LAUMAS LCB User manual

ENGLISH ENGLISH ENGLISH ENGLISH
User Manual
version 1.02
LCB
2014/30/UE
EN55022:2010 EN61000-6-2:2005 EN61000-6-4:2007
SYSTEM IDENTIFICATION

KEY TO SYMBOLS
Below are the symbols used in the manual to draw the reader's attention:
Warning! This operation must be performed by skilled workers.
Read the following indications carefully.
Further information.
GUARANTEE
24 months from the delivery document date. The guarantee covers only defected parts and includes the replacement
parts and labour. All shipping and packing costs are paid by the customer. It is possible to have the repair in
guarantee on condition that the returned product has not been transformed, damaged or repaired without
authorization. No guarantee is applicable on returned products without the original label and/or serial number. No
guarantee against misuse.
Batteries: Laumas provides 1 year guarantee from the date of delivery note, against material defects or battery
manufacturing faults.
DISPOSAL
This symbol on the product or on the packaging indicates that:
-This is an electrical/electronic device that should be disposed of separately from the municipal solid waste via
designated collection facilities
-Improper use or disposal can be harmful to human health and the environment
-Non-compliance with these guidelines will be sanctioned according to the laws in force in the destination country
-It is recommended to dispose of packaging in accordance with local regulations

TABLE OF CONTENTS
INTRODUCTION TO THE USER MANUAL ........................................................................... 1
LCB MANAGER ....................................................................................................................... 1
LEDs AND KEYS FUNCTIONS.........................................................................................................1
COMMUNICATION INTERFACE ............................................................................................. 2
GROSS WEIGHT (GW) - NET WEIGHT (NW)...................................................................................3
EXCHANGE REGISTERS (R1, R2, W1, W2) ....................................................................................3
STATUS REGISTER 1 (SR1) ............................................................................................................3
STATUS REGISTER 2 (SR2) ............................................................................................................3
INSTRUMENT STATUS (IS)..............................................................................................................3
EXECUTION REGISTER (EXR) ........................................................................................................4
INPUTS (INS) AND OUTPUTS (OUTS) STATUS..............................................................................4
DIGITAL OUTPUTS COMMAND (CMDOUT)....................................................................................5
POSSIBLE COMMANDS TO BE SENT TO THE COMMAND REGISTER (CMDR) .........................5
INSTRUMENT COMMISSIONING.......................................................................................... 7
MENU MAP ............................................................................................................................ 8
SETPOINT................................................................................................................................ 8
SYSTEM PARAMETERS (BASE program)............................................................................. 8
SYSTEM PARAMETERS (LOAD program)............................................................................. 9
BATCHING CONSTANTS (LOAD program) ........................................................................... 9
PROGRAMMING OF SYSTEM PARAMETERS .................................................................. 10
INSTRUMENT INFORMATION .............................................................................................. 10
THEORETICAL CALIBRATION ............................................................................................. 10
THEORETICAL FULL SCALE.........................................................................................................11
SENSITIVITY ...................................................................................................................................11
DIVISION .........................................................................................................................................12
MAXIMUM CAPACITY (BASE program)........................................................................................13
TARE WEIGHT ZERO SETTING.....................................................................................................13
ZERO VALUE MANUAL ENTRY.....................................................................................................13
REAL CALIBRATION (WITH SAMPLE WEIGHTS)............................................................... 14
FILTER ON THE WEIGHT...................................................................................................... 16
ANTI-PEAK......................................................................................................................................16
ZERO PARAMETERS............................................................................................................ 17
RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES ..........................................17
AUTOMATIC ZERO SETTING AT POWER-ON..............................................................................17
ZERO TRACKING ...........................................................................................................................18
SETTING UNITS OF MEASURE............................................................................................ 18
DISPLAY COEFFICIENT (BASE program) ....................................................................................19
OUTPUTS AND INPUTS CONFIGURATION (BASE program)............................................. 20
OUTPUTS AND INPUTS CONFIGURATION (LOAD program) ............................................ 22
SEMI-AUTOMATIC TARE (NET/GROSS).............................................................................. 22
PRESET TARE (SUBTRACTIVE TARE DEVICE) ................................................................. 23
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS) ............ 23
PEAK (BASE program) ......................................................................................................... 24

