Luminor LB4-031/2 User manual

OWNERS MANUAL
Operation & Installation Instructions
LB4-031/2, LB4-061/2, LB4-101/2, LB4-201/2, LB5-031/2, LB5-061/2, LB5-101/2,
LB5-201/2, LB6-061/2, LB6-101/2, LB6-201/2
illuminating technologies for life

Congratulations on purchasing this ultraviolet disinfection system. By purchasing a LUMINOR UV
isinfection system you are receiving not only a high quality product but also peace of mind. Protecting
your water supply with a UV system gives you reassurance that your family will have access to safe
drinking water throughout your entire home with no chance of microbiological contamination. This is
chemical free process which is simple in its concept and effective in its abilities to inactivate
microorganisms present in the water supply. Simple maintenance, continuous disinfection and
ultimately safe water, LUMINOR makes it that easy.

3 | P a g e
TABLE OF CONTENTS
Safety Considerations ............................................................................................................................ Page 4
Before You Begin ................................................................................................................................... Page 4
Water Quality Parameters ..................................................................................................................... Page 4
Assembly ................................................................................................................................................ Page 5
System Sizing ......................................................................................................................................... Page 6
Lo ation ................................................................................................................................................. Page 6
Installation ............................................................................................................................................. Page 6
System Disinfe tion ............................................................................................................................... Page 10
System Disinfe tion ............................................................................................................................... Page 10
Cleaning (quartz sleeve & UV sensor ..................................................................................................... Page 11
Operation (4.0 Controller Overview) ..................................................................................................... Page 12
Operation (5.0 & 6.0 Controller Overview) ........................................................................................... Page 13
Operation (5.0 & 6.0 Power-up) ............................................................................................................ Page 13
Operation (5.0 Controller S reens) ........................................................................................................ Page 14
Operation (6.0 Controller S reens) ........................................................................................................ Page 15
Operation (5.0 & 6.0 UV Intensity) ........................................................................................................ Page 15
Operation (5.0 & 6.0 Lamp Countdown) ............................................................................................... Page 16
Operation (5.0 & 6.0 Lamp Countdown Reset) ..................................................................................... Page 16
Operation (5.0 & 6.0 Failure Modes) ..................................................................................................... Page 17
Operation (5.0 & 6.0 QR Codes) ............................................................................................................ Page 18
Expansion Modules ................................................................................................................................ Page 18
Warranty Registration ........................................................................................................................... Page 20
BLACKCOMB4.0 Spe ifi ations ................................................................................................................ Page 21
BLACKCOMB5.0 Spe ifi ations ................................................................................................................ Page 22
BLACKCOMB6.0 Spe ifi ations ................................................................................................................ Page 23
Limited Warranty Statement ................................................................................................................. Page 24

4 | P a g e
Safety Consideration:
Although your UV system has been manufa tured to the highest safety standards, extreme are must be followed
when operating and/or maintaining your system.
1. Whenever you are servi ing this equipment, always dis onne t the power ord from the ele tri al outlet.
2. As the energy given off by the UV lamp an be harmful to your eyes and skin, NEVER look dire tly at an
illuminated UV lamp without adequate eye prote tion and always prote t your skin from dire t exposure
to the UV light.
3. To ensure the system provides adequate disinfe tion, ALWAYS repla e any omponent (lamp and sleeve)
with a genuine manufa turer’s repla ement part.
4. Do not operate the unit if it has any damaged or missing omponents.
5. To avoid possible ele tri al sho k, use only with a properly grounded ele tri al outlet.
6. Never perform any maintenan e to the system unless you are omfortable in doing so. Conta t the
manufa turer for servi e instru tions if required.
7. Do not use this system for any purpose other than what it was intended for. Misuse of this system ould
potentially ause harm to the user.
8. Your system is intended to be installed indoors and away from leaking plumbing. DO NOT plug the unit in
if the system or any of the omponents are wet.
9. The disinfe tion system should be dire tly installed into a ground fault ir uit interrupter (GFCI). If the use
of an extension ord is required, the ord must be manufa tured with a minimum of 16 gauge wire and
are should be taken to avoid potential tripping hazards.
10. We re ommend that a li ensed plumber or ertified te hni ian perform the installation
Before You Begin:
Before you begin, you’ll need the following:
Tools
•Pipe utter, ha ksaw or other spe ialized tools required to ut into your existing plumbing (e.g. if you have
PEX piping)
•Soldering tools (tor h, flux, emery loth and solder)
•Wren h (for tightening fittings)
Other Materials
•Inlet/outlet onne tions
•Teflon™ tape
Water Quality Parameters:
UV disinfe tion is extremely effe tive against mi roorganisms but only if the UV light an pass through the water it
needs to treat. This means that the quality of your water is very important in order to ensure omplete
disinfe tion.
It is imperative that you have your water tested for at the least the parameters listed below. If the water ex eeds
the listed parameters LUMINOR strongly re ommends that appropriate pretreatment equipment be installed
(pretreatment will depend on parameters being treated):

