LogTag Recorders TREL30-16 Specification sheet

www.logtagrecorders.com
TREL30-16
Ultra-Low
Temperature
Recorder with
Display
Product User Guide
Document Release Version: 1.1
Published 22. November 2020
Copyright © LogTag Recorders, 2004-2020
LogTag Recorders

TREL30-16 User Guide 2
Contents
Introduction 6
Features 6
Case 6
Buttons 7
Display 7
External Probes 7
What You Need 8
Required Equipment 8
Optional Items 8
Configuring the TREL30-16 for logging 9
Standard Configuration Options 10
TREL30-16 Start Options 11
Audible Alarm 11
Alarm Configuration Options 13
Advanced Configuration Settings 14
Display Overview 16
Real Time Clock 18
Adjusting the Display Clock 18
Starting the Logger 19
Push button start 19
Push Button Start with Start Delay 21
Automatic date/time start 21
During Recording 23
Marking a reading with an inspection mark 24
Clearing an Alarm 25
Paused Readings 25
Probe is disconnected 26
Power Save 27
Total Number of Days Statistics Collected 28

TREL30-16 User Guide 3
Reviewing Day Statistics directly on the TREL30-16 29
Trip Minimum/Maximum Temperatures 33
Resetting the Trip Minimum/Maximum Temperatures 33
Stopping the TREL30-16 35
Automatically 35
Push-button stop 35
Resetting the Logger 37
Hibernating a TREL30-16 37
Technical Specifications 40
Appendix 1 - Battery Replacement 42
Appendix 2 - Glossary 43

TREL30-16 User Guide 4
Safety Information
The TREL30-16 temperature logger contains a Lithium Battery.
When the battery indicates “LOW”, you can replace it as per the
instructions in Battery Replacement on page42.
The empty battery should be recycled or disposed of according to
your local regulations.
Do not expose the logger to extreme temperatures as it may lead to
the destruction of the battery and may cause injuries.
Keep out of the reach of children.
Liability
LogTag Recorders’ standard warranty terms apply. A copy can be requested by emailing
support@logtagrecorders.com. In addition, LogTag Recorders shall not be held liable:-
lif the device was used beyond LogTag Recorders’ stated limitations
lfor any claims due to the improper storage and use of the device
lfor any problems with refrigeration units
lfor the bad quality of the monitored goods, if any
lfor incorrect readings if the device was used with a low battery
lfor consequential loss
Battery Life
The main battery in the TREL30-16 is designed to power the device for up to 12 months of
operation, provided:-
lthe device was not stored for more than 12 months prior to activation
la fresh battery from a reputable manufacturer is used
lthe device is not downloaded excessively to a PC (more than once a week)
lthe recording interval is not shorter than 5 minutes
lthe acoustic alarm is not active over long periods
lthe device is stored and operated according to LogTag Recorders’ recommendations
Disclaimer
The TREL30-16 monitors temperature exposure and not the quality of the goods it
accompanies. Its purpose is to signal if product quality evaluation/testing is required.

TREL30-16 User Guide 5
Typographical Conventions
Text in this font refers to buttons on the TREL30-16.
Text in this font refers to option settings, dialogue boxes or actions to be taken in
LogTag® Analyzer.
Text in this font describes features of the product.
This text describes certain aspects of the product, where incorrect use of a feature
may lead to inadvertent loss of data.
This text contains important information for the correct operation of
your TREL30-16.
This text contains information that explains some aspects of a feature in more
detail.
This text contains tips that help you get the best out of your TREL30-16 logger

