Macurco TX-6-AM User manual

Macurco™ Ammonia Detector, Controller and Transducer TX-6-AM
User Instructions
Important: Keep these User Instructions for reference

2
TABLE OF CONTENTS
GENERALSAFETYINFORMATION 3
Intended Use 3
List of Warnings and Cautions 3
USE INSTRUCTIONS AND LIMITATIONS 4
UseFor 4
DoNotUseFor 4
General Description 4
Features 5
Specifications 5
INSTALLATION AND OPERATING INSTRUCTIONS 5
Location 6
Installation 6
Garage diagram 7
4-20 mA Output diagram 7
Multiple Device diagram 8
Alarm Control Panel diagram 8
DVP-120 Control Panel diagram 9
Alternate Alarm Panel 10
PowerUp 10
Operation 10
Default – Factory Settings 10
Display setting 11
Buzzer setting 11
Alarm Relay setting 12
Alarm Relay Configuration 12
Fan Relay setting 12
Fan Relay Delay setting 13
Fan Minimum Runtime setting 13
Fan Relay Latching setting 13
4-20mA Output setting 13
On Board Diagnostics 13
Sensor Poisons 14
MAINTENANCE 14
End-of-Warranty Signal 14
Sensor Replacement 15
Sensor Life Reset 15
Cleaning 16
Testing 16
Operation Test 16
Ammonia Test 18
Field Calibration Procedure 20
MACURCO GAS DETECTION PRODUCTS WARRANTY 22

3
GENERAL SAFETY INFORMATION
Intended Use
The Macurco TX-6-AM is alow voltage, dual relay Ammonia (NH3) detector, controller and transducer. The TX-6-AM has selectable4-
20 mA output, buzzer and digital display options. It is an electronic detection system used to measure the concentration of NH3 and
provide feedback and automatic exhaust fan control to help reduce NH3 concentrations in industrial refrigeration, warehouses, hockey
rinks or other commercial applications. The TX-6-AM is a lowlevel meter capableof displaying from0-100 ppm of Ammonia. The TX-6-
AM is factory calibrated and 100% tested for properoperation, but can also be calibrated in the field.
List of Warnings and Cautions within these User Instructions
Each person using this equipment must read and understand the information in these User Instructions before use. Use of this
equipment by untrained or unqualified persons, or use that is not in accordance with these User Instructions, may adversely affect
product performance and result in sickness or death.
Use only for monitoring the gas which the sensor and instrument are designed to monitor. Failure to do so may result in exposures
to gases not detectable and cause sickness or death. For proper use, see supervisor or User Instructions, or call Technical
Service at 1-877-367-7891.
This equipment may not function effectively below 0F or above 125F (-18C or above 52C). Using the detector outside of this
temperature range may adversely affect product performance andresult in sickness ordeath.
This detector helps monitor for the presence and concentration level of a certain specified airborne gas. Misuse may produce an
inaccurate reading, which means that higher levels of the gas being monitored may be present and could result in overexposure
and cause sickness or death. For proper use, see supervisor or User Instructions, or call Technical Serviceat 1-877-367-7891.
High voltage relay terminals (120/240 VAC) are located within this detector, presenting a hazard to service technicians. Only
qualified technicians should open the detector case and service the internal circuits. Ensure power is removed from the detector
relays prior to servicing the unit. Failure to doso may result in sickness or death.
Do not disassemble unit or attempt to repair or modify any component of this instrument. This instrument contains no user
serviceable parts, and substitution of components may impair product performance and result in sickness ordeath.
Using a certified gas with a concentration other than the one listed for this detector when conducting a calibration or calibration
verification test (bump test) will produce inaccurate readings. This means that higher levels of the gas being monitored may be
present and could result in overexposure and cause sickness or death. For proper use, see supervisor or User Instructions, or
call Technical Service at 1-877-367-7891
The following steps must be performed when conducting a calibration or calibration verification test (bump test) to ensure proper
performance ofthe monitor.Failure to do so mayadversely affect product performance and result in sickness or death.
When performinga calibration or calibration verification test (bump test) only use certified calibration gas at the
required concentration level. Do not calibrate with expired calibration gas.
If the instrument cannot be calibrated, do not use until the reason can be determined and corrected.
Do not cover or obstruct display or visual alarm
Ensure sensor inlets are unobstructed and is free of debris
! WARNING