ANALOG OUTPUT (LCBRS485ANA only) ........................................................................... 24
TEST....................................................................................................................................... 25
SETPOINT PROGRAMMING (BASE program) .................................................................. 26
BATCHING (LOAD program).............................................................................................. 27
INTRODUCTION TO THE OPERATION ................................................................................ 27
BATCHING SEQUENCE ........................................................................................................ 27
BATCHING SEQUENCE PROGRAMMING ....................................................................................28
BATCHING START .........................................................................................................................28
BATCHING EXECUTION.................................................................................................................29
WAITING PHASE ............................................................................................................................29
CYCLE END PHASE .......................................................................................................................30
OPERATION SETTINGS........................................................................................................ 30
PROGRAMMING OF BATCHING CONSTANTS ................................................................... 31
MINIMUM WEIGHT..........................................................................................................................31
MAXIMUM WEIGHT ........................................................................................................................32
SAFE EMPTYING TIME ..................................................................................................................32
WAITING TIME ................................................................................................................................32
NO COMPARISON TIME.................................................................................................................33
NO PRODUCT LOAD TIME.............................................................................................................33
NO PRODUCT UNLOAD TIME .......................................................................................................33
FALL................................................................................................................................................34
TOLERANCE...................................................................................................................................35
SLOW ..............................................................................................................................................35
TAPPING FUNCTION......................................................................................................................36
AUTOTARE .....................................................................................................................................36
AUTOTARE DELAY ........................................................................................................................37
STABLE TARE ................................................................................................................................37
CONSUMPTION FOR EACH FORMULA ........................................................................................37
WAITING CONFIRMATION FROM PC (SLAVE) ............................................................................38
FORMULAS PROGRAMMING............................................................................................... 38
DELETING FORMULAS..................................................................................................................39
INSTRUMENT MANAGEMENT DURING THE BATCHING................................................... 39
INFORMATION ABOUT THE BATCHING IN PROGRESS.............................................................39
BATCHING STOP AND PAUSE......................................................................................................39
ALARMS MANAGEMENT DURING THE BATCHING ....................................................................40
BATCHING DATA READING ..........................................................................................................40
CONSUMPTION MANAGEMENT .......................................................................................... 41
CONSUMPTION READING .............................................................................................................41
CONSUMPTION DELETION ...........................................................................................................42
ALARMS .............................................................................................................................. 42
FIELDBUSES....................................................................................................................... 44
MODBUS-RTU (LCBRS485 and LCBRS485ANA only)........................................................ 44
INSTRUMENT SETUP.....................................................................................................................44
PC/PLC SETUP ...............................................................................................................................44
FUNCTIONS SUPPORTED IN MODBUS........................................................................................45
COMMUNICATION ERROR MANAGEMENT .................................................................................46
COMMUNICATION EXAMPLES .....................................................................................................47