Introduction TREL30-16 User Guide 6
Introduction
The LogTag® TREL30-16 low temperature logger features a data logging memory storing
up to 15,905 temperature readings and a separate statistical memory, storing maximum
and minimum reading as well as alarm duration for each of the last 30 days.
During recording the display shows the temperature of the most recent reading, whether
or not this is within or outside the acceptance range, an alarm trigger summary of the last
30 days (today and 29 days previous), the current time and battery status.
Alarm events can be triggered when a number of readings are outside preset Alarm
thresholds and a “day alarm indicator” appears on the display.
Logged temperature data can be downloaded via a standard LogTag® Interface to the free
companion software LogTag® Analyzer, where you can display data in chart, list or
summary formats. The software also allows electronic archiving and exporting or
transmitting data in support of sophisticated data management systems, including
LogTag®Online.
Features
The TREL30-16 temperature logger features the familiar LogTag® case layout.
Figure 1: TREL30-16 features
Case
lMounting lug for secure fastening of logger to fixtures
lGold-plated, high-quality temperature external probe socket
lRobust polycarbonate case, IP61
lDurable communications contacts

Features TREL30-16 User Guide 7
Buttons
lSTART/CLEAR/STOP button; can be used to start and stop the unit or to clear an alarm. It
is also used to exit the statistics review.
lREVIEW/MARK button; can be used to enter the statistics review and to scroll through
the statistical data directly on the display. It is also used to place an inspection mark
in the data listing.
Display
The display shows ‘at a glance’ if alarm events have occurred for both the current day and
up to 29 days in the past. Details of any alarm event can be checked directly on the unit by
inspecting the statistics history on the logger's display or in more detail by downloading
the logged data.
External Probes
The TREL30-16 will accept any external probe from the ST10 product range.

What You Need TREL30-16 User Guide 8
What You Need
Required Equipment
In addition to your LogTag® TREL30-16 temperature logger you will need the following
items:
lA USB interface for communication to the PC
lAn external probe of the ST10 series
lFor configuration - a PC running Windows 7 or later and LogTag® Analyzer installed
Optional Items
In addition to the above, following items are useful accessories:
lA wallmount bracket, which can be used, for example, to attach the TREL30-16 to
the side of a cabinet
lA silica sand temperature buffer, which simulates environmental behaviour of a
vaccine vial

Configuring the TREL30-16 for logging TREL30-16 User Guide 9
Configuring the TREL30-16 for logging
Before a TREL30-16 logger can be deployed, it must be configured with the parameters
required for starting and recording temperature values.
Loggers can be purchased unconfigured, or pre-configured, ready to be started, using one
of a number of different profiles that are available.
You can check if your logger is already configured or not by briefly pressing the
REVIEW/MARK button.
The logger is already configured to start
If your logger is configured to start it will show either
if the logger was pre-configured for a push-button start, or
if the logger was pre-configured for a Date/Time start.
If your logger shows one of these screens you may skip the configuration process.
The logger is unconfigured
An unconfigured TREL30-16 logger is delivered to you hibernated (i.e. in a state of low
power consumption), and pressing the REVIEW/MARK shows:
and then

Standard Configuration Options TREL30-16 User Guide 10
If your logger is unconfigured you must set it up with the parameters required for:
lstarting and recording temperature values
ltriggering alarms
This is done using LogTag® Analyzer software, which can also be used for downloading and
analyzing data.
Standard Configuration Options
The standard configuration options include settings such as User ID, start method, pre-
start recording, logging interval and duration, start delay, and password.
For detailed information about each parameter please read the section about Configuring a
LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.

