Leica MATS User manual

Leica MATS
User manual

Leica MATS – Contents
Contents
2
Page
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Assembly
The components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Use
Operating controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Switching on and off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Temperature control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Dimensions and weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

Leica MATS – Description 3
MATS = Microscope-stage Automatic
Thermocontrol System
The Leica thermocontrol system MATS is a heating system for
microscopes and stereomicroscopes and allows the viewing of
sensitive microscopic specimens under accurate temperature
conditions.
The present user manual describes the assembly and
operation of the Leica thermocontrol system MATS.
Your instrument is supplied with a printed user manual in
English.
Additional languages and information can be found on the inter-
active CD-ROM that is supplied with your Leica stereomicro-
scope or macroscope. User manuals and updates are available
for download on our homepage at www.stereomicroscopy.com.
Detailed information about the microscope, stereomicroscope
or macroscope can be found in the separate user manual to
each instrument.
Purpose
User manual
Description

Before assembling and commissioning the Leica
thermocontrol system MATS, read the present user
manual and observe all safety notes.
Adhere to general and local regulations relating to accident
prevention and environmental protection.
The Leica thermocontrol system MATS is constructed in
accordance with the latest technologies and is provided with a
statement of conformity with EC requirements.
The Leica thermocontrol system MATS is a heating system for
microscopes and stereomicroscopes and intended for viewing
sensitive microscopic specimens under accurate temperature
conditions. The thermoplate heating stage can be heated up to
50 °C.
The Leica thermocontrol system MATS consists of a control
unit, a power cable and various thermoplate stage types for
Leica stereomicroscopes (transmitted light stands) and micro-
scopes (upright & inverted). Thermoplate and control unit are
matched to each other.
Always use thermoplate and control unit from the same
packaging, never from different packagings.
If the Leica thermocontrol system MATS differently than
described in this user manual, it may lead to personal injuries or
property damage.
• Do not fit different plugs or cables.
• Do not change or convert the Leica thermocontrol system
MATS.
• Do not dismantle any part of the motor-focus system unless
instructions for doing so are given in the user manual.
• The Leica thermocontrol system MATS may not be opened
and maintained by unauthorized personnel.
Legal requirements
Conformity with
European Community
directive
Permitted uses
Prohibited
uses
4Leica MATS – Safety concept
Safety concept

5
Leica MATS – Safety concept
The Leica thermocontrol system MATS may only be used in
closed facilities. The following installation sites must be
excluded:
• Where flammable gas, corrosive gas oil mist and particles
that can deteriorate electrical insulation are generated or are
abundant.
• Where highly intense vibration or impact is generated or
transferred.
• Near high voltage power lines or where inductive interfer-
ence can affect the operation of the Leica thermocontrol
system MATS.
• Sites which create condensation water or humidity or are
exposed to direct sunlight.
The process value displayed at the control unit corresponds to
the temperature of the table surface. If the specimen in a petri
dish or on a slide glass indicates a lower temperature
compared to the displayed process value, adjust the set value
to obtain a proper specimen temperature. Otherwise, the
specimen could be damaged.
If the sensor malfunctions, the surface may become
very hot.
•Dielectric strength: 2 sec. at 1350 V AC between power
supply terminal and ground terminal
•Insulation resistance: 500 V Mega more than 100 MΩ
minimum between power supply connection and ground
connection
•Fire protection: Temperature cannot be set higher than 50 °C
•Fuse: T AC250V 1.6 A
•The Leica thermocontrol system MATS is electromagneti-
cally compatible in accordance with the directive 89/336/EEC
and can be operated at the same time as other electrical
appliances.
Place of use
Temperature control
Protective measures
of the manufacturer
Safety regulations

