MED Associates ENV- 112CM User manual

instrumentation and software for research
RETRACTABLE RESPONSE LEVER
ENV-112CM
PRODUCT MANUAL
DOC-050
Rev. 1.4
Copyright ©2017
All Rights Reserved
Med Associates Inc.
P.O. Box 319
St. Albans, Vermont 05478
Phone: 802.527.2343
Fax: 802.527.5095
www.med-associates.com

- ii -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
Table of Contents
Chapter 1 | Introduction................................................................................................1
Specs............................................................................................................................................... 1
Chapter 2 | Connections ................................................................................................2
Operate Connector........................................................................................................................ 2
Response Connector ..................................................................................................................... 2
*Optional Lever Position Connector ........................................................................................... 2
Chapter 3 | Adjustable Tensioner ................................................................................3
Chapter 4 | Maintenance ...............................................................................................3
Cleaning.......................................................................................................................................... 3
Cleaning Supplies/Tools ............................................................................................................... 3
Instructions .................................................................................................................................... 4
Appendix A | Contact Information...............................................................................5

MED ASSOCIATES INC. RETRACTABLE LEVER
- 1 -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
CHAPTER 1 | INTRODUCTION
The ENV-112CM is a low profile stainless steel retractable response lever used in experiments
where the user wants to deny subjects access to a response.
The ENV-112CM permits placement just 2.1cm above the grid floor and can be raised as needed
with chamber filler panel inserts. This lever is designed to work with ENV-007, ENV-008, and
ENV-009 series chambers, and includes (3) SG-216A 18” connection cables.
Figure 1.1 – ENV-112CM
Figure 1.2 – SG-216A 18” Cable
Specs
•Panel: 3" W x 3.25" H (7.6 x 8.3 cm)
•Lever: 1.88" W x 0.75" D (4.8 x 1.9 cm)
•Lever Protrusion: 0.75" (1.9 cm)
•Position Above Floor Grid: 0.835" (2.1 cm)
•Lever Tension: Factory set to 25 grams

MED ASSOCIATES INC. RETRACTABLE LEVER
- 2 -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
CHAPTER 2 | CONNECTIONS
The lever extension motor is activated with a single output control line. By turning the output
bit “ON” the lever will extend into the chamber. When the output bit is turned “OFF” the lever
will retract out of the chamber. When the animal depresses the lever the signal is received on
the input bit.
Operate Connector
Operate connects to any available Output on a Standard MED Connection Panel via an included
SG-216A cable.
Molex Pin # Function Wire Color
1 28 Volt Ground Black
2 Extend/Retract Lever White
3 +28 Volts Red
Response Connector
Response connects to any available Input on a Standard MED Connection Panel via an included
SG-216A cable.
Molex Pin # Function Wire Color
1 28Volt Ground Black
2 Lever Response White
3 + 28 Volts Red
*Optional Lever Position Connector
A lever position feature has been added to the circuitry of this low profile lever. When the lever
is extended, a signal is sent to an Input bit. The signal is turned off when the lever is retracted.
Lever Position connects to any available Input on a Standard MED Connection Panel via an
included SG-216A cable.
Molex Pin # Function Wire Color
1 28Volt Ground Black
2 Lever Position White
3 + 28 Volts Red

MED ASSOCIATES INC. RETRACTABLE LEVER
- 3 -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
CHAPTER 3 | ADJUSTABLE TENSIONER
The adjustable lever tension is factory set to 25 grams but may be adjusted from 15 to 100
grams. Turn the tensioner clockwise to increase lever tension and counter clockwise to decrease
lever tension. The tension should not be adjusted without consulting Med Associates Support.
Figure 3.1 – Tensioner
CHAPTER 4 | MAINTENANCE
Cleaning
It is very important and necessary to schedule periodic cleaning of the ENV-112CM Retractable
Response Lever to ensure reliable operation and response detection. This should be done once a
week or once a month based on use, to reduce buildup of dander, dust, or debris which can lead
to unreliable operation and response detection. Humid environments may require more
frequent cleaning. Med-Associates recommends that sticky substances such as sucrose, peanut
butter, etc. should not be placed on the paddles of retractable levers as it can cause damage to
mechanical and electrical components.
NOTE: Items returned for repair that have not been properly cleaned will be not covered
under warranty. Please contact Med Associates (see Appendix A | Contact Information) if
you would like to send your Retractable Response Lever to us for a routine cleaning
service. We will be happy to send you a quote for this service.
Cleaning Supplies/Tools
• #1 Philips Screwdriver
• Compressed Air
• Swab and Wipes
• Quatricide or 10% Ethanol

MED ASSOCIATES INC. RETRACTABLE LEVER
- 4 -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
Instructions
1. Remove Retractable Lever from chamber and disconnect the cables from back of lever.
2. Using a #1 Philips screwdriver, remove the 4 screws (2 on each side) shown below.
3. Slide the case backwards, making sure not to pull completely off as there are wires
connecting the circuit board to the lever assembly.
4. Spray compressed air into lever, below the paddle, also spraying into all gaps and
crevices to loosen any debris and allowing debris to fall out the lever.

MED ASSOCIATES INC. RETRACTABLE LEVER
- 5 -
DOC-050 Rev 1.4 Copyright © 2017
Med Associates, Inc.
5. Using a swab soaked with cleaning solution, clean the lever paddle and any other areas
with build-up or residue as needed.
6. Spray compressed air (see step 4) again to blow out any remaining debris.
7. Slide the case forward and reattach using the 4 screws removed in step 2. Reinstall lever
in chamber.
APPENDIX A | CONTACT INFORMATION
Please contact Med Associates, Inc. for information regarding any of our products.
For Technical questions, email
For Sales questions, email
support@med-associates.com.
Visit our website at
sales@med-associates.com.
Phone: 802.527.2343
www.med-associates.com.
Fax: 802.527.5095