CANOPEN (LCBCANOPEN only) ......................................................................................... 49
INSTRUMENT SETUP.....................................................................................................................49
PC/PLC SETUP ...............................................................................................................................49
STATUS LED...................................................................................................................................49
IO-LINK (LCBIOLINK only).................................................................................................... 50
PC/PLC SETUP ...............................................................................................................................50
STATUS LED...................................................................................................................................50
ETHERCAT (LCBETHERCAT only) ...................................................................................... 51
PC/PLC SETUP ...............................................................................................................................51
STATUS LED...................................................................................................................................51
ETHERNET/IP (LCBETHEIP only)......................................................................................... 52
INSTRUMENT SETUP.....................................................................................................................52
PC/PLC SETUP ...............................................................................................................................52
STATUS LED...................................................................................................................................54
MODBUS/TCP (LCBMODBUSTCP only) .............................................................................. 55
INSTRUMENT SETUP.....................................................................................................................55
PC/PLC SETUP ...............................................................................................................................55
STATUS LED...................................................................................................................................56
POWERLINK (LCBPOWERLINK only) ................................................................................. 57
INSTRUMENT SETUP.....................................................................................................................57
PC/PLC SETUP ...............................................................................................................................57
STATUS LED...................................................................................................................................57
PROFINET-IO (LCBPROFINETIO only) ................................................................................ 58
INSTRUMENT SETUP.....................................................................................................................58
PC/PLC SETUP ...............................................................................................................................58
STATUS LED...................................................................................................................................59
SERCOSIII (LCBSERCOSIII only)......................................................................................... 60
INSTRUMENT SETUP.....................................................................................................................60
PC/PLC SETUP ...............................................................................................................................60
STATUS LED...................................................................................................................................61
SERIAL PROTOCOLS (LCBRS485 and LCBRS485ANA only) ......................................... 61
SERIAL COMMUNICATION SETTING .................................................................................. 61
FAST CONTINUOUS TRANSMISSION PROTOCOL ............................................................ 62
CONTINUOUS WEIGHT TRANSMISSION TO REMOTE DISPLAYS PROTOCOL............... 63
ASCII BIDIRECTIONAL PROTOCOL .................................................................................... 64
SERIAL PROTOCOLS ALARMS ........................................................................................... 69
RESERVED FOR THE INSTALLER..................................................................................... 70
DATA DELETION AND PROGRAM SELECTION ................................................................. 70
DECLARATION OF CONFORMITY - EU............................................................................. 72

- 1 -
INTRODUCTION TO THE USER MANUAL
This manual explains the operation of the LCB instrument through the use of the LCB Manager
software and/or the communication interface.
For example, DIVISION section: the first part explains what the division is and how to modify it by
using LCB Manager; the second part (COMMUNICATION INTERFACE) explains how to
perform the same operation by using the communication interface commands.
For a better understanding of the contents, it is assumed the knowledge of the following topics:
-management of weight values via the communication interface (see section DIVISION);
-alarm identification (see section ALARMS);
-registers and data of the communication interface (see section COMMUNICATION INTERFACE);
-LCB Manager software (see section LCB MANAGER).
LCB MANAGER
The LCB Manager software replicates on the PC monitor the Laumas instrument
interface with 7-segment display, signalling LEDs and keyboard. Any reference to
the display, the keys and the LEDs is related to the LCB Manager software.
For further information, refer to the LCB Manager manual.
LEDs AND KEYS FUNCTIONS
LED
Function
NET
net weight (semi-automatic tare or preset tare)
>0<
zero (deviation from zero not more than ±0.25 divisions)
><
stability
kg
unit of measure: kg
g
unit of measure: g
W1
not used
W2
W3

- 2 -
KEY
Function
Keyboard shortcut
Enter in menu/return to the weight displaying
CTRL+M
Confirm or enter in submenu
ENTER
Select figure to be modified or go to previous menu item
LEFT ARROW
Modify selected figure or go to previous menu item
UP ARROW
Select figure to be modified or go to next menu item
RIGHT ARROW
Modify selected figure or go to next menu item
DOWN ARROW
Cancel or return to previous menu
ESC
Into menus LEDs light up in sequence to indicate that it is not displaying a weight.
COMMUNICATION INTERFACE
Data exchanged by the instrument can be input data or output data. For specific information regarding
the communication interface used, see section FIELDBUS.
Output data from instrument (reading)
Abbreviation*
Dimension (byte)
Gross weight
GW
4
Net weight
NW
4
Exchange register R1
R1
4
Exchange register R2
R2
2
Status Register 1
SR1
2
Status Register 2
SR2
1
Instrument status
IS
1
Execution register
EXR
2
Digital inputs status
INS
1
Digital outputs status
OUTS
1
Input data to instrument (writing)
Abbreviation*
Dimension (byte)
Command Register
CMDR
2
Digital outputs command
CMDOUT
2
Exchange register W1
W1
4
Exchange register W2
W2
2
*In the user manual, the abbreviation of the registers will be used instead of the
extended name.