4
USE INSTRUCTIONS AND LIMITATIONS
Each person using this equipment must read and understand the information in these User Instructionsbefore use. Use ofthis
equipment by untrained or unqualified persons, oruse that isnot in accordance with these User Instructions, may adversely affect
product performance and result in sickness or death.
Use For
The TX-6-AM provides Ammonia detection and automatic exhaust fan control for industrial refrigeration, warehouses, hockey rinks or
other commercial applications. Ammonia or NH3, is a colorless gas with a pungent, suffocating odor. Note: Shipped as a liquefied
compressed gas - easily liquefied under pressure. The TX-6-AM is designed to meet the NIOSH and OSHA standards for Ammonia
exposure limits. TX-6-AM can be used stand alone, with the Macurco DVP-120 Detection and Ventilation Control Panel, other 12 VAC
or 24 VDC fire/security panels or building automation systems.
Use only for monitoring the gas which the sensor and instrument are designed to monitor. Failure to do so mayresult in exposures to
gases not detectable and cause sickness or death. For proper use, see supervisor or User Instructions, or call TechnicalService at 1-
877-367-7891.
Do Not Use For
The TX-6-AM is not intended for use in hazardous locations. Do not mount the TX-6-AM where the normal ambient temperature is
below 0°F or exceeds 125°F (below -18C or above 52C). The TX-6-AM mounts on a 4x4 electrical box supplied by the contractor. Do
not install the TX-6-AM inside anotherbox unless it has good air flow through it.
This equipment may not function effectively below 0F or above 125F (-18C or above 52C). Using the detector outside of this
temperature range may adversely affect product performance andresult in sickness ordeath.
General Description
The TX-6-AM is a low voltage, dual relay Ammonia (AM) detector and automatic ventilation controller. The TX-6-AM uses a
microcomputer controlled, electronic system to measure the concentration of Ammonia, actuate relays and provide a 4-20mA output.
The TX-6-AM has a field replaceable, electrochemical sensor (expected life of 2+ years) and optional gas test and calibration kits. The
TX-6-AM is a low level meter capable of displaying from0-100ppm of Ammonia.
! WARNING
! WARNING
! WARNING

5
Features
ETL Listed to UL 61010-1
Low level meter capable of displaying from 0-100 ppm Ammonia
The TX-6-AM meets NIOSH and OSHA standardsfor gas exposure
Selectable fan and alarm relay activation
5 A SPDT fan relay controls starters of exhaust fans
0.5 A N.O. or N.C. alarm relay connectsto warning devices or control panels
4-20 mA Current Loop
TX-6-AM mounts ona standard 4x4 electrical box and becomes cover forthe box
Supervised system: any internal detector problem will cause the fan & alarm relay to activate
Calibration kit is available. One screw allows access for calibration or gas test
Specifications
Power: 3 W (max) from 12 to 24 VAC or 12 to 32 VDC
Current @ 24 VDC:75 mA in alarm (two relays), 50 mA (fan relay only) and23 mA stand by
Shipping Weight:1 pound (0.45kg)
Size: 4 1/2 x 4 x 21/8 in. (11.4 X 10.2 X 5.4 cm)
Color: Dark gray
Connections: plugs/terminals
Mounting box: (not included) 4x4 electric
Fan relay: 5 A, 240VAC, pilot duty, SPDT, latching or non-latching
Fan relay actuation: selectable at diS (disable), 15, 20, 25 (default), 30, 35, 40, 45, 50, 55, 60, 65, 70, or 75 ppm NH3
Fan Delay Settings of 0, 1, 3 (default), 5 and 10 minutes
Fan Minimum Run Time settings are 0 (default), 3, 5,10 or 15 minutes
Fan relay latching ornot latching (default) selectable
Alarm relay: 0.5A 120 V, 60 VA
Alarm relay actuation: selectable N.O. (default) or N.C.
Alarm relay settings: diS, 20, 25, 30, 35, 40, 45, 50, 55, 60, 65, 70, 75 (default), 80, 85, 90, 95 or 100 ppm NH3
Current Loop: 4-20 mA for 0-100ppm NH3, selectable to off or on (default)
Buzzer: 85 dBA at 10cm settable to off (default) or on
Digital display: 3 digit LED selectable to off (default) or on.
Operating environ:0°F to 125°F (-18C to 52C).10 to 90% RH
INSTALLATION AND OPERATING INSTRUCTIONS
The following instructions are intended to serve as a guideline for the use of the Macurco TX-6-AM Ammonia Detector. It is not to be
considered all-inclusive, nor is it intended to replace the policy and procedures for each facility. If you have any doubts about the
applicability of the equipment to your situation, consultan industrial hygienist or call Technical Service at 1-877-367-7891.

