Lafayette Instrument 5030P1 User manual

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com
Model 5030P1MAN32269
Copyright © 2020 by Lafayette Instrument Company, Inc. All Rights Reserved- Rel. 6.11.20
Lafayette Hydraulic
Pinch Gauge
User’s Manual
1

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
2
Introduction
The Lafayette Hydraulic Pinch Gauge was designed as a diagnostic tool to measure the strength of
different pinch applications. The pinch gauge has a real time force needle and a peak hold needle.
The real time force needle displays the current force being applied to the gauge. The peak hold needle
displays the maximum force that was applied during the trial.
Operation
The therapist holds the rounded end of the gauge supporting the weight of the instrument with the
pressure pad positioned toward the patient. The gauge must be “zeroed” before each trial. To zero the
instrument turn the center knob counter-clockwise until the peak hold needle returns to zero. With the
therapist still supporting the weight of the instrument, the patient applies pressure to the pinch block
and pressure pad with the thumb and nger being tested.

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 3
Applications
The patient should be seated with the test arm at the side of the elbow exed at 90 degrees with the
forearm and wrist in the neutral position.
The various pinch styles include
1. Tip Pinch: Test involving the tip of the thumb and the tip of each or any of the ngers making
certain that the other ngers are not involved.
2. Key Pinch: Test involving the pad of the thumb and the PIP joint of the index nger.
3. Palmar Pinch: Test involving the pad of the thumb and the pads of the index and middle ngers.
Note: Lafayette Instrument Company recommends placing the patient’s thumb on the pressure
pad with the index and pointer ngers on the back when doing the Palmar Pinch test.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
4
7
6
9
15
4
3
2
8

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 5
Key
1. Peak Hold Needle
2. Crystal
3. Pound Scale
4. Kilogram Scale
5. Peak-Hold Knob
6. Pinch Block
7. Pressure Pad
8. Zero Adjustment Screw
9. Real Time Force Needle
Service Tips
The Lafayette Hydraulic Pinch Gauge is designed to provide reliable service, with minimal
maintenance, over a period of many years. To make sure the instrument is reading accurately, it is
recommended to make a few checks, occasionally, as listed below.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
6
Zero-Adjustment Screw
Calibration
The instrument is calibrated at the factory using a
custom calibration device. To ensure proper operation,
the instrument should be returned to the manufacturer
for calibration on a regular, preferably, annual basis.
If the instrument has been dropped or there is some
particular reason to suspect the calibration is in error,
the instrument should be serviced by the manufacturer
immediately.
Zero Reading
The gauge needle may be above or below the zero
reading upon receipt of the instrument. The gauge on
this instrument is sensitive to pressure changes due
to changes in altitude and will most likely require an

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 7
adjustment to re-zero the instrument. Follow these steps to return the gauge needle to zero:
1. Unscrew (counter clockwise) and remove the lens from the gauge (the red peak-hold needle is
attached to the lens)
2. Remove the handle (to prevent applying pressure during this process)
3. Place the instrument on a sturdy surface
4. Using a small slotted screwdriver, rotate the slotted screw counter clockwise in the gauge dial
face until the needle is below zero (if it is already zero, turn the cam until it is slightly farther below
zero)
5. Turn the screw in the reverse direction until the gauge needle is almost touching the zero line
6. Reattach the gauge lens and the handle
7. Pick up and squeeze to apply pressure to the instrument a few times
8. Place the instrument back on a sturdy surface and return the red peak-hold needle to zero by
turning the center knob
9. Check to see that the red peak-hold needle is at zero
10. If the red peak-hold needle is not on zero, repeat the steps until the red peak-hold needle is at zero

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
8
Red Peak-Hold Needle
Check for excessive friction in the red peak-hold
assembly by turning the peak-hold knob in the lens
counterclockwise. If the peak-hold needle deects
the gauge needle, return the instrument for service.
Likewise, if the red peak-hold needle moves too
freely, (i.e. spins without pressure applied) return the
instrument for service.
Peak-Hold Needle

