manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Milwaukee
  6. •
  7. Measuring Instrument
  8. •
  9. Milwaukee 2205-20 User manual

Milwaukee 2205-20 User manual

OPERATOR'S MANUAL
TO REDUCE THE RISK OF INJURY, USER MUST READ AND UNDERSTAND OPERATOR'S
MANUAL.
Cat. No.
2205-20
2206-20
FORK METERS
23
IMPORTANT SAFETY INSTRUCTIONS
Symbology
WARNING READ ALL SAFETY WARNINGS AND INSTRUCTIONS.
Failure to follow the warnings and instructions may result in electric shock, fire and/or serious
injury, as well as instrument damage and/or damage to the equipment being tested.
Save these instructions - This operator’s manual contains important safety and operating
instructions for the MILWAUKEE Fork Meters. Before using, read this operator’s manual and
all labels on the Fork Meters.
DANGER
Never make measurement on a circuit in which voltage over AC1000V exists. Use only leads
rated 1000V or better.
Do not attempt to make measurement in the presence of flammable gases. Otherwise, the use
of the instrument may cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measuring range.
Only test on unenergized circuits unless absolutely necessary.
Checktoolfunctionalityonaknowncircuitfirst.Neverassumetoolisworking.Assumecircuits
are live until they can be proven de-energized.
Do not ground yourself while measuring. Avoid body contact with earthed or grounded sur-
faces such as pipes, radiators, ranges and refrigerators.
Never open the Battery cover during a measurement.
This instrument is to be used only in its intended applications or conditions. Otherwise, the
instrument's safety functions may not work, resulting in serious personal injury and instru-
ment damage.
To reduce the risk of injury from shock and arc blasts, always wear personal protective
equipment where live conductors are exposed.
WARNING
Never attempt to make measurement if any abnormal conditions, such as broken case and
exposed metal parts are found on the instrument.
Do not rotate the Rotary Dial while the test leads are being connected.
Verify proper operation on a known source before use or taking action as a result of the indi-
cation of the instrument.
Do not install substitute parts or make any modification to the instrument. For repair or re-
calibration, return the tool to a factory Service/Sales Support Branch or authorized service
station.
Do not try to replace the batteries if the surface of the instrument is wet.
Disconnect all the cords and cables from the object under test and power off the instrument
before opening the Battery Cover for Battery replacement.
This tool is designed to be powered by 2-AA batteries properly inserted into the MILWAUKEE
Fork Meters. Do not attempt to use with any other voltage or power supply.
Install battery according to polarity (+ and –) diagrams.
Do not leave batteries within the reach of children.
Do not mix new and used batteries. Do not mix brands (or types within brands) of batteries.
Properly dispose of used batteries.
Do not incinerate or dismantle batteries.
Under abusive conditions, liquid may be ejected from the battery, avoid contact. If contact
accidentally occurs, flush with water. If liquid contacts eyes, additionally seek medical help.
Liquid ejected from the battery may cause irritation or burns.
CAUTION
Set the Dial to an appropriate position before starting measurement.
Firmly insert the test leads.
Disconnect the test leads from the instrument for current measurement.
Do not expose the instrument to the direct sun, high temperature and humidity or dewfall.
Altitude 2000m or less. Appropriate operating temperature is within -10ºC and 50ºC.
Keep away from excess dust and water.
Be sure to power off the instrument after use. When the instrument will not be in use for a long
period, place it in storage after removing the batteries.
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use
abrasives or solvents.
Read Operator’s Manual
Double insulation
Risk of electric shock
Indicates that this instrument can
clamp on bare conductors when
measuring a voltage correspond-
ing to the applicable measurement
category, which is marked next to
this symbol.
Earth
Danger, Warning, or Caution -
Consult the operators manual for
additional safety information.
Battery compartment
European Conformity Mark
Underwriters Laboratories, Inc.,
United States and Canada
Cat III Classification of transient overvolt-
ages, based on nominal line voltage
to earth.
Cat IV
Do not dispose of this product as
unsorted municipal waste.
Accuracy is specified for 1 year after calibration, at
operating temperatures of 18°C to 28°C (64°F to
82°F), with relative humidity at 0 % to 85 %.
Maximum voltage between any terminal and
earth ground.... 1000 V
