LAUMAS TLM8 User manual

ENGLISH ENGLISH ENGLISH ENGLISH
User Manual
version 1.04
TLM8
2014/30/UE
EN55022:2010 EN61000-6-2:2005 EN61000-6-4:2007
SYSTEM IDENTIFICATION
CE-MAPPROVED EN45501:2015-2014/31/EU-OIML R76:2006

KEY TO SYMBOLS
Below are the symbols used in the manual to draw the reader's attention:
Warning! Risk of electrocution.
Warning! This operation must be performed by skilled workers.
Read the following indications carefully.
Further information.
GUARANTEE
24 months from the delivery document date. The guarantee covers only defected parts and includes the replacement parts and
labour. All shipping and packing costs are paid by the customer. It is possible to have the repair in guarantee on condition that the
returned product has not been transformed, damaged or repaired without authorization. No guarantee is applicable on returned
products without the original label and/or serial number. No guarantee against misuse.
Batteries: Laumas provides 1 year guarantee from the date of delivery note, against material defects or battery manufacturing faults.
Disposal of Waste Equipment by Users in Private Households in the European Union
This symbol on the product or on its packaging indicates that this product must not be disposed of with
your other household waste. It is your responsibility to dispose of your waste equipment by handing it
over to a designated collection point for the recycling of waste electrical and electronic equipment. The
separate collection and recycling of your waste equipment at the time of disposal will help preserve
natural resources and protect human health and the environment. For more information about where you
can drop off your waste equipment for recycling, please contact your local waste disposal Authority or
the equipment retailer.

TABLE OF CONTENTS
USER WARNINGS................................................................................................................. 1
RECOMMENDATIONS FOR CORRECT INSTALLATION OF WEIGHING INSTRUMENTS. 1
RECOMMENDATIONS FOR CORRECT INSTALLATION OF THE LOAD CELLS ............... 1
LOAD CELL INPUT TEST (QUICK ACCESS) ......................................................................... 3
LOAD CELL TESTING............................................................................................................. 3
MAIN SPECIFICATIONS OF THE INSTRUMENT.................................................................. 4
TECHNICAL SPECIFICATIONS ............................................................................................ 5
KEYS AND SYMBOLS FUNCTIONS..................................................................................... 6
MENU MAP ............................................................................................................................ 7
SETPOINT................................................................................................................................ 7
SYSTEM PARAMETERS ......................................................................................................... 7
DISPLAY VISUALIZATIONS UPON INSTRUMENT SWITCH-ON......................................... 9
PROGRAMMING OF SYSTEM PARAMETERS .................................................................... 9
FILTER ON THE WEIGHT........................................................................................................ 9
ANTI PEAK......................................................................................................................................10
ALIBI MEMORY (ONLY FOR INSTRUMENTS WHERE THIS OPTION IS AVAILABLE)...... 11
ALIBI MEMORY READING .............................................................................................................11
OUTPUTS AND INPUTS CONFIGURATION ......................................................................... 12
SEMI-AUTOMATIC TARE (NET/GROSS).............................................................................. 13
PRESET TARE (SUBTRACTIVE TARE DEVICE) ................................................................. 14
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS) ............ 14
PEAK...................................................................................................................................... 15
ANALOG OUTPUT................................................................................................................. 15
SERIAL COMMUNICATION SETTING .................................................................................. 16
RS485 SERIAL COMMUNICATION ................................................................................................18
DIRECT CONNECTION BETWEEN RS485 AND RS232 WITHOUT CONVERTER.......................18
AUTOMATIC DIAGNOSTICS OF LOAD DISTRIBUTION...................................................... 19
TEST....................................................................................................................................... 20
EVENTS LOG......................................................................................................................... 21
INFO MENU............................................................................................................................ 21
SETPOINT PROGRAMMING............................................................................................... 22
USE WITH W SERIES INSTRUMENTS ............................................................................... 23
CONNECTION TO THE WEIGHT INDICATOR ...................................................................... 23
ADDITIONAL MENU MAP...................................................................................................... 24
TLM8 REMOTE CONTROL.................................................................................................... 25
REMOTE KEYPAD LOCKING.........................................................................................................25
ADDITIONAL ALARMS.......................................................................................................... 25
ALARMS .............................................................................................................................. 26
PRINTING EXAMPLES........................................................................................................ 28
INFORMATION ON TYPE-APPROVED INSTRUMENTS IN THE EUROPEAN UNION ...... 30

