Meditech Europe Multiwave Oscillator User manual

Operating Instructions
MultiWave Oscillator
(MWO)
Original instructions
Edition EN V1 2023/01

Edition EN V1 2023/01
2/17
Project Data : January 2023
Trade name : Meditech Europe B.V.
Product name : MultiWave Oscillator (MWO)
Product number : 100012
Model : 220/240 White
Power :220/240 V / 35 mA
Manufacturer : Meditech Europe B.V.
Autor : J.K.
Address : Daalder 14
8305 BE Emmeloord
The Netherlands
Phone: 0031 527 292331
www.meditecheurope.nl
info@meditecheurope.nl
VAT NL859592480B01
Revision index : V1
Revision date : 2023/01

Edition EN V1 2023/01
3/18
Table of Contents
1About these operating instructions......................................................5
1.1 General......................................................................................................5
1.2 Warning signs............................................................................................5
2General safety regulations.....................................................................6
2.1 Principles...................................................................................................6
2.2 Intended use..............................................................................................6
2.3 Foreseeable misuse..................................................................................6
2.4 Selection and qualification of personnel ...................................................7
2.5 Position of safety devices..........................................................................7
2.6 Function of safety devices.........................................................................7
2.7 Safety regulations......................................................................................7
2.8 Expansion and conversion........................................................................8
2.9 Measures in an emergency.......................................................................8
3Scope of supply.......................................................................................9
4Possible accessories (available from Meditech Europe) .................10
5Technical description ...........................................................................11
5.1 Overview..................................................................................................11
5.2 Interfaces.................................................................................................11
5.3 Type plate................................................................................................11
6Transport................................................................................................12
6.1 Before the transport.................................................................................12
6.2 Transporting the machine........................................................................12
6.3 Unpacking the machine...........................................................................12

Edition EN V1 2023/01
4/18
7Operating the machine .........................................................................13
7.1 Switching the machine on .......................................................................13
7.2 Operating the machine............................................................................13
7.3 Switching the machine off .......................................................................14
8Troubleshooting....................................................................................15
9Decommissioning..................................................................................16
9.1 Decommissioning....................................................................................16
9.2 Disposal and recycling ............................................................................16
10 Guarantee...............................................................................................17
11 EC Declaration of Conformity..............................................................18

Edition EN V1 2023/01
5/18
1About these operating instructions
These operating instructions should be read in full before operating
the Multiwave Oscillator (MWO) for the first time or if you have been
authorised to perform other tasks using the MWO. Please contact
Meditech Europe in case of any questions.
In particular, chapter 2 „General safety regulations“must be
observed.
1.1 General
These instructions are intended to facilitate familiarization with the
MWO and utilisation of its intended application capabilities.
The operating instructions contain important notes which allow the
MWO to be operated safely and properly. Observing them will assist
in:
◼Avoiding risk of harm
◼Avoiding repair costs and breakdown times
◼Increasing the reliability and lifespan of the product
Persons authorised to perform tasks using the MWO must be at least
21 years of age and must read and apply these operating instructions.
In addition to these operating instructions, the regulations
concerning accident prevention and environmental protection
applicable in the respective country of use and on the site of
application must be observed.
1.2 Warning signs
Special safety notes are listed at the relevant points where
applicable. They are identified with the following symbols.
Power current
This sign indicates that the device is connected to a voltage of 230
volts ac. This requires the device to be de-energised if accessories are
being connected.
Non-ionising radiation
This means that the device will emit low-energy radiation from radio
waves at radio frequencies.

Edition EN V1 2023/01
6/18
2General safety regulations
2.1 Principles
In case of any doubt about the correct setup, please contact
Meditech Europe.
Use a Faraday cage (available at Meditech Europe) when the MWO is
in use to prevent disruption of (mobile) networks, telephones,
laptops and/or other electrical devices and networks in your vicinity.
Do not open the MWO device. This may only be done by a specialist
from Meditech Europe. Opening the MWO device by anyone other
than a Meditech Europe specialist will render the guarantee void.
To avoid damage to the device, do not use the MWO without
connected accessories (available from Meditech Europe).
The MWO is an experimental relaxation device and not a medical
device. No rights may be derived and/or claims made from the use of
the MWO. The use of the MWO is entirely at the risk and expense of
the user.
Cleaning: unplug the MWO. Use a dry cloth. A damp cloth can be
used for stubborn stains. Be careful not to get the connectors wet.
When in doubt about how to use, switch on and/or set up the MWO,
always contact Meditech Europe BV via telephone number 0031 527
292331 or by e-mail at info@meditecheurope.nl.
Meditech Europe is not liable for misuse, incorrect use or damage
caused by any use other than as described in this manual. In the
event of any dispute, Dutch law shall apply.
2.2 Intended use
The energy field can be used to provide a wellness experience for the
human body. It can also be used on animals, plants, and food.
2.3 Foreseeable misuse
This MWO does not replace regular medical care. Always see a
medical specialist when needed.

