Milwaukee MI 413 User manual

USER MANUAL
MI 413
Free & Total Chlorine
High Range Meter
Dear Customer,
Thank you for choosing a Martini product.
This manual will provide you with the
necessary information for the correct use of
the instrument.
Please read it carefully before using the
meter. This instrument is in compliance with
directives.
SPECIFICATIONS:
Range 0.00 to 10.00 mg/L Cl2
Resolution 0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision ±0.10 mg/L @ 5.00 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell and
525 nm narrow band interference filter
Method Adaptation of the USEPA
method 330.5 and Standard Method
4500-Cl G.
Environment 0 to 50°C (32 to 122°F);
100% RH max.
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing
ON/OFF.
2• When the LCD displays “- - -”, the
meter is ready.
3• Fill the cuvet with 10
mL of sample, up to the
mark, and replace the
cap.
4• Place the cuvet into the
holder and ensure that
the notch on the cap is
positioned securely into
the groove.
5• Press ZERO and “SIP”
will blink on the display.
6• After a few seconds the
display will show “-0.0-
”. The meter is now
zeroed and ready for
measurement.
7• Empty the cuvet !
8• For free chlorine measurement:
Add 5 mL of Distiled/Deionized water
by using a syringe
9• Fill the cuvet up to the 10 mL mark
with 5 mL of unreacted sample using
a pipette.
10 • Add the content of one packet
of MI526-100 reagent.
11• Replace the cap and swirl
gently for 20 seconds
12• Reinsert the cuvet into the
holder and ensure that the
notch on the cap is
positioned securely into the groove.
13• Wait 1 minute for free chlorine
and press READ. “SIP” will blink
during measurement.
14• The instrument directly displays
concentration in mg/L of chlorine.
15• For total chlorine measurement,
Prepare new 5 mL of Distiled /
Deionized water by using a
syringe
16• Fill the cuvet up to the 10 mL mark
with 5 mL of unreacted sample
using a pipette
17• Add the content of one packet of
MI524-100 reagent
18• Replace the cap and swirl
gently for 20 seconds
19• Reinsert the cuvet into the holder
and ensure that the notch on the
cap is positioned securely into the
groove.
20• Wait for 2 minutes and 30
seconds then
press READ
and “SIP” will
blink during
measurement.
21• The instrument directly displays
concentration in mg/L of total
chlorine.
Interferences
• Bromine (positive error)
• Chlorine dioxide (positive error)
• Iodine (positive error)
• Oxidized Manganese and Chromium
(positive error)
• Ozone (positive error)
• Alkalinity above 250 mg/L CaCO3 or
acidity above 150 mg/L CaCO3 will
not reliable develop the full amount of
color or it may rapidly fade. To resolve
this, neutralize the sample with diluted
HCl or NaOH.
GUIDE TO DISPLAY CODES:
This prompt appears for 1 second each
time the instrument is turned on.
The dashes “- - -”
indicates that the
meter is in a ready
state and zeroing can
be performed.
Sampling In Progress.
Flashing “SIP” prompt
appears each time the
meter is performing a
measurement.
“-0.0-”, the meter is in
a zeroed state and
measurement can be
performed.
The blinking "BAT"
indicates that the
battery voltage is
getting low and the
battery needs to be
replaced.
“-bA-”, the battery is
dead and must be
replaced. Once this
indication is displayed,
the meter will lock up.
Change the battery
and restart the meter.
“Conf”, the meter has
lost its configuration. Contact your
dealer or the nearest Martini Customer
Service Center.
ERROR MESSAGES
On zero reading
• Blinking “-0.0-” indicates that the
zeroing procedure failed due to a low
signalto-noise ratio.
In this case
press ZERO again.
• “no L”, the
instrument can not
adjust the light
level. Please check
that the sample
does not contain
any debris.
• “L Lo”, there is
not enough light to
perform a
measurement.
Please check the
preparation of the
zero cuvet.
• “L Hi”, there is too
much light to
perform a
measurement.
Please check the preparation of the zero
cuvet.
On sample reading
• “-SA-”, there is too
much light for the
sample
measurement.
Please check if the
right sample cuvet is
inserted.
• “Inv”, the sample
and the zero cuvet
are inverted.
• “ZErO”, a zero
reading was not
taken. Follow the instruction in the
measurement procedure for zeroing the
meter.
• Under range. A blinking “0.00”
indicates that the
sample absorbs less
light than the zero
reference.
Check the procedure
and make sure you
use the same cuvet
for reference (zero)
and measurement.
• A flashing value of
the maximum
concentration
indicates an over
range condition.
The concentration of the sample is
beyond the programmed range: dilute
the sample and re-run the test.
BATTERY REPLACEMENT
Battery replacement must only take
place in a non-hazardous environment.
Simply rotate the battery cover on the
back of the meter.
Detach the
battery from
the terminals
and attach a
fresh 9V
battery while
paying
attention to
the correct
polarity.
Insert the battery and replace the cover.
ACCESSORIES:
MI513-045 Free & Total Chlorine
reagent set (45 tests)
MI0001 Glass cuvets (2 pcs)
MI0002 Caps for cuvets (2 pcs)
MI0003 Stoppers for cuvets (2 pcs)
MI0004 Tissue for wipping cuvets
(4 pcs)
MI0005 9V battery (1 pc)
WARRANTY
This instrument is warranted against defects
in materials and manufacturing for a period of
two years from the date of purchase.
If during this period the repair or replacement
of parts is required, where the damage is not
due to negligence or erroneous operation by
the user, please
return the meter to either distributor or our
office in the original packing and the repair
will be free of charge.
Damages due to accidents, misuse,
tampering or lack of prescribed maintenance
are not covered.