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 9
Wrist Strap
The Lafayette Hydraulic Pinch Gauge is equipped with a wrist strap. Simply place your hand through
the strap before pinching the pressure pad to protect the instrument from damage if dropped.
Cleaning
Cleaning with mild antibacterial soap and water is recommended. Do not soak or saturate the device
with water. Handle can also be cleaned with an isopropyl alcohol wipe. Do not use alcohol on the
plastic gauge faceplate or cover.
Disclaimer: The cleaning instructions for Lafayette Instrument products are a recommendation of
compatible cleaning materials only. Product end users are responsible for instituting an appropriate
cleaning regimen utilizing best practices and techniques. Lafayette Instrument assumes no
responsibility for the cleanliness or sanitation of the products after initial use nor makes any claim that
the use of the recommended cleaning materials mitigates all risk of potential cross infection.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
10
Hydraulics
Any sign of hydraulic oil is an indication of damage to the instrument which requires service by the
manufacturer.
Hydraulic Fluid Exposure
Lafayette hand and pinch dynamometers are lled with a biodegradable, plant based hydraulic uid.
Under normal circumstances, the uid is contained within the device and does not come into contact
with the user or patient. If the device is damaged, the uid can leak out and human exposure to the
uid is a possibility. The uid is non-toxic and contains no chemicals known to cause cancer, birth
defects or other reproductive harm. Despite the relative safety of the uid, precautions should be
taken to limit exposure. If exposure occurs, the following guidelines from the Material Safety Data
Sheet (MSDS) of the uid should be followed.
ORAL: DO NOT INDUCE VOMITING. If conscious, give 2 glasses of water. Get immediate medical
attention.

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 11
EYE: Flush with water at least 15 minutes. Get medical attention if eye irritation develops or persists.
SKIN: Wash with soap and water. Remove contaminated clothing. Get medical attention if irritation
develops. Launder contaminated clothing before reuse.
INHALATION: Remove exposed person to fresh air if adverse effects are observed. If breathing is
labored, administer oxygen. If breathing has stopped, apply articial respiration. If irritation persists,
get medical attention.
Warranty
The Lafayette Hydraulic Pinch Gauge has a 1 year parts and labor warranty. Please contact our service
attempt to service or calibrate outside of Lafayette Instrument will void the warranty.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
12
Terms and Conditions
LIC Worldwide Headquarters
Toll-Free: (800) 428-7545 (USA only)
Phone: (765) 423-1505
Fax: (765) 423-4111
Email: [email protected]
Mailing Address
Lafayette Instrument Company
3700 Sagamore Pkwy N
Lafayette, IN 47904, USA
Lafayette Instrument Europe
Phone: +44 1509 817700
Fax: +44 1509 817701
Email: [email protected]
Phone, Fax, Email or Mail-in Orders
All orders need to be accompanied by a hard copy of your
purchase order. All orders must include the following information:
• Quantity
• PartNumber
• Description
• Yourpurchaseordernumberormethodofpre-payment
• Yourtaxstatus(includetax-exemptnumbers)
• Shippingaddressforthisorder
• Billingaddressfortheinvoicewe’llmailwhenthisorder
is shipped
• Signatureandtypednameofpersonauthorizedtoorder
these products
• Yourtelephonenumber
• Youremailaddress
• YourFAXnumber

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 13
Domestic Terms
There is a $50 minimum order. Open accounts can be extended
to most recognized businesses. Net amount due 30 days from
thedateofshipmentunlessotherwisespeciedbyus.Enclose
payment with the order; charge with VISA, MasterCard, American
Express, or pay COD. We must have a hard copy of your purchase
order by mail, E-mail or fax. Students, individuals and private
companies may call for a credit application.
International Payment Information
There is a $50 minimum order. Payment must be made in
advance by: draft drawn on a major US bank; wire transfers to
our account; charge with VISA, MasterCard, American Express,
orconrmedirrevocableletterofcredit.Proformainvoiceswill
be provided upon request.
Exports
If ordering instrumentation for use outside the USA, please
specify the country of ultimate destination, as well as the power
requirements (110V/60Hz or 220V/50Hz). Some model numbers
for220V/50Hzwillhavea“*C”sufx.
Quotations
Quotationsaresuppliedupon request. Writtenquotationswill
include the price of goods, cost of shipping and handling, if
requested,andestimateddeliverytimeframe.Quotationsare
good for 30 days, unless otherwise noted. Following that time,
prices are subject to change and will be re-quoted at your request.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
14
products must be in saleable condition, and credit is subject to
inspection of the merchandise.
Repairs
Instrumentation may not be returned without prior authorization
by Lafayette Instrument Company and a completed Return Form.
When you complete the Form, or call Lafayette Instrument, you
willreceiveaReturn#.YourReturn#numberwillbegoodfor
30 days. Address the shipment to:
Lafayette Instrument Company
3700 Sagamore Parkway North
Lafayette, IN 47904, USA.
ShipmentscannotbereceivedattheLICPOBox.Itemsshould
be packed well, insured for full value, and returned along with a
copy of the Return Form or the Return #. An estimate of repair
willbegivenpriortocompletionONLYifrequestedinanenclosed
cover letter. We must have a completed purchase order by mail
or fax, or repair work cannot commence for non-warranty repairs.
Cancellations
Orders for custom products, custom assemblies or instruments
builttocustomerspecicationswillbesubjecttoacancellation
penalty of 100%. Payment for up to 100% of the invoice value
of custom products may be required in advance. Cancellation
for a standard Lafayette Instrument manufactured product
once the product has been shipped will normally be assessed a
charge of 25% of the invoice value, plus shipping charges. Resell
items, like custom products, will be subject to a cancellation
penalty of 100%.
Exchanges and Refunds
Please see the cancellation penalty as described above. No
item may be returned without prior authorization of Lafayette
Instrument Company and a completed Return Form. A copy of
the Return Form or your assigned Return # (you will receive this
via email after submitting the form) must be included with the
returned goods. The merchandise should be packed well and
fully insured. Unopened merchandise may be returned prepaid
within thirty (30) days after receipt of the item and in the original
shipping carton. Collect shipments will not be accepted. Returned