6
This detector helps monitor for the presence and concentration level of a certain specified airborne gas. Misuse may produce an
inaccurate reading, which means that higher levels of the gas being monitored may be present and could result in overexposure and
cause sickness or death. For proper use, see supervisor or User Instructions, or call Technical Service at 1-877-367-7891.
Location
A TX-6-AM is normally mounted high in the room on a wall or column one foot below the ceiling in a central area where air movement is
generally good. Use the same spacing as for smoke detectors, 30-foot centers, 900 square feet per detector (465 sq. meters). The
coverage depends on air movement within the room or facility. Extra detectors may be needed near any areas where people work or
where the air is stagnant. The TX-6-AM mounts on a 4x4 electrical box supplied by the contractor. Do not install the TX-6-AM inside
another box unless it has good air flow through it. Do NOT mount the TX-6-AM where the normal ambient temperature is below below
0°F or exceeds 125°F (below -18°C or above 52°C).
High voltage relay terminals (120/240 VAC) are located within this detector, presenting a hazard to service technicians. Only qualified
technicians should open the detector case and service the internal circuits. Ensure power is removed from the detector relays prior to
servicing the unit. Failure to do so may result in sickness or death.
Installation
1. The TX-6-AM mounts on a 4” square (or 4x4) electrical box supplied by the contractor. Do not mount the TX-6-AM inside
another box, unless it has good air flow through it.
2. Connect the TX-6-AM to Class 2 power supply only. It is suggested to use a separate transformer for powering the unit or units
because of possible interference’s fromother devices on the same power supply.
3. Connect the TX-6-AM to the control cables with terminal plugs.When making connections, make surethe power is off.
4. There are two terminals for Power: 12 to 24 VAC or12 to 32 VDC, with no polaritypreference.
5. There are two terminals forthe dry alarm relay contacts, againwith no polaritypreference. Thealarm relay can switch up to 0.5
A 120 V, or 60 VA. The alarm relay is activated if gas reaches or exceeds the alarm settings. See OPERATION section of
these User Instructions for details on relay settings.
6. The alarm relay can be configured to normally open (default) (N.O.) or normally closed (N.C.) and will activate if the gas
concentration exceeds alarm set point. It will deactivate once the gas concentration drops below the alarm set point. Note that
the “disable” setting will cause the alarm relay not to engage at all.
7. The dry contact, SPDT fan relay has three terminals. The common (COM.), normally open (N.O.) and the normally closed
(N.C.) contact. The fan relay can switch up to 5.0 A up to 240 VAC. See OPERATION section of these User Instructions for
details on relay settings.
! WARNING
! WARNING

7
8. The Fan Relay can be configured for latching or non-latching (default) when activated (when the gas concentration exceeds fan
relay set point).
9. The Fan Relay will engage if the fan setting Ammonia concentration is exceeded for longer than the Fan Relay Delay time.
Unless it is configured for latching, the fan relay will disengage once both of these conditions have been met:
Ammonia concentration has dropped below fan setting
Fan Relay Run timehas been exceeded
Note that the “disable”fan setting will cause the fan relay to not engage.The fan relay will engage in trouble fault condition and
will disengage once trouble fault condition is cleared.
10. The Current Loop is4 mA in clean air and 20 mAfor 100 ppm Ammonia

8

9

10
Power Up
The TX-6-AM steps through an internal self-test cycle for the first 1 minute that it is powered. The unit will execute the test cycle any
time power is dropped and reapplied (i.e. power failure). During the self-test cycle the unit will display the Firmware Version number,
then count down from 60 to 0 and finally go into normal operation. The Fan and Alarm relay will be activated for the power-up cycle
unless the “Power Up Test” (PUt) option is OFF. The indicator light (LED) will flash green during the self-test cycle. At the end of the 1
minute cycle, the unit will take its first sample ofthe air and the indicator light will turn solid green.
Operation
1. With the display function turned “On”, the TX-6-AM will show the current concentration of NH3 ppm or “0” (zero) in clean air.
When the NH3 concentration reaches the Fan Relay setting (25 ppm, for example) the display will flash back and forth between
“FAn” and “25”. With the display function turned ”OFF”, the display does not show the gas concentration, but will show “FAn” as
long as the fan relay is activated.
2. With the display function turned “On” and the NH3 concentration reaching the Alarm Relay setting, (75ppm, for example) the
display will flash back and forth between “ALr” and “75”. The buzzer will sound indicating “Alarm” if the buzzer is turned “On”.
With the display function turnedoff the display does not show the NH3 concentration, but will show “ALr” when the Alarm relay
is activated.
3. With the 4-20 mA function turned “On” and the NH3 concentration climbing the 4-20 mA signal will ramp up corresponding to
the concentration (0-100 ppm).
Default Configuration – Factory Settings
The default Power Up Test setting is On
The default Display setting is OFF