5 | P a g e
Hardness: <7 gpg (120 mg/L) – if hardness level is 7 gpg or slightly below the uartz
sleeve must be cleaned periodically in order to ensure efficient UV
penetration; if above the water must be softened
Iron (Fe): <0.3 ppm (0.3 mg/L)
Manganese (Mn): <0.05 ppm (0.05 mg/L)
Turbidity: < 1 NTU
Tannins (organics): <0.1 ppm (0.1 mg/L)
UVT (transmittance): >85% (Please contact LUMINOR if water has a UVT that is less than 80% for
pre-treatment recommendations)
You can have your water tested at a private analytical laboratory or by your local dealer. It is always
recommended to install pre-filtration of at least 5 microns prior to a LUMINOR UV disinfection system.
Assembly:
Step 1: Unpack the system and ensure all the components are included with the system. Your system is shipped
with the following components:

6 | P a g e
System Sizing:
All LUMINOR UV systems are rated for a specific flow rate under specific water uality parameters. The e uipment
has been designed to ensure that the appropriate dose will be delivered provided all parameters are followed.
PL AS NOT that the flow rate for each system can be higher or lower than what the system is rated for however
this will change the dose level that the system will deliver. A lower flow is not a concern as the dose will increase
however a higher flow rate will decrease the dose and therefore compromise the microorganism inactivation.
It is important to know the maximum flow rate that your water system delivers. If you do not have this
information you can simply fill a 1 gallon bucket with water and time how long it takes to fill up. This will be the
maximum flow rate for the home. Choose a BLACKCOMB system that is suitable.
PL AS NOT : It is always better to oversize your system then to undersize. For example, if your pump delivers 8
gpm it is recommended to install any of the BLACKCOMB 10 gpm systems. There will come a time where your
home is using water at the pump’s maximum capacity.
Location:
Step 1: Find a suitable location to mount the UV
reactor and the accompanying controller. In
choosing your location ensure the controller is
located within 5 feet of a ground fault circuit
interrupter (GFCI) and that there is easy access
to the main cold water line prior to any branch
lines and before the hot water heater. If you
have any other water treatment e uipment,
such as a softener or water filter, ensure that
the UV is the last piece of treatment e uipment.
PLEASE NOTE: All LUMINOR UV disinfection
systems are intended for indoor use only as they
should not be exposed to the elements.
Step 2: Your system includes mounting
hardware for both the UV reactor and the
controller. If the supplied fasteners are not
compatible with the structure in which you are
mounting the device, please ensure you use the
correct fasteners.
Step 3: To facilitate lamp removal, ensure there
is enough space at the lamp connector end to
safely remove the UV lamp and/or uartz sleeve
(a space e ual to the length of the unit will
suffice) (see Figure 1.).
Figure 1. Lamp Removal Spacing

