Lovibond ORPTester10 User manual

Before you begin:
To condition electrode, immerse electrode in electrode storage solution or tap water for at least
30 minutes before use. DO NOT use de-ionized water.
Calibration
Calibration is not necessary unless exact readout agreement with a work standard and
at a specific ORP value is needed. To achieve this, use the following procedure:
1. Switch unit on (ON/OFF button)
2. Select a known solution sample from the actual process as near the critical ORP value as
possible, for example, 470 mV.
3. Dip the conditioned electrode into this solution for 2 to 5 minutes until the readings stabilize.
Ensure the whole sensor is immersed.
4. Press CAL button to enter Calibration mode. The screen will switch to a dual display with the
upper display flashing the relative mV value to be calibrated and the lower display showing the
default measurement value of the sample. The ‘CAL’ indicator will also be shown on the screen
replacing ‘MEAS’ to indicate that the tester is in calibration mode.
Note: To abort calibration, press the ‘CAL’ button to return to the measurement mode.
5. Immediately press and hold the HOLD/ENT button to scroll the reading to match value of
the known solution sample. The offset adjustment is up to +/- 150mV from the default mea-
surement.
6. To complete the calibration, release the HOLD/ENT button and wait for 5 seconds for automatic
confirmation. The primary display will stop flashing and the secondary display will automatically
show ‘CO’ before the screen return to the single display measurement mode. The mV annun-
ciator will now changed to ‘R.mV’ indicating the value displayed is a relative mV value of the
default measurement.
7. Rinse electrode in tap water before measuring other solutions.
ORP Measurement
1. Remove cap from the electrode and press the ON/OFF button to switch the tester on.
2. Dip the electrode about 2 to 3 cm into the test solution. Stir once and let the reading stabilize
for 2 to 5 minutes.
3. Note the ORP value or press HOLD/ENT button to freeze the reading. Press HOLD/ENT again
to release the reading.
4. Press ON/OFF to turn off tester. If you do not press a button for 8.5 minutes, the tester will
automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation
1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be displayed and
the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be displayed
anymore indicating the held measurement is released.
User Reset
You can reset the ORP calibration to the factory default by using the user reset function.
1. Switch off the tester.
2. While pressing the ‘Cal’ button, press and release the ON/OFF button to enter the ‘User
Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with a flashing ‘nO’
selection.
3. Use the ‘CAL’ button to toggle between ‘nO’ and ‘YES’ selection.
• nO deactivates reset selection
• YES activates the reset selection
4. Press the HOLD/ENT button to confirm the selection made.
Electrode Maintenance
1. Rinse the electrode with tap water or electrode storage solution after each measurement.
2. In aggressive chemicals, dir ty or viscous solutions, and solutions with heavy metals or proteins,
take readings quickly and rinse electrode immediately afterward.
3. If possible, keep a small piece of paper or sponge in the electrode cap – moistened with
clean water or electrode storage solution (NOT de-ionized water) – and close the cap over the
electrode.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity (shown in diagram below).
GB
• Large Screen
• Waterproof
ORP / Redox
Tester
• Double
Junction
Instruction
manual
Redox/ORP 10
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new tester.
When the tester fails to calibrate or gives fluctuating readings in calibration standards, you need
to change the electrode.
1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the collar counter
clockwise (see picture A). Save the ribbed tester collar and O-ring for later use.3. Align the four
tabs on the new module so that they match the four slots on the tester (see picture B).
2. Pull the old electrode module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on the tester (see
picture B).
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto
the new electrode module. Push the collar over the module and thread it into place by firmly
twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after an electrode
replacement.
Applications
Water quality testing • pools • spas • cooling towers • water purication • chromate reduc-
tion • cyanide oxidation • boilers • pulp bleaching • ozone systems • water and wastewater
treatment • and more!
Warranty
The waterproof ORPTester10 are warranted to be free from manufacturing defects for 2 years
and electrode module for 12 months. If repair, adjustment or replacement is necessary and has
not been the result of abuse or misuse within the time period, please return the tester – freight
prepaid – and correction will be made without charge. Out of warranty products will be repaired
on a charge basis.
Return of Items
Authorization must be obtained from your distributor before returning items for any reason.
When applying for authorization, please include information regarding the reason the item(s)
are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance of
products without notice. Prices are subject to change without notice.
Accessories:
19 46 66 Replacement sensor MicroDirect ORP10