11
The default Buzzer setting is OFF
The default Alarm Relay Setting is activation at 75 ppm
The default Alarm Relay Configuration is Normally Open
The default Fan Relay Setting is activation at 25 ppm
The default Fan Relay Delay setting is 3 minutes
The default Fan Relay Runtime setting is 0
The default Fan Relay Latching condition is OFF
The default 4-20mA Output setting is On
To change settings, remove the Philips screw on the front of the TX-6-AM. Pull off the front cover of the unit.
Selecting Default Configuration – “dEF”
To select the Default Configuration, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push the
Enter button to enter the Con menu. The first selection is the “dEF” or Default setting. Push Enter. If it is already in Default
configuration, there will be no action. If it is not already in Default configuration, “no” will be displayed. Push Next to change it to “yES”
(flashing) then push Enter to confirm the change (solid) and push Enter again to return to “dEF” in the con menu. Push Next until “End”
is displayed then push Enter to get back to normal operation.

12
Selecting Power Up Test Option – “PUt”
To select the Power Up Test Configuration, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then
push the Enter button to enter the Con menu. Then push the Next button to get to the second selection “PUt” or Power Up Test
setting. Push Enter. If the test is “On” push Next to turn it “OFF” (flashing) then push Enter to confirm the change (solid) and push
Enter again to return to “PUt”in the Con menu. Push Next until “End” is displayed then push Enter to get back to normal operation.
Selecting Display Option – “dSP”
To select the Display Configuration, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push the
Enter button to enter the Con menu. Then push the Next button to get to the third selection “dSP” or Display setting. Push Enter. If
the display is “On” push Next to turn it “OFF” (flashing) then push Enter to confirm the change (solid) and pushEnter again to return to
“dSP” in the Con menu. Push Next until “End” is displayed then push Enter to get back to normal operation.
Selecting Buzzer Option – “bUZ”
To select the Buzzer Configuration, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push the
Enter button to enter the Con menu. The forth selection is the “bUZ” or Buzzer setting. Push Next three times to get to “bUZ” then
Enter. If the display is “On” push Next to turn it “OFF” (flashing) then push Enter to confirmthe change (solid) and push Enter again to
return to “bUZ”in the Con menu.Push Next until “End” is displayed then push Enter to get back to normal operation.
Selecting Alarm Relay Setting – “ArS”
To select the Alarm Relay Setting, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push the
Enter button to enter the Con menu. The fifth selection isthe “ArS” or Alarm Relay Setting. Push Next four times to get to “ArS” then
Enter. If the display is “diS” (disabled) push Next to change it to 20, 25, 30, 35, 40, 45, 50, 55, 60, 65, 70, 75 (default), 80, 85, 90, 95 or
100ppm NH3 (flashing) then push Enter to confirm the change (solid) and push Enter again to return to “ArS” in the Con menu. Push
Next until “End” is displayed then push Enter to get back to normal operation.
Selecting Alarm Relay Configuration – “Arc”
To select the Alarm Relay Configuration, in normalmode, push the Next button to get to “Con” or the Configuration menu. Thenpush
the Enter button to enter the Con menu. The sixth selection is the “Arc” or Alarm Relay Configuration. Push Next five times to get to
“Arc” then Enter. If the relayis “nO” (normally open) push Next to turn it to “nC” (flashing) then pushEnter to confirm the change (solid)
and push Enter again to return to “Arc” in the Con menu. Push Next until “End” is displayed then push Enter to get back to normal
operation.
Selecting Fan Relay Settings – “FrS”
To select the Fan Relay setting, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push theEnter
button to enterthe Con menu. The seventh selection isthe “FrS” or Fan Relay setting. Push Next six times to get to“FrS” then Enter.
If the fan relay is “diS” (disabled) push Next to change it to 15, 20, 25 (default), 30, 35, 40, 45, 50, 55, 60, 65, 70, or 75 ppm NH3