- 1 -
USER WARNINGS
RECOMMENDATIONS FOR THE PROPER USE OF WEIGHING INSTRUMENT
-Keep away from heat sources and direct sunlight
-Repair the instrument from rain (except special IP versions)
-Do not wash with water jets (except special IP versions)
-Do not dip in water
-Do not spill liquid on the instrument
-Do not use solvents to clean the instrument
-Do not install in areas subject to explosion hazard (except special Atex versions)
RECOMMENDATIONS FOR CORRECT INSTALLATION OF WEIGHING INSTRUMENTS
The terminals indicated on the instrument’s wiring diagram to be connected to earth must
have the same potential as the weighed structure (same earthing pit or earthing system). If
you are unable to ensure this condition, connect with an earthing wire the terminals of the
instrument (including the terminal –SUPPLY) to the weighed structure.
The cell cable must be individually led to its panel input and not share a conduit with other cables;
connect it directly to the instrument terminal strip without breaking its route with support terminal strips.
Use “RC” filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid inverters in the instrument panel; if inevitable, use special filters for the inverters and separate
them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch
etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
MAXIMUM CABLE LENGTHS
-RS485: 1000 metres with AWG24, shielded and twisted cables
-Analog current output: up to 500 metres with 0.5 mm2 cable
-Analog voltage output: up to 300 metres with 0.5 mm2cable
RECOMMENDATIONS FOR CORRECT INSTALLATION OF THE LOAD CELLS
INSTALLING LOAD CELLS: The load cells must be placed on rigid, stable in-line structures; it
is important to use the mounting modules for load cells to compensate for misalignment of the
support surfaces.
PROTECTION OF THE CELL CABLE: Use water-proof sheaths and joints in order to protect the
cables of the cells.

- 2 -
MECHANICAL RESTRAINTS (pipes, etc.): When pipes are present, we recommend the use of
hoses and flexible couplings with open mouthpieces with rubber protection; in case of hard pipes,
place the pipe support or anchor bracket as far as possible from the weighed structure (at a distance
at least 40 times the diameter of the pipe).
WELDING: Avoid welding with the load cells already installed. If this cannot be avoided, place the
welder ground clamp close to the required welding point to prevent sending current through the load
cell body.
WINDY CONDITIONS - KNOCKS - VIBRATIONS: The use of weigh modules is strongly
recommended for all load cells to compensate for misalignment of the support surfaces. The system
designer must ensure that the plant is protected against lateral shifting and tipping relating to:
shocks and vibration; windy conditions; seismic conditions in the installation setting; stability of the
support structure.
EARTHING THE WEIGHED STRUCTURE: By means of a copper wire with suitable cross-section,
connect the cell upper support plate with the lower support plate, then connect all the lower plates to
a single earthing system. Electrostatic charges accumulated because of the product rubbing against
the pipes and the weighed container walls are discharged to the ground without going through or
damaging the load cells. Failure to implement a proper earthing system might not affect the
operation of the weighing system; this, however, does not rule out the possibility that the cells and
connected instrument may become damaged in the future. It is forbidden to ensure earthing system
continuity by using metal parts contained in the weighed structure.
FAILURE TO FOLLOW THE INSTALLATION RECOMMENDATIONS WILL BE CONSIDERED
A MISUSE OF THE EQUIPMENT
OK OK
NO NO
NO
OK

- 3 -
LOAD CELL INPUT TEST (QUICK ACCESS)
From the weight display, press for 3 seconds: the top of the
display shows the gross weight; the bottom shows the response
signal of each load cell expressed in mV with three decimals.
Example: a load cell with 2.000 mV/V sensitivity provides a
response signal between 0 and 10 mV.
LOAD CELL TESTING
Load cell resistance measurement (use a digital multimeter):
-Turn off the instrument.
-The value between the positive signal wire and the negative signal wire must be equal or similar
to the one indicated in the load cell data sheet (output resistance).
-The value between the positive excitation wire and the negative excitation wire must be equal or
similar to the one indicated in the load cell data sheet (input resistance).
Load cell voltage measurement (use a digital multimeter):
-Turn on the instrument.
-Take out the load cell to be tested from underneath the container, or alternatively, lift the
container support.
-Make sure that the excitation of two wires of the load cell connected to the instrument (or
amplifier) is 5 VDC ±3%.
-Measure the response signal between the positive and the negative signal wires by directly
connecting them to the tester, and make sure that it is comprised between 0 and 0.5 mV.
-Apply load to the cell and make sure that there is a signal increment.
IF ONE OF THE ABOVE CONDITIONS IS NOT MET, PLEASE CONTACT THE TECHNICAL
ASSISTANCE SERVICE.
3 s