Edition EN V1 2023/01
7/18
2.4 Selection and qualification of personnel
The MWO is produced by Meditech Europe located in Emmeloord in
the Netherlands. Meditech Europe employs only qualified personnel.
2.5 Position of safety devices
Fig. 2.5 Red Circle = Fuse
2.6 Function of safety devices
Fuse –a safety system that ensures no short circuit can endanger
operation of the device. The fuse will cut the electrical supply in the
event of a short circuit.
2.7 Safety regulations
Never use the MWO if you are fitted with a pacemaker, an
Implantable Cardioverter Defibrillator (ICD) or an insulin pump.
Hearing devices should be removed.
Always contact a physician or medical specialist when in doubt about
use of the MWO in your particular case.
Always use the MWO in a dry and well-ventilated area. Place the
MWO with at least 30 centimetres of free space on all sides where

Edition EN V1 2023/01
8/18
the air can be properly ventilated, and no other materials may come
into contact with the device.
Never block the vents of the MWO.
Make sure the MWO is not in contact with heat sources such as
electrical stoves, gas stoves, open fire, or other heat sources.
Never use the MWO outdoors.
The MWO must not be left in operation and unattended.
Do not touch the cable connectors when the MWO is switched on.
If the power cord is damaged it should be replaced with a special
power cord available exclusively from Meditech Europe. Make sure
the power cord hangs freely.
2.8 Expansion and conversion
The device needs to be de-energised to connect any accessories.
2.9 Measures in an emergency
In the event of fire use a fire extinguisher for electronic devices.

Edition EN V1 2023/01
9/18
3Scope of supply
MultiWave Oscillator:
Power cord:

Edition EN V1 2023/01
10/18
4Possible accessories (available from Meditech Europe)
-Faraday Cage
-Lakhovsky Antennas
-Rife Therapy Tube
-Therapy pillow
-Printplate Antenna
For a total overview, please visit our website:
www.meditecheurope.nl
All available Meditech Europe accessories are connected to the MWO
at the customer's own risk.
Meditech Europe is not liable for misuse, incorrect use or damage
caused by any use other than as described in this manual.
The use of accessories not supplied by Meditech Europe will render
the guarantee void.
Availability of accessories may change. For up-to-date information
visit our website: www.meditecheurope.nl.

Edition EN V1 2023/01
11/18
5Technical description
5.1 Overview
The MWO creates high-frequency resonances. The device requires
220V. The 220V passes through a transformer, a spark gap, and the
Tesla coil to create high-frequency resonances and voltages.
5.2 Interfaces
Accessories may be connected to the device using cables. The device
will transmit the voltage through the cables.
5.3 Type plate
Fig. 5.3 Type Plate
Manufacturer : Meditech Europe
Details MWO : type and power
Serial number : unique

Edition EN V1 2023/01
12/18
6Transport
The MWO should be handled with care.
6.1 Before the transport
Meditech Europe will pack the MWO carefully for transport or pick-
up.
6.2 Transporting the machine
The device must be transported in the horizontal position.
6.3 Unpacking the machine
The MWO should be unpacked with care. Do not use any sharp tools.

Edition EN V1 2023/01
13/18
7Operating the machine
7.1 Switching the machine on
Plug in the power cord.
On/Off button - Turns the MWO on.
The fans will be heard immediately.
7.2 Operating the machine
Pulse Strength - This button adjusts the power of the MWO which
makes it possible to determine the strength of the energy field.
Turning it clockwise opens the spark gap and makes the spark
between the contact points sound louder. Turning the button further
(always turn clockwise) reduces the gap between the contact points
and makes the spark sound softer.
Each device is made by hand. This also makes the pulse settings for
each device unique.
Once the pulse strength is set as desired, proceed to ‘Fine Tuning’.
Important: The Pulse Strength button should only be turned
clockwise!
Fine Tuning - Once you have set the pulse strength the MWO needs
to be fine-tuned to make the energy field stable. Turn this button
clockwise until the sound is stable and as constant as possible.
Whenever the pulse strength is adjusted, the fine tuning must also
be adjusted to obtain a stable and constant sound.
Time Set - This sets the duration of the treatment and activates the
MWO. By turning the timer to the desired number of minutes the
MWO will start the treatment session. When the desired number of
minutes has passed the MWO will automatically switch off.
Accessories can be connected to the MWO via the connectors.