Other Milwaukee Measuring Instrument manuals

Milwaukee MC310 User manual

Milwaukee

Milwaukee MC310 User manual

Milwaukee EC40 User manual

Milwaukee

Milwaukee EC40 User manual

Milwaukee Mi451 User manual

Milwaukee

Milwaukee Mi451 User manual

Milwaukee pH54 User manual

Milwaukee

Milwaukee pH54 User manual

Milwaukee MW190 MAX User manual

Milwaukee

Milwaukee MW190 MAX User manual

Milwaukee MA888 User manual

Milwaukee

Milwaukee MA888 User manual

Milwaukee 2235-20 User manual

Milwaukee

Milwaukee 2235-20 User manual

Milwaukee 2231-20 User manual

Milwaukee

Milwaukee 2231-20 User manual

Milwaukee Heavy Duty 2270-20 User manual

Milwaukee

Milwaukee Heavy Duty 2270-20 User manual

Milwaukee MI408 User manual

Milwaukee

Milwaukee MI408 User manual

Milwaukee MW306 MAX User manual

Milwaukee

Milwaukee MW306 MAX User manual

Milwaukee C65 User manual

Milwaukee

Milwaukee C65 User manual

Milwaukee L4 CLL User manual

Milwaukee

Milwaukee L4 CLL User manual

Milwaukee MI407 User manual

Milwaukee

Milwaukee MI407 User manual

Milwaukee MW500 User manual

Milwaukee

Milwaukee MW500 User manual

Milwaukee MW170 MAX User manual

Milwaukee

Milwaukee MW170 MAX User manual

Milwaukee MC110 User manual

Milwaukee

Milwaukee MC110 User manual

Milwaukee MW801 User manual

Milwaukee

Milwaukee MW801 User manual

Milwaukee CALIBRATION 2238-20 Operating instructions

Milwaukee

Milwaukee CALIBRATION 2238-20 Operating instructions

Milwaukee LDM 30 User manual

Milwaukee

Milwaukee LDM 30 User manual

Milwaukee MC510 User manual

Milwaukee

Milwaukee MC510 User manual

Milwaukee MI415 User manual

Milwaukee

Milwaukee MI415 User manual

Milwaukee MC125 User manual

Milwaukee

Milwaukee MC125 User manual

Milwaukee PRO MI415 User manual

Milwaukee

Milwaukee PRO MI415 User manual

Popular Measuring Instrument manuals by other brands

King Instrument 7480 Series installation instructions

King Instrument

King Instrument 7480 Series installation instructions

ESD SYSTEMS SPl-20686 Operation installation and maintenance

ESD SYSTEMS

ESD SYSTEMS SPl-20686 Operation installation and maintenance

Swan Analytical Instruments AMU Inducon Operator's manual

Swan Analytical Instruments

Swan Analytical Instruments AMU Inducon Operator's manual

WSTECH CS DataVario manual

WSTECH

WSTECH CS DataVario manual

CPS TEMP-SEEKER TM360A owner's manual

CPS

CPS TEMP-SEEKER TM360A owner's manual

Tektronix WVR6020 Technical reference

Tektronix

Tektronix WVR6020 Technical reference

SPL-Laboratory Wireless Bass Meter user manual

SPL-Laboratory

SPL-Laboratory Wireless Bass Meter user manual

Ditel KOSMOS JUNIOR-C user manual

Ditel

Ditel KOSMOS JUNIOR-C user manual

ALFAMATIC Visual Point installation manual

ALFAMATIC

ALFAMATIC Visual Point installation manual

Rice Lake 480 Legend Series Operation manual

Rice Lake

Rice Lake 480 Legend Series Operation manual

Testomat CL operating instructions

Testomat

Testomat CL operating instructions

YOKOGAWA PX8000 user manual

YOKOGAWA

YOKOGAWA PX8000 user manual

MEASUREMENT TECHNOLOGY LCI-90 installation manual

MEASUREMENT TECHNOLOGY

MEASUREMENT TECHNOLOGY LCI-90 installation manual

YOSensi YO Analog user guide

YOSensi

YOSensi YO Analog user guide

Dostmann Electronic P705 Operation manuals

Dostmann Electronic

Dostmann Electronic P705 Operation manuals

BRUEL & KJAER 2212 Instructions and applications

BRUEL & KJAER

BRUEL & KJAER 2212 Instructions and applications

Katronic KATflow 150 Quick start manual

Katronic

Katronic KATflow 150 Quick start manual

HBK Z6R Mounting instructions

HBK

HBK Z6R Mounting instructions