Warranty and calIbratIon
The MicroShot Injector is warranted against defects in materials and workmanship for one year from date of purchase.
Please contact your distribution partner for service or return policies.
Each MicroShot Injector is calibrated to deliver a stroke of the plunger slide within +/- .025 mm of nominal.
Most syringes are calibrated by the manufacturer therefore, please refer to your syringe manufacturer
for calibration requirements for your particular syringe.
saFety and cleanIng
Please use all applicable laboratory safety procedures when using sharp objects or chemical solvents.
Laboratory goggles should be worn at all times when working with the MicroShot Injector.
The injector can be cleaned with warm solution of laboratory-grade detergent and rinsed with deionized water.
Dry the injector completely before use.
Half Moon Bay, CA | Newark, DE
PreParIng For an InjectIon
1. With the Plunger Stop in the released position, rinse
and fill the syringe with sample using the Plunger Slide.
2. Keeping the Plunger Slide raised, arm the Plunger Stop.
3. Lower the Plunger Slide slowly until the Volume Rod
comes to rest on the Plunger Stop.
4. Wipe the needle.
5. The MicroShot Injector is now filled and armed for
injection.
MaKIng an InjectIon
1. Align the syringe needle with the injection port.
2. Push the needle all the way into the injection port and then
withdraw the MicroShot Injector. The injection occurs when
the Trigger is displaced by the top of the injection port.
3. Start the instrument.
tIPs For best PerForMance
For split injections use a small amount of glass wool in the inlet liner to prevent the sample from reaching the bottom of the
liner, which would cause vaporization past the split point. Other appropriate inlet liner designs will also serve this same function.
Glass wool also helps wick low volatility compounds from the tip of the syringe needle.
An easy way to verify the MicroShot Injector is functioning properly is as follows:
• Load and arm the syringe.
• Hold the needle tip approximately 2 cm above a piece of paper.
• Manually push the trigger and listen for an audible click.
• Look for a small spot of liquid on the paper.
ProbleM Probable cause solutIon
Syringe Plunger end doesn’t go into
Plunger Slide.
Plunger Bolt is in the closed position. Move Plunger Bolt to the open position
(Figure 3).
Sample doesn’t fill syringe. Syringe plunger not connected to Plunger Slide. Close Plunger Bolt (Figure 4).
Syringe needle is plugged. Remove and clean or replace syringe.
Microshot Injector fails to make
injection.
Plunger Stop not armed before injection. Arm Plunger Stop prior to injection (Figure 5).
Trigger is not getting pressed in all the way. Make sure Trigger contacts septum nut.
Plunger Slide or syringe plunger sticking. Remove and clean syringe. Verify free plunger
movement.
troubleshootIng
Figure 5
Step 3
Step 2
Step 1
MERLIN INSTRUMENT COMPANY 2010-07