Standard Configuration Options TREL30-16 User Guide 11
TREL30-16 Start Options
During configuration with LogTag® Analyzer you can decide when the TREL30-16 starts
taking temperature readings:
Push button start
The logger will start taking temperature readings as soon as you have pressed the
START/CLEAR/STOP button (see Starting the Logger on page19).
When you choose the push button start option, you can Enable pre-start logging (or
disable it) and also Begin recording after a delay.
Pre-start readings
If you enable pre-start readings, the TREL30-16 starts recording as soon as it is configured
and will continue to do so until you start the unit via the button. No alarms are processed
while pre-start readings are being taken. Using pre-start readings is a good way to avoid
data loss if you forget to start the unit, as you can still access the data using LogTag®
Analyzer.
Start delay
If you configure the TREL30-16 to start after a delay period, the logger will not
immediately record temperature readings after you have pressed START/CLEAR/STOP, but
start a countdown timer instead, and record readings only after the timer has ended. The
value for the timer is set during configuration. If pre-start readings are enabled, these will
continue to be recorded during the delay period.
Date/Time start
The logger will start taking temperature readings at the date and time you enter during
configuration (local time). You cannot combine a date/time start with pre-start readings or
the start delay function.
Audible Alarm
The TREL30-16 is fitted with a buzzer. You can choose to activate the buzzer when an
alarm event has been triggered to provide extra feedback. This is enabled or disabled in the
Advanced Settings when configuring the logger with LogTag® Analyzer.
Please note, that continual activation of the audible alarm will reduce the working
life of the battery. When an alarm event is triggered, the alarm should be cleared
as soon as possible.

Standard Configuration Options TREL30-16 User Guide 12
The alarm will sound once every four seconds for the first 24 hours, then sound less
frequently to preserve battery life1 until the alarm is cleared, the unit stops or is re-
configured.
Note: The interval will increase, even if the alarm is re-triggered!
Please see more information about re-triggering alarms in
The buzzer will temporarily turn off when you are reviewing data.
1 The interval at which the alarm sounds will increase to 8 seconds when the clock passes midnight for a second time (i.e. the
alarm will sound every 4 seconds between 24 and 48 hours, depending on when during the day the alarm was first triggered).
When the clock passes midnight a third time, the interval changes to 12 seconds until the alarm is cleared.

Standard Configuration Options TREL30-16 User Guide 13
Alarm Configuration Options
The TREL30-16 can display an alarm if either of the configured alarm trigger conditions
have been met. This is indicated on the display by showing the Alarm Indicator ( ) and the
Day Alarm Marker for
today ( ).
Each alarm trigger condition consists of a threshold temperature value, an activation type
(which can be instant, consecutive or accumulative2) and a delay time, if it is not an
instant alarm.
If an alarm trigger condition requires readings to exceed an upper threshold temperature it
is called an
upper alarm
. If an alarm trigger condition requires readings to go below a
lower threshold it is called a
lower alarm
.
All alarm trigger conditions are configured in the Alarm Settings tab during configuration
of the logger with LogTag® Analyzer.
The screen shows an example where:
lthe upper alarm is triggered when the temperature is -70.0°C or above for an
accumulative time of 10 hours.
lthe lower alarm is triggered when the temperature is -80.0°C or below continuously
for 1 hour.
Once an alarm has triggered, the alarm indicator ( ) remains shown until the alarm is
cleared (see Clearing an Alarm on page25) or the unit is reconfigured. The day alarm
marker remains shown until midnight, then it turns off and the marker for the
2
lInstant = one temperature reading is above (below) the threshold
lConsecutive = temperature readings are above (below) the threshold for the time defined in the activation delay without
interruption
lAccumulative = temperature readings are above (below) the threshold for the total time defined in the activation delay
time, but may not necessarily be sequential