7 | P a g e
Installation:
Step 1: The UV disinfe tion system should always be the last pie e of treatment before the water bran hes off to
the hot and old water lines if the intent is for Point of Entry (POE) (see figure 2). If the intent for the UV system is
for Point of Use (POU) the system should be the final step before the fau et.
Step 2: LUMINOR strongly re ommends that a 5 mi ron filter be installed before the UV system for a final polishing
step before the water is disinfe ted.
Step 3: The rea tor an be installed either horizontally or verti ally using the lamps provided, however verti al
installation is the preferred method with the inlet at the bottom (lamp onne tion at the top) as it allows any air
that may be in the lines to be easily purged from the system.
Figure 2. Recommended PO Installation
Step 4: If you do not know the flow rate of the water supply in your home it is re ommended that you use a flow
restri tor so that the rated flow of your parti ular LUMINOR BLACKCOMB system is not ex eeded and the UV dose
is not ompromised. The flow restri tor should be installed on the outlet port of the rea tor.
Step 5: It is strongly re ommended to have a li ensed plumber onne t the UV rea tor to the water supply and
may be a requirement depending on where you are lo ated. If you are attempting this yourself, ensure you have
all the ne essary tools and fittings to a omplish this task.
Step 6: Although there are many methods of installation, this manual will provide a re ommended pro edure using
opper plumbing and standard soldering methods. LUMINOR re ommends the use of unions, a by-pass assembly
and shut-off valves as this will allow you to isolate and remove the UV rea tor if ne essary (this method is a

8 | P a g e
recommended method only however it allows for the maximum convenience but it re uires extra components and
more time for installation. Please refer to Figure 3 for the recommended installation.
Figure 3. Recommended Installation
Step 7: Before you cut into the cold water line, measure and cut all piping as per the recommended layout. Once
all the components are ready, start by installing the female adapters onto the ports of the reactor. To ensure a
proper seal, the use of Teflon™ tape is recommended on all threaded connections.
Step 8: Next solder all the assembly together, including ball valves and unions and finally connected the by-pass
assembly to the cold water feed line (water in and water out).
Step 9: You can now gently remove the uartz sleeve from its packaging being VERY careful not to touch the length
with your hands. The use of cotton gloves (not included) is recommended for this procedure as oils from our hands
can leave residue on the sleeve and lamp which can ultimately block the UV light from getting to the water. In the
package, you will find a lubricated o-ring. Place the o-ring over the open-end of the sleeve as shown in Figure 4.
Figure 4. Quartz Sleeve Installation
Step 10: Carefully slide the sleeve into the reactor until you can feel it hit the opposite end of the reactor. Slightly
push the sleeve in to feel it lock into the spring inside the reactor. Ensure that the o-ring is butted up against the
reactor. Assemble the provided gland nut onto the threaded end of the reactor and tighten. The gland nut has a
positive stop to avoid over-tightening, hand tighten ONLY. Install the provided stainless steel compression spring
inside the uartz sleeve. This spring simply sits in the bottom of the uartz sleeve and works with the lamp and

9 | P a g e
LUMI-Loc™ connector to create the proper lamp alignment. PL AS NOT : DO NOT install a UV lamp inside the
uartz sleeve without the sleeve spring in place.
Step 11: (Applies only if you have a UV sensor…BLACKCOMB6.0) The UV sensor for the system is packaged in a
separate plastic bag. Carefully remove the sensor from its packaging and insert the sensor into the UV sensor port
(remove the protective cap on the UV reactor first). The sensor can only go in one way. Ensure that the flat portion
of the UV sensor matches up with the half metal lip on the sensor port (flat portion should face the lamp
connection end) (see Figure 5.). Ensure that the sensor is fully seated in the sensor port and then tighten (turning
clockwise) the sensor nut. PL AS NOT : DO NOT over tighten the nut as this could damage the Teflon sensor
body. Plug the male connector into the IEP port located on the right side of the controller, Figure 6, (make sure the
controller is not plugged in as the sensor MUST be attached before power is applied to the controller.)
Figure 5. UV Sensor Installation Figure 6. I P Connection
Step 12: The reactor is now ready for water flow. When all plumbing connections have been completed you should
check for any possible leaks. Slowly turn on the water supply and check for leaks. Make sure the by-pass valves are
functioning properly and that the water is flowing through the reactor. The most common leak is from the o-ring
not making a proper seal on the reactor. If this is the case turn the water off, drain the reactor, remove the o-ring,
dry it and reapply silicon grease. Replace o-ring ensuring that it is properly sealed against the reactor and check
again for leaks.
Step 13: The controller can now be mounted on the wall. The controller should always be above or beside the
reactor to ensure that no moisture can deposit on any of the connections (see Figure 2.). Always mount the
controller vertically. For safety purposes the controller should be connected to a ground fault circuit interrupter
(GFCI) (also known as a ground fault interrupter (GFI)).
Step 14: You can now remove the UV lamp from its packaging being careful not to touch the lamp uartz with your
hands. Again, the use of cotton gloves is recommended to avoid deposited oils on the lamp glass. Always hold the
lamp by the ceramic ends. With the lamp outside of the reactor, affix the UV lamp to the lamp connector as shown
in Figure 7.
Figure 7. UV Lamp Connection

