Handling Instructions
Please note some other handling instructions to consider when Operating your PhotonMaster Luminometer &
Bluetooth Module:
PhotonMaster Luminometer
•Do not move or tilt the units during measurement and ensure you are working on a level surface. This may
cause fluctuations in the measurement data.
•Always remove the cuvette after taking a measurement to prevent liquid spilling inside the chamber and
causing a malfunction with the PhotonMaster Luminometer.
•Keep this unit at least 10 cm away from stirrers, mixers or any device that may generate electromagnetic noise.
•Do not expose this unit to liquids, reagents or organic solvents. Doing so may cause malfunction, discoloration
or deformation. If the unit was exposed to such liquids, reagents or solvents, then wipe them away immediately.
•Avoid using this unit in locations: where the temperature fluctuates rapidly; directly exposed to steam; where
excessive condensation occurs; where corrosive gases are generated; subject to vibration; with intense
magnetic fields; or in locations exposed to excessive dust or debris.
•Before using this unit, wash your hands thoroughly. This unit is extremely sensitive, so the ATP (Adenosine
Triphosphate) may become contaminated through contamination from hands.
•Do not leave this unit in unstable place or hazardous locations. Do not apply excessive shock to this unit and
do not drop it. Doing so will cause permanent damage.
•Be careful to prevent electrostatic charge buildup on the sample tube. Static electricity may cause an increase
the measurement data.
PhotonMaster Bluetooth Module
•Do not leave this unit in unstable places or hazardous locations. Do not apply excessive shock to this unit and
do not drop it. Doing so may cause permanent damage.
•Do not expose this unit to excessive water or submerge. This unit is not water-proof, excessive water or
moisture may cause an internal short or damage to the operation of the device.
•Do not expose this unit to liquids, reagents or organic solvents. Doing so may cause malfunction, discoloration
or deformation. If the unit was exposed to such liquids, reagents or solvents, then wipe them away immediately.
•Avoid using this unit in locations: where the temperature fluctuates rapidly; directly exposed to steam; where
excessive condensation occurs; where corrosive gases are generated; subject to vibration; with intense
magnetic fields; or in locations exposed to excessive dust or debris.
•Customers can return their used PBM for recycling by contacting us at www.luminultra.com and use our contact
form. Please provide date of purchase and serial number of the unit you wish to dispose of and a
representative will contact you within 1-2 business days with further instructions. Please do not return any
items to us without prior written authorization.