Ensure that
• the Leica thermocontrol system MATS is operated, main-
tained and repaired only by authorized and trained personnel.
• the operating personnel has read and understood this user
manual for the stereomicroscope or microscope and particu-
larly all safety regulations and applies them.
• Repairs may be performed only by service technicians
trained by Leica Microsystems.
• Only original Leica Microsystems spare parts may be used.
• Disconnect the power supply for maintenance and repair
work, otherwise there is a risk of injury through electric
shock.
• Only use the power cables listed on page 16.
• Inspect the power cable regularly for damage. If it is defec-
tive, replace it immediately, because it could cause the
instrument and ancillary equipment to become energized and
thereby endangering personnel.
• Observe proper cable routing so that no one can become
entangled in it or fall over it. The instrument could tip over
and be damaged.
• Avoid pulling at the power cable.
Be careful when handling liquids. Spilled liquids on electrical
devices can energize the instrument and ancillary equipment,
possibly causing personal injuries or damage to instruments.
Requirements
for the user
Repairs and servicing
Power cable
Liquids
6Leica MATS – Safety concept

7
Leica MATS – Safety concept
Cleaning, care Handle your Leica thermocontrol system MATS with care and
avoid anything that may scratch the thermoplate.
• Disconnect the power cable before cleaning the instrument.
• Never open or dismantle the control unit.
• Respect the warning signs relating to the handling of liquids.
• Do not soak into water or into solvent.
• Never use any solvents such as gasoline or thinner to wipe
the instrument. The color of the surface could change or
flake, or imprinted characters could be removed.
In case of dirt accumulation, carefully clean the control
unit and thermoplate with a soft piece of cloth moistened
with a solution of neutral cleaning agent dilluted in water.

1 Control unit
2 Thermoplate type MATS-MZ367 LT for
Leica M stereomicroscopes with transmitted light base HL
3 Thermoplate type MATS-LZ LT for Leica M stereomicroscope
with transmitted light base (lightfield/darkfield)
4 Thermoplate type MATS-URS LT for inverted microscopes
Leica DMIRB (3-plate x/y stage)
5 Thermoplate type MATS-UDM LT for inverted microscopes
Leica DMIRB & DMIL
6 Thermoplate type MATS-U55S LT with adapter for upright
microscopes, general
7 Thermoplate type MATS-STD LT standard
The components
8Leica MATS – Assembly
Assembly

9
Leica MATS – Assembly
MATS - MZ367 LT
MATS - LZ LT
MATS - URS LT
MATS - UDM LT
MATS - U55S LT
MATS - STD LT

Assembly
Remove the existing glass insert from the base of a stereomi-
croscope to reach the correct temperature.
Place the thermoplate on the object stage of the microscope/
stereomicroscope according to Fig. 1-6.
Heating stage type
MATS-MZ367 LT for Leica M
stereomicroscopes with trans-
mitted light base HL
Tighten attachment screws on
both sides of the thermoplate.
Heating stage type
MATS-LZ LT for Leica M
stereomicroscopes with
transmitted light base
(lightfield/darkfield)
Heating stage type
MATS-URS LT for
inverted microscopes
Leica DMIRB
(3-plate x/y stage)
Thermoplate
Fig. 1:
Fig. 2:
Fig. 3:
10 Leica MARTS – Installation

Leica MATS – Installation 11
Fig. 4:
Fig. 5:
Fig. 6:
Control units
Heating stage type
MATS-UDM LT for inverted
microscopes
Leica DMIRB & DMIL
Heating stage type
MATS-U55S LT for upright
microscopes, general
Attach heating stage with
adapter according to the
figure.
Heating stage type
MATS-STD LT standard
Connect the thermoplate
with the control unit.
Plug the power cable
connector into the input
receptacle of the heating
unit.
Approved power cables
see p. 16.

1 Power switch
2 Temperature control
3 Setting the temperaturs: Up/Down keys
4 PV display shows the process value and various symbols
5 SV display shows different settings values
Operating controls
12 Leica MATS – Operation
Use
4
5
2
3
1

Before commissioning the Leica thermocontrol
system MATS, read the safety regulations on p. 4-7.
Press the power switch (1).
When you finish operating, turn off the power switch and
disconnect the plug of power supply cord.
The thermoplate heating stage can be heated up to 50 °C.
The process value displayed at the control unit corresponds to
the temperature of the table surface. If the specimen in a petri
dish or on a slide glass indicates a lower temperature com-
pared to the displayed process value, adjust the set value to
obtain a proper specimen temperature. Otherwise, the spec-
imen could be damaged.
With stereomicroscopes it must be ensured that the
existing glass insert is removed from the base before the
heating stage is placed. Otherwise, the correct tempera-
ture is not guaranteed.
To adjust the temperature, press the Up/Down keys. When the
key is pushed for 1 sec or more, the indicated temperature
changes continuously. 2 seconds after above adjustment, the
temperature controller starts operating to obtain set tempera-
ture.
Set the temperature control to the desired temperature using
the keys (2).
Wait 10 minutes until the temperature of the thermoplate
stabilizes.
Switching on and off
Control of the
Temperature
Setting the
temperature
13
Leica MATS – Operation