- 4 -
MAIN SPECIFICATIONS OF THE INSTRUMENT
Weight transmitter with 6-wire load cells inputs suitable for assembly on back panel fitted
Omega/DIN rail. Dimensions: 148x92x60 mm. Backlit LCD graphic display, 128x64 pixel resolution,
60x32 mm viewing area. 5-key keypad.
16 bit analog output (tension or current).
RS485 serial port for connection to: PC/PLC up to 32 instruments (max 99 with line repeaters) by
ASCII Laumas or ModBus R.T.U. protocol, remote display, printer. Optional: integrated Profibus DP,
DeviceNet, CANopen, Profinet IO, Ethernet/IP, Ethernet TCP/IP, Modbus TCP, RS232/RS485,
SERCOS III, PowerLink, CC-Link output.
8 independent channels:automatic detection of connected load cells.
Digital equalization: load cell response uniformed via software.
Load distribution: graph showing the weight percentage on each load cell.
Automatic diagnostics: load distribution check to detect any faults.
Events log: storage of the last 50 events: calibrations, zero-settings, errors, equalizations.
Alarm Relay: outputs can be set to switch in case of alarm.

- 5 -
TECHNICAL SPECIFICATIONS
POWER SUPPLY and CONSUMPTION (VDC)
12/24 VDC ±10%; 5 W
NO. OF LOAD CELLS IN PARALLEL and SUPPLY
max 16 (350 ohm); 5 VDC / 240 mA
LINEARITY
< 0.01% F.S.
ANALOG OUTPUT LINEARITY
< 0.01% F.S.
THERMAL DRIFT
< 0.0005% F.S./°C
ANALOG OUTPUT THERMAL DRIFT
< 0.003 % F.S./°C
A/D CONVERTER
8 channels, 24 bit (16000000 points), 4.8 kHz
DIVISIONS
(with measurement range ±10 mV = sens. 2 mV/V)
±999999
MEASUREMENT RANGE
±39 mV
MAX SENSITIVITY OF USABLE LOAD CELLS
±7 mV/V
MAX CONVERSIONS PER SECOND
600 conversions/second
DISPLAY RANGE
±999999
NO. OF DECIMALS / DISPLAY INCREMENTS
0÷4 / x 1 x 2 x 5 x 10 x 20 x 50 x 100
DIGITAL FILTER / READINGS PER SECOND
11 levels / 5÷600 Hz
RELAY OUTPUTS
N. 5 - max 115 VAC; 150 mA
DIGITAL INPUTS
N. 3 - optoisolated 5 - 24 VDC PNP
SERIAL PORTS
RS485
BAUD RATE
2400, 4800, 9600, 19200, 38400, 115200
HUMIDITY (non condensing)
85%
STORAGE TEMPERATURE
-30°C +80°C
WORKING TEMPERATURE
-20°C +60°C
ANALOG OUTPUT
16 bit - 65535 divisions
0÷20 mA; 4÷20 mA (max 400 ohm);
0÷10 V; 0÷5 V (min 2 kohm)
METROLOGICAL SPECIFICATIONS OF TYPE-APPROVED INSTRUMENTS
APPLIED STANDARDS
2014/31/EU; OIML R76:2006;
EN45501:2015
OPERATION MODE
single interval, multi-interval, multiple range
ACCURACY CLASS
III or IIII
MAX NUMBER OF SCALE VERIFICATION DIVISIONS
10000 (class III); 1000 (class IIII)
MINIMUM INPUT SIGNAL FOR SCALE VERIFICATION
DIVISION
0.2 μV
WORKING TEMPERATURE
-10°C +40°C

