V. Precaution
1. When the meter can’t complete power on self-calibration, the interface will prompt “E30”. Possible
causes:
A. The ambient light is too strong.
B. Test sample is placed in the test slot, just take away the test sample and turn on the meter
again.
C. Dust enters the test hole and blown off the dust with a dust blower.
D. The meter is faulty and needs to be returned to the factory for maintenance.
2. When the meter is not in use, turn off the power.
3. Avoid to contact with corrosive materials and be away from high temperature and high humidity
environment.
4. Place the test sample in the test slot and the sample should be as close as possible to the left side
of the test slot.
5. After using for a long time, due to the luminous efficiency of the LED light source (the temperature of
the LED light source increases, the efficiency decreases), the data of UVR, IRR and VLT may not be
displayed as “0.0, 0.0, 100” when there is no test sample. At this time, turn off the meter and then
the meter will restart and self-calibrate. The measurement accuracy and normal use of the meter will
not be affected.
VI. Standard packing list
No. Description Quantity Unit
1 Solar Film Transmission Meter 1 set
2 User Manual 1 pcs
3 AA Battery 4 pcs
4 Certificate / Warranty Card 1 pcs
VII. Service
1. The meter has one-year warranty. If the meter works abnormally, please send the whole meter to
the company for maintenance.
2. Provide users with spare parts and lifelong maintenance services.
3. Provide the users with the meter inspection service for free.
4. Free technical support for long term.