Control unit
– Weight: about 1.3kg
– Dimensions: width 113 mm, length 210 mm, height 128-74 mm
Thermoplate type MATS-MZ367 LT for Leica M stereomicro-
scope with transmitted light base HL
– Weight: about 420 g
Thermoplate type MATS-LZ LT for Leica M stereomicroscope
with transmitted light base (lightfield/darkfield)
– Weight: about 540 g
Thermoplate type MATS-URS LT for inverted microscopes
Leica DMIRB (3-plate x/y stage)
– Weight: about 100 g
Thermoplate type MATS-UDM LT for inverted microscopes
Leica DMIRB &DMIL
– Weight: about 80 g
Thermoplate type MATS-U55S LT for upright microscopes,
general
– Weight: about 100 g
Thermoplate type MATS-STD LT standard
– Weight: about 300 g
14 Leica MATS – Dimensions and weight
Dimensions and weight

15
Leica MATS – Dimensions and weight
202
174
257
265
10
MATS - MZ367 LT
145
113
165
61
100
80
6
116
80
142
115
ø92
5
205
180
7
160
185
MATS - LZ LT
MATS - STD LT
MATS - U55S LT
MATS - URS LT
MATS - UDM LT
146
220
196
332 10
3
132
150
132
150
Dimensions in mm

Leica thermocontrol system MATS
100-240 V ±10 % AC, 50/60 Hz, 0.68 A, protection class I
– indoor use only
– temperature: 5 °C to 40 °C
– maximum relative humidity: 35% to 80%
– altitude: up to 2000 m max.
– environmental conditions: installation category II according
to IEC 664, degree of pollution 2
For mains supplies with 100 to 120 V, use only the following
power cables:
– UL listed, detachable cord set, 3 conductor grounding type
SVT No.18 AWG rated at 125 V, 7 A minimum
– if an extension cord is used, use only a power supply cable
with PE (protective earth) connector
For mains supplies with 220 V to 240 V:
– inside the EU, use only 3-wire power cable, plug and
connector
– according to EU/EN standards
– Class 1 equipment should be connected to PE
(protective earth) terminal
– if an extension cord is used, use only a power supply cable
with PE (protective earth) connector
Power supply
Operating Environ-
mental Conditions
Power cable
16 Leica MATS – Technical Data
Technical Data

– method: The temperature control is based on the
PID controller with solid state relay
– achieved accuracy: T (actual temperature) in the middle of
the heating stage lies within a range of [t > T > t-1] °C,
whereby t represents the set temperature (under the
condition that the set value t = 37 °C)
– step: 0.1 °C
– setting method: with Up and Down key
– setting range: room temperature to 50 °C
– achievable temperature accuracy: ±0.3 °C
(with indicator temperature)
– sensor: thermocouple
– connection to heating stage: 4-pole connector
(cable length: 1000 mm)
– instructions method: 7-segment digital display and
single-lighting indicator
– step: 0.1 °C
– indicating accuracy: ±0.5%
50 °C within 10 minutes
Control of the
Temperature
Temperature Indicator
Duration
17
Leica MATS – Technical Data