Edition EN V1 2023/01
14/18
7.3 Switching the machine off
On/Off button - Turns the MWO off.
The fans will switch off.

Edition EN V1 2023/01
15/18
8Troubleshooting
Turning the Pulse Strength button anti-clockwise may damage the
spark gap. In this case the MWO will no longer respond. Contact
Meditech Europe and do not try to repair it yourself.
After the MWO has been used for several years the spark gap
requires replacing. This may be indicated by a metallic odour and
noise. Fine tuning the MWO sound may also no longer be possible.
Contact Meditech Europe and do not try to replace it yourself.

Edition EN V1 2023/01
16/18
9Decommissioning
9.1 Decommissioning
Make sure that the MWO is switched off using the On/Off button.
The fans should not be audible after switching off.
Unplug the machine.
Disconnect any accessories from the connectors.
9.2 Disposal and recycling
Contact Meditech Europe for disposal and recycling of the MWO and
accessories. Meditech Europe is pleased to receive all used products
for disposal and recycling according to environmental regulations.
Please contact your local authorities should you wish to dispose of
the product yourself.

Edition EN V1 2023/01
17/18
10 Guarantee
Meditech Europe will guarantee this product for 2 years.
Opening the MWO device by anyone other than a Meditech Europe
specialist will render the guarantee void.
Meditech Europe is not liable for misuse, incorrect use or damage
caused by any use other than as described in this manual.
Service Address
Meditech Europe B.V.
Daalder 14
8305 BE Emmeloord
The Netherlands
Phone: 0031 527 292331
www.meditecheurope.nl
info@meditecheurope.nl

Edition EN V1 2023/01
18/18
EC DECLARATION OF CONFORMITY
This declaration is issued under the complete responsibility of the manufacturer
MANUFACTURER
Company name: Meditech Europe
Full address: Daalder 14
Postcode: 8305 BE
City: Emmeloord
Country: Netherlands
Person authorized to
compile the technical fil
e
: Ms. J. Kraaij / Operational Director
MACHINE DESCRIPTION AND IDENTIFICATION
Generic name: Multiwave Oscillator
Use: The energy fie l
d can pr oduce a wel lness exper ience for
the human body, animals, plants and food
Model: Multiwave Oscillator
Type: 220/240 White
Commercial name: Multiwave Oscillator
AGREEMENT
The manufacturer declares that the above machine complies with all applicable provisions
of the
Low Voltage Directive 2014/35/EU
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU & Delegated Directive (EU) 2015/863
Whereby the following harmonized standards and where appropriate technical standards
and specific
a
t ions have been used
EN ISO 12100:2010; EN IEC 61000-3-2:2019; EN IEC 61000-3-3:2013;
IEC 61000-6-4:2019; EN IEC 61010-1:2011; EN IEC 63000:2018;
EN 55016-2-1:2014; EN 55016-2-3:2017; EN 55011:2016;
EN IEC 61326-1:2013
City: Emmeloord, Name: Mr R. Frankhuisen
Netherlands Position: Owner
Date: 26 January 2023 Signature:
11 EC Declaration of Conformity
Other manuals for Multiwave Oscillator
1
This manual suits for next models
1
Table of contents
Popular Measuring Instrument manuals by other brands

Motrona
Motrona touchMATRIX PN350 operating manual

Bellingham + Stanley Ltd.
Bellingham + Stanley Ltd. E-Line 90 instruction manual

Delta
Delta LTL-XL quick guide

FRONIUS
FRONIUS Smart Meter IP quick start guide

Digi-Sense
Digi-Sense 20250-31 user manual

Aqualytic
Aqualytic CHECKIT Comparator D55 instruction manual

Kyoritsu Electrical Instruments Works, Ltd.
Kyoritsu Electrical Instruments Works, Ltd. 6300 instruction manual

SUPER NOVA
SUPER NOVA DV-41 operating manual

Almaco
Almaco UNI-2 user manual

Promax
Promax TV EXPLORER II user manual

ABQINDUSTRIAL
ABQINDUSTRIAL DELMHORST J-2000 owner's manual

In-situ
In-situ VuLink CI Operator's manual