- 6 -
KEYS AND SYMBOLS FUNCTIONS
KEY
Short press
Long press
(3 s)
Into menus
Semi-automatic zero
Cancel or return to previous
menu
Gross Net
Net Gross
Select figure to be modified
or go to previous menu item.
Weight print
mV load cell test
Modify selected figure or go
to next menu item
Setting setpoint and hysteresis
Confirm or enter in
submenu
+
Setting general parameters (press
immediately followed by )
+
Setting preset tare (press
immediately followed by )
Load distribution display
SYMBOL
Function
net weight (semi-automatic tare or preset tare)
zero (deviation from zero not more than ±0.25 divisions)
stability
unit of measure: kg
unit of measure: g
current range in multiple range or multi-interval instruments
The symbols are activated in sequence within the menus to indicate that the display is
not showing a weight.

- 7 -
MENU MAP
Into menus changes are applied right after pressing the ENTER key (no further confirmation is
required).
SETPOINT
SYSTEM PARAMETERS
…
…
+
optional

- 8 -
+
3 s

- 9 -
DISPLAY VISUALIZATIONS UPON INSTRUMENT SWITCH-ON
Upon switch-on, the display shows in sequence:
- → (ONLY in case of approved program);
-instrument model (e.g.: );
- followed by the software code (e.g.: );
-program type: (base);
-followed by the software version (e.g.: );
- followed by the hardware code (e.g.: );
-serial number (e.g.: );
PROGRAMMING OF SYSTEM PARAMETERS
From the weight display, press simultaneously keys MENU and ESC to access the parameter setting.
MENU/ENTER: to enter a menu/confirm the data entry.
: to modify the displayed figure or menu item.
: to select a new figure or modify the displayed menu item.
ESC: to cancel and return to the previous menu.
FILTER ON THE WEIGHT
Setting this parameter allows a stable weight display to be obtained.
To increase the effect (weight more stable) increase the value (from 0 to 9, default 4).
As seen in the diagram:
-By confirming the message, the currently programmed filter value is displayed.
-By changing and confirming the value, the weight is displayed and it will be possible to
experimentally verify its stability.
-If stability is not satisfactory, confirming brings back the message and the filter may be
modified again until an optimum result is achieved.
+

- 10 -
The filter enables to stabilise a weight as long as its variations are smaller than the corresponding
“response time”. It is necessary to set this filter according to the type of application and to the full
scale value set.
FILTER VALUE
Response times
[ms]
Display and serial port refresh
frequency
[Hz]
0
12
300
1
150
100
2
260
50
3
425
25
4 (default)
850
12.5
5
1700
12.5
6
2500
12.5
7
4000
10
8
6000
10
9
7000
5
A
6
600
The "A" filter can only be set if the instrument is connected to one load cell only.
ANTI PEAK
When the weight is stable, the anti peak filter removes any sudden disturbances with a maximum duration
of 1 second. Confirm the filter on the weight with ENTER and select one of the following options:
-: anti peak filter enabled (default);
-: anti peak filter disabled.

- 11 -
ALIBI MEMORY (ONLY FOR INSTRUMENTS WHERE THIS OPTION IS AVAILABLE)
To save a weight to the alibi memory press the key for less than 3 seconds or close the
dedicated input or send the command via serial protocol (see the Protocols manual).
If the net function is active, the net weight is stored; otherwise, the gross weight is stored.
Saving to the alibi memory is only possible if:
-the alibi memory is enabled;
-the weight is stable;
-the weight is above the minimum weight (20 e);
-the weight is below the max capacity;
-the weight has changed since the latest storage to alibi.
An identifying number is automatically assigned to each record stored in the memory, it increases
for stored records from 1 to 999999, and is automatically reset to 1 after reaching 999999.
WARNING: the alibi memory cannot store all the 999999 records; the number of records that
can be stored depends on the instrument type and on the program used.
The alibi memory is used in a “circular” manner: when the memory end is reached, the system will
restart from the memory beginning by overwriting the first record; an alarm message is displayed
().
ALIBI MEMORY READING
(identification): enter the ID number of your required record and confirm, the display will show:
-Gross weight () or net weight (); confirm with ENTER;
-Tara weight () or preset tare (), only if the recalled weight is a net weight; confirm
with ENTER;
The parameter provides the ID of the last record stored in Alibi Memory by default.
If the entered identifier does not match any record, the message is displayed.
+