- 3 -
GROSS WEIGHT (GW) - NET WEIGHT (NW)
The weight values are expressed as positive integer numbers, including decimal figures, but without
decimal point. Read the Status Register to obtain information about sign and possible errors on the
weight.
EXCHANGE REGISTERS (R1, R2, W1, W2)
These registers are used for instrument management operations. Two read and two write exchange
registers are available.
STATUS REGISTER 1 (SR1)
Bit 0
Load cell error
Bit 8
Net weight negative sign
Bit 1
AD converter malfunction
Bit 9
Peak weight negative sign
Bit 2
Maximum weight exceeded by 9 divisions
Bit 10
Net display mode
Bit 3
Gross weight higher than 110%
of full scale
Bit 11
Weight stability
Bit 4
Gross weight beyond 999999
or less than -999999
Bit 12
Weight within ±¼ of a division
around ZERO
Bit 5
Net weight beyond 999999
or less than -999999
Bit 13
Bit 6
Bit 14
Bit 7
Gross weight negative sign
Bit 15
STATUS REGISTER 2 (SR2)
Bit 0
Preset tare enabled
Bit 4
Bit 1
Instrument ready
Bit 5
Bit 2
Bit 6
Bit 3
Bit 7
INSTRUMENT STATUS (IS)
This register is used for the LOAD program only.
0
Instrument at rest (weight displaying)
13
alarm
1
Batching not possible/programming mode
14
alarm
2
Batching phase
15
alarm
3
Waiting phase
16
alarm
4
Cycle end phase
17
alarm
5
Batching paused
18
alarm

- 4 -
6
alarm
19
alarm
7
alarm
20
8
alarm
21
9
22
10
alarm
23
11
24
Temporary message
12
EXECUTION REGISTER (EXR)
This register allows you to monitor the progress of a command previously sent to CMDR.
Register contents
Description
Command code sent to CMDR
Command executed correctly
0
No command running
1
Command running
4
Error: the command requires a qualified access
5
Error: non-existent command
0xFFFF
Error: command not executed
It may take several seconds to execute some commands.
INPUTS (INS) AND OUTPUTS (OUTS) STATUS
DIGITAL INPUTS STATUS (INS)
DIGITAL OUTPUTS STATUS (OUTS)
Bit 0
INPUT 1 status
Bit 0
OUTPUT 1 status
Bit 1
INPUT 2 status
Bit 1
OUTPUT 2 status
Bit 2
Bit 2
OUTPUT 3 status
Bit 3
Bit 3
Bit 4
Bit 4
Bit 5
Bit 5
Bit 6
Bit 6
Bit 7
Bit 7
Bit=1: high input; Bit=0: low input
Bit=1: output is closed; Bit=0: output is open