13
(flashing) then push Enter to confirm the change (solid) and push Enter again to return to “FrS” in the Con menu. Push Next until “End”
is displayed then push Enter to get back to normal operation.
Selecting Fan Relay Delay– “Frd”
To select the Fan Relay Delay setting, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then push the
Enter button to enter the Con menu. The eighth selection is the “Frd” or Fan Relay Delay. Push Next seven times to get to “FrD” then
Enter. If the delay is “0” (disabled) push Next to change it to 1, 3, 5, or 10 minutes (flashing) then push Enter to confirm the change
(solid) and push Enter again to return to “Frd” in the Con menu. Push Next until “End” is displayed then push Enter to get back to
normal operation.
Selecting Fan Relay Run Time – “Frr”
To select the Fan Minimum Runtime setting, in normal mode, push the Next button to get to “Con” or the Configuration menu. Then
push the Enter button to enterthe Con menu. The ninth selection is the “Frr” or Fan Minimum Run Time. Push Next eight times to get
to “Frr” then Enter. If the runtime is “0” (disabled) push Next to change it to 3, 5, 10or 15 minutes (flashing) then push Enter to confirm
the change (solid) and push Enter again to return to “Frr” in the Con menu. Push Next until “End” is displayed then push Enter to get
back to normal operation.
Selecting Fan Relay Latching Option – “FrL”
To select the Fan Relay Latching Option, in normal mode, push the Next button to get to “Con” orthe Configuration menu. Then push
the Enter button to enter the Con menu. The tenth selection is the “FrL” or Fan Relay Latching Option. PushNext nine times to get to
“FrL” then Enter. If latching is “OFF” push Next to turn it to “On” (flashing) then push Enter to confirm the change (solid) and push
Enter again to return to “FrL”in the Con menu. Push Next until “End” is displayed then push Enter to get back to normal operation.
Selecting 4-20mA Output Option – “420”
To select the 4-20mA Output Option, in normal mode, push the Next button to get to “Con” or theConfiguration menu.Then push the
Enter button to enterthe Con menu. The eleventh selection is the “420” or 4-20mA Output Option. Push Next ten times to get to “420”
then Enter. If the 4-20mA is “On” push Next to turn it to “OFF” (flashing) then push Enter to confirm the change (solid) and push Enter
again to returnto “420” in the Con menu. Push Next until “End” is displayed then push Enter to get back to normal operation.
Onboard Diagnostics
The TX-6-AM monitors all critical functions of the unit through software diagnostics that continuously test and verify unit operations. If a
problem is found, the unit will switch to a fail-safe/error mode or trouble condition. In this error mode, the Fan and Alarm relays will be
activated, the 4-20 mA current loop will go to 24 mA, the unit will display the error code and the buzzer will chirp intermittently. This is a
safety precaution. To clear this mode, simply turn off power to the unit for a few seconds, or push the ENTER/TEST switch (inside the
unit). This will cause the unit to restart the 1 minute self-test cycle.

14
The 4-20 mA signal can be used for troubleshooting:
0 mA is most likelya connection problem
4-20 mA is normal gas reading range (0-100 ppm)
24 mA indicates a Trouble condition
Error Codes
t01 Sensor is missing
t02 Temperature compensation failed
t04 Bad EEPROM checksum
t10 Bad EEPROM
t20 Bad calibration
t40 Factory calibration wasnot done
t80 ADC reading failed
t100 Under range sensor
t200 Sensor warranty expired
NOTE: For trouble codes over 080 the display will alternatebetween t_1 and t00for t100 and between t_2 and t00 for t200.
If the error mode repeats frequently, check for continuous power and proper voltage. If power is not the problem and a unithas
repeating errorconditions, it may need to be returned toMacurco for service, per these User Instructions.
If the error mode indicates “Sensor warranty expired”see the Sensor Replacement section of these User Instructions.
Sensor Poisons
The gas sensor in the detector is designed with extreme sensitivity to the environment. As a result, the sensing function may be
deteriorated if it is exposed to adirect spray from aerosols such as paints, silicone vapors, etc., or to a high density of corrosive gases
(such as Ammonia, sulfur dioxide) for an extended period of time.
MAINTENANCE
The TX-6-AM requires periodic maintenance. The unit uses an electrochemical sensor with a 2-3 year life expectancy that can be
tested, calibrated and replaced in the field. The TX-6-AM replacement sensor is available through your local representative or from
Macurco. All other maintenance and repair of products manufactured by Macurco are to be performed at the appropriate Macurco
manufacturing facility. Macurco doesnot sanction any third-party repair facilities.
End-of-Warranty Signal
Two years after the TX-6-AM is installed the sensor end-of-warranty signal will be activated indicating that the TX-6-AM sensor has
reached the end of its warranty period. The end-of-warranty signal will cause an error code t200 “Sensor warranty expired”. See Error
Codes section. In addition to the end-of-warranty signal the sensor itself has a tamper proof warranty expiration sticker dated 2 years
from the unit manufacture date. See the Macurco Fixed Gas Detection Products Limited Warranty section.

