, 2.5A Φ5x20 mm) 1
Accessories
Cat. No. Description Quantity
I-4000-A Block A, 40 x 1.5 mL 1
I-4000-B Block B, 54 x 0.5 mL 1
I-4000-C Block C, 96 x 0.2 mL 1
I-4000-D Block D, 24 x 15 mm 1
I-4000-E Block E, water bath (ID: 115 x 73 x 38 mm)
(not suggested to use when mixing function is on)
1
I-4000-G Block G, 26 x 0.5 mL and 24 x 1.5 mL 1
I-4000-H Block H, 40 x 2.0 mL 1
I-4000-J Block J, 96-well ELISA plate 1

10 | LABNET MICROTUBE SHAKING INCUBATOR
11.0 Limited Warranty
Corning Incorporated (Corning) warrants that this product will be free from defects in material
and workmanship for a period of one (1) year from date of purchase. CORNING DISCLAIMS ALL
OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE. Corning’s sole obligation shall be
to repair or replace, at its option, any product or part thereof that proves defective in material or
workmanship within the warranty period, provided the purchaser notifies Corning of any such defect.
Corning is not liable for any incidental or consequential damages, commercial loss or any other damages
from the use of this product.
This warranty is valid only if the product is used for its intended purpose and within the guidelines
specified in the supplied instruction manual. This warranty does not cover damage caused by accident,
neglect, misuse, improper service, natural forces or other causes not arising from defects in original
material or workmanship. This warranty does not cover motor brushes, fuses, light bulbs, batteries or
damage to paint or finish. Claims for transit damage should be filed with the transportation carrier.
In the event this product fails within the specified period of time because of a defect in material or
workmanship, contact Corning Customer Service at: USA/Canada 1.800.492.1110, outside the U.S.
+1.978.442.2200, visit www.corning.com/lifesciences, or contact your local support office.
Corning’s Customer Service team will help arrange local service where available or coordinate a return
authorization number and shipping instructions. Products received without proper authorization will
be returned. All items returned for service should be sent postage prepaid in the original packaging or
other suitable carton, padded to avoid damage. Corning will not be responsible for damage incurred by
improper packaging. Corning may elect for onsite service for larger equipment.
Some states do not allow limitation on the length of implied warranties or the exclusion or limitation
of incidental or consequential damages. This warranty gives you specific legal rights. You may have other
rights which vary from state to state.
No individual may accept for, or on behalf of Corning, any other obligation of liability, or extend the
period of this warranty.
For your reference, make a note of the serial and model number, date of purchase, and supplier here.
Serial No. __________________________________________ Date Purchased _______________________________________
Model No. _________________________________________ Supplier _____________________________________________
12.0 Equipment Disposal
According to Directive 2012/19/EU of the European Parliament and of the Council of 4 July 2012
on waste electrical and electronic equipment (WEEE), this product is marked with the crossed-out
wheeled bin and must not be disposed of with domestic waste.
Consequently, the buyer shall follow the instructions for reuse and recycling of waste electronic and
electrical equipment (WEEE) provided with the products and available at www.corning.com/weee.

To request certificates, please contact us at www.labnetlink.com.
Warranty/Disclaimer: Unless otherwise specified, all products are for research use only. Not intended
for use in diagnostic or therapeutic procedures. Corning Life Sciences makes no claims regarding the
performance of these products for clinical or diagnostic applications.
© 2020 Corning Incorporated. All rights reserved. 11/20 CLSLN-AN-1001DOC REV1
For a listing of trademarks, visit www.corning.com/clstrademarks. All other trademarks are the property of their respective owners.
For additional product or technical information, visit www.corning.com/lifesciences or call 800.492.1110.
Outside the United States, call +1.978.442.2200 or contact your local Corning sales ofce.
Corning Incorporated
Life Sciences
836 North St.
Building 300, Suite 3401
Tewksbury, MA 01876
t 800.492.1110
t 978.442.2200
f 978.442.2476
www.corning.com/lifesciences
ASIA/PACIFIC
Australia/New Zealand
t 61 427286832
Chinese Mainland
t 86 21 3338 4338
f 86 21 3338 4300
India
t 91 124 4604000
f 91 124 4604099
Japan
t 81 3-3586 1996
f 81 3-3586 1291
Korea
t 82 2-796-9500
f 82 2-796-9300
Singapore
t 65 6572-9740
f 65 6735-2913
Taiwan
t 886 2-2716-0338
f 886 2-2516-7500
EUROPE
LATIN AMERICA
Brazil
t 55 (11) 3089-7400
Mexico
t (52-81) 8158-8400
www.labnetlink.com
Other manuals for I-4001-HCS
1
This manual suits for next models
1
Table of contents
Other Labnet Accessories manuals

Labnet
Labnet I5110A User manual

Labnet
Labnet VorTemp 56 User manual

Labnet
Labnet I-5222-DS User manual

Labnet
Labnet 311D Quick guide

Labnet
Labnet VorTemp 56 User manual

Labnet
Labnet I5110 User manual

Labnet
Labnet VorTemp 1550 User manual

Labnet
Labnet 222DS User manual

Labnet
Labnet 211DS User manual

Labnet
Labnet Accublock D1301-230V User manual
Popular Accessories manuals by other brands

Atim Cloud Wireless
Atim Cloud Wireless ACW/SF8-THM-I user guide

A&D
A&D FS-D Setup instructions

Atom Longboards
Atom Longboards Atom Electric B Series owner's manual

FedEx
FedEx SenseAware PT300D quick start guide

MobilePal
MobilePal Qi user manual

ROSE DISPLAYS
ROSE DISPLAYS ANOSKYBOX-TOP LOADING 4 WAY manual