Notes:
• If no valid buffer is detected,
the LCD will show the
“WRNG” message.
• For exiting the calibration after the first point
is accepted, press the “ON/OFF” button: the
LCD will show the “OK 1” message for 1
second and then returns to the normal mode.
If using pH 4.01 or 10.01 (or 9.18) buffer
for the first calibration point, the meter will
exit the calibration mode after recognizing
the buffer value and showing the “OK 1”
message for 1 second.
• To quit the calibration procedure and return
to the last data, press “ON/OFF” after en-
tering the calibration mode and before the
first point is accepted. The secondary LCD
shows the “ESC” message for 1 second
and the meter returns to the normal mode.
• To clear a previous calibration and reset to
the default value, press the “SET” key after
entering the calibration mode and before
the first point is accepted. The secondary
LCD shows the “CLR” message for 1 sec-
ond, the meter resets to the default calibra-
tion and the “CAL” tag on the LCD
disappears.
BATTERY REPLACEMENT
The meter shows the remaining battery per-
centage every time it is switched on. When the
battery level is below 5%, the battery symbol
on the bottom left of the LCD lights up to
indicate a low battery condition.
The meter is also provided with BEPS (Battery
Error Prevention System), which automatically
switches the meter off if the battery level is so
low to cause erroneous readings. It is recom-
mended to replace immediately the battery.
Turn the meter off, re-
move the battery compart-
ment cover from the rear
of the meter and replace
the rundown 9V battery
with a new one.
Install the battery while
paying attention to its po-
larity and reattach the
cover.
SPECIFICATIONS
RANGE (*) -2.00 to 16.00 pH
-5.0 to 105.0 ºC or 23.0 to 221.0 °F
RESOLUTION 0.01 pH
0.1 °C or 0.1 °F
ACCURACY (@25°C) ±0.02 pH
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140 °F; ±2°F outside
TYPICAL EMC DEVIATION ±0.02 pH
±0.2 °C or ±0.4 °F
TEMPERATURE COMPENSATION
Automatic, from -5 to 80°C
pH CALIBRATION Automatic, 1 or 2-point
PROBE (included) MA914BR/1
amplified pH/temperature probe
ENVIRONMENT 0 to 50°C; 100% RH max.
BATTERY 1 x 9V alkaline (included)
BATTERY LIFE approx. 500 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 200×85×50 mm
WEIGHT 260 g (with battery)
(*) The temperature range is limited to 80°C (176°F) if
using the MA914BR/1 probe.
ACCESSORIES
MA914BR/1Combination amplified pH/
Temperature probe with
BNC & RCA connectors and
1 m cable
M10004B pH4.01 buffer solution, 20
ml sachet, 25 pcs.
M10006B pH6.86 buffer solution, 20
ml sachet, 25 pcs.
M10007B pH7.01 buffer solution, 20
ml sachet, 25 pcs.
M10009B pH9.18 buffer solution, 20
ml sachet, 25 pcs.
M10010B pH10.01 buffer solution, 20
ml sachet, 25 pcs.
MA9004 pH4.01 buffer solution, 230
ml bottle
MA9006 pH6.86 buffer solution, 230
ml bottle
MA9007 pH7.01 buffer solution, 230
ml bottle
MA9009 pH9.18 buffer solution, 230
ml bottle
MA9010 pH10.01 buffer solution, 230
ml bottle
MA9015 Electrode storage solution,
230 ml bottle
MA9016 Electrode cleaning solution,
230 ml bottle
M10000B Electrode rinse solution, 20
ml sachet, 25 pcs.