This manual suits for next models

1

Other Lovibond Measuring Instrument manuals

Lovibond SD 60 User manual

Lovibond

Lovibond SD 60 User manual

Lovibond SWIMTEC MD100 User manual

Lovibond

Lovibond SWIMTEC MD100 User manual

Lovibond MD200 User manual

Lovibond

Lovibond MD200 User manual

Lovibond Tintometer MD100 User manual

Lovibond

Lovibond Tintometer MD100 User manual

Lovibond MD 610 User manual

Lovibond

Lovibond MD 610 User manual

Lovibond SensoDirect Oxi200 User manual

Lovibond

Lovibond SensoDirect Oxi200 User manual

Lovibond EC 2000 Gardner Assembly instructions

Lovibond

Lovibond EC 2000 Gardner Assembly instructions

Lovibond Tintometer MD100 User manual

Lovibond

Lovibond Tintometer MD100 User manual

Lovibond RT300 User manual

Lovibond

Lovibond RT300 User manual

Lovibond Photometer-System MD100 User manual

Lovibond

Lovibond Photometer-System MD100 User manual

Lovibond SD 400 Oxi L User manual

Lovibond

Lovibond SD 400 Oxi L User manual

Lovibond MD200 User manual

Lovibond

Lovibond MD200 User manual

Lovibond SensoDirect pH110 User manual

Lovibond

Lovibond SensoDirect pH110 User manual

Lovibond TR 515 Assembly instructions

Lovibond

Lovibond TR 515 Assembly instructions

Lovibond Fx Assembly instructions

Lovibond

Lovibond Fx Assembly instructions

Lovibond TB300 IR User manual

Lovibond

Lovibond TB300 IR User manual

Lovibond MD200 User manual

Lovibond

Lovibond MD200 User manual

Lovibond Photometer-System MD100 User manual

Lovibond

Lovibond Photometer-System MD100 User manual

Lovibond MD200 User manual

Lovibond

Lovibond MD200 User manual

Lovibond SD 300 pH User manual

Lovibond

Lovibond SD 300 pH User manual

Lovibond SD 90 User manual

Lovibond

Lovibond SD 90 User manual

Lovibond Scuba User manual

Lovibond

Lovibond Scuba User manual

Lovibond RT 400 User manual

Lovibond

Lovibond RT 400 User manual

Lovibond MD 100 User manual

Lovibond

Lovibond MD 100 User manual

Popular Measuring Instrument manuals by other brands

Vericom Profile X1 user guide

Vericom

Vericom Profile X1 user guide

Simpson Falcon F45 Series manual

Simpson

Simpson Falcon F45 Series manual

PASCO ME-9837 instruction manual

PASCO

PASCO ME-9837 instruction manual

SRAM Quarq DZero Basic Troubleshooting

SRAM

SRAM Quarq DZero Basic Troubleshooting

Pintek hvp-40 instruction manual

Pintek

Pintek hvp-40 instruction manual

Chauvin Arnoux MEMO TD80 quick start

Chauvin Arnoux

Chauvin Arnoux MEMO TD80 quick start

Chauvin Arnoux C.A 1110 user manual

Chauvin Arnoux

Chauvin Arnoux C.A 1110 user manual

LTB MNL X00 Series operating manual

LTB

LTB MNL X00 Series operating manual

netvox R716S user manual

netvox

netvox R716S user manual

Agilent Technologies 4338B Service manual

Agilent Technologies

Agilent Technologies 4338B Service manual

OPHIR vega user manual

OPHIR

OPHIR vega user manual

TSI Instruments P-TRAK Ultrafine Operation and service manual

TSI Instruments

TSI Instruments P-TRAK Ultrafine Operation and service manual

GREISINGER 4016138956801 manual

GREISINGER

GREISINGER 4016138956801 manual

ero electronic DPL instruction manual

ero electronic

ero electronic DPL instruction manual

Ruskin TDFi-RT Technical bulletin

Ruskin

Ruskin TDFi-RT Technical bulletin

RayTek GP Series operating instructions

RayTek

RayTek GP Series operating instructions

R&S ZVAX24 quick start guide

R&S

R&S ZVAX24 quick start guide

Goodwe GM3000C Quick installation guide

Goodwe

Goodwe GM3000C Quick installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.