10 | P a g e
Step 15: Carefully insert the UV lamp into the reactor sliding it inside the uartz sleeve located inside the reactor
(do not drop the lamp into the reactor). Affix the LUMI-Loc™ connector into the gland nut by inserting the
connector into the nut and turning the connector approximately ¼ turn to lock the connector to the gland nut.
Figure 8. LUMI-Loc™ Connector
Step 16: Affix the captive ground screw to the ground lug on the UV reactor to ensure proper grounding continuity.
Figure 9. Ground Screw Connection
Step 17: Your system is now ready to be plugged into the appropriate GFCI protected outlet. Plug the unit in and
refer to the “System Disinfection” section before any water is allowed to flow through the system.
System Disin ection:
Now that the system is installed you will need to disinfect the lines throughout the entire home or facility. To
accomplish this, please follow these simple steps:
Step 1: Before performing this task, check to make sure that there are no “dead ends” in the lines throughout the
home as these can harbor bacteria. Also ensure that the UV system is powered-up and ready for operation before
you start this disinfection process.

11 | P a g e
Step 2: Remove the artridge from the filter sump and fill with 1-2 ups of household blea h (most household
blea hes are 5.25% hlorine). Repla e filter sump and slowly turn on the water supply.
Step 3: Go to ea h lo ation in the home or fa ility (in luding outdoor taps) and run some water until hlorine an
be dete ted (by smell). Ensure all fau ets, dishwasher, shower heads, washing ma hine, toilets, showers,
refrigerators, et . are all he ked. On e the hlorine is dete ted at ea h lo ation lose all fau ets and let the
hlorine sit in the lines for a minimum of 30 minutes.
Step 4: After 30 minutes, reinstall the filter artridge into the filter sump and then flush the hlorine solution by
opening all fau ets and let them run until hlorine an no longer be dete ted (by smell). Your home has now been
ompletely disinfe ted with your LUMINOR BLACKCOMB UV system ready to ina tivate any mi roorganisms that
enter the home.
PL AS NOT : This pro edure must be performed following any time the UV is shut down for servi e, without
power, or is inoperative for some reason in order to avoid possible ontamination of the water lines.
Cleaning the Quartz Sleeve:
Depending on the water quality, the quartz sleeve may require periodi leaning. At a minimum, the quartz sleeve
should be leaned on an annual basis. The following steps outline a basi leaning pro edure.
Step 1: If the system has an inlet shut-off valve, shut this valve off to prevent water flow through the system. If
there is no inlet shut-off valve, turn off main water inlet valve (and turn of water pump if you have one).
Step 2: Dis onne t power ord of UV system from ele tri al outlet.
Step 3: Release water pressure by opening a downstream fau et and lose the outlet shut-off valve if the
installation in udes one (if there is no outlet shut-off valve, be prepared for water leakage from the system as the
head pressure in the system will ause water to flow ba k through the outlet plumbing and through the rea tor).
Step 4: Remove the aptive ground s rew from the ground lug on the UV rea tor.
Step 5: Remove the lamp onne tor from the rea tor (gland nut) by pushing the LUMI-lo onne tor in and
turning it ¼ turn ounter- lo kwise (gently pull on the lamp able to ensure the onne tor is seated properly).
Step 6: Pull lamp out of rea tor still atta hed to onne tor.
Step 7: Remove the gland nut from the rea tor exposing the end of the quartz sleeve.
Step 8: Carefully pull the quartz sleeve out of the rea tor and remove the o-ring
Step 9: Using a soft, lint-free loth or towel wipe the sleeve down using CLR® or LIME-A-WAY® (or other
ommer ial s ale leaner) to remove any s aling or iron deposits that may be on the outside of the quartz sleeve.
Be areful not to get any moisture or liquids on the inside of the sleeve.
Step 10: Wipe the sleeve with separate dry loth.