15
The end-of-warranty signal can be silenced for 48 hours by pressing the "ENTER/TEST" button or by temporarily dropping power to the
unit. The end-of-warranty signal provides the user an opportunity to test and/or calibrate the sensor assuring that it is still performing
within acceptable parameters though the sensor is nearing the end of its 2-3 year expected life. The silence function will continue to be
available for 29 days after the TX-6-AM initiates the initial end-of-warranty signal. After this29 day period the TX-6-AM can no longer be
silenced and the sensor must be replaced or the sensor life reset.
Sensor Replacement
1. Remove power to the unit
2. Remove the Philips screw on the front of the TX-6-AM. Pull the front cover of the unit off.
3. Remove the sensor by pulling it gently from the three pronged socket.
4. Remove the Shorting Spring from the new sensor if installed and insert thenew sensor intothe socket.
5. Power up the unit. The TX-6-AM steps through an internal self-test cycle for the first 1 minute that it is powered. During theself-test
cycle the unit will display the Firmware Version number, then count down from 60 to 0 and finally go into normal operation. The
indicator light (LED) will flash green during the self-test cycle. At the end of the 1 minute cycle, the unit will take its first sample of
the air and the indicator light will turn solid green.
6. Let the new sensor stabilize for at least 5 minutes then refer to the FIELD CALIBRATION PROCEDURE section to zero and
calibrate the unit.
7. After the successful calibration is complete, reset the sensor life.
Sensor Life Reset
1. Remove the Philips screw on the front of the TX-6-AM. Pull the front cover of the unit off.
2. To reset the sensor life (rSt), from normal or warm-up mode, press the Next button four times to get toSEn or Sensor Mode.
3. Then press the Enter button to getto “rSt” - Reset Sensor Mode.
4. Press the Enter button again to see the sensor reset status. If the sensor life has already been reset, done “don” will be displayed.
If it has not already been reset, “no” will be displayed. Push Next to change it to “yES” (flashing) then push Enter to confirm the
change (solid) and push Enter again to return to “rSt” in the SEn menu. Push Next until “End” is displayed then push Enter to get
back to normal operation.
NOTE: If the sensor is reset and not replaced it is necessary to test and/or calibrate the sensor to assure that it is still performing within
acceptable specifications though the sensor is nearing the end of its 2-3 year expected life. There will be no other indication of sensor
performance.
! WARNING

16
Do not disassemble unit orattempt to repair or modify any component of this instrument. This instrument contains no user serviceable
parts, and substitution of components may impair intrinsic safety, which may adversely affect product performance and result in
sickness or death.
CAUTION
Avoid the use of harsh cleaning materials, abrasives and other organic solvents. Such materials maypermanently scratch the surfaces
and damage thedisplay window, labels, sensor or instrument housing.
Cleaning
Cleaning of the external surfaces is best carried out using a damp cloth with a mild detergent or soap. Use a vacuum cleaner with soft
brush to remove dust or contamination under the cover. Do not blow outthe sensorwith compressed air.
TESTING
Using a certified gas with a concentration other than the one listed for this detector when conducting a calibration or calibration
verification test (bump test) will produce inaccurate readings. This means that higher levels of the gas beingmonitored may be present
and could result in overexposure and cause sickness or death. For proper use, see supervisor or User Instructions, or call Technical
Service at 1-877-367-7891
General
All TX-6-AM units are factory calibrated and 100% tested for proper operation. The unit also performs a regular automatic self-test
during normal operation. If the unit detects an improper voltage or inoperable component, it will default into Error mode. In this error
mode, the Fan and Alarm relays will be activated, the 4-20 mA output will go to 24 mA, the unit will display the error code and the
buzzer will chirp intermittently.
Operation Test
Normally this will be the only test required for the TX-6-AM and is the recommended way to test the unit or units after installation. Check
that the green TX-6-AM operating LED light is illuminated continuously. If not, do not proceed with the tests. Ifthe unit is in error mode
contact your local representative or Macurco technical service representative for information on resolving the problem.
1. Remove the single screw in the middle of the front cover of the TX-6-AM.
2. Remove the front cover.
3. Locate the switch labeled ENTER/TEST onthe left side of the printed circuit board.
4. Observe the LED light on the frontof the TX-6-AM.
5. If the light is solid green proceed to step 7.
6. If the light is offor flashingGreen, refer to the General section above.
! WARNING