Model 5030P1 User’s Manual
www.lafayetteinstrument.com . info@lafayetteinstrument.com 15
Damaged Goods
Damaged instrumentation should not be returned to Lafayette
Instrument prior to a thorough inspection. If a shipment arrives
damaged, note damage on delivery bill and have the driver sign
it to acknowledge the damage. Contact the delivery service, and
theywillle an insuranceclaim.Ifdamageisnot detected at
the time of delivery, contact the carrier/shipper and request an
inspection within 10 days of the original delivery. Please call the
Lafayette Instrument Customer Service Department for repair
or replacement of the damaged merchandise.
Limited Warranty
Lafayette Instrument Company warrants equipment
manufactured by the company to be free of defects in
material and workmanship for a period of one year from the
date of shipment, except as provided hereinafter. The original
manufacturer’swarrantywillbehonoredbyLafayetteInstrument
for items not manufactured by Lafayette Instrument Company,
i.e. resell items. This assumes normal usage under commonly
accepted operating parameters and excludes consumable
products.
Warranty period for repairs or used instrumentation purchased
from Lafayette Instrument is 90 days. Lafayette Instrument
Company agrees either to repair or replace, at its sole option
and free of part charges to the customer, instrumentation
which, under proper and normal conditions of use, proves to
be defective within the warranty period. Warranty for any parts
of such repaired or replaced instrumentation shall be covered
under the same limited warranty and shall have a warranty period
of 90 days from the date of shipment or the remainder of the
original warranty period whichever is greater. This warranty and
remedy are given expressly and in lieu of all other warranties,
expressedorimplied,ofmerchantabilityortnessforaparticular
purpose and constitutes the only warranty made by Lafayette
Instrument Company.

Lafayette Hydraulic Pinch Gauge
3700 Sagamore Pkwy N Lafayette, IN 47904 . Phone: (765) 423-1505
Lafayette Instrument Company neither assumes nor authorizes
any person to assume for it any other liability in connection
with the sale, installation, service or use of its instrumentation.
Lafayette Instrument Company shall have no liability whatsoever
for special, consequential, or punitive damages of any kind from
any cause arising out of the sale, installation, service or use of
its instrumentation.
All products manufactured by Lafayette Instrument Company are
testedandinspectedpriortoshipment.Uponpromptnotication
by the Customer, Lafayette Instrument Company will correct
any defect in warranted equipment of its manufacture either,
at its option, by return of the item to the factory, or shipment of
a repaired or replacement part. Lafayette Instrument Company
will not be obliged, however, to replace or repair any piece of
equipment, which has been abused, improperly installed, altered,
damaged, or repaired by others. Defects in equipment do not
include decomposition, wear, or damage by chemical action or
corrosion, or damage incurred during shipment.
Limited Obligations Covered by this Warranty
1. In the case of instruments not of Lafayette Instrument
Companymanufacture,theoriginalmanufacturer’swarranty
applies.
2. Shipping charges under warranty are covered only in one
direction. The customer is responsible for shipping charges
to the factory if return of the part is required.
3. This warranty does not cover damage to components due to
improper installation by the customer.
4. Consumable and or expendable items, including but not
limited to electrodes, lights, batteries, fuses, O-rings, gaskets,
and tubing, are excluded from warranty.
5. Failure by the customer to perform normal and reasonable
maintenance on instruments will void warranty claims.
6. If the original invoice for the instrument is issued to a
company that is not the company of the end user, and not
an authorized Lafayette Instrument Company distributor,
then all requests for warranty must be processed through
the company that sold the product to the end user, and not
directly to Lafayette Instrument Company.
16
Table of contents
Other Lafayette Instrument Measuring Instrument manuals

Lafayette Instrument
Lafayette Instrument J00105 User manual

Lafayette Instrument
Lafayette Instrument NICHOLAS MMT 01160 User manual

Lafayette Instrument
Lafayette Instrument 01165A User manual

Lafayette Instrument
Lafayette Instrument ACUMAR User manual

Lafayette Instrument
Lafayette Instrument J00111 User manual

Lafayette Instrument
Lafayette Instrument ACU100 User manual