- 5 -
DIGITAL OUTPUTS COMMAND (CMDOUT)
This register is used for the BASE program only; it allows to control the outputs set to mode
(see section OUTPUTS AND INPUTS CONFIGURATION (BASE program)).
Bit 0
OUTPUT 1 status
Bit 4
Bit 8
Bit 12
Bit 1
OUTPUT 2 status
Bit 5
Bit 9
Bit 13
Bit 2
OUTPUT 3 status
Bit 6
Bit 10
Bit 14
Bit 3
Bit 7
Bit 11
Bit 15
Force outputs
Bit=1: output is closed; Bit=0: output is open
Setting bit 15 to 1 on the PLC, the master takes control of all the outputs, whatever their
setting.
POSSIBLE COMMANDS TO BE SENT TO THE COMMAND REGISTER (CMDR)
Instrument information
1220
SW, HW and REV reading
6040
Reference number reading
1221
Year of production
and serial number reading
Calibration and system parameters
7
Semi-automatic tare activation
6025
Filter reading
8
Semi-automatic zero
6026
Filter writing
9
Semi-automatic tare deactivation
6027
Automatic zero setting at power on reading
87
Preset tare reading
6028
Automatic zero setting at power on writing
88
Preset tare writing
6029
Unit of measure reading
100
Zero-setting for calibration
6030
Unit of measure writing
101
Sample weight storage
6033
Display coefficient reading
102
Sample weight reading
6034
Display coefficient writing
103
Sample weight writing
6043
Theoretical zero reading
130
Preset tare application
6044
Theoretical zero writing
6000
Theoretical full scale writing
6101
Resettable weight
for small weight changes reading
6001
Theoretical full scale reading
6102
Resettable weight
for small weight changes writing
6002
Calibration points deleting
6103
Zero tracking reading
6005
Calibration points reading
6104
Zero tracking writing
6006
Calibration points writing
6127
Anti-peak activation reading
6007
Sensitivity reading
6128
Anti-peak activation writing
6008
Sensitivity writing
6129
Display coefficient activation reading
6009
E/divisions reading
6130
Display coefficient activation writing
6010
E/divisions writing
6131
Peak activation and peak weight reading
6015
Maximum capacity reading
6132
Peak activation writing
6016
Maximum capacity writing

- 6 -
Inputs/outputs settings and test
1122
Inputs configuration reading
1124
Outputs configuration reading
1123
Inputs configuration writing
1125
Outputs configuration writing
6137
Start mV test
Setpoint and hysteresis
90
Setpoint 1 reading
162
Hysteresis 1 reading
91
Setpoint 2 reading
163
Hysteresis 2 reading
92
Setpoint 3 reading
164
Hysteresis 3 reading
93
Setpoint 1 writing
165
Hysteresis 1 writing
94
Setpoint 2 writing
166
Hysteresis 2 writing
95
Setpoint 3 writing
167
Hysteresis 3 writing
99
Data storage to EEPROM
Batching operation, constants and formulas
1000
Maximum weight reading
1032
No product load time reading
1001
Maximum weight writing
1033
No product load time writing
1002
Minimum weight reading
1034
No product unload time reading
1003
Minimum weight writing
1035
No product unload time writing
1004
Safe emptying time reading
1042
Fall mode reading
1005
Safe emptying time writing
1043
Fall mode writing
1006
Waiting time reading
1044
Fall divisions reading
1007
Waiting time writing
1045
Fall divisions writing
1012
Manual fall reading
1052
Consumption activation reading
1013
Manual fall writing
1053
Consumption activation writing
1016
Tolerance reading
1062
Slave configuration reading
1017
Tolerance writing
1063
Slave configuration writing
1020
Slow reading
1096
No comparison time reading
1021
Slow writing
1097
No comparison time writing
1022
Slow ON reading
1132
Operation settings reading
1023
Slow ON writing
1133
Operation settings writing
1024
Slow OFF reading
2035
Formula set reading
1025
Slow OFF writing
2036
Formula set writing
1026
Autotare reading
2037
Formula preset reading
1027
Autotare writing
2038
Formula preset writing
1028
Autotare delay reading
2040
Formulas deleting
1029
Autotare delay writing
2030
Formula number
and cycles to be batched reading
1030
Stable tare reading
2031
Formula number
and cycles to be batched writing
1031
Stable tare writing
Commands and data related to the batching in progress
201
Batching start
2032
Running cycle reading
202
Batching pause
2100
Batching data reading
203
Batching resume
2101
Batching data reading: actual batched
204
Batching stop
2102
Batching data reading: theoretical batched
205
Accept alarm during the batching
2103
Batching data reading: starting tare