12 | P a g e
Step 11: Once the sleeve is cleaned replace the o-ring and slide the sleeve back into the reactor following the
steps outlined on page 8 of the manual.
Cleaning the UV Sensor:
Depending on the water uality, the UV sensor may re uire periodic cleaning. At a minimum, the UV sensor should
be cleaning on an annual basis. The following steps outline a basic cleaning procedure.
Step 1: If the system has an inlet shut-off valve, shut this valve off to prevent water flow through the system. If
there is no inlet shut-off valve, turn off main water inlet valve (and turn of water pump if you have one).
Step 2: Disconnect power cord of UV system from electrical outlet.
Step 3: Release water pressure by opening a downstream faucet and close the outlet shut-off valve if the
installation incudes one (if there is no outlet shut-off valve, be prepared for water leakage from the system as the
head pressure in the system will cause water to flow back through the outlet plumbing and through the reactor).
Step 4: Place something under the reactor to catch any water that may come out of the reactor during the
removal of the UV sensor.
Step 5: Unscrew (counterclockwise) sensor nut from the reactor and pull the sensor slowly out of the sensor port.
Step 6: Holding the sensor in your hand wipe the flat portion (sensor face) of the sensor with isopropyl alcohol
using a clean lint-free cloth.
Step 7: Replace sensor as per instructions on pages 8-9 of the manual
Operation:
The LUMINOR BLACKCOMB system comes with a feature laden controller that incorporates both the lamp driver
(ballast) and control features in one water-tight case. Two controllers are available for the BLACKCOMB systems
(depending on your model). CAUTION: Prior to performing any maintenance on your UV system, you must always
disconnect the power.
BLACKCOMB 4.0 CONTROLLER
This controller features a power factor corrected, constant current lamp
driver with a universal power input. Simplistic in operation, this system
features two LED’s that show lamp status. When the UV lamp is on, the
green LED will be illuminated. When the UV lamp is not on, the red LED
will be illuminated and an audible buzzer will be sounding. To remedy
this condition, the UV lamp must be replaced with a new UV lamp.
Please Note: While the red LED is illuminated and the buzzer is sounding
the water from the system should NOT be consumed. If any water does
pass through the system during this period, please follow the

13 | P a g e
disinfection procedure as outlined in this manual before the water is
consumed. It should also be noted that even though this system has a
visual and audible warning built into the controller, the green LED does
not necessarily indicate that the water coming from this system is in fact
potable (safe to drink). This system does not measure the level of
disinfection; it simply measures the “on-off” status of the lamp. Please
have your water checked for microbiological contaminants on a regular
basis.
BLACKCOMB5.0 & BLACKCOMB6.0 CONTROLLER
This controller features a power factor corrected, constant current lamp
driver with a universal power input. A full colour LCD screen provides the
user with a detailed description of the system’s performance in addition
to providing any applicable fault messages and system diagnostics. The
controllers used in both the BLACKCOMB5.0 and BLACKCOMB6.0 are
identical. The difference is that the BLACKCOMB6.0 series of products
includes a UV intensity monitor. All BLACKCOMB5.0 and BLACKCOMB6.0
controllers include an “infinite expandability port” located on the right
side of the controller. Simply plug in an optional UV sensor module into
the expandability port of a BLACKCOMB5.0 controller and the system will
now monitor the UV intensity of the system!
BLACKCOMB5.0 & BLACKCOMB6.0 Power-up Sequence
Upon start up, the 5.0 & 6.0 controller will run through a diagnostic start-up and the se uence will be displayed as
follows on the colour LCD:
The controller next checks for the installation of any optional modules that may be installed on the system. It first
checks to see if a module is installed and then either initializes the module to function with the controller or
returns a “not detected” screen and moves on to the next module. The screens will appear as follows:
UV Sensor
Module Chec