Standard Configuration Options TREL30-16 User Guide 14
previous day is shown ( ) to indicate the alarm was registered against what is now the
previous day. When midnight passes next, this marker will move to and so on.
Alarm Re-triggering
A note on alarm re-triggering: As soon as an alarm is triggered, the corresponding
delay time (but not any others) resets to zero and alarm processing starts again.
The alarm processing for all other alarm delays is not affected. Therefore,
accumulative or consecutive alarms will re-trigger, if the alarm conditions are met
again, and the Alarm Indicator ( ) and the Day Alarm Marker ( ) will be shown,
even if any previous alarm was cleared. Clearing an Alarm does not reset any of
the delay values.
Advanced Configuration Settings
Select Advanced Settings for additional configuration settings. These settings decide how
some of the elements are displayed on the unit's own display and set certain options
specific to the TREL30-16.
Figure 2: TREL30-16 advanced configuration screen in LogTag® Analyzer 3
Parameters that influence the appearance of the display are:
lPause alarm/statistics processing (see Paused Readings on page25)
lTemperature display unit (see Display Overview on page16)
lSwitch off display after 30 seconds (Power save, see Power Save on page27)
lshowing the number of days on the display, for which statistics were collected (see
Total Number of Days Statistics Collected on page28)
lallowing a user to reset the trip's minimum and maximum values on the display
during recording (see Resetting the Trip Minimum/Maximum Temperatures on
page33)
Parameters influencing specific behavior of the TREL30-16 are:
lclearing and resetting alarms when the START/CLEAR/STOP button is pressed (see
Clearing an Alarm on page25)

Standard Configuration Options TREL30-16 User Guide 15
lleaving the alarm turned on, even if readings return to the normal temperature range
again
lallowing the user to stop the logger with the REVIEW/MARK button
lallowing the user to reset the logger with the START/CLEAR/STOP button
lenabling the buzzer for the audible alarm
For detailed information about each parameter please read the section about Configuring a
LogTag® for logging in LogTag® Analyzer’s User Guide or press F1 for help.

Display Overview TREL30-16 User Guide 16
Display Overview
Figure 3: Display Overview with all segments turned on
Day Alarm Markers
This grid shows 3 rows of 10 markers, named
Today
( ) to
Day -29
( ), which are switched on
when an alarm event occurred on that day.
Temperature Value
This shows the most recently recorded temperature while the TREL30-16 is recording. Once the
logger has stopped, nothing will be displayed. During review, this will show minimum or maximum
temperatures.
Day Number
During Review, this shows the day number of the currently displayed day statistic. Today is 00
DAYS, days in the past are represented between yesterday -01 DAYS and -29 DAYS. During
configuration of the logger in LogTag® Analyzer you can also enable this to show the total number
of days for which statistics are available.
Battery Low
A battery test is performed hourly. The battery low symbol will appear if the TREL30-16's
battery is low and requires changing. Please follow the instructions provided in Battery
Replacement on page42. If the symbol is not shown while the display is turned on, the battery is
still OK.
ALARM/OK indicator
The symbol is shown as soon as the TREL30-16 has registered an alarm event.While there are
no alarms, the symbol is shown.
Recording indicators
The recording indicators show what the TREL30-16 is currently recording.

Display Overview TREL30-16 User Guide 17
lIf READY is shown, the TREL30-16 is ready to be started with the START/CLEAR/STOP button.
Depending on the configuration it may already record pre-start readings.
lIf STARTING is shown, the logger has been started, and a start delay is active. The word
DELAY is also shown, together with the time in hours and minutes until the start.
lIf is shown, the TREL30-16 is recording temperature at the sample interval defined
during configuration.
lIf is shown together with the word PAUSED, the product is also recording, but the
recorded values are not taken into account when calculating alarm events and durations.
lIf the word STOPPED is shown, the TREL30-16 has finished recording temperature data.
Time Value and Time Indicators
The time value display is used to show one of the following:
lThe current time
lThe time remaining until the logger starts recording (for a delayed start)
lA duration, for example of an alarm
The time indicators identify which of those is displayed as follows:
lIf TIME is shown, the time value represents the current time in hours and minutes (24- hour
format).
lIf DELAY is shown, the time value represents a start delay, or the time remaining until a
date/time start will occur.
lIf DURATION is shown, the time value represents the duration, for example the time above
the upper alarm limit.
The word boot appears in place of the clock if new firmware is being uploaded to the TREL30-16.
The word USb appears when the unit is plugged into a USB port.
The word dtstappears in place of a time value if the logger is configured for a date/time start, and
the start time has not yet passed.
Reading Type
The word CURRENT is shown when the temperature on the display represents the last recorded
temperature.
The word MAX is shown in Review mode, when the temperature on the display represents the
maximum recorded temperature for the day displayed.
The word MIN is shown in Review mode, when the temperature on the display represents the
minimum recorded temperature for the day displayed.
Above/Below Threshold Arrows
The up-arrow is shown when the temperature displayed (i.e. last recorded) is above the
specified upper temperature threshold. The down-arrow is shown when the temperature
displayed is below the specified lower temperature threshold.
Temperature Units
Depending on the selected PDF temperature units, this shows either°F or°C.