- 7 -
206
Cancel alarm
2104
Batching data reading: final gross
207
Cancel alarm
2105
Batching data reading: alarms
209
Continue batching
1135
Totals deleting
250
Batching data reading complete
2020
Totals reading
1114
Batching data status reading
Certain data, when specifically indicated, will be written directly in the EEPROM type memory. This
memory has a limited number of writing operations (100000), therefore it is necessary to pay
particular attention to not execute useless operations on said locations. The instrument in any case
makes sure that no writing occurs if the value to be memorised is equal to the value in memory.
INSTRUMENT COMMISSIONING
Check that the display shows the weight and that when loading the load cells there is an increase in
weight. If there is not check and verify the connections and correct positioning of the load cells.
-If the instrument has already been theoretical CALIBRATED (plant system identification tag
present on the instrument and on the cover: load cell’s rated data already entered):
▫reset to zero (see section TARE WEIGHT ZERO SETTING);
▫check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
-If the instrument HAS NOT BEEN CALIBRATED (missing plant system identification tag)
proceed with calibration:
▫if load cells data are unknown, follow the procedure in section REAL CALIBRATION (WITH
SAMPLE WEIGHTS);
▫enter the rated data of load cells following the procedure given in section THEORETICAL
CALIBRATION;
▫reset to zero (see section TARE WEIGHT ZERO SETTING);
▫check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
-If you use the analog output, set the desired analog output type and the full scale value (see
section ANALOG OUTPUT).
BASE PROGRAM ONLY
-If setpoint are used, set the required weight values and the relevant parameters (see sections
SETPOINT PROGRAMMING and OUTPUTS AND INPUTS CONFIGURATION).
LOAD PROGRAM ONLY
Required settings for the first batching:
-set the minimum weight value (see section MINIMUM WEIGHT);
-set the parameter (see section OPERATION SETTINGS);
-set the formula 01 (see section FORMULAS PROGRAMMING);
-start the batching by closing the START contact or remotely.

- 8 -
MENU MAP
Into menus changes are applied right after pressing the key (no further confirmation is required).
SETPOINT
SYSTEM PARAMETERS (BASE program)
…
…
LCBRS485ANA only

- 9 -
SYSTEM PARAMETERS (LOAD program)
BATCHING CONSTANTS (LOAD program)
LCBRS485ANA only
…
…
it appears only if =
it appears only if is set
it appears only if is set

- 10 -
PROGRAMMING OF SYSTEM PARAMETERS
INSTRUMENT INFORMATION
The LCB Manager software shows, in the Information area of the main screen, some data
concerning the connected instrument (see the LCB Manager manual).
COMMUNICATION INTERFACE
Command sent to CMDR
Register to read
Register contents
1220
R1 (16 most significant bit)
software code
R1 (16 least significant bit)
firmware version
R2
hardware code
1221
R1 (16 most significant bit)
year of production
R1 (16 least significant bit)
serial number
R2
program code*
*0=BASE; 13=LOAD
THEORETICAL CALIBRATION
This function allows the load cell rated values to be set.
To perform the theoretical calibration, set in sequence the values of theoretical full scale, sensitivity
and division.
By modifying the theoretical full scale, the sensitivity or the division:
-the real calibration is cancelled and the theoretical calibration only is considered valid;
-the system parameters containing a weight value will be set to default values.

- 11 -
THEORETICAL FULL SCALE
(default: ): the system full scale is given by one cell capacity multiplied by the
number of cells used. Example: 4 cells of 1000 kg FULL SCALE = 1000 x 4 = 4000.
The instrument is supplied with a theoretical full scale value corresponding to 10000.
To restore factory values, set 0 as full scale.
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6000 to CMDR.
Send command 6001 to CMDR.
Read the value in R1.
SENSITIVITY
(default: 2.00000 mV/V): the sensitivity is a load cell rated parameter expressed in mV/V.
Set the average sensitivity value indicated on the load cells. It’s possible to set a value between
0.50000 and 7.00000 mV/V. Example: 4-cell system with sensitivity 2.00100, 2.00150, 2.00200,
2.00250; enter 2.00175, calculated as (2.00100 + 2.00150 + 2.00200 + 2.00250) / 4.
COMMUNICATION INTERFACE
WRITING
READING
Multiply the value by 100000.
Write the value in W1.
Send command 6008 to CMDR.
Send command 6007 to CMDR.
Read the value in R1.
Divide the value by 100000.
Example: to set to 2.00175, write 200175 in W1 and send command 6008 to CMDR.