17
7. Push the Test switchonce.
8. The TX-6-AM will step through a cycle test:
a) The display progresses through the BUZ (buzzer test), Art (alarm relay test), Frt (fan relay test) then 42t (4-20
mA output test).
b) During the first 3 seconds ofthe test cycle, the Buzzer will sound
c) During the next 5 seconds of the test cycle, the Alarm relay will be closed, so any devices connected to that
relay will be tested.
d) The Fan relay will be activated for the next 1 minute of the test, so if the fan circuits are wired in the normal
manner, the fan should run.
e) The 4-20mA output will then ramp up from 4 to 20 mA over the next 130 seconds of the test, so if the circuit is
wired in the normal manner, the control panel or building automation system should respond.
f) At the end of the test cycle, the light will turn green and be on steady (Normal Operation), the Fan & Alarm
relay will be in standby mode and the 4-20 mA output will return to 4 mA (in clean air).
9. When testing is completed reassemble the unit or units.
Manual Operation Test
This option gives the user the opportunity to manually initiate an individual test for each relay, the analog output and the sensor
response to gas. From normal operation mode press the Next button 2 times to get to the Test Mode (tSt). Press the Enter button once
to get into the Test Menu. Press the Next button to scroll through the five test options and press Enter to initiate the selected test. Note
that if the relay or 4–20 mA output has been disabled, the test selection will not be displayed in the test menu.
BUZ – Buzzer Test, 3 seconds
Art - Alarm Relay Test, 5 seconds
Frt - Fan Relay Test, 60 seconds
42t - 420 loop test, 130 seconds
gtS - Gas Test, 3 minutes (no output to the panel during the gas test)
The display will flash during the test, or in the case of the gas test, the gas level will alternate with gtS. Once the test is complete, the
display will return to steady display. To exit the test menu, press the Next button until “End” is displayed, then press Enter to return to
normal mode.

18
Ammonia Gas Test
The following steps must be performed when conducting a calibration or calibration verification test (bump test) to ensure proper
performance ofthe monitor.Failure to do so mayadversely affect product performance and result in sickness or death.
When performinga calibration or calibration verification test (bump test) only use certified calibration gas at the
required concentration level. Do not calibrate with expired calibration gas.
If the instrument cannot be calibrated, do not use until the reason can be determined and corrected.
Do not cover or obstruct display or visual alarm cover.
Ensure sensor inlets are unobstructed and is free of debris
General
The TX-6-AM can be bump-tested or calibrated with the AM-FCK with Ammonia gas, regulator and test hood, available through your
local representative or from Macurco.
Contents of the FCK
AM-FCK: (1) Gas Cylinder, 25 ppmAmmonia gas in air
Gas regulator with about two feet of plastic tubing
Gas test hood
FCK Information
Several detectors can be calibrated with one FCK. The only limitation is the amount of gas in the cylinder and the flow of the regulator.
The 34 liter cylinder for example with a 0.5LPM regulator has approximately 68 minutes of continuous calibration run time.
Replacement cylinders are available. The gas cylinder should be replaced when the pressure gauge on the regulator shows 25-psi or
less.
Note: For optimum test results it is suggestedthat the unit be in clean air (green light on) andbe in a low ambient air flow
Gas Testing
Testing the Fan Relay –
1. Remove the Philips screw on the front of the TX-6-AM. Remove the front cover.
2. Open the FCK. Connect the 25 ppm gas cylinder to the regulator.
3. Check the pressure gauge on the regulator. If you have 25-psi or less you will need to replace the gas canister.
4. Assemble regulator, hose and test hood and place theTest Hood over the gas sensor.
Note: The time to activate the Fan relay depends on the delay setting.
! WARNING