14 | P a g e
Solenoid
Module Chec
4-20 mA
Module Chec
Ethernet
Module Chec
Remote Alarm
Module Chec
Once the controller has finished checking for the installation of all available modules a single screen is displayed
showing which specific modules are installed and activated on this particular system. Please Note: If the
corresponding module icon is not displayed on this screen, then that module is not installed on this system. If you
believe this module is installed correctly, recheck the applicable connections to ensure a solid connection point
and then restart the controller. The controller then continues to optimize the lamp efficiency by waiting 60
seconds to allow the lamp to reach its optimum output. Finally, a final “start-up complete” screen is displayed.
These screens will appear as follows:

15 | P a g e
BLACKCOMB5.0 Operational Screens
On BLACKCOMB5.0 systems (without the UV monitor), the default screen shows the LUMINOR Home Screen. At
any point during operation the user is able to scroll through the LUMINOR Home Screen, Lamp life remaining and
QR Code/Contact Info screens by simply pressing the button located on the front of the controller.
BLACKCOMB6.0 Operational Screens
On systems that have the UV module installed (and on all BLACKCOMB6.0 systems), the default screen shows the %
UV Intensity. At any point during operation the user is able to scroll through the % UV Intensity, Lamp life
remaining and QR Code screens by simply pressing the button located on the front of the controller.
% UV Intensity
If your system is e uipped with a UV sensor the % UV Output screens will indicate the level of UV intensity that is
being detected within the reactor by the sensor. This will always be visible on the Home Screen of the controller.
Things that can affect the % UV Output are poor water uality, scaling of the uartz sleeve and/or sensor, lamp
failure, expired lamp life and sensor failure. The following screens show the % UV Output dropping on a monitored
system.
Once the UV output drops below 56%, the numbers and warning sign switch to red and a 15 second intermittent
audible chirp is given off by the ballast. When the output drops below 51%, the display switches to a solid red and
button
button

16 | P a g e
a constant audible signal is provided. The screens alternate between this solid red screen and another screen
indicating “water may be unsafe for consumption”. At this point, the controller also provides a signal to de-activate
the solenoid valve, shutting off the flow of water if a solenoid valve is installed on the system.
BLACKCOMB5.0 & BLACKCOMB6.0 Lamp Countdown Sequence
The BLACKCOMB5.0 & BLACKCOMB6.0 counts down the number of days until a lamp change is re uired. The
operation is as follows:
Once the lamp has seven days remaining until a lamp change is re uired, the screen changes to a yellow caution
screen. At this point, an audible chirp is given off by the ballast to draw your attention to the pending lamp change
condition. When the controller passes the zero day threshold, the screen changes to solid red and cycles between
a red “lamp expired” screen and a “water may be unsafe for consumption” screen. The same intermittent audible
chirp is heard throughout this lamp expired se uence.
At any point during this se uence, the audible chirp can be deferred by holding the controller button down for a
period of five seconds. The screen shown below will appear for five seconds indicating that the audible defferal has
been activated and then the system switches back to the current lamp change or lamp expired screen. This audible
deferral will last for seven days after which then the alarm will sound once again. This audible deferral feature can
be activated as many times as you wish. PL AS NOT : During the condition of lamp expiration and audible
deferral, the water may be unsafe for consumption and should not be consumed without another form of
disinfection.