Jaw Opening (maximum conductor size) ....
approx. 5/8” (16 mm)
Temperature ....
Operating: -10°C to 50°C (14°F to 122°F)
Storage: -40°C to 60 °C (-40°F to 140°F)
Temperature Coefficient .... 0.1 x (specified
accuracy)/°C (<18°C or >28°C)
Operating Altitude.... 2,000 meters
IP Rating
(International Dust and Water Protection).... IP54
Drop Test .... 1 Meter
Battery .... 2 AA, NEDA 15 A,IEC LR6
Battery Life .... Approx. 26 Hours all lights on.
Safety Compliances .... EN61010-1, UL 61010-1,
EN61010-031 (Probes),
IEC 61010-2-32 (Clamp Assemblies),
IEC/EN 61010-1 2nd Edition for measurement
Category IV 600V, Category III 1000V,
Pollution Degree 2, EMC EN61326-1
Certifications .... cULus, CE
General Specifications
45
Dial Position Range Resolution Accuracy
Voltage AC/DC 400V
1000V 0.1V/1V : ±(1.5% + 5 dgt) 45-500 Hz
: ±(1.0% + 2 dgt)
Current AC 200A 0.1A ±(3.0% + 3 dgt) 45-60 Hz
Resistance
Continuity
400.0
4.000k
40.00k
400.0k
4.000M
40.00M
0.1
0.001k
0.01k
0.1k
0.001M
0.01M
±(1.0% + 5 dgt)
±(1.0% + 2 dgt)
±(1.0% + 2 dgt)
±(1.0% + 2 dgt)
±(1.0% + 2 dgt)
±(2.0% + 5 dgt)
Capacitance (2206-20) 1000F1F±(1.9% + 2 dgt)
100.0F 0.1F
Current DC (2206-20) 400.0A
1000A0.1A ±(1.0% + 2 dgt)
Temperature C° / F°
(2206-20) -40ºC ~ 400ºC
-40ºF ~ 752ºF 0.1°C
0.2°F C°:±(1% + 2C°)
F°:±(1% + 3 F°)
Lo-Z
Lo-Z Low Input Impedance
(2205-20) 400.0V
1000V 0.1V/1V ±(2.0% + 3 dgt) AC:45~500Hz
Functions
FUNCTIONAL DESCRIPTION
1. Light On/Off button
2. Hold button
3. Jaws
4. Light
5. NCV indicator
6. Dial
7. Display
8. Terminal inputs
4
2
1
3
6
7
8
Cat. No.
2206-20
Lo-Z
6
Cat. No.
2205-20
5
ASSEMBLY
Selecting C° or F°
1. TurnRotaryDialto OFFand
disconnect the test leads.
2. Unscrew and remove bat-
tery door and remove any
installed batteries.
3. Set the C°/F° switch to the
desired position.
4. Replacebatteriesaccording
to “Loading/Changing the Batteries.
5. Close the battery door and tighten screw se-
curely.
Loading/Changing the Batteries
Replace batteries when the Low
Battery indicator is displayed.
1. TurnRotary DialtoOFF anddiscon-
nect the test leads.
2. Unscrew and remove battery door.
3. Inserttwo(2)AAbatteries,according
to the polarity marked in the battery
compartment
4. Close the battery door and tighten screw
securely.
WARNING
To avoidan electricalhazard, turntheRotary
Dial to OFF and disconnect the test leads
before opening battery compartment or
replacing batteries.
OPERATION
Before Use
Confirmthe RotaryDial is settothecorrectposition,
the instrument is set to the correct measurement
mode, and the Data hold function is disabled. Oth-
erwise, desired measurement cannot be made.
LCD Backlight
The LCD backlight willturn off after about3 minutes
of inactivity. Push any button or turn the rotary dial
to turn the backlight on.
Making a Measurement
Voltage
DANGER
To avoid electrical shock:
Never make measurement on a circuit in
which voltage over AC 1000V or DC 1000V
exists.
Do not use with the Battery Cover removed.
Keep fingers behind the guards and away
from test lead tips during measurements.
1.Set the Dial to posi-
tion.
2.Connect the red test lead
to the Vterminal and the
black test lead to the COM
terminal.
3.AC: Connect the test leads
to the circuit under test. The
reading is displayed.
DC: Connect the red test
lead to the positive (+)
side and black test leads
to the negative (-) side of
the circuit under test. The
reading is displayed. A reversed connection is
indicated as a negative value.
Current
1.Set the Dial to
position.
2.Run theconduc-
tor through the
jaws. The read-
ing is displayed.
NOTE: Do not
place 2 or more
wires between
jawsatthe same
time. Place wire between arrows. Otherwise,
irregular results will occur.
CAUTION
Maximum conductor size is approx 5/8”
diameter.
DANGER
To avoid electrical shock:
Never make measurement on a circuit in
which voltage over AC 1000V or DC 1000V
exists. Jaws are designed not to short the
circuitunder test.If equipmentunder testhas
exposed conductive parts, however, extra
precaution should be taken to minimize the
possibility of shorting.
Do not use with the Battery Cover removed.
Disconnectthetestleadsfromtheinstrument
for current measurement.
Correct
Incorrect
Incorrect
Lo-Z Low Input Impedance
(Cat. No. 2205-20 only)
Automatic voltage detection
(AC or DC).
1.Set the Dial to Lo-Z posi-
tion.
2.Connect the red test lead
to the Vterminal and the
black test lead to the COM
terminal.
3.AC: Connect the test leads
to the circuit under test. The
reading is displayed.
DC: Connect the red test
lead to the positive (+) side
and black test leads to the negative (-) side of
the circuit under test. The reading is displayed.
Areversedconnectionisindicated as anegative
value.
WARNING
Only use MILWAUKEE test leads with the
MILWAUKEE Fork Meters.