Real Time Clock TREL30-16 User Guide 18
Real Time Clock
The time shown on the recording display is linked to the logger’s internal real time clock.
A day change occurs when the display time rolls through midnight (i.e. 00:00), which
triggers the statistical data to be finalized for the day, and a new day to be started when
the next reading is taken.
Each time the logger is configured with LogTag® Analyzer the display clock value is set to
the PC's current local time (or timezone).
Note that the logger’s internal real time clock value is only updated when the
recorder is configured with LogTag® Analyzer. This prevents the data logging
becoming discontinuous, which would be the case if the real time clock were to be
changed together with the display clock.
LogTag® Analyzer can display the logged readings in a number of different time zones,
regardless of where it was configured.
Adjusting the Display Clock
The display clock of the logger can be set to the current local time, either by using
LogTag® Analyzer software or directly on the unit, using the buttons.
To set the clock, press and hold the START/CLEAR/STOP ( ) button, then immediately press
and hold the REVIEW/MARK ( ) button. Keep holding both buttons together
continuously for a period of 8 seconds.
Press the buttons firmly!
During this period the CLOCKADJ icon flashes. Release the buttons when the flashing
stops. The clock can now be adjusted.
Initially, the minutes digits flash.
Press the button to increment the minutes digit (once
it reaches 59 it rolls back to 00 on the next press).
Accept the minutes value by pressing . The hours
value now flashes.
Press the button to increment the hours digits (once
it reaches 23 it rolls back to 00 on the next press).

Starting the Logger TREL30-16 User Guide 19
Accept the hours value by pressing . The new clock
value is now stored, and the display shows the normal
screen.
NOTE: It is advised that the real time clock is only adjusted when the product is
not recording (STOPPED or READY) to avoid large shifts in day boundary data.
If a display clock adjustment is made while recording data, the next log taken will
be identified in the downloaded data with a time change mark.
Changes to the display clock do not affect the internal real time clock value, so the logged
data does not show time gaps.
Starting the Logger
Push button start
The logger's display must show READY for it to be started. The current time is also shown.
Press and hold the START/CLEAR/STOP button. First, STARTING is shown in addition to
READY:
Then READY disappears:

Starting the Logger TREL30-16 User Guide 20
Once READY disappears, release the button within two seconds. STARTING will also
disappear, and the symbol will be shown. The TREL30-16 now records temperature
data.
The logger will
not
start if you:
lRelease the button before READY disappears
lKeep holding the button for more than 2 seconds after READY disappears
Table of contents
Other LogTag Recorders Data Logger manuals

LogTag Recorders
LogTag Recorders UTRED30-16-WiFi Specification sheet

LogTag Recorders
LogTag Recorders UTREL30-WiFi Specification sheet

LogTag Recorders
LogTag Recorders USRIC-8 User manual

LogTag Recorders
LogTag Recorders UTRED30-WiFi Specification sheet

LogTag Recorders
LogTag Recorders UTRED30-16 Specification sheet

LogTag Recorders
LogTag Recorders UTREL-16 Specification sheet

LogTag Recorders
LogTag Recorders UTREL30-16 Specification sheet

LogTag Recorders
LogTag Recorders USRIC-8 Specification sheet

LogTag Recorders
LogTag Recorders TRIX-8 User manual

LogTag Recorders
LogTag Recorders TRID30-F User manual