- 12 -
DIVISION
:the division is the minimum weight increment value which can be displayed. It is automatically
calculated by the system according to the performed calibration, so that it is equal to 1/10000 of full
scale. It can be changed and be variable between 0.0001 and 100 with x1 x2 x5 x10 increments.
DIVISIONS
Index
Division
Active decimals
Index
Division
Active decimals
0
100
0
10
0.05
2
1
50
0
11
0.02
2
2
20
0
12
0.01
2
3
10
0
13
0.005
3
4
5
0
14
0.002
3
5
2
0
15
0.001
3
6
1
0
16
0.0005
4
7
0.5
1
17
0.0002
4
8
0.2
1
18
0.0001
4
9
0.1
1
COMMUNICATION INTERFACE
WRITING
READING
Write the index* in W1.
Send command 6010 to CMDR.
Send command 6009 to CMDR.
Read the index* in R1.
*see DIVISIONS table
WARNING: when a parameter is expressed in weight value, it is necessary to consider the number
of divisions and active decimals set on the instrument: the value must be multiplied or divided by 10n
(n=active decimals, see DIVISIONS table) and rounded to the set divisions.
Examples: writing a value
Weight value
to set
Division
Active
decimals
Value to write
in the register
Value rounded
to the division
100
0.1
1
1000 given by 100x101
100.0
12.00
0.05
2
1200 given by 12.00x102
12.00
33
5
0
33 given by 33x100
35
20.123
0.002
3
20123 given by 20.123x103
20.122
Examples: reading a value
Weight value read
by the instrument
Division
Active
decimals
Corresponding weight value
1000
0.1
1
100.0 given by 1000/101
1200
0.05
2
12.00 given by 1200/102
35
5
0
35 given by 35/100
20122
0.002
3
20.122 given by 20122/103

- 13 -
MAXIMUM CAPACITY (BASE program)
: maximum displayable weight (from 0 to full scale; default: 0). When the gross weight exceeds
this value by 9 divisions, the display shows . To disable this function, set 0.
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6016 to CMDR.
Send command 6015 to CMDR.
Read the value in R1.
TARE WEIGHT ZERO SETTING
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
-confirm the message by pressing ;
-the weight value to be set to zero is displayed. In this phase all of the LEDs are flashing;
-confirming once again, the weight is set to zero (the value is stored to the permanent memory);
-press to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
COMMUNICATION INTERFACE
To reset the tare weight, send command 100 to CMDR.
ZERO VALUE MANUAL ENTRY
WARNING: perform this procedure only if it’s not possible to reset the weighed structure tare, for
example because it contains product that cannot be unloaded.
Set in this parameter the estimated zero value (from 0 to 999999; default: 0).
COMMUNICATION INTERFACE
WRITING
READING
Write the value in W1.
Send command 6044 to CMDR.
Send command 6043 to CMDR.
Read the value in R1.

