Shenzhen Linshang Technology Co., Ltd.
2020-10-26 Page 5 of 8
In the “STOP” interface, long press the “POWER” button to power off the meter.
2) In the “STOP” interface, the meter will automatically power off in 3 minutes without any operation.
3) In the automatic trigger measurement “Ready” state, the longest waiting time is 50 minutes. If the
measurement cannot be triggered within 50 minutes, it will automatically turn off.
3. Measurement
In the measurement mode, three operation states exist:
READY: Ready state; in the auto trigger mode, this means the meter is waiting for triggering to begin a
measurement.
RUN: Measuring state; this means the meter is measuring data.
STOP: Measurement is over.
In measurement interface:
M: Manual measurement mode
A: Auto measurement mode
1) Manual measurement mode: in the STOP interface. Short press the "POWER" button and then short
press the "SELECT" button to confirm the new measurement, clear the historical data (display "-----"),
automatically enter the measurement interface (RUN state) after 1s. The interface dynamically
displays energy E, real-time power value P, real-time temperature value T, measurement time. Short
press the "POWER" button or after 32 minutes, themeasurement is over and enter the STOPinterface.
The interface displays the measurement result energy E, maximum power value PMAX, maximum
temperature value TMAX and measurement time.
2) Automatic mode: After pressing the button to confirm the new measurement, the meter enter the
READY state to clear the historical data and wait for the trigger condition (Trigger power) to be met,
the energy display "----" and flashing. When the power value is greater than the set trigger power, the
trigger condition is met, enter the measurement interface (RUN state). The interface dynamically
displays energy E, real-time power value P, real-time temperature value T and measurement time.
When the recording time reaches 32 minutes or the power value is less than the set trigger power, the
measurement is automatically ended and enters the STOP interface. The interface displays the
measurement result energy E, maximum power value PMAX, maximum temperature value TMAX and
measurement time.