19
5. Turn on the regulatorto start the gasflow and wait with the gas applied continuously.
6. With the display function turned “On”, the TX-6-AM will show the current concentration of gas or “0” (zero) in clean air. When
the gas concentration reaches the Fan Relay setting (20 ppm, for example) the display will flash back and forth between “FAn”
and “20”. With the displayfunction turned ”OFF”, the display does not show the gas concentration, but will show “FAn” as long
as the fan relay is activated.
Note: If the Fan relay does not close within 2 minutes, there are five possibilities:
a. Gas cylinder is empty, check the pressure gauge. Replace thegas cylinder if 25psior less.
b. Unit needs to be re-calibrated (go through recalibration and re-test).
c. Detector is in need of servicing (return unit to factory for servicing).
d. Detector has fan relay set to disable (diS) or higher than 20. Set fan relay to 20 ppmand repeat the test.
e. Detector has fan relay delay set to 3 minutes. Set fan relay to 0 minutes and repeat the test.
7. Remove the gas from the sensor. Proceed to Test the Alarmrelay or replace thetop cover. Fan relay test is complete.
Testing the Alarm Relay
Note: The gas concentration to activate the Alarm relay depends on the setting.
1. Connect the 25 ppm cylinder of Ammonia to the regulator.
2. Check the pressuregauge. If there is 25psi or less the cylinder shouldbe replaced.
3. Place the test hood over the gas sensor. Turn on the regulator to start the gas flow.
4. The Fan relay should activate accordingto the settings.
5. With the display function turned “On” and the gas concentration reaching the Alarm Relay setting, (20 ppm, for example) the
display will flash back and forth between “ALr” and “20”. The buzzer will sound indicating “Alarm” if the buzzer is turned “On”.
With the display function turned off the display does not show the gas concentration, but will show “ALr” when the Alarm relay
is activated.

20
Note: If the Alarm relay fails to operate within 2 minutes, there are four possibilities:
a. Gas cylinder is empty, check the pressure gauge. Replace thegas cylinder if 25-psi or less.
b. Unit needs to be re-calibrated (go through recalibration and re-test).
c. Detector is in need of servicing (return unit to factory for servicing).
d. Detector has Alarm relay set to disable (diS) or higher than 20. Set Alarm relay to 20 ppm and repeat the test.
6. Remove the gas from the sensor after Test. Proceed to Test the 4-20mA output or replace the top cover. Alarm relay test is
complete.
Testing the 4-20mA current loop
1. Connect the 25 ppm cylinder of Ammonia to the regulator.
2. Check the pressuregauge. If there is 25-psi or less thecylinder should be replaced.
3. Place the test hood from the regulator over the gas sensor. Turn on the regulator to start the gas flow.
4. The Fan relay should activate accordingto the settings.
5. The Alarm relay should activate according tothe settings.
6. The 4-20 mA output should ramp up from 4mA in clean air to 8 mA at 25 ppm. See 4-20 mA diagram on page 7.
Note: If the 4-20mA output does not ramp up within 2 minutes, there are four possibilities:
a. Gas cylinder is empty, check the pressure gauge. Replace thegas cylinder if 25-psi or less.
b. Unit needs to be re-calibrated (go through recalibration and re-test).
c. Detector is in need of servicing (return unit to factory for servicing).
d. Detector has 4-20 mAoption set to “OFF”. Set 4-20mA option to “On”and repeat the test.
7. Remove the gas from the sensor. Re-assemble the TX-6-AM (make sure the LED is aligned with the front case hole). 4-20
current loop test is complete.
FIELD CALIBRATION PROCEDURE
Note: For optimum calibration results the unit should be in clean air and be in a low ambient air flow.
Zero the Sensor
1. Ensure that the ambient air has less than 0.1 ppm NH3 before proceeding with Zeroing the sensor. Remove the Philips screw
on the front of the TX-6-AM. Pull the front cover of theunit off.
2. To select Calibration Zero Mode (000), from normal mode, press the Next button three times to get to CAL or Calibration
Mode.
3. Then press the Enter button to get to “000” - Calibration Zero Mode.
4. Press the Enter button and the display will read the current gas level alternating with 000 (blinking) indicating zero calibration
in progress (max 165 sec).
5. If the process is successful, the display will read 0 alternating with PAS (blinking) Zero Calibration complete.
Other manuals for TX-6-AM
1
Table of contents
Other Macurco Gas Detector manuals

Macurco
Macurco CM-6 User manual

Macurco
Macurco CM-E1 User manual

Macurco
Macurco TX-6-HS User manual

Macurco
Macurco RD-6 User manual

Macurco
Macurco TX-6-ND User manual

Macurco
Macurco TX-6-AM User manual

Macurco
Macurco 6 Series User manual

Macurco
Macurco ND-2 User manual

Macurco
Macurco DMK-1 User manual

Macurco
Macurco GD-6 User manual