- 14 -
REAL CALIBRATION (WITH SAMPLE WEIGHTS)
After having performed the THEORETICAL CALIBRATION and the TARE WEIGHT ZERO
SETTING, this function allows correct calibration to be done using sample weights of known
value and, if necessary, any deviations of the indicated value from the correct value to be
corrected.
Load onto the weighing system a sample weight, which must be at least 50% of the maximum
quantity to be weighed. By confirming the message the flashing value of the weight
currently on the system is displayed. In this phase all of the indicator LEDs are off. Adjust the value
on display by using the arrow keys if necessary. After confirming, the new set weight will appear with
all the indicator LEDs flashing. After an additional confirmation, the message will be
restored and it is possible to return to the weight displaying.
Example: for a system of maximum capacity 1000 kg and 1 kg division, two sample weights are
available, one of 500 kg and the other one of 300 kg. Load both weights onto the system and correct
the indicated weight to 800. Now remove the 300 kg weight, the system must show 500; remove the
500 kg weight too; the system must read zero. If this does not happen, it means that there is a
mechanical problem affecting the system linearity.
WARNING: identify and correct any mechanical problems before repeating the procedure.
-If theoretical full scale and recalculated full scale in real calibration are equal, it means
that the theoretical calibration is currently in use; otherwise, the real calibration based on
sample weights is in use.
-If the correction made changes the previous full scale for more than 20%, all the
parameters with settable weight values are reset to default values.
LINEARISATION OPTION ON MAX 5 POINTS:
It is possible to perform a linearisation of the weight repeating the above-described procedure
up to a maximum of five points, using five different sample weights. The procedure ends by
pressing the button or after entering the fifth value; at this point it will no longer be possible to
change the calibration value, but only to perform a new real calibration. To perform a new calibration,
should return to the weight display and then re-entering into the calibration menu.
By pressing after having confirmed the sample weight that has been set, the full scale appears,
recalculated according to the value of the maximum sample weight entered and making reference to
the cell sensitivity set in the theoretical calibration ().

- 15 -
COMMUNICATION INTERFACE
-Load onto the weighing system a sample weight, which must be at least 50% of the maximum
quantity to be weighed.
-Write the sample weight value in W1.
-Write 1 in W2.
-Send command 6006 to CMDR.
To perform a new real calibration, repeat the procedure.
LINEARISATION OPTION ON MAX 5 POINTS:
It is possible to perform a linearisation of the weight up to a maximum of five points, using five
different sample weights.
The calibration points must be consecutive and the weights must be in ascending order. If
one or both conditions are not met, the calibration point is not inserted and the instrument
responds with the error code "command not executed".
CALIBRATION POINTS MANAGEMENT
WRITING
READING
Write the weight value in W1.
Write the calibration point index (from 1
to 5) in W2.
Send command 6006 to CMDR.
Write the calibration point index (from 1
to 5) in W2
Send command 6005 to CMDR.
Read the weight value in R1.
Read the calibration point validity
(1=valid; 0=not valid) in R2.
Example: calibrate a system on 3 points using 3 sample weights.
CALIBRATION A
Calibration point
1
2
3
OK
Sample weight
100 kg
500 kg
800 kg
CALIBRATION B
Calibration point
1
3
2
NO
Sample weight
100 kg
500 kg
800 kg
CALIBRATION C
Calibration point
1
2
3
NO
Sample weight
100 kg
800 kg
500 kg
-CALIBRATION A: correct, the calibration points and sample weights are in ascending order; the
procedure ends successfully.
-CALIBRATION B: incorrect, the calibration points are not consecutive; to continue, set the
calibration point 2 before 3.
-CALIBRATION C: incorrect, the sample weights are not in ascending order; to continue, use a
weight greater than 800 kg for calibration point 3.
Before performing a multi-point calibration again, reset the value of all calibration
points by sending command 6002 to CMDR.
Other manuals for LCB
1
Table of contents
Popular Industrial Equipment manuals by other brands

Digga
Digga Mini Bigfoot Trencher Operator's manual

Simplex
Simplex Full-Size Fuelport manual

KMC Controls
KMC Controls KMCConquest HLO-1050 installation guide

Eaton
Eaton HMH36 user manual

Steelwrist
Steelwrist Tiltrotator Front pin lock X04 user manual

Siemens
Siemens SITRANS F M Verificator operating instructions