- 12 -
OUTPUTS AND INPUTS CONFIGURATION
OUTPUTS
The outputs are set by default as follows: / / / /.
Possible operation modes:
- (normally open): the relay is de-energised and the contact is open when the weight is
lower than the programmed setpoint value; it closes when the weight is higher than or equal to
the programmed setpoint value.
- (normally closed): the relay is energised and the contact is closed when the weight is
lower than the programmed setpoint value; it opens when the weight is higher than or equal to
the programmed setpoint value.
-: the contact will switch on the basis of weight, according to setpoint (see section SETPOINT
PROGRAMMING).
-: the contact will not switch on the basis of weight, but is controlled by remote protocol
commands.
-: relay switching occurs when the weight is stable.
-: relay switching occurs when one of the following alarms is triggered: , ,
, , , , ; the operation mode is forced to (normally
closed).
If the operation mode is selected, the following options are also active:
-: the contact will switch on the basis of gross weight.
-: the contact will switch on the basis of net weight (If the net function is not active, the
contact will switch on the basis of gross weight).
-: relay switching occurs for both positive and negative weight values.
-: relay switching occurs for positive weight values only.
-: relay switching occurs for negative weight values only.
By confirming with ENTER the setpoint operation can be set to the value 0:
-: relay switching will not occur if the setpoint value is 0.
-:
-setpoint = 0 and switching = : relay switching occurs when the weight is 0; the
relay will switch again when the weight is different from zero, taking hysteresis into
account (both for positive and for negative weights).
-setpoint = 0 and switching = : relay switching occurs for a weight higher than or equal
to 0, the relay will switch again for values below 0, taking hysteresis into account.
-setpoint = 0 and switching = : relay switching occurs for a weight lower than or equal
to 0, the relay will switch again for values above 0, taking hysteresis into account.
+
…
…

- 13 -
INPUTS
Default: input 1 = input 2 = input 3 =
Possible operation modes:
- (NET/GROSS): by closing this input for no more than one second, it’s making an
operation of SEMI-AUTOMATIC TARE and the display will show the net weight. To display the
gross weight again, hold the NET/GROSS input closed for 3 seconds.
-: by closing the input for no more than one second, the weight is set to zero (see section
WEIGHT ZERO-SETTING FOR SMALL VARIATIONS (SEMI-AUTOMATIC ZERO)).
-: keeping the input closed the maximum weight value reached remains on display. Opening
the input the current weight is displayed.
-: closing the input no operation is performed, the input status may however be read remotely
by way of the communication protocol.
-: closing the input for max one second the weight is transmitted over the serial
connection according to the fast continuous transmission protocol only once (only if is
set in the item ).
-: when the input is closed the weight is displayed based on the set coefficient (see setting
of the units of measure and coefficient), otherwise the weight is displayed.
-: when the input is closed the data are sent for printing if in the communication protocol
of either serial port the parameter is set. If the alibi memory is active, data storage is
carried out too.
SEMI-AUTOMATIC TARE (NET/GROSS)
THE SEMI-AUTOMATIC TARE OPERATION IS LOST UPON INSTRUMENT POWER-
OFF.
To perform a net operation (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press the
TARE key for less than 3 seconds. The instrument displays the net weight (just set to zero) and the
NET symbol lights up. To display the gross weight again, keep the NET/GROSS input closed or
press TARE for 3 seconds.
This operation can be repeated many times by the operator to allow the loading of several products.
Example:
Put the box on the scale, the display shows the box weight; press TARE, the display shows the net
weight to zero; introduce the product in the box, the display shows the product weight. This
operation can be repeated several times.
While the net weight is displayed, keep pressed to display gross weight. When
the key is released the net weight will be displayed again.
The semi-automatic tare operation is not allowed if the gross weight is zero or negative or
unstable or above the maximum capacity.

