Switches instrument on/off.Long key press Over 2 Seconds while
powering on is to activate the instrument without Auto-off function.
Switches backlighting on/off
Short key press to display reference level of present test wavelength. Long
key press to set a new reference level of preset test wavelength.
Short key press selects measurement wavelengths.
Long key press SELECTS measurementAWI
Switches measurement units among dBm,dB and mw
Switches instrument on/off.Long key press Over 2 Seconds while
powering on is to activate the instrument without Auto-off function
Short key press selects measurement wavelengths.
Modulated Wavelength Shifting Key:Switches modulated wavelength
and continuous wavelength
Long key press SelectsAWI
Maintenance:
1、please disconnect the AC adapter/charger and cover the protective dust cap
once you finish using.
2.It is a good idea to clean the connector and the instrument when they get dirty
through use.Optical cleaning Pads and anhydrous alcohol is recommended.And
Please be careful not to get the detergent inside the instrument.
3.To ensure the measurement accuracy,please send the instrument to Service
Center for calibration once a year.
User self calibration:
1、Open state At the same time Press the "λ" and "dBm/dB/mw" keys,Long key
press Over 2 Seconds while
2、Press the“dBm/dB/mw”keys +0.1dBm;Press the“REF”keys -0.1dBm; If the
calibration is correct Press the switchkey to restart ,Restore normal test state.
(Note: use the fiber to connect with optical power meter to calibrate the value
until getting the desired value for calibration)