17 | P a g e
BLACKCOMB5.0 & BLACKCOMB6.0 Lamp Countdown Reset Sequence
After changing the lamp on the BLACKCOMB5.0 & BLACKCOMB6.0 systems, the controller must be reset in order for
the system to begin its countdown function on the newly installed lamp. To perform this reset, you must firmly
depress the button on the front of the controller and then manually plug the power cord back into the wall outlet
initiating power to the unit. Keep holding down the button for five seconds until you hear an audible chirp
indicating the controller has reset the internal timer. Release the switch and the lamp countdown feature has now
been reset. The following two screens will be displayed during this process.
BLACKCOMB5.0 & BLACKCOMB6.0 Failure Modes
The BLACKCOMB controller continuously monitors your UV system and if there is a problem with the system the
controller will provide both a visual and audible signal indicating the specific fault that may be adversely affecting
the operation of your system. The fault conditions are listed in a priority se uence as follows:
LAMP FAILURE: If at any time during the operation of the system, the UV lamp fails
to be illuminated, the controller will return both an audible and visual signal indicating
lamp failure. In addition, a constant audible buzzer will sound during lamp failure.
UV SENSOR FAILURE: Assuming the system has a UV Sensor installed, if at any time
during the operation of the system, a UV sensor fails, the controller will return both an
audible and visual signal indicating sensor failure. For sensor failure, the controller will
cycle between “sensor failure” and “water may be unsafe for consumption” screens as
an added precaution. In addition, a constant audible buzzer will sound during sensor
failure.
SOLENOID MODULE FAILURE: Assuming the system has an optional Solenoid
Module installed, if at any time during the operation of the system, a solenoid valve fails
to operate, the controller will return both an audible and visual signal indicating
solenoid failure. In addition, an intermittent (15 sec) audible buzzer will sound during
solenoid failure.
4-20 mA MODULE FAILURE: Assuming the system has an optional 4-20 mA
Module installed, if at any time during the operation of the system, a 4-20 mA module
fails, the controller will return both an audible and visual signal indicating a 4-20 mA
failure. In addition, an intermittent (15 sec) audible buzzer will sound during a 4-20 mA
failure.

18 | P a g e
REMOTE ALARM MODULE FAILURE: Assuming the system has an optional
Remote Alarm Module installed, if at any time during the operation of the system, a
remote alarm module fails, the controller will return both an audible and visual signal
indicating remote alarm failure. In addition, an intermittent (15 sec) audible buzzer will
sound during remote alarm failure.
ETHERNET MODULE FAILURE: Assuming the system has an optional Ethernet
Alarm Module installed, if at any time during the operation of the system, an Ethernet
module fails, the controller will return both an audible and visual signal indicating
Ethernet failure. In addition, an intermittent (15 sec) audible buzzer will sound during
Ethernet failure.
BLACKCOMB5.0 & BLACKCOMB6.0 QR Codes
A QR code (abbreviated from Quick Response code) is a type of matrix barcode (or two-dimensional code) first
designed for the automotive industry. LUMINOR uses the QR code to store a link to a specific page on our website.
Users with a camera phone e uipped with the correct reader application can scan the image of the QR code and
over a wireless network connect to a LUMINOR web page in the cellular's browser. LUMINOR’s QR webpage has
information on how to purchase replacement components as well as a helpful video directory on system servicing
(i.e. How to change a UV lamp or uartz sleeve). To access the QR code on the BLACKCOMB5.0 or BLACKCOMB6.0
controller simply press the control button twice and the QR code screen will appear as follows:
BLACKCOMB5.0 & BLACKCOMB6.0 Expansion Modules
As shown earlier in this manual, the BLACKCOMB5.0 and BLACKCOMB6.0 controller incorporate an “Infinite
xpandability Port” (I P) on the controller. It is through this port and via a custom communication protocol that
the UV sensor and all expanison modules are connected. Each module (including the sensor) comes with both a
male and female connection. To activate the device, you simply need to connect the first device (does not matter
which device it is) to the controller and all subse uent devices are then connected into the female end of the other
device. This “daisy chain” method allows for a truly modular approach and is extremely user friendly. An example
of this “daisy chain” is shown below.