Inspect test leads before each use. Use me-
ter to run a continuity test.
67
Resistance
1.Set the Dial to position.
2.Connect the red test lead
to the Vterminal and the
black test lead to the COM
terminal.
Confirm “OL” is indicated
on the display, and then
short-circuit the tips of test
leadstomakethe indication
zero.
3.Connect the test leads to
thebothendsof the resistor
under test.
4.The reading is displayed.
Continuity
1.Set the Dial to position.
2.Connect the red test lead
to the Vterminal and the
blacktestleadto the COM
terminal.
Confirm “OL” is indicated
on the display, and then
short-circuit the tips of
test leads to make the
indication zero. A buzzer
will sound.
3.Connect the test leads
to the both ends of the
conductor under test. If
the resistance under test
is 30 or less, the buzzer
will sound.
Capacitance
(Cat. No. 2206-20 only)
1.Set the Dial to position.
2.Connect the red test lead
to the Vterminal and the
black test lead to the COM
terminal.
3.Discharge capacitor.
4.Connect the test leads to the
both ends of the capacitor
under test.
5.The reading is displayed.
CAUTION
When current flows from the display side to
the underside of the meter, the polarity is
positive; flow from underside to display side,
the polarity is negative.
Current
Flame Rectification Circuit Test
(Cat. No. 2206-20 only)
1.Set the Rotary Dial to
position. DC mark
is displayed.
2.Connect the red test
lead to the Vterminal
and the black test lead
to the COM terminal.
Contact the red test
lead to the flame sen-
sor probe and the black
test lead to the control
module.
3.Turn on the heating unit. The reading is dis-
played.
DANGER
To avoid electrical shock:
Never make measurement on a circuit in
which voltage over AC 1000V or DC 1000V
exists.
Do not use with the Battery Cover removed.
Flame
sensor
probe
Control
Module
DANGER
To reduce the risk of electric shock for
Resistance, Continuity, and Capacitance
measurements, never use the meter on an
energized circuit. Make sure a capacitor is
fully discharged before touching or attempt-
ing to make a measurement.
Do not use with the Battery Cover removed.
CAUTION
After shorting the test leads, the displayed
value may not be zero due to the resistance
of test leads themselves.
WARNING
Never connect the Temperature Probe to an
energized circuit.
CAUTION
When the dial is set to , the room tempera-
ture should be displayed. If anything else is
displayed, something may wrong with the
meter. Stop using the meter immediately.
Temperature (Cat. No. 2206-20 only)
CAUTION
TheDataHoldreadingsarereleasedwhenthe
meter enteres Sleep Mode.
Non-Contact Voltage Detection (NCVD)
Sleep Mode
The meter is automatically powered off in about 20
min after the last Rotary Dial or button operation.
To reset, turn the Rotary Dial to OFF. If the display
is still blank when a new Rotary Dial setting is
selected, replace the batteries.
The meter does use battery power in sleep mode.
Be sure to switch the tool to OFF to conserve bat-
tery power.
ACCESSORIES
WARNING Always remove batteries
before changingor removing accessories.Only
useaccessories specifically recommendedfor
this tool. Others may be hazardous.
DANGER
The LED may not be displayed due to instal-
lation condition of electrical circuit or equip-
ment. Never touch the circuit under test to
avoid possible danger even if the LED for
NCVD is not displayed.
Check the functionality of LED on a well-
known power supply prior to measurement.
When the LED doesn’t light up, do not make
measurement.
NCVD indication is affected by external volt-
age, and how the meter is held or placed.
HOLD Key
Data Hold Function - Freezes the value on the
display. Press the “HOLD” button to freeze the
reading. The reading will be held regardless of
subsequent variation in input. HOLD is displayed
withthereading.To exit DataHoldmode,press the
HOLD button again.
SMART HOLD: The meter will beep continuously
andthedisplaywillflash ifthemeasuredsignalis 50
counts larger than the display reading. (However,
it can not detect across the AC and DC Voltage/
Current)
When the meter is on in any
function, the non-contact volt-
agedetectorwill indicatewitha
Red LED on the display when
anelectricfieldexceeding 90V
is detected. Place the edge of
either the jaw near the electric
field.
Worklight LED ON/OFF
To turn the light on and off, press the button.
Over-flow indication
Any time the input exceeds the measuring range
“OL” or “-OL” is displayed.
1.SettheRotaryDial to posi-
tion.
2.Connect the K-type Tem-
perature Probe to the input
terminal. The positive (+)
side of Probe should be
connected to V.
3.Place the probe sensor in
the desired location.
4.The reading is displayed.
Discover various automotive tools in our online store.