18 Leica MATS – User Manual
Notes

19
Leica MATS – User Manual
Notes

Illustrations, descriptions and technical data are not binding and may be changed without notice.
M2-227-0en • © Leica Microsystems (Switzerland) Ltd • CH-9435 Heerbrugg, 2002 • Printed in Switzerland – 10.6.2002 – RDV
Leica Microsystems – the brand
for outstanding products
Leica Microsystems – an international company
with a strong networ of customer services
Australia: Gladesville, NSW Tel. +1 800 625 286 Fax +61 2 9817 8358
Austria: Vienna Tel. +43 1 486 80 50 0 Fax +43 1 486 80 50 30
Canada: Richmond Hill/Ontario Tel. +1 905 762 20 00 Fax +1 905 762 89 37
China: Hong Kong Tel. +8522 564 6699 Fax +8522 564 4163
Denmark: Herlev Tel. +45 44 5401 01 Fax +45 44 5401 11
France: Rueil-Malmaison
Cédex Tel. +33 1 4732 8585 Fax +33 1 4732 8586
Germany: Bensheim Tel. +49 6251 1360 Fax +49 6251 136 155
Italy: Milan Tel. +39 02 57 486 1 Fax +39 02 5740 3273
Japan: Tokyo Tel. +81 3 543 596 09 Fax +81 3 543 596 15
Korea: Seoul Tel. +82 2 514 6543 Fax +82 2 514 6548
Netherlands: Rijswijk Tel. +31 70 41 32 130 Fax +31 70 41 32 109
Portugal: Lisbon Tel. +35 1 213 889 112 Fax +35 1 213 854 668
Singapore: Tel. +65 6 77 97 823 Fax +65 6 77 30 628
Spain: Barcelona Tel. +34 93 494 9530 Fax +34 93 494 9532
Sweden: Sollentuna Tel. +46 8 625 45 45 Fax +46 8 625 45 10
Switzerland: Glattbrugg Tel. +41 1 809 34 34 Fax +41 1 809 34 44
United Kingdom: Milton Keynes Tel. +44 1908 666 663 Fax +44 1908 609 992
USA: Bannockburn/Illinois Tel. +1 800 248 0123 Fax +1 847 405 0164
and representatives of Leica Microsystems
in more than 100 countries.
Leica Microsystems’ Mission is to be the world’s first-choice provider of innovative
solutions to our customers’ needs for vision, measurement, lithography and analysis
of microstructures.
Leica, the leading brand for microscopes and scientific instruments, has developed
from five brand names, all with a long tradition: Wild, Leitz, Reichert, Jung and
Cambridge Instruments. Leica symbolizes not only tradition, but also innovation.
The companies of the Leica Microsystems
Group operate internationally in five business
segments, where we rank with the market
leaders.
Microscopy
Our expertise in microscopy is the basis for all
our solutions for visualization, measurement
and analysis of microstructures in life sciences
and industry.
Specimen Preparation
We specialize in supplying complete solutions
for histology and cytopathology.
Imaging Systems
With confocal laser technology and image
analysissystems,weprovidethree-dimensional
viewing facilities and offer new solutions for
cytogenetics, pathology and material sciences.
Medical Equipment
Innovative technologies in our surgical micro-
scopes offer new therapeutic approaches in
microsurgery. With automated instruments for
ophthalmology, we enable new diagnostic
methods to be applied.
Semiconductor Equipment
Our automated, leading-edge measurement
and inspection systems and our E-beam
lithography systems make us the first choice
supplier for semiconductor manufacturers all
over the world.
The Business Units in Leica Microsystems hold the management system certificates
for the international standards ISO 9001 and ISO 14001 relating to quality management,
quality assurance and environmental management.
Leica Microsystems (Switzerland) Ltd
Business Unit SM
CH-9435 Heerbrugg
Telephone +41 71 726 33 33
Fax +41 71 726 33 99
www.leica-microsystems.com
www.stereomicroscopy.com
Table of contents
Popular Heating System manuals by other brands

Vents
Vents VUT 300 E2V EC user manual

Innova
Innova HRW DOMO Installation, use and maintenance

Vent-Axia
Vent-Axia HR200WJ Installation and maintenance instructions

Viessmann
Viessmann VITOTRONIC 100 operating instructions

Herz
Herz firematic PELLET 120 operating instructions

Aqua-Hot
Aqua-Hot 200 Series installation manual

Gree
Gree Ultra Heat GMV Multi VRF Heat Recovery... Service manual

Lennox
Lennox CompleteHeat AM30 User's guide & installation instructions

therm-ic
therm-ic Integration Mono button BLE manual

widmer
widmer 93485 operating instructions

Greentek
Greentek LCH-4H Installation, operation and maintenance manual

flowair
flowair LEO D S Technical documentation operation manual