19 | P a g e
The following optional expansion modules are available for use on your BLACKCOMB5.0 and BLACKCOMB6.0
controllers. Contact your authorized distributor for purchasing information.
REMOTE ALARM MODULE: Allows for a signal to be connected to a
remote monitor such as a buzzer, light, alarm system, PLC, etc., via a pair
of contacts. In normal operation the OK and COM contacts will be
connected, and in a fault condition (Low UV, Lamp fail, Power Fail), the
Fault and COM contacts will be connected. Maximum Contact Rating is 1A-
120VAC/VDC (uses 16-22 AWG).
SOLENOID CONNECTION MODULE: Allows the integration of a
NORMALLY CLOSED line voltage solenoid valve. Simply affix the module's
supplied bare lead cord to the solenoid valve, connect the power cable to
the electrical outlet, and the male IEP connector to the female IEP port on
the controller. PL AS NOT : When this device is connected to a
BLACKCOMB5.0 system (without the UV module), then the solenoid will
close on a lamp failure mode only. When used on a BLACKCOMB6.0 system,
the solenoid will close when the UV level drops below 50%. Also note that
in cases where emergency use of untreated water is re uired, the
controller can be placed into a manual override mode allowing for the flow
of water in an alarm condition.
4-20 mA MODULE: Allows for a 4-20mA signal transfer of the UV
output to a remote device such as a data logger or computer. Simply plug
the supplied cable into any female IEP connections. Two screw terminals
are supplied to connect the wire that leads to the receiving end of the
signal. As the distance re uired is unknown, no wire is supplied with this
system, however 16-22 AWG is re uired.
ETHERNET MODULE: Allows for all controller functions to be
connected to a computer via an Ethernet cable. This module is currently in
development and is not available at this point-in-time.

20 | P a g e
Warranty Registration
It is imperative that you complete the warranty registration process. This not only registers your UV disinfection
system for the provided manufacturer’s warranty, but also allows the factory to provide you with any important
product updates or technical bulletins concerning your product. The registration process is a simple process and
can be done online at http://www.luminoruv.com/support_prodreg.php or by filling out the attached warranty
card. Please ensure that ALL information is filled in, including a valid e-mail address. PL AS NOT : This
information is for the sole purpose of technical support for your disinfection system and will not be used, or sold,
to any other organization for any other purpose. Please refer to LUMINOR’s Privacy Policy statement at
http://www.luminoruv.com/main_copyright.php for further information.
LUMINOR EQUIPMENT SPECIFICATIONS
BLACKCOMB
4.0
, Residential UV systems, modular non-
monitored
MODEL LB4-031 LB4-032 LB4-061 LB4-062 LB4-101 LB4-102 LB4-201 LB4-202
Flow Rate (@ 16 mJ/cm
2
)
6.1 GPM 11.0 GPM 20 GPM 39.2 GPM
23 lpm 42 lpm 77 lpm 150 lpm
1.4 m
3
/hr. 2.5 m
3
/hr. 4.6 m
3
/hr. 8.9 m
3
/hr.
Flow Rate (@ 30 mJ/cm
2
)
3.1 GPM 5.8 GPM 11 GPM 21 GPM
12 lpm 22 lpm 41 lpm 79 lpm
0.7 m
3
/hr. 1.3 m
3
/hr. 2.5 m
3
/hr. 3.6 m
3
/hr.
This manual suits for next models
13
Table of contents
Other Luminor Water Filtration System manuals
Popular Water Filtration System manuals by other brands

Katadyn
Katadyn MyBottle owner's manual

Insignia
Insignia NS-A700PF6 Quick setup guide

Zip
Zip HydroTap G4 installation instructions

Trox
Trox TFP-TC-S Installation and maintenance manual

Parker
Parker Racor FBO-10 Installation, operation and service instructions

Watex
Watex CMS 8 TWIN Installation and operating manual