- 14 -
PRESET TARE (SUBTRACTIVE TARE DEVICE)
It is possible to manually set a preset tare value to be subtracted from the display
value provided that the ≤ max capacity condition is verified. In multi-
interval instruments, the max permitted value is Max1 (max capacity of range 1).
By default the instrument shows the last programmed preset tare value: to apply it press and
then ENTER.
After setting the tare value, going back to the weight display, the display shows the net weight
(subtracting the preset tare value) and the NET symbol lights up to show that a tare has been
entered. To delete a preset tare and return to gross weight display, hold down TARE for about 3
seconds or keep the NET/GROSS input (if any) closed for the same length of time (3 seconds). The
preset tare value is set to zero. The NET symbol is turned off when the gross weight is displayed
once again.
While the net weight is displayed, keep pressed to display the preset tare. When
the key is released the net weight will be displayed again.
-IF A SEMI-AUTOMATIC TARE (NET) IS ENTERED, IT IS NOT POSSIBLE TO
ACCESS THE ENTER PRESET TARE FUNCTION.
-IF A PRESET TARE IS ENTERED, IT’S STILL POSSIBLE TO ACCESS THE SEMI-
AUTOMATIC TARE (NET) FUNCTION. THE TWO DIFFERENT TYPES OF TARE ARE
ADDED.
ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE
LOST WHEN THE INSTRUMENT IS TURNED OFF.
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing
the key for less than 3 seconds, the message is displayed for 3 seconds, by
pressing ENTER the weight is set to zero. The zero-setting will be lost upon instrument power-off.
This function is only allowed when:
-the weight is stable; otherwise the alarm appears;
-the combined effect of the semi-automatic zero and zero tracking does not exceed 2% of the
maximum capacity, otherwise the alarm appears.
0
+

- 15 -
PEAK
By keeping the PEAK input closed the maximum weight value reached remains displayed. By
opening the input the current weight is displayed.
Peak values have no legal validity.
ANALOG OUTPUT
-: it selects the analog output type (4÷20 mA, 0÷20 mA, 0÷10 V, 0÷5 V; default: 4÷20 mA).
-: choice of a weight followed by the analog output: gross () or net (). If the net
function is not active, the analog output varies according to gross weight.
- : set the weight value for which you wish to obtain the minimum analog output value.
Only set a value different from zero if you wish to limit the analog output range; for
instance: for a full scale value of 10000 kg you require an 4 mA signal at 5000 kg and
20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.
- : set the weight value for which you wish to obtain the maximum analog output value; it
must correspond to the value set in the PLC program (default: calibration full scale). E.g.: if I am
using a 4÷20 mA output and in the PLC program I wish to have 20 mA = 8000 kg, I will set the
parameter to 8000.
- : analog output correction to zero: if necessary adjust the analog output, allowing the PLC
to indicate 0. The sign “-“ can be set for the last digit on the left. E.g.: if I use a 4÷20 mA output
and, with the minimum analog setting, the PLC or tester read 4.1 mA, I must set the parameter to
3.9 to obtain 4.0 on the PLC or tester.
- : correction of analog output to full scale: if necessary permit modification of the analog
output by allowing PLC to indicate the value set in the parameter . E.g. if I am using a
4÷20 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA, I must set
the parameter to 20.1 to get 20.0 on the PLC or tester.
Minimum and maximum values which can be set for zero and full scale corrections:
ANALOG OUTPUT TYPE
Minimum
Maximum
0÷10 V
-0.150
10.200
0÷5 V
-0.150
5.500
0÷20 mA
-0.200
22.000
4÷20 mA
-0.200
22.000
+

- 16 -
NOTE: the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero ( ) may be greater than the weight set for the analog full scale
( ). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000 = 0 analog output 0÷10 V
Weight = 0 kg analog output = 10 V
Weight =5000 kg analog output = 5 V
Weight =10000 kg analog output = 0 V
All analog outputs of the instrument are ACTIVE and SINGLE ENDED type, therefore
they can be connected only to PASSIVE receiver devices. The minimum load allowed for
voltage outputs is 2 kohm, the maximum load allowed for current outputs is 400 ohm.
SERIAL COMMUNICATION SETTING
-: it disables any type of communication (default).
-: MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-: ASCII bidirectional protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
-
-: continuous weight transmission protocol (see Communication protocols manual), at the
frequency set in item (from 10 to 300).
- (set: = , = ).
- (set: = , = ).
-: continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series remote
displays; the remote display shows the net weight or gross weight according to its settings (set:
= , = , = ).
-: continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series remote
displays; the remote display shows the net weight or gross weight according to its settings (set:
= , = , = ).
ACTIVE
ANALOG
OUTPUT
PASSIVE
RECEIVER
Voltage or current analog signal
-COMMON
+
Other manuals for TLM8
1
Table of contents
Other LAUMAS Transmitter manuals