  9. Martini Instruments MI 406 User manual

Martini Instruments MI 406 User manual

Other Martini Instruments Measuring Instrument manuals

Martini Instruments MI 306 User manual

Martini Instruments

Martini Instruments MI 306 User manual

Martini Instruments pH 58 User manual

Martini Instruments

Martini Instruments pH 58 User manual

Martini Instruments MI 806 User manual

Martini Instruments

Martini Instruments MI 806 User manual

Martini Instruments MI 170 User manual

Martini Instruments

Martini Instruments MI 170 User manual

Martini Instruments Mi 160 User manual

Martini Instruments

Martini Instruments Mi 160 User manual

Martini Instruments MI 415 User manual

Martini Instruments

Martini Instruments MI 415 User manual

Martini Instruments pH 55 User manual

Martini Instruments

Martini Instruments pH 55 User manual

Martini Instruments MI 415 User manual

Martini Instruments

Martini Instruments MI 415 User manual

Martini Instruments MI 805 User manual

Martini Instruments

Martini Instruments MI 805 User manual

Martini Instruments Mi 106 User manual

Martini Instruments

Martini Instruments Mi 106 User manual

Martini Instruments pH 56 User manual

Martini Instruments

Martini Instruments pH 56 User manual

Martini Instruments MI 605 User manual

Martini Instruments

Martini Instruments MI 605 User manual

Martini Instruments MI 413 User manual

Martini Instruments

Martini Instruments MI 413 User manual

Martini Instruments MI 408 User manual

Martini Instruments

Martini Instruments MI 408 User manual

Martini Instruments MI 404 User manual

Martini Instruments

Martini Instruments MI 404 User manual

Martini Instruments MI 405 User manual

Martini Instruments

Martini Instruments MI 405 User manual

Martini Instruments MI 411 User manual

Martini Instruments

Martini Instruments MI 411 User manual

Martini Instruments Mi 105 User manual

Martini Instruments

Martini Instruments Mi 105 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

USER MANUAL
MI 406
Free Chlorine Meter
Dear Customer,
Thank you for choosing a Martini product. This
manual will provide you with the necessary infor-
mation for the correct use of the instrument.
Please read it carefully before using the meter.
This instrument is in compliance with directives.
Range 0.00 to 5.00 mg/L Cl2
Resolution 0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision ±0.04 mg/L @ 1.50 mg/L
LightSource Tungstenlamp
LightDetector Silicon Photocell and 525 nm
narrowbandinterferencefilter
Method Adaption of the USEPA
method 330.5 and Standard
Method 4500-Cl G.
Environment 0 to 50°C (32 to 122°F);
100% RH max.
Battery Type 1 x 9 volt
Auto-Shutoff After 10' of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
SPECIFICATIONS:
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays “- - -”,
the meter is ready.
3• Fill the cuvet with 10 mL of
unreacted sample, up to the
mark, and replace the cap.
4• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
5• Press ZERO and “SIP” will blink
on the display.
6• Wait a few seconds and the
display will show “-0.0-”. The
meter is now zeroed and
ready for measurement.
9• Replace the cuvet into the holder and
ensure that the notch on the cap is posi-
tioned securely into the groove.
10• Press READ and “SIP” will blink
during measurement.
11• The instrument directly displays
the concentration in mg/L of free chlorine.
Interferences
• Positive error:
- Bromine
- Chlorine Dioxide
- Iodine
- Oxidized Manganese and Chromium
- Ozone
• Alkalinity above 250 mg/L CaCO3 or acidity
above 150 mg/L CaCO3will not reliably
develop the full amount of color or it may
rapidly fade.
To resolve this, neutralize the sample with
diluted HCl or NaOH.
• In case of water with hardness greater
than 500 mg/L CaCO3, shake the sample
for approximately 2’ after adding the
powder reagent.
7• Remove the cuvet from the meter.
8• Add 3 drops of DPD 1 indicator and 3 drops
of DPD 2 buffer to an empty cuvet.
Immediately add 10 mL of unreacted sample.
Replace the cap and shake gently.
ISTMI406 03/05
GUIDETODISPLAYCODES:
This prompt appears for 1
second each time the instru-
ment is turned on.
The dashes “- - -” indicates
that the meter is in a ready
state and zeroing can be per-
formed.
Sampling In Progress. Flash-
ing “SIP” prompt appears each
time the meter is performing a
measurement.
“-0.0-”, the meter is in a ze-
roed state and measurement
can be performed.
The blinking "BAT" indicates
that the battery voltage is get-
ting low and the battery needs
to be replaced.
“-bA-”, the battery is dead and
must be replaced. Once this
indication is displayed, the
meter will lock up. Change the
battery and restart the meter.
“Conf”, the meter has lost its
configuration. Contact your
dealer or the nearest Martini
Customer Service Center.
ERROR MESSAGES
On zero reading
•Blinking “-0.0-” indicates
that the zeroing procedure
failed due to a low signal-
to-noise ratio. In this case
press ZERO again.
•“no L”, the instrument can-
not adjust the light level.
Please check that the
sample does not contain
any debris.
•“L Lo”, there is not enough
light to perform a measure-
ment. Please check the
preparation of the zero
cuvet.
•“L Hi”, there is too much
light to perform a measure-
ment. Please check the
preparation of the zero
cuvet.
On sample reading
•“-SA-”, there is too much
light for the sample mea-
surement. Please check if
the right sample cuvet is
inserted.
•“Inv”, the sample and the
zero cuvet are inverted.
•“ZErO”, a zero reading was
not taken. Follow the in-
struction in the measure-
ment procedure for zeroing
the meter.
•Under range. A blinking
“0.00” indicates that the
sample absorbs less light
than the zero reference.
Check the procedure and
make sure you use the same
cuvet for reference (zero)
and measurement.
•A flashing value of the maxi-
mum concentration indicates
an over range condition.
The concentration of the
sample is beyond the pro-
grammed range: dilute the
sample and re-run the test.
BATTERY REPLACEMENT
Battery replacement must only take place in a
non-hazardous environment.
Simply rotate the battery cover on the back of
the meter.
ACCESSORIES:
MI506-100 Free Chlorine (100 tests)
MI0001 Glass cuvets (2 pcs)
MI0002 Caps for cuvets (2 pcs)
MI0003 Stoppers for cuvets (2 pcs)
MI0004 Tissue for wipping cuvets
(4 pcs)
MI0005 9V battery (1 pc)
WARRANTY
This instrument is warranted against defects in mate-
rials and manufacturing for a period of two years
from the date of purchase.
If during this period the repair or replacement of parts
is required, where the damage is not due to negli-
gence or erroneous operation by the user, please
return the meter to either distributor or our office in
the original packing and the repair will be free of
charge.
Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.
Detach the battery
from the terminals
and attach a fresh
9V battery while
paying attention to
the correct polarity.
Insert the battery
and replace the