Other manuals for 2205-20

3

This manual suits for next models

1

Other Milwaukee Measuring Instrument manuals

Milwaukee CD97 User manual

Milwaukee

Milwaukee CD97 User manual

Milwaukee C65 User manual

Milwaukee

Milwaukee C65 User manual

Milwaukee Mi451 User manual

Milwaukee

Milwaukee Mi451 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee MW605 MAX User manual

Milwaukee

Milwaukee MW605 MAX User manual

Milwaukee PH55 User manual

Milwaukee

Milwaukee PH55 User manual

Milwaukee MW101 User manual

Milwaukee

Milwaukee MW101 User manual

Milwaukee M12 3632-20 User manual

Milwaukee

Milwaukee M12 3632-20 User manual

Milwaukee MC730 User manual

Milwaukee

Milwaukee MC730 User manual

Milwaukee pH600 User manual

Milwaukee

Milwaukee pH600 User manual

Milwaukee MI408 User manual

Milwaukee

Milwaukee MI408 User manual

Milwaukee MW13 User manual

Milwaukee

Milwaukee MW13 User manual

Milwaukee MAX MW160 User manual

Milwaukee

Milwaukee MAX MW160 User manual

Milwaukee MI413 PRO User manual

Milwaukee

Milwaukee MI413 PRO User manual

Milwaukee pH 58 User manual

Milwaukee

Milwaukee pH 58 User manual

Milwaukee MC110 User manual

Milwaukee

Milwaukee MC110 User manual

Milwaukee PRO MI415 User manual

Milwaukee

Milwaukee PRO MI415 User manual

Milwaukee MW190 MAX User manual

Milwaukee

Milwaukee MW190 MAX User manual

Milwaukee MI404 User manual

Milwaukee

Milwaukee MI404 User manual

Milwaukee PST MW100 PRO User manual

Milwaukee

Milwaukee PST MW100 PRO User manual

Milwaukee MW 14 User manual

Milwaukee

Milwaukee MW 14 User manual

Milwaukee C12 CMH User manual

Milwaukee

Milwaukee C12 CMH User manual

Milwaukee EC40 User manual

Milwaukee

Milwaukee EC40 User manual

Milwaukee MW700 PRO User manual

Milwaukee

Milwaukee MW700 PRO User manual

Popular Measuring Instrument manuals by other brands

Xylem WTW SenTix 950 operating manual

Xylem

Xylem WTW SenTix 950 operating manual

Pentax 36141 - Digital Spot Meter operating instructions

Pentax

Pentax 36141 - Digital Spot Meter operating instructions

Jula 730-028 operating instructions

Jula

Jula 730-028 operating instructions

Watermark 950R Installation and operating instructions

Watermark

Watermark 950R Installation and operating instructions

OpenEye OE-MM Hardware manual

OpenEye

OpenEye OE-MM Hardware manual

Lumel ND03 user manual

Lumel

Lumel ND03 user manual

TLV EF200R-C technical information

TLV

TLV EF200R-C technical information

Vernier Go Direct GDX-PYR manual

Vernier

Vernier Go Direct GDX-PYR manual

Trans instruments WalkLAB HP 9010 Operation manual

Trans instruments

Trans instruments WalkLAB HP 9010 Operation manual

Global Industrial 412606 instruction manual

Global Industrial

Global Industrial 412606 instruction manual

Kobold HND-PS01 operating instructions

Kobold

Kobold HND-PS01 operating instructions

J.ASSY VISUM Monitor Operator's manual

J.ASSY

J.ASSY VISUM Monitor Operator's manual

Endress+Hauser Micropilot FMR67 technical information

Endress+Hauser

Endress+Hauser Micropilot FMR67 technical information

Danfoss PR-SC4K Modbus installation guide

Danfoss

Danfoss PR-SC4K Modbus installation guide

ICP DAS USA PM-3133-MTCP user manual

ICP DAS USA

ICP DAS USA PM-3133-MTCP user manual

GW Instek GDS-1000B Series user manual

GW Instek

GW Instek GDS-1000B Series user manual

KAM Simple Precision PKF user manual

KAM

KAM Simple Precision PKF user manual

Hioki FT3424 instruction manual

